Bipolar Coagulation Forceps have been designed to be used with endoscopes to cauterize and coagulate or to perform hemostasis using high-frequency current within the digestive tract.
Device Story
Bipolar Coagulation Forceps are sterile, single-use endoscopic accessories; used with electrosurgical generators to deliver bipolar high-frequency current for cauterization, coagulation, or hemostasis in the digestive tract. Device features two electrodes of opposite polarity at the sheath tip; current flows between electrodes, eliminating need for a patient neutral electrode. Operated by clinicians during endoscopic procedures; output is thermal energy for tissue effect. Benefits include localized hemostasis and tissue coagulation. Device is compatible with endoscope working channels.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including thermal effect studies and animal studies to demonstrate substantial equivalence to the predicate. Biocompatibility testing performed per ISO 10993-5, 10993-10, and 10993-11.
Technological Characteristics
Materials: PTFE, Stainless Steel 304, Stainless Steel N, PEEK450-GL30, Ceramic, Stainless Steel 630. Energy: Bipolar Radio Frequency. Dimensions: 2.5mm sheath, 4.0-7.0mm span, 1650-2300mm length. Connectivity: Standalone, connects to electrosurgical generator. Sterilization: ISO 11135. Safety: IEC 60601.
Indications for Use
Indicated for use with endoscopes to cauterize, coagulate, or perform hemostasis via high-frequency current within the digestive tract.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 5, 2021
Hangzhou AGS MedTech Co., Ltd. Yanping Fu RA Supervisor Building 5, Building 6, No. 597 Kangxin Road Yuhang District Hangzhou, Zhejiang 311106 CHINA
Re: K210406
> Trade/Device Name: Bipolar Coagulation Forceps Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: Class II Product Code: KGE Dated: August 20, 2021 Received: August 26, 2021
Dear Yanping Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) / K210406
Device Name Bipolar Coagulation Forceps
Indications for Use (Describe)
Bipolar Coagulation Forceps have been designed to be used with endoscopes to cauterize and coagulate or to perform hemostasis using high-frequency current within the digestive tract.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
{3}------------------------------------------------
K210406 Page 1 of 6
Image /page/3/Picture/1 description: The image shows the logo for AGS Medtech. The logo consists of a circular design with three curved shapes in blue and green. To the right of the circular design, the text "AGS MEDTECH" is written in English, followed by the same text in Chinese characters. The logo appears to be for a medical technology company.
Section 5 510(k) Summary Bipolar Coagulation Forceps
# 510(k) Summary
We submit this 510(k) Summary as per 21 CFR 807.92, it meets the content and format regulatory requirements.
# 5.1 Submitter
| Submitted by/Owner: | Hangzhou AGS MedTech Co., Ltd.<br>Building 5, Building 6, No.597 Kangxin Road Yuhang<br>District, 311106 Hangzhou, Zhejiang, China |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration Number: | 3010288205 |
| Registration Status: | Active |
| Contact Person: | Yanping Fu<br>Phone: 0086-15958493282<br>Fax: 0086-0571-87671225<br>Email: fuyp@bioags.com |
| Date Prepared: | Feb. 07, 2021 |
# 5.2 Proposed Device
| Trade Name: | / |
|-------------------------|--------------------------------------------------|
| Device Name: | Bipolar Coagulation Forceps |
| Common Name: | Bipolar Coagulation Forceps |
| Regulation class: | Class II |
| Regulation Number: | 876.4300 |
| Regulation Description: | Endoscopic electrosurgical unit and accessories. |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | KGE |
| Product Code Name: | Forceps, Biopsy, Electric |
### 5.3 Predicate Device
| Trade Name: | Coagrasper™ |
|-------------------------|--------------------------------------------------|
| Device Name: | Single-use Electrosurgical Hemostatic Forceps |
| Common Name: | Electrosurgical Hemostatic Forceps Series |
| 510(k) Number: | K062517 |
| Regulation class: | Class II |
| Regulation Number: | 876.4300 |
| Regulation Description: | Endoscopic electrosurgical unit and accessories. |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | KGE |
| Product Code Name: | Forceps, Biopsy, Electric |
### 5.4 Device Description
The Bipolar Coagulation Forceps described in this submission are a sterile, single use
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "K210406" on the first line and "Page 2 of 6" on the second line. The text is in a simple, sans-serif font and is likely part of a document or report. The text suggests that this is page 2 of a 6-page document with the identifier K210406.
Image /page/4/Picture/1 description: The image contains the logo for AGS Medtech. The logo consists of a circular design with three blue curved shapes forming a triangle, with a green triangle in the center. To the right of the circular design, the text "AGS MEDTECH" is written in a sans-serif font, with the Chinese characters for "Anjie Si Medical" written below.
Section 5 510(k) Summary Bipolar Coagulation Forceps
devices compatible with the working channel of endoscope. The device is used with endoscopes to cauterize and coagulate or to perform hemostasis using high-frequency current within the digestive tract.
### 5.5 Indication for use statement
Bipolar Coagulation Forceps have been designed to be used with endoscopes to cauterize and coagulate or to perform hemostasis using high-frequency current within the digestive tract.
### 5.6 Comparison of Technology Characteristics
Our proposed device Bipolar Coagulation Forceps is substantially equivalent to the predicate devices. The differences between the Bipolar Coagulation Forceps and the predicate devices do not raise any questions regarding its safety and effectiveness. The differences are listed in the table below:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for AGS Medtech. The logo consists of a blue and green circular design on the left, with the text "AGS MEDTECH" in black on the right. Below the English text is the same company name in Chinese characters. The logo is simple and professional, and it is likely used to represent the company's brand.
K210406
Page 3 of 6
Section 5 510(k) Summary
Bipolar Coagulation Forceps
| Table 5.6 Comparison of technical characteristics | | | | |
|---------------------------------------------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Item | Proposed device | Predicate device | Comparison |
| | Common name | Bipolar Coagulation Forceps | Single-use electrosurgical hemostatic forceps | / |
| | Trade name | / | Coagrasper™ | / |
| | Model number | 5461, 5462, 5463, 5464 | KD-410LR | / |
| | 510(k) submitter | Hangzhou AGS MedTech<br>Co., Ltd. | Olympus Medical Systems<br>Corporation | / |
| | 510(k) number | / | K062517 | / |
| Technical | Principles of<br>operation | Bipolar Coagulation Forceps<br>is an applied part of<br>electrosurgical generator,<br>using bipolar high-frequency<br>current delivered by the<br>electrosurgical generator to<br>cauterize and coagulate or to<br>perform hemostasis within the<br>digestive tract. We designed<br>two electrodes (two forceps<br>cup) of opposite polarity<br>attached to the tip of its<br>sheath. The high-frequency<br>electricity flows from one<br>forceps cup to another forceps<br>cup. No neutral electrode is<br>needed. | Monopolar Single-use electrosurgical hemostatic forceps is an applied part of electrosurgical generator, using monopolar high-frequency current delivered by the electrosurgical generator to cauterize and coagulate or to perform hemostasis within the digestive tract. The high-frequency electricity flows from the active electrode (forceps) to the neutral electrode placed on patient skin. | Similar.<br>Both Bipolar Coagulation Forceps and Monopolar Single-use Electrosurgical Hemostatic Forceps are applied part of electrosurgical generator.<br>Bipolar Coagulation Forceps has two electrodes (two forceps cup) of opposite polarity attached to the tip of its sheath.<br>Thermal effect study and animal study had been conducted to demonstrate substantial |
| Bipolar Coagulation Forceps | | | | |
| | Item | Proposed device | Predicate device | Comparison |
| | | | | equivalence,<br>please refer to<br>Section 18.2<br>Thermal Effect<br>study and 19<br>Animal study of<br>this submission. |
| | Structure | Bipolar Coagulation Forceps<br>consists of Handle part, plug,<br>sheath and forceps part. | Electrosurgical<br>hemostatic<br>forceps consists of handle<br>part, plug, sheath, forceps<br>part, and A cord. | Similar.<br>Electrosurgical<br>hemostatic<br>forceps has a A<br>cord while<br>Bipolar<br>coagulation<br>forceps don't<br>have. |
| | Energy Use | Bipolar Radio Frequency<br>Current | Monopolar Radio<br>Frequency Current | Different.<br>Thermal effect<br>study and<br>animal study<br>had been<br>conducted to<br>demonstrate<br>substantial<br>equivalence,<br>please refer to<br>Section 18.2<br>Thermal Effect<br>study and 19<br>Animal study of<br>this submission. |
| | Sheath tube | 2.5mm | 2.75mm | Similar. |
| | Span length | 4.0mm, 5.0mm, 6.5mm,<br>7.0mm | 5.0mm | Similar.<br>The difference<br>will not raise<br>any question of<br>safety and<br>effectiveness.<br>Our proposed<br>device has more<br>choice on span |
| Item | Proposed device | Predicate device | Bipolar Coagulation Forceps<br>Comparison | |
| | | | length for<br>operators,<br>operators could<br>choose based on<br>their clinical<br>needs. | |
| Minimum<br>Working<br>Channel: | 3.2mm | 2.8mm | Similar. | |
| | Working length | 1650mm, 1950mm, 2300mm | 1650mm; | Similar.<br>The difference<br>will not raise<br>any question of<br>safety and<br>effectiveness.<br>Our proposed<br>device has more<br>choice on<br>working length<br>for operators,<br>operators could<br>choose based on<br>their clinical<br>needs. |
| Biological | Materials or<br>substances in<br>contact with the<br>same human<br>tissue or body<br>fluids | Plastic PTFE, Stainless<br>Steel 304, Stainless Steel<br>N, PEEK450-GL30,<br>Ceramic, Stainless Steel<br>630. | No exact information. | Different.<br>Biocompatibility<br>tests have been<br>done for the<br>difference.<br>Biological risks<br>are acceptable. |
| | Biocompatibility | In Vitro Cytotoxicity Test:<br>ISO 10993-5: 2009;<br>Skin sensitization Test: ISO<br>10993-10: 2010;<br>Intracutaneous Reactivity<br>Test: ISO 10993-10: 2010;<br>Acute Systemic Toxicity Test:<br>ISO 10993-11:2017;<br>Pyrogenicity: ISO<br>10993-11:2017; | No exact information. | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for AGS Medtech. The logo consists of a circular design with three curved shapes in blue and green. To the right of the logo, the text "AGS MEDTECH" is written in English, followed by the same name written in Chinese characters. The logo appears to represent a company in the medical technology field.
K210406
Page 4 of 6
Section 5 510(k) Summary
Bipolar Coagulation Forceps
{7}------------------------------------------------
K210406
Page 5 of 6
Image /page/7/Picture/1 description: The image contains a logo for AGS Medtech. The logo consists of a circular design with three blue curved shapes forming a triangle, with a green triangle in the center. To the right of the logo, the text "AGS MEDTECH" is written in English, followed by Chinese characters that translate to "Anjie Si Medical".
Section 5 510(k) Summary
Bipolar Coagulation Forceps
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo for AGS Medtech. The logo consists of a circular symbol with three curved shapes in blue and green. To the right of the symbol, the text "AGS MEDTECH" is written in English, followed by the same name written in Chinese characters. The logo appears to be for a medical technology company.
K210406 Page 6 of 6
Section 5 510(k) Summary Bipolar Coagulation Forceps
### 5.7 Applicable Guidance Document NA
# 5.8 Performance Data
The Bipolar Coagulation Forceps meets all design specifications and medical device standards for electrosurgical safety (IEC 60601), biocompatibility (ISO 10993) and sterility (ISO 11135). The non-clinical performance meets the design specification and shows substantial equivalence to the predicated device.
# 5.9 Clinical Test
No Clinical test is included in this submission.
# 5.10 Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Hangzhou AGS MedTech Co., Ltd has demonstrated that proposed device Bipolar Coagulation Forceps is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.