← Product Code [LZA](/productcode/LZA) · K210369

# Powder Free Nitrile Examination Glove (K210369)

_Pastel Glove Sdn Bhd · LZA · May 13, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210369

## Device Facts

- **Applicant:** Pastel Glove Sdn Bhd
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** May 13, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Disposable nitrile examination glove; worn on hand/finger; prevents contamination between patient and examiner. Used in medical settings; operated by clinicians or patients (OTC). Device acts as physical barrier; manufactured from nitrile latex compound. Performance verified via barrier testing, physical property testing, and biocompatibility assessments. Output is a protective barrier; aids in infection control; benefits patient/examiner by reducing pathogen transmission risk.

## Clinical Evidence

No clinical data. Bench testing only. Compliance verified against ASTM D6319-10 (nitrile gloves), ASTM D5151-06 (holes), ASTM D6124-06 (residual powder), ISO 10993-10 (irritation/sensitization), and ISO 10993-11 (systemic toxicity).

## Technological Characteristics

Material: Nitrile (Acrylonitrile-butadiene). Form: Ambidextrous, finger-textured, blue. Standards: ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, ISO 10993-10, ISO 10993-11, ISO 2859-1. Non-sterile. Powder-free.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Powder Free Blue Nitrile Examination Glove ([K153562](/device/K153562.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 13, 2021

Pastel Glove Sdn Bhd Terence Lim RA Specialist PT285724, Jalan Pusing, Taman Perindustrian Perabot Negeri Park Pusing, Perak 31550, Malaysia

## Re: K210369

Trade/Device Name: Powder Free Nitrile Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: April 20, 2021 Received: April 26, 2021

Dear Terence Lim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Ryan Ortega Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210369

Device Name POWDER FREE NITRILE EXAMINATION GLOVE

Indications for Use (Describe)

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Pastel Glove Sdn Bhd

PT 285724, Jalan Pusing, Taman Perindustrial Perabot Negeri Perak, Pusing 31550, Perak, Malaysia

### Attachment 2

## 510 (K) SUMMARY (K210369)

1.0 Submitter Pastel Glove Sdn Bhd PT 285724, Jalan Pusing, Taman Perindustrial Perabot Negeri Perak Pusin 31550, Perak, Malaysia

| Tel:                      | +60-19 552 5000       |
|---------------------------|-----------------------|
| Name of Contact Person:   | Terence Lim           |
| Email Address:            | limsinkooi@@gmail.com |
| Date of Summary Prepared: | April 20, 2021        |

#### 2.0 Name of Device:

| 510(k) number          | K210369                               |
|------------------------|---------------------------------------|
| Trade Name:            | Powder Free Nitrile Examination Glove |
| Classification Name:   | Polymer Patient Examination Glove     |
| Device Classification: | I                                     |
| Regulation Number:     | 21 CFR 880.6250                       |
| Panel:                 | General Hospital                      |
| Product Code:          | LZA                                   |

#### 3.0 Identification of The Legally Marketed Device

Predicate Device Name: Powder Free Blue Nitrile Examination Glove

Predicate 510(K) Number: K153562

Manufacturer's Name: VIETGLOVE CORPORATION

#### 4.0 Description of Device

Powder Free Nitrile Examination Glove meets all the current specifications listed under the ASTM Specification D6319-10. Standard Specification for Nitrile Examination Gloves for Medical Application. The principle operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.

#### 5.0 The Intended Use of Glove

A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner

#### 6.0 Summary of the Technological Characteristic of the Device compared to the Predicate Device for Substantial Equivalent Discussion.

There is no difference in technology characteristic compared to the predicate device. Gloves are made from nitrile latex compound. Non-Sterile, Powder Free Nitrile Examination Gloves has the below technological characteristic compared to ASTM or Equivalent standards.

{4}------------------------------------------------

# Pastel Glove Sdn Bhd

PT 285724, Jalan Pusing, Taman Perindustrial Perabot Negeri Perak, Pusing 31550, Perak, Malaysia

# Comparison between Proposed & Predicate Device

| Characteristics and<br>Parameters                              | Proposed Device -<br>Powder Free Nitrile<br>Examination Gloves<br>(K210369)                                                                                                                     | Predicate Device -<br>Powder Free Blue<br>Nitrile Examination<br>Gloves (K153562 )                                                                                                                 | Discussion                                                   |
|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| Product Code                                                   | LZA                                                                                                                                                                                             | LZA                                                                                                                                                                                                | Same product Code                                            |
| Intended Use                                                   | A patient examination<br>glove is a disposable<br>device intended for<br>medical purposes that<br>is worn on the hand or<br>finger to prevent<br>contamination between<br>patient and examiner. | A patient examination<br>glove is a disposable<br>device intended for<br>medical purposes that<br>is worn on the hand or<br>finger to prevent<br>contamination<br>between patient and<br>examiner. | Same Intended Use.                                           |
| Classification                                                 | Class 1                                                                                                                                                                                         | Class 1:                                                                                                                                                                                           | Same Class.                                                  |
| Raw Rubber Material                                            | Nitrile (Acrylonitrile-<br>butadiene)                                                                                                                                                           | Nitrile (Acrylonitrile-<br>butadiene)                                                                                                                                                              | Same synthetic rubber<br>material.                           |
| Surface Appearance                                             | 1. Blue<br>2. Ambidextrous<br>3. Finger Textured                                                                                                                                                | 1. Blue<br>2. Ambidextrous<br>3. Finger Textured                                                                                                                                                   | Same color, ambidextrous<br>design and same texture<br>area. |
| Overall Length<br>Minimum 230mm                                | Average : 242 mm                                                                                                                                                                                | 240 – 250 mm<br>(specification sheet)                                                                                                                                                              | Similar                                                      |
| Width<br>S: 75mm - 95mm<br>M: 85mm - 105mm<br>L: 100mm - 120mm | Average :<br>S : 84 mm<br>M: 94mm<br>L 103mm:                                                                                                                                                   | Meeting<br>specification                                                                                                                                                                           | Similar                                                      |
| Palm Thickness<br>(Minimum 0.05mm)                             | Average : 0.06                                                                                                                                                                                  | More than 0.05mm                                                                                                                                                                                   | Similar                                                      |
| Finger Thickness<br>(Minimum 0.05mm)                           | Average : 0.08                                                                                                                                                                                  | More than 0.05mm                                                                                                                                                                                   | Similar                                                      |
| Tensile Strength<br>(before age)<br>Minimum 14 MPa             | Average : 17.75 MPa                                                                                                                                                                             | More than 14 MPa<br>(Specification Sheet)                                                                                                                                                          | Similar                                                      |
| Tensile Strength<br>(After Age)<br>Minimum 14 MPa              | Average : 16.07 MPa                                                                                                                                                                             | More than 14 MPa<br>(Specification Sheet)                                                                                                                                                          | Similar                                                      |
| Ultimate Elongation<br>before age<br>(Minimum 500%)            | Average : 560%                                                                                                                                                                                  | 500% - 600%<br>(Specification Sheet)                                                                                                                                                               | Similar                                                      |
| Ultimate Elongation<br>after age<br>(Minimum 400%)             | Average : 510%                                                                                                                                                                                  | 400% - 500%<br>(Specification Sheet)                                                                                                                                                               | Similar                                                      |
| Characteristics and<br>Parameters                              | Proposed Device -<br>Powder Free Nitrile<br>Examination Gloves<br>(K210369)                                                                                                                     | Predicate Device -<br>Powder Free Blue<br>Nitrile Examination<br>Gloves (K153562 )                                                                                                                 | Discussion                                                   |
| Freedom of Holes<br>Meet AQL 2.5 at G1                         | Meet AQL 1.5 with<br>G1                                                                                                                                                                         | Meet AQL 1.5 with<br>G1.                                                                                                                                                                           | Similar                                                      |
| Residual powder test<br>(Less than<br>2mg/glove)               | Average powder<br>residue for each size:<br>S : 0.45 mg/glove<br>M 0.43 mg/glove<br>L 0.27 mg/glove                                                                                             | Contained less than<br>2mg/glove                                                                                                                                                                   | Similar                                                      |
| Primary Skin Irritation                                        | Under the conditions<br>of study, not an<br>irritant                                                                                                                                            | Under the conditions<br>of study, not an<br>irritant                                                                                                                                               | Similar                                                      |
| Dermal Sensitization                                           | Under the conditions<br>of study, not a<br>sensitizer.                                                                                                                                          | Under the conditions<br>of study, not a<br>sensitizer.                                                                                                                                             | Similar                                                      |
| Acute Systemic<br>Toxicity                                     | Not induce systemic<br>toxicity                                                                                                                                                                 | Not done                                                                                                                                                                                           | Meeting the requirements<br>per ISO 10993-11                 |

{5}------------------------------------------------

## Pastel Glove Sdn Bhd

PT 285724, Jalan Pusing, Taman Perindustrial Perabot Negeri Perak, Pusing 31550, Perak, Malaysia

### Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed

Device complies with the following standards:

- ISO 10993-10:2010 Biological Evaluation of Medical Devices Part 10: Tests For Irritation And ● Skin Sensitization.
- ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic . toxicity
- ASTM D6124-06,Standard Test Method for Residual Powder on Medical Gloves ●
- ASTM D5151-06, Standard Test Method for Detection of Holes in Medical Gloves. ●
- . ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application
- ISO 2859-1:1999, "Sampling Procedures for Inspection by Attributes Part I: Sampling Plans ● Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection.

### Clinical Test Conclusion

No clinical test is included in this submission

### Conclusion

The conclusion drawn from the nonclinical tests demonstrates the subject device in 510(K) submission K210369. Powder Free Nitrile Examination Glove is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K153562.

---

**Source:** [https://fda.innolitics.com/device/K210369](https://fda.innolitics.com/device/K210369)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
