s-Clean Pre-Milled Abutment is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. s-Clean Pre-Milled Abutment is intended for use with the s-Clean SQ-SL Fixtures in the chart. All digitally designed abutments for use with s-Clean Pre-Milled Abutment are intended to be manufactured at a Dentis validated milling center.
Device Story
Pre-milled titanium alloy blanks used to fabricate patient-specific dental abutments; processed via CAD/CAM technology at Dentis-validated milling centers. Clinician prescribes patient-specific design; abutment supports single or multiple tooth prostheses in maxilla or mandible. Device connects to s-Clean SQ-SL dental implants via hex or non-hex interface; secured with provided abutment screw. Used in dental clinics; provides structural support for prosthetic restorations; benefits patients by enabling customized, stable tooth replacement.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including fatigue testing per ISO 14801:2016 on worst-case configurations. Biocompatibility and sterilization validation leveraged from previously cleared Dentis devices (K200099, K111364).
Indicated for use with dental implants as support for single or multiple tooth prostheses in maxilla or mandible of partially or fully edentulous patients. Compatible with s-Clean SQ-SL Fixtures.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized representation of a human figure. The FDA name is written in blue, with the words "U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.
August 19, 2021
Dentis Co., Ltd. % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620
Re: K210362
Trade/Device Name: s-Clean Pre-Milled Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: July 13, 2021 Received: July 19, 2021
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K210362
Device Name s-Clean Pre-Milled Abutment
#### Indications for Use (Describe)
s-Clean Pre-Milled Abutment is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
| No. | Implant System Compatibility | Implant Diameter (mm) | Platform Diameter (mm) |
|-----|------------------------------|-----------------------|------------------------|
| 1 | s-Clean SQ-SL Fixture | $Ø5.8$ | $Ø4.3$ |
| 2 | s-Clean SQ-SL Fixture | $Ø6.8$ | $Ø4.5$ |
s-Clean Pre-Milled Abutment is intended for use with the s-Clean SQ-SL Fixtures in the chart. All digitally designed abutments for use with s-Clean Pre-Milled Abutment are intended to be manufactured at a Dentis validated milling center.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
#### Submitter
Dentis Co., Ltd. Gyu Ri Kim 99, Seongseoseo-ro, Dalseo-gu Daegu, 42718 Korea Email: kgr1026@dentis.co.kr Tel. +82-53-589-3541 Fax. +82-53-289-7922
# Device Information
- Trade Name: s-Clean Pre-Milled Abutment
- Common Name: Dental Abutment System
- Classification Name: Endosseous dental implant abutment
- Product Code: NHA
- Panel: Dental
- Regulation Number: 872.3630
- Device Class: Class II ●
- Date prepared: 08/18/2021
# Predicate Devices:
Primary Predicate K181037, DIO CAD/CAM Abutment by DIO CORPORATION
Reference Device K171694, s-Clean TiN Coating Abutment by Dentis Co., Ltd. K111364, HAPTITE Coating Implant System K200099, s-Clean SQ-SL Implant System Mini
# Indication for Use:
s-Clean Pre-Milled Abutment is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
| No. | Implant System Compatibility | Implant Diameter (mm) | Platform Diameter (mm) |
|-----|------------------------------|-----------------------|------------------------|
| 1 | s-Clean SQ-SL Fixture | Ø5.8 | Ø4.3 |
| 2 | | Ø6.8 | Ø4.5 |
s-Clean Pre-Milled Abutment is intended for use with the s-Clean SQ-SL Fixtures in the chart. All digitally designed abutments for use with s-Clean Pre-Milled Abutment are intended to be manufactured at a Dentis validated milling center.
# Official Correspondent
Withus Group Inc. April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
{5}------------------------------------------------
#### Device Description:
Patient-specific abutment is made from titanium alloy conforming to ASTM F136 titanium abutment to be used in fabricating patient-specific abutments. The subject abutments are indicated for cemented or "Screw-and Cement-Retained Prosthesis(SCRP)" restorations. Each patient-specific abutment is individually prescribed by the clinician.
The diameters of patient-specific abutment are 5.8, 6.8mm and two connection designs (Hex, Non-hex).
Patient-Specific Abutment is compatible with following Implant Systems:
| Proprietary Name | Dentis s-Clean s-Line |
|---------------------------------------------|-------------------------------|
| Compatible Implants (K number) | K210134 |
| Implant diameter size | 5.8/6.8 |
| Implant Interface Connection Type/Size (mm) | Internal Connection type/ 2.5 |
| Type of Implant-Abutment Connection | Hex/Non-Hex |
Patient-Specific Abutments are supplied with s-Clean abutment screw and provided non-sterile.
| Patient-Specific Abutment design Limitation (Unit :mm) | | | |
|--------------------------------------------------------|------------------|----------------|---------------|
| Model Name | Range (Diameter) | Range (Length) | Range (Angle) |
| DSCBA10H | 4.5-6.5 | 7.3-17.9 | 0-30° |
| DSCBA14H | | | |
| DSCBA10N | | | |
| DSCBA14N | | | |
| DSCBM10H | 4.5-6.5 | 7.3-17.9 | 0-30° |
| DSCBM14H | | | |
| DSCBM10N | | | |
| DSCBM14N | | | |
#### Materials:
- . S-Clean Pre-Milled Abutment and S-Clean Abutment Screw is fabricated from Ti-6A1-4V ELI (Conforming to ASTM Standard F136).
{6}------------------------------------------------
# K210362
# Summaries of Technology Characteristics
The subject device is substantially equivalent to the current cleared devices. They are substantialy equivalent in intended use, material and connection interfaces to the implants are identical diameter and connection type. Comparison demonstrating Substantial Equivalence follows:
| | Subject Device | | | | Primary Predicate Device | | | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|-----------------------|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|-------------------------------------|-------------------------------------|
| Applicant | Dentis Co., Ltd. | | | | DIO Corporation | | | |
| Trade Name | s-Clean Pre-Milled Abutment | | | | DIO CAM/CAM Abutment | | | |
| 510(k) No. | K210362 | | | | K181037 | | | |
| Classification Name | Endosseous Dental Implant, Abutment (872.3630) | | | | Endosseous Dental Implant, Abutment (872.3630) | | | |
| Product Code | NHA | | | | NHA | | | |
| Class | II | | | | II | | | |
| Material | Ti-6AL-4V ELI (ASTM F136) | | | | Ti-6AL-4V ELI (ASTM F136) | | | |
| Diameter (mm) | CAD/CAM Patient-Specific Abutment:<br>5.8/6.8 | | | | CAD/CAM Patient-Specific Abutment:<br>3.0/3.3/3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0 | | | |
| Sterile | Steam Sterilization by user<br>(Provided Non-Sterile) | | | | Steam Sterilization by user<br>(Provided Non-Sterile) | | | |
| Type of Retention | Screw-retained or cement retained | | | | Screw-retained or cement retained | | | |
| Abutment Seat | Sits on Taper | | | | Sits on Taper | | | |
| Anatomical Site | Oral Cavity | | | | Oral Cavity | | | |
| Constructions | Machined | | | | Machined | | | |
| | s-Clean Pre-Milled Abutment is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. | | | | DIO CAD/CAM Abutment is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. | | | |
| | | Implant System Compatibility | Implant Diameter (mm) | Platform Diameter (mm) | | Implant System Compatibility | Implant Diameter (mm) | Platform Diameter (mm) |
| Indications For Use/<br>Intended Use | 1 | s-Clean SQ-SL Fixture | Ø5.8 | Ø4.3 | | UF(II) Narrow Implant System | 3.0/3.3 | 3.0/3.3 |
| | 2 | | Ø6.8 | Ø4.5 | | UF Sub merged Implant System | 3.8/4.0/4.5/5.0/5.5<br>/6.0/6.5/7.0 | 3.8/4.0/4.5/5.0/5.5<br>/6.0/6.5/7.0 |
| | s-Clean Pre-Milled Abutment is intended for use with the s-Clean SQ-SL Fixtures in the chart. All digitally designed abutments for use with s-Clean Pre-Milled Abutment are intended to be manufactured at a Dentis validated milling center. | | | | | UF(II) Implant System | 3.8/4.0/4.5/5.0/5.5 | 3.8/4.0/4.5/5.0/5.5 |
<Patient Specific Abutment>
{7}------------------------------------------------
# K210362
| | Patient specific abutment is intended for use with the UF implant systems<br>provided in the chart. All digitally designed abutments for use with DIO<br>CAD/CAM Abutments are intended to be manufactured at a DIO<br>Corporation validated milling center. |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial<br>Equivalence<br>Comparison | The subject patient specific abutment is substantially equivalent in designs, dimensions, material, indications, abutment seat, screw seat,<br>anatomical site, connection, and technological characteristics with the identified primary predicate device. The patient specific abutment<br>is similar in fundamental scientific technology to the predicate device in that they all have been designed, manufactured and tested in<br>compliance with FDA's Class II special controls guidance document root-food endosseous dental implants and endosseous dental<br>implant abutments.<br>The Indications for Use of the subject and primary predicate device are identical other than the compatible implant bodies. This<br>difference is mitigated by fatigue testing, and identification of reference device for compatible implant bodies. Both the predicate and<br>subject devices are intended to be milled into patient specific abutments using CAD/CAM technology under the manufacturing control<br>of the sponsor. Any differences in technology characteristics are accompanied by information that demonstrated the device is<br>substantially equivalent. |
# <s-Clean Abutment Screw>
| | Subject Device | Reference Device |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| 510(k) No. | K210362 | K171694 |
| Trade Name | s-Clean Pre-Milled Abutment | s-Clean TiN Coating Abutments |
| Manufacturer | Dentis Co., Ltd | Dentis Co., Ltd |
| Product Name | s-Clean Abutment Screw | s-Clean Abutment Screw |
| Diameter | $ø2.32$ | $ø2.32$ |
| Length | 9.4 | 9.8 |
| Material | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Sterility | End-User Sterile | End-User Sterile |
| Brief Comparison | The Subject Device and Reference Device(K171694) has same manufacturer, Diameter, Indication for use, material and Sterility. Only<br>length is different but this difference is not important factor for performance. Both devices are substantially equivalent. | |
{8}------------------------------------------------
# Non-Clinical Testing
The subject device was tested to evaluate its substantial equivalence according to the following standards.
- Fatigue Tests on subject device under the worst case scenario according to ISO 14801:2016 .
Below tests were performed for predicate devices and leveraged for the subject device:
- Biocompatibility testing according to ISO 10993-1:2009 abutments referenced in K200099 ●
- . End User Sterilization Validation Test Report according to ANSI/AAMI ST79, ISO 17665-1, ISO 17665-2, ISO 11737-1, ISO 11737-2, and ISO 11138-1 referenced in K111364
The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device.
Non-clinical test data was conducted in accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments", and it consisted of testing finished assembled implant/abutment systems of the worst-case scenario, through fatigue testing. The result say that device performance was substantial equivalent.
Clinical testing was not necessary to establish substantial equivalency of the device.
# Conclusion
The s-Clean Pre-Milled Abutment constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate devices. Therefore, the s-Clean Pre-Milled Abutment and its predicates are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.