K210352 · Great Group Medical Co, Ltd. · CAH · Mar 27, 2022 · Anesthesiology
Device Facts
Record ID
K210352
Device Name
GGM Breathing Circuit Bacterial Filter
Applicant
Great Group Medical Co, Ltd.
Product Code
CAH · Anesthesiology
Decision Date
Mar 27, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5260
Device Class
Class 2
Indications for Use
The Bacterial Filter is intended to reduce the transmission of bacteria and viruses to a patient during anesthesia. For use with ventilators, anesthesia machines and open flow systems where filtration of inspired gases is desired. The Bacterial Filter is single use device for use on a single patient for up to 24hrs. The Bacterial Filter is designed to be used in hospital environments by trained personnel. HME (Heat and Moisture Exchange) Filter is intended to reduce the transmission of bacteria and viruses to/from a patient, and to maintain moisture levels in the patient's respiratory tract during anesthesia, artificial respiration and other types of assisted ventilation. HME Filter is normally positioned at the patient end of the breathing system between the circuits Y-piece and the catheter mount or patient airway device. HME Filter is a single use device for use on a single patient for up to 24hrs. HME Filter is designed to be used in hospital environments by trained personnel.
Device Story
GGM Breathing Circuit Bacterial Filter comprises two variants: Bacterial Filter (filtering only) and HME Filter (filtering plus heat/moisture exchange). Input: respiratory gases from ventilators/anesthesia machines. Operation: Bacterial Filter uses electrostatic cotton to trap pathogens; HME Filter uses electrostatic cotton for filtration and paper roll for passive humidification/heat preservation. Output: filtered, conditioned gas delivered to patient airway. Used in hospital settings by trained clinicians. Device placed between breathing circuit Y-piece and patient airway interface (ET tube/LMA). Benefits: reduces pathogen transmission to/from patient; maintains airway moisture during ventilation. Single-use, disposable (max 24 hours).
Clinical Evidence
No clinical data. Bench testing only. Performance validated via ISO 9360-1 (moisture loss, pressure drop, leakage, compliance), ASTM F2101 (BFE/VFE), and ISO 18562 (particulate/VOC emissions). Biocompatibility confirmed via ISO 10993 series (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation).
Technological Characteristics
Materials: Polypropylene/K-Resin housing, electrostatic polypropylene cotton, paper HME roll. Connectivity: Standard 22/15mm connections (ISO 5356-1) and luer ports (ISO 594-2/ISO 80369-7). Form factor: In-line breathing circuit filter. Sterilization: Ethylene Oxide (EO). Single-use, disposable.
Indications for Use
Indicated for adult patients requiring anesthesia or assisted ventilation. Bacterial filters (VF-2160, VF-2160-1) for patients with tidal volumes 250ml-1500ml. HME filters (VH-3110) for patients with tidal volumes 250ml-1500ml; (VH-3210) for patients with tidal volumes 250ml-1000ml. Designed for hospital use by trained personnel.
Regulatory Classification
Identification
A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.
Predicate Devices
HepaShield Bacterial Viral Breathing System Filter (K191909)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 27, 2022
Great Group Medical Co., Ltd. Kiki Hung Regulatory Affairs No. 168, Xingong 2nd Rd., Tianzhong Township Changhua County, 520 Taiwan
Re: K210352
Trade/Device Name: GGM Breathing Circuit Bacterial Filter Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: Class II Product Code: CAH Dated: March 18, 2022 Received: March 23, 2022
Dear Kiki Hung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-
combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K210352
#### Device Name
GGM Breathing Circuit Bacterial Filter
## Indications for Use (Describe)
GGM Breathing Circuit Bacterial Filter contains two types of bacterial Filters, namely "Bacterial Filter" with only filtering function and "HME Filter" with heat and moisture exchanger function.
Intended patient population: VF-2160, VF-2160-1 and VH-3110 are intended for adult patients with the tidal volume between 250ml-1500ml; VH-3210 is intended for adult patients with the tidal volume between 250ml-1000ml.
## Bacterial Filter (VF-2160, VF-2160-1) -
The Bacterial Filter is intended to reduce the transmission of bacteria and viruses to a patient during anesthesia. For use with ventilators, anesthesia machines and open flow systems where filtration of inspired gases is desired. The Bacterial Filter is single use device for use on a single patient for up to 24hrs. The Bacterial Filter is designed to be used in hospital environments by trained personnel.
## HME Filter (VH-3110, VH-3210) -
HME (Heat and Moisture Exchange) Filter is intended to reduce the transmission of bacteria and viruses to/from a patient, and to maintain moisture levels in the patient's respiratory tract during anesthesia, artificial respiration and other types of assisted ventilation.
HME Filter is normally positioned at the patient end of the breathing system between the circuits Y-piece and the catheter mount or patient airway device. HME Filter is a single use device for use on a single patient for up to 24hrs. HME Filter is designed to be used in hospital environments by trained personnel.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
## K210352
- 5.1 Type of Submission: Traditional
- 5.2 Date of Summary: 03/18/2022
| 5.3 | Submitter: | Great Group Medical Co., Ltd. |
|-----|------------|-----------------------------------------------|
| | Address: | No. 168, Xingong 2nd Rd., Tianzhong Township, |
| | | Changhua County 520, Taiwan (R.O.C.) |
| | Phone: | +886-4-875-8181 |
| | Fax: | +886-4-875-6161 |
| | Contact: | Kiki Hung |
| | | (liting@greatgroup.com.tw) |
#### 5.4 Identification of the Device:
| Proprietary/Trade name: | GGM Breathing Circuit Bacterial Filter |
|------------------------------|----------------------------------------|
| Classification Product Code: | CAH |
| Regulation Number: | 868.5260 |
| Regulation Description: | Breathing circuit bacterial filter |
| Review Panel: | General Hospital |
| Device Class: | II |
#### 5.5 Identification of the Predicate Device I:
| Predicate Device Name: | HepaShield Bacterial Viral Breathing<br>System Filter | |
|------------------------------|-------------------------------------------------------|--|
| Manufacturer: | Flexicare Medical Limited | |
| Classification Product Code: | CAH | |
| Regulation number: | 868.5260 | |
| Device Class: | II | |
| 510(k) Number: | K191909 | |
{4}------------------------------------------------
ર્ડ.(
| Identification of the Predicate Device II: | |
|--------------------------------------------|---------------------------|
| Predicate Device Name: | ThermoShield HME Filter |
| Manufacturer: | Flexicare Medical Limited |
| Classification Product Code: | CAH |
| Regulation number: | 868.5260 |
| Device Class: | II |
| 510(k) Number: | K163300 |
# 5.7 Indications for Use of the Device
GGM Breathing Circuit Bacterial Filter contains two types of bacterial filters, namely "Bacterial Filter" with only filtering function and "HME Filter" with heat and moisture exchanger function.
Intended patient population: VF-2160, VF-2160-1 and VH-3110 are intended for adult patients with the tidal volume between 250ml-1500ml; VH-3210 is intended for adult patients with the tidal volume between 250ml-1000ml.
Bacterial Filter (VF-2160, VF-2160-1) -
The Bacterial Filter is intended to reduce the transmission of bacteria and viruses to a patient during anesthesia. For use with ventilators, anesthesia machines and open flow systems where filtration of inspired gases is desired.
The Bacterial Filter is single use device for use on a single patient for up to 24hrs. The Bacterial Filter is designed to be used in hospital environments by trained personnel.
HME Filter (VH-3110, VH-3210) -
HME (Heat and Moisture Exchange) Filter is intended to reduce the transmission of bacteria and viruses to/from a patient, and to maintain moisture levels in the patient's respiratory tract during anesthesia, artificial respiration and other types of assisted ventilation.
HME Filter is normally positioned at the patient end of the breathing system between
{5}------------------------------------------------
the circuits Y-piece and the catheter mount or patient airway device. HME Filter is a single use device for use on a single patient for up to 24hrs.
HME Filter is designed to be used in hospital environments by trained personnel.
# 5.8 Device Description
GGM Breathing Circuit Bacterial Filter contains two types of bacterial filters, namely "Bacterial Filter" with only filtering function and "HME Filter" with heat and moisture exchanger function.
Bacterial Filter (models VF-2160, VF-2160-1):
Bacterial Filter is a single-used medical device, which is used for filtering the bacteria and virus in the respiratory gas to the patient during the respiratory therapy. VF-2160 and VF-2160-1 are all intended for adult patients with the tidal volume between 250ml-1500ml.
HME Filter (models VH-3110, VH-3210):
HME Filter is a kind of breathing circuit bacterial filter with the function of passive heat and moisture exchanger (artificial nose). It is a breathing circuit bacterial filter that contains electrostatic cotton for filtration and contains heat preserved and moisture absorbed paper roll for airway heat preservation and humidification. The Heat and Moisture Exchanger Filter (HMEF) should be installed on the patient's end based on the principle of passive humidification treatment. VH-3110 is intended for adult patients with the tidal volume between 250ml-1500ml, and VH-3210 is intended for adult patients with the tidal volume between 250ml-1000ml.
{6}------------------------------------------------
#### 5.9 Technological Characteristic Comparison Table
The GGM Breathing Circuit Bacterial Filter, including Bacterial Filters and HME Filters, submitted in this 510(k) file is substantially equivalent in intended use, has similar technology/principles of operation, and similar performance to the cleared HepaShield Bacterial Viral Breathing System Filter (K191909) and ThermoShield HME Filter (K163300).
| Item<br>Feature | Subject device | Predicate device | Comparison | | |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|------------|------------|
| Submitter | Great Group Medical Co., Ltd. | Flexicare Medical Limited | | | |
| Proprietary Name | GGM Breathing Circuit Bacterial<br>Filter - Bacterial Filters | HepaShield Bacterial Viral Breathing<br>System Filter | | | |
| 510(k) No. | K210352 | K191909 | | | |
| Models | VF-2160<br>VF-2160-1 | N/A | N/A | | |
| Intended Use | The Bacterial Filter is intended to<br>reduce the transmission of bacteria<br>and viruses to a patient during<br>anesthesia. For use with ventilators,<br>anesthesia machines and open flow<br>systems where filtration of inspired<br>gases is desired.<br><br>The Bacterial Filter is single use<br>device for use on a single patient for<br>up to 24hrs. The Bacterial Filter is<br>designed to be used in hospital<br>environments by trained personnel. | Flexicare's HepaShield Bacterial<br>Viral Breathing System Filters are<br>intended to reduce the transmission<br>of bacteria and viruses to/from a<br>patient during anesthesia. For use<br>with ventilators, anesthesia machines<br>and open flow systems where<br>filtration of inspired and/or expired<br>gases is desired.<br><br>Flexicare's HepaShield Bacterial<br>Viral Breathing System Filters are<br>single use devices for use on a single<br>patient for up to 24hrs and are<br>available in Adult size. Flexicare's<br>HepaShield Bacterial Viral Breathing | Same | | |
| Item<br>Feature | Subject device | Predicate device | | | |
| Submitter | Great Group Medical Co., Ltd. | Flexicare Medical Limited | Comparison | | |
| Proprietary Name | GGM Breathing Circuit Bacterial<br>Filter - Bacterial Filters | HepaShield Bacterial Viral Breathing<br>System Filter | | | |
| 510(k) No. | K210352 | K191909 | | | |
| Models | VF-2160<br>VF-2160-1 | N/A | N/A | | |
| Intended patient<br>population | For Adult patients with the tidal<br>volume between 250ml-1500ml | System Filters are designed to be<br>used in hospital environments by<br>trained personnel. | | | |
| Type of use | Prescription Use | Adult | Same | | |
| | | Prescription Use | Same | | |
| Components | Filter housing top<br>Filter housing bottom<br>Filter cotton (electrostatic cotton)<br>Luer port cap | Filter housing top<br>Filter housing bottom<br>Filter Media pack<br>Tethered luer port cap<br>Outer shrink sleeve | Similar | | |
| Assembly Method | Ultrasonic welded housing | Ultrasonic welded housing | Same | | |
| Emergency Use | Yes | Yes | Same | | |
| Environment of<br>Use | Hospital | Hospital | Same | | |
| Patient use/<br>Duration of use | Single use, disposable, <24hrs | Single use, disposable, <24hrs | Same | | |
| Supplied sterile | Sterile | Both Non-sterile & sterile variants | Similar | | |
| Patient Connection | ET tube, Laryngeal Mask Airway,<br>Catheter mount, Breathing circuit | ET tube, Laryngeal Mask Airway,<br>Catheter mount, Breathing circuit | Same | | |
| Colors | Clear/ colorless, Green | Clear/ colorless, Blue | Different | | |
| Item<br>Feature | Subject device | | Predicate device | Comparison | |
| Submitter | Great Group Medical Co., Ltd. | | Flexicare Medical Limited | | |
| Proprietary Name | GGM Breathing Circuit Bacterial<br>Filter - Bacterial Filters | | HepaShield Bacterial Viral Breathing<br>System Filter | | |
| 510(k) No. | K210352 | | K191909 | | |
| Models | VF-2160 | VF-2160-1 | N/A | N/A | |
| Standard 22/15mm<br>connections in<br>compliance with<br>ISO 5356-1 | Yes | | Yes | Same | |
| Luer port for gas<br>sampling in<br>compliance with<br>ISO 594-2/ISO<br>80369-7 | Yes | | Yes | Same | |
| Configurations | Straight without<br>luer port | Straight with luer<br>port | Straight with luer port | Similar | |
| Filtration | Electrostatic | | Mechanical | Different | |
| Placement with<br>Circuit | Machine side | | Patient side<br>Machine side | Similar | |
| Item<br>Feature | Subject device | | Predicate device | | |
| Submitter | Great Group Medical Co., Ltd. | | Flexicare Medical Limited | Comparison | |
| Proprietary Name | GGM Breathing Circuit Bacterial<br>Filter - Bacterial Filters | | HepaShield Bacterial Viral Breathing<br>System Filter | | |
| 510(k) No. | K210352 | | K191909 | | |
| Models | VF-2160 | VF-2160-1 | N/A | N/A | |
| Weight | 21g | 23g | 42g | Different | |
| Internal Volume/<br>Dead space as per<br>ISO 9360-1 | 30ml | | 47ml | Similar | |
| Bacterial/ Viral<br>Filtration<br>Efficiency/ Filter<br>Integrity (Fresh) | BFE - 99.9%<br>VFE - 99.9% | | BFE- 99.99999%<br>VFE- 99.9999% | Similar | |
| Salt Method<br>Filtration<br>Efficiency | 99.61% |…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.