← Product Code [DZE](/productcode/DZE) · K210336

# Neodent Implant System - Zirconia Implant System (K210336)

_Jjgc Industria E Comercio DE Materiais Dentarios S.A. · DZE · Aug 13, 2021 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210336

## Device Facts

- **Applicant:** Jjgc Industria E Comercio DE Materiais Dentarios S.A.
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** Aug 13, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusal loading. Multiple teeth applications can be rigidly splinted. The CR Abutment is an abutment placed over Neodent Zirconia Implants in order to provide support for prosthetic restorations, such as copings or crowns. It may be used for single-unit restorations that are cement-retained in esthetical areas over implants installed in maxilla or mandible.

## Device Story

Neodent Implant System (Zirconia) consists of endosseous dental implants and CR abutments; used by dentists/oral surgeons in clinical settings. Implants are Y-TZP (Yttrium-stabilized zirconium dioxide) with sand-blasted, acid-etched surface; apically tapered with trapezoidal threads. CR abutments connect to implants via ZiLock internal interface; secured with titanium alloy screws (ASTM F136). System supports prosthetic restorations (crowns/copings) to restore chewing function. Clinicians select appropriate implant/abutment size based on patient anatomy; surgical placement into jawbone; immediate or delayed loading based on primary stability. Benefits include restoration of dental function and esthetics. Device is single-use, sterile (Ethylene Oxide).

## Clinical Evidence

Bench testing only. No clinical data. Testing included ISO 14801 dynamic fatigue, static torsional loading, insertion torque in sawbones (Types I-IV), and screw torque testing. Biocompatibility assessed per ISO 10993-1 and ISO 10993-18. Sterilization validated per ISO 11135-1 (SAL 10^-6). Bacterial endotoxin testing per ANSI/AAMI ST72:2011 (<0.05 EU/device). Shelf life validated per ASTM F1980 (5 years).

## Technological Characteristics

Material: Y-TZP (Yttrium-stabilized zirconium dioxide); Abutment screw: Titanium alloy (ASTM F136). Design: Apically tapered, trapezoidal threads, ZiLock internal indexer. Surface: Sand-blasted, acid-etched. Sterilization: Ethylene Oxide (SAL 10^-6). Dimensions: Lengths 10-13mm, Diameter 3.75mm. Connectivity: Mechanical interface (non-networked).

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- Neodent Implant System - Zirconia Implant System ([K201491](/device/K201491.md))

## Reference Devices

- PURE Ceramic Implants - Institut Straumann AG ([K151328](/device/K151328.md))
- Neodent Implant System – CM Line ([K101945](/device/K101945.md))
- MRI Compatibility For Existing Neodent Implant System ([K182620](/device/K182620.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

JJGC Industria e Comercio de Materiais Dentarios SA % Jennifer Jackson Director of Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810

Re: K210336

Trade/Device Name: Neodent Implant System - Zirconia Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: July 14, 2021 Received: July 16, 2021

Dear Jennifer Jackson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

### 510(k) Number (if known) K210336

#### Device Name

Neodent Implant System - Zirconia Implant System

Indications for Use (Describe) Indications for Use for Zirconia Implants:

The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusal loading. Multiple teeth applications can be rigidly splinted.

Indications for Use for CR Abutment for Zirconia System:

The CR Abutment is an abutment placed over Neodent Zirconia Implants in order to provide support for prosthetic restorations, such as copings or crowns. It may be used for single-unit restorations that are cement-retained in esthetical areas over implants installed in maxilla or mandible.

| Type of Use (Select one or both, as applicable)                                                                                              |                                                                                                                              |
|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)  </span> | <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)  </span> |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K210336 510(k) Summary

### ADMINISTRATIVE INFORMATION

| Sponsor                                    | JJGC Indústria e Comércio de Materiais Dentários SA<br>(dba Neodent)<br>Av. Juscelino Kubitschek de Oliveira, 3291<br>Curitiba, Parana, Brazil 81270-200<br>Registration No.: 3008261720<br>Owner/Operator No.: 10031702                                                                         |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person                             | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs,<br>Straumann USA<br>E-mail: jennifer.jackson@straumann.com<br>Telephone (978) 747-2509                                                                                                                                                |
| Date Prepared                              | 13/Aug/2021                                                                                                                                                                                                                                                                                      |
| Preparer / Alternate Contact               | Mariana Soares Hartmann<br>Regulatory Affairs Analyst<br>JJGC Indústria e Comércio de Materiais Dentários SA<br>E-mail: mariana.hartmann@neodent.com                                                                                                                                             |
| DEVICE NAME AND CLASSIFICATION             |                                                                                                                                                                                                                                                                                                  |
| Trade/ Proprietary Name<br>Common Name     | Neodent Implant System – Zirconia Implant System<br>Endosseous dental implant<br>Endosseous dental implant abutment                                                                                                                                                                              |
| Classification Name                        | Implant, Endosseous, Root-Form<br>Endosseous dental implant abutment                                                                                                                                                                                                                             |
| Classification Regulations<br>Product Code | 21 CFR 872.3640 / 21 CFR 872.3630, Class II<br>DZE/NHA                                                                                                                                                                                                                                           |
| Classification Panel<br>Reviewing Branch   | Dental Products Panel<br>Dental Devices Branch                                                                                                                                                                                                                                                   |
| PREDICATE DEVICE INFORMATION               |                                                                                                                                                                                                                                                                                                  |
| Primary Predicate Device                   | K201491 – Neodent Implant System – Zirconia Implant<br>System, JJGC Indústria e Comércio de Materiais<br>Dentários S.A                                                                                                                                                                           |
| Reference Predicate Devices                | K151328 - PURE Ceramic Implants - Institut Straumann<br>AG<br>K101945 - Neodent Implant System – CM Line, JJGC<br>Indústria e Comércio de Materiais Dentários S.A<br>K182620 - MRI Compatibility For Existing Neodent<br>Implant System, JJGC Indústria e Comércio de Materiais<br>Dentários S.A |

{4}------------------------------------------------

### INDICATIONS FOR USE

### Indications for Use for Zirconia Implants:

The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage surgical procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with physiological occlusal loading. Multiple teeth applications can be rigidly splinted.

#### Indications for Use for CR Abutment for Zirconia System:

The CR Abutment is an abutment placed over Neodent Zirconia Implants in order to provide support for prosthetic restorations, such as copings or crowns. It may be used for single-unit restorations that are cement-retained in esthetical areas over implants installed in maxilla or mandible.

### SUBJECT DEVICE DESCRIPTIONS

- Intended for single use;
- Provided sterile via Ethylene Oxide;
- . All devices of Zirconia Implant System are manufactured of Y-TZP (Yttrium-stabilized zirconium dioxide);
- . The CR Abutment for Zirconia system is supplied along with a titanium alloy screw (ASTM F136);
- ZiLock prosthetic interface with internal indexer;
- Apically tapered implant with trapezoidal thread profile;
- Implant provided with sand-blasted, acid etched surface finish;
- . The subject Implants are compatible with the Zirconia Bases (already cleared per market per K201491) and CR Abutments (subject of this submission);
- . The subject CR Abutments are compatible with the Zirconia Implants 4.3 mm (already cleared per market per K201491) and Zirconia Implants 3.75 mm (subject of this submission).

{5}------------------------------------------------

Neodent Implant System – Zirconia Implant System

### TECHNOLOGICAL CHARACTERISTIC COMPARISON TABLES

|                                   | SUBJECT DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | PRIMARY PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | REFERENCE PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                      |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
|                                   | Neodent Implant System - Zirconia Implant<br>System<br>Implants<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A.                                                                                                                                                                                                                                                                                                                                                                                                          | K201491<br>Neodent Implant System - Zirconia Implant<br>System<br>Implants<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A.                                                                                                                                                                                                                                                                                                                                                                                               | K151328<br>PURE Ceramic Implants<br>Institut Straumann AG                                                                                                                                                                                                                                                                                                                                                                                   | Equivalence<br>Discussion                                                                                            |
| Indications for<br>Use            | The Neodent Implant System is intended to<br>be surgically placed in the bone of the upper<br>or lower jaw to provide support for<br>prosthetic devices, such as artificial teeth, to<br>restore chewing function. It may be used<br>with single-stage or two-stage surgical<br>procedures, for single or multiple unit<br>restorations,<br>and<br>may<br>be<br>loaded<br>immediately when good primary stability is<br>achieved and with physiological occlusal<br>loading. Multiple teeth applications can be<br>rigidly splinted. | The Neodent Implant System is intended to<br>be surgically placed in the bone of the upper<br>or lower jaw to provide support for<br>prosthetic devices, such as artificial teeth, to<br>restore chewing function. It may be used<br>with single-stage or two-stage surgical<br>procedures, for single or multiple unit<br>restorations,<br>and<br>may<br>be<br>loaded<br>immediately when good primary stability is<br>achieved and with physiological occlusal<br>loading. Multiple teeth applications can be<br>rigidly splinted. | The Straumann® PURE Ceramic Implant<br>(Monotype) is indicated for restoration in<br>single tooth gaps and in an edentulous or<br>partially edentulous jaw. The prosthetic<br>restorations used are single crowns, fixed<br>partial or full dentures, which are connected<br>to the implants through the corresponding<br>components.<br>The Ø3.3 mm reduced diameter implants<br>are recommended for central and lateral<br>incisors only. | Identical<br>The Indications for Use of the subject<br>devices are the same as for the primary<br>predicate devices. |
| Implant-<br>Abutment<br>interface | Straight internal connection<br>indexing features (Zilock)                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Straight internal connection<br>indexing features (Zilock)                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical<br>Subject devices and Primary Predicate<br>devices present the same implant-<br>abutment interface.       |
| Design                            | Apically Tapered format<br>Trapezoidal threads profile                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Apically Tapered format<br>Trapezoidal threads profile                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical<br>Subject devices and Primary Predicate<br>devices present the same design.                               |
| Reusable                          | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | No                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical<br>The subject devices and the predicate devices<br>are indicated for single use.                          |

#### Table 1. Technological Characteristic Comparison Table - Zirconia Implants

{6}------------------------------------------------

# Traditional 510(k) Submission

| Neodent Implant System - Zirconia Implant System |  |  |  |  |
|--------------------------------------------------|--|--|--|--|
|--------------------------------------------------|--|--|--|--|

|                         | SUBJECT DEVICE                                                                                                              | PRIMARY PREDICATE DEVICE                                                                                                               | REFERENCE PREDICATE DEVICE                                |                                                                                                                                                                                                                                                         |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                         | Neodent Implant System - Zirconia Implant<br>System<br>Implants<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | K201491<br>Neodent Implant System - Zirconia Implant<br>System<br>Implants<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | K151328<br>PURE Ceramic Implants<br>Institut Straumann AG | Equivalence<br>Discussion                                                                                                                                                                                                                               |
| Length (mm)             | 10; 11.5; 13                                                                                                                | 8; 10; 11.5; 13                                                                                                                        | 8; 10; 12; 14 mm                                          | Equivalent<br>Range of lengths for subject devices is<br>within the range of lengths for the predicate<br>devices.                                                                                                                                      |
| Diameter (Ø)<br>(mm)    | 3.75 mm                                                                                                                     | 4.3 mm                                                                                                                                 | 3.3; 4.1 mm                                               | Equivalent<br>The diameter of the subject devices is<br>smaller than the diameter of the primary<br>predicate devices, but greater than the<br>reference predicate devices. Dynamic<br>Fatigue Testing were performed to ensure<br>the system strength. |
| Material                | Yttrium-stabilized zirconium dioxide<br>(Y-TZP).                                                                            | Yttrium-stabilized zirconium dioxide<br>(Y-TZP).                                                                                       | Yttrium-stabilized zirconium dioxide<br>(Y-TZP).          | Identical<br>Subject devices and predicate devices are<br>manufactured of the same raw material.                                                                                                                                                        |
| Sterilization<br>Method | Provided sterile via Ethylene Oxide to an<br>SAL of 10-6.                                                                   | Provided sterile via Ethylene Oxide to an SAL of<br>10-6.                                                                              | Provided sterile via Ethylene Oxide to an SAL<br>of 10-6. | Identical<br>Subject devices and predicate devices a<br>provided sterilized by the same method.                                                                                                                                                         |

{7}------------------------------------------------

# Traditional 510(k) Submission

Neodent Implant System – Zirconia Implant System

|                                   | SUBJECT DEVICE                                                                                                                                                                                                                                                                                                       | PRIMARY PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                                               | REFERENCE PREDICATE DEVICE                                                                                                                                                                                                                                                                                                                                                                                                                         | EQUIVALENCE DISCUSSION                                                                                                                                                                                                                                                                                                                                                                                                                         |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                   | Neodent Implant System - Zirconia Implant<br>System<br>Implants<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A.                                                                                                                                                                                          | K201491<br>Neodent Implant System - Zirconia Implant<br>System<br>Implants<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A.                                                                                                                                                                                                                                                                                                                                 | K101945<br>Neodent Implant System - CM Line<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A.                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Indications for<br>Use            | The CR Abutment is an abutment placed<br>over Neodent Zirconia Implants in order to<br>provide support for prosthetic restorations,<br>such as copings or crowns. It may be used<br>for single-unit restorations that are cement-<br>retained in esthetical areas over implants<br>installed in maxilla or mandible. | The Zirconia Base is an abutment placed<br>over Neodent Zirconia Implants in order to<br>provide support for custom-made<br>prosthetic restorations, such as copings or<br>crowns. It may be used for cement- or<br>screw-retained single unit restorations. All<br>digitally designed copings and/or crowns to<br>be used with the Neodent Zirconia Base<br>Abutment System are intended to be sent to<br>Straumann for manufacture at a validated<br>milling center. | The Neodent Implant System is intended to<br>be surgically placed in the bone of the upper<br>or lower jaw to provide support for<br>prosthetic devices, such as artificial teeth,<br>to restore chewing function. It may be used<br>with single-stage or two-stage procedures,<br>for single or multiple unit restorations, and<br>may be loaded immediately when good<br>primary stability is achieved and with<br>appropriate occlusal loading. | Equivalent<br>The Indications for the subject devices are<br>included within the scope of Indications for the<br>primary predicate devices. Both are indicated to<br>be placed over implants to support prosthetic<br>restorations such as copings or crowns and the<br>differences between the Indications for Use are<br>due to the fact that the predicate device is<br>digitally designed while the subject device is a<br>stock abutment. |
| Implant-<br>Abutment<br>interface | Straight internal connection<br>indexing features (Zilock)                                                                                                                                                                                                                                                           | Straight internal connection<br>indexing features (Zilock)                                                                                                                                                                                                                                                                                                                                                                                                             | Cone morse internal connection                                                                                                                                                                                                                                                                                                                                                                                                                     | Identical<br>Subject devices and Primary Predicate devices<br>present the same implant-abutment interface.                                                                                                                                                                                                                                                                                                                                     |

### Table 2. Technological Characteristic Comparison Table – Zirconia CR Abutments

{8}------------------------------------------------

### Traditional 510(k) Submission

Neodent Implant System – Zirconia Implant System

|                         | SUBJECT DEVICE<br>Neodent Implant System - Zirconia Implant<br>System<br>Implants<br>JJGC Indústria e Comércio de Materiais | PRIMARY PREDICATE DEVICE<br>K201491<br>Neodent Implant System - Zirconia Implant<br>System<br>Implants<br>JJGC Indústria e Comércio de Materiais | REFERENCE PREDICATE DEVICE<br>K101945<br>Neodent Implant System - CM Line<br>JJGC Indústria e Comércio de Materiais<br>Dentários S.A. | EQUIVALENCE DISCUSSION                                                                                                                                                                                                                               |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                         | Dentários S.A.                                                                                                              | Dentários S.A.                                                                                                                                   |                                                                                                                                       |                                                                                                                                                                                                                                                      |
| Design                  | Angulation: 0; 17°<br>Gingival Height: 1.5; 2.5 mm<br>Platform: NP; SP                                                      | Angulation: 0<br>Gingival Height: 1.5; 2.5 mm<br>Platform: NP; SP                                                                                | Angulation: 0, 17 and 30°<br>Gingival Height: 1.5; 2.5 and 3.5 mm                                                                     | Equivalent<br>The angulation and gingival height of the<br>subject devices are within the range of<br>angulation and gingival height of the<br>reference predicate device (K101945) the<br>platform is identical of the primary<br>predicate device. |
| Reusable                | No                                                                                                                          | No                                                                                                                                               | No                                                                                                                                    | Identical<br>The subject devices and the predicate devices<br>are intended for single use.                                                                                                                                                           |
| Material                | Abutment: Yttrium-stabilized zirconium<br>dioxide<br>(Y-TZP).<br>Screw: Titanium alloy ASTM F136.                           | Abutment: Yttrium-stabilized zirconium<br>dioxide<br>(Y-TZP).<br>Screw: Titanium alloy ASTM F136.                                                | Abutment and Screw: Titanium alloy ASTM<br>F136.                                                                                      | Identical<br>The subject devices have the same material of<br>construction as the primary predicate devices.                                                                                                                                         |
| Sterilization<br>Method | Provided sterile via Ethylene Oxide to an SAL of<br>$10^{-6}$                                                               | Provided sterile via Ethylene Oxide to an SAL of<br>$10^{-6}$                                                                                    | Provided sterile via Ethylene Oxide to an SAL of<br>$10^{-6}$                                                                         | Identical<br>The subject devices have the same sterilization<br>method and same SAL as the predicate<br>devices.                                                                                                                                     |

{9}------------------------------------------------

The subject implants have the same indications for use and an equivalent range of lengths as the primary predicate devices, being contemplated within the range of lengths of the predicate devices.

Subject and predicate implants have the same implant-to-abutment interface. Both present sand-blast and acid-etch surface finishes and same sterile barrier system. The subject implants are manufactured of the same materials and are sterilized using the same sterilization method as the primary predicate devices.

The subject CR Abutments have equivalent indications for use of the primary predicate devices and an equivalent range of gingival height and angulation as the reference predicate devices, being contemplated within the range of gingival height and diameter of the reference predicate devices. They also present the same sterilization method and are made of the same material as the primary predicate devices.

Overall, the subject devices are equivalent to the predicate devices as follows:

- same intended use,
- . same operating principle,
- incorporate the same basic design,
- incorporate the same materials, and
- have same packaging and are sterilized using the same materials and processes

# PERFORMANCE DATA

# Biocompatibility

Representative samples of each of the subject devices was subjected to the following:

- Biological Safety Assessment guided by ISO 10993-1,
- Chemical characterization was performed per ISO 10993-18.

An assessment was made to define if new biocompatibility tests were needed or whether prior testing applies to the subject devices. The subject devices do not represent a new worst case and we are therefore relying on the results of the tests provided in K201491.

# Mechanical testing

Dynamic fatigue test per ISO 14801 was performed to determine the fatigue strength for the dental implant system, according to FDA Guidance. The tested subject devices exhibit a level of performance equivalent to that reviewed for the predicate devices.

Torsion Test was performed to evaluate the Zirconia Implant System under static torsional loading.

Insertion test was performed to evaluate the insertion torque of the Zirconia Implant System when inserted into sawbones material representing bone type I, II, III and IV.

SEM images made after fatigue loading test were included to demonstrate that the subject devices present similar wear in the implant inner thread compared to the Zirconia Implant (K201491).

Screw torque testing was performed with torque application and visual analysis before and after use to assess whether the torque would damage the items. No visible failure was identified after applying torque 

{10}------------------------------------------------

and removing items from the implant. The subject screws are considered equivalent to those already cleared per K201491.

# Sterilization validation

The subject devices are sterilized by Ethylene Oxide, according to ISO 11135-1 via the over-kill method. The method achieved a Sterility Assurance Level of 1x10-6.

An assessment was made to determine if the subject devices represent new worst-cases for sterilization validation and since they do not represent a new worst-case, they are being adopted into the already the same sterilization validation method already cleared under K201491.

The Subject devices are not represented to be "pyrogen free".

Bacterial Endotoxin Testing are performed on representative samples selected and tested based upon the raw material, manufacturing processes and sterilization process, according to ANSI/AAMI ST72:2011, Bacterial Endotoxins – Test Methods, Routine Monitoring and Alternatives to Batch Testing. The obtained results were <0,05 EU/device.

# MR Conditional Labeling

The MR Conditional Labeling was leveraged from K182620.

# Shelf Life

Product and package stability have been validated per ASTM F1980. The subject device's shelf life was determined to be 5 years.

# CONCLUSION

The subject devices and the predicate devices have equivalent intended use, design, and technological characteristics. Equivalent range of overall dimensions and sterilization method. The data included in this submission demonstrate that the subject devices are substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K210336](https://fda.innolitics.com/device/K210336)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
