K210282 · Cardiovascular Systems, Inc. · NTE · Mar 3, 2021 · Cardiovascular
Device Facts
Record ID
K210282
Device Name
WIRION Embolic Protection System
Applicant
Cardiovascular Systems, Inc.
Product Code
NTE · Cardiovascular
Decision Date
Mar 3, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The WIRION® is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/debris) while performing atherectomy in calcified lesions of the lower extremities. The diameter of the vessel at the site of filter basket placement should be between 3.5mm to 6.0mm. WIRION® may be used with commercially available 0.014" guide wires.
Device Story
WIRION is a temporary embolic protection system (EPS) used during lower extremity atherectomy; filters distal to intervention site to contain/remove thrombus and debris. Rapid exchange, pre-crimped filter design; compatible with standard 0.014" guide wires. Operated by physicians in clinical settings; provides embolic capture to reduce distal embolization risks. Device includes filter delivery and retrieval components; fluoroscopic markers assist placement. Output is physical capture of embolic material; enables safer atherectomy procedures.
Clinical Evidence
Bench testing only. Biocompatibility testing, simulated use, bond and torque strength testing, and package testing were performed to verify that modified materials and processes met pre-determined specifications.
Technological Characteristics
Rapid exchange embolic protection system; pre-crimped filter; 0.014" guidewire compatible. Features fluoroscopic markers for visualization. Single-use device. Sterilization method and SAL identical to predicate.
Indications for Use
Indicated for patients undergoing atherectomy in calcified lower extremity lesions. Suitable for vessel diameters 3.5mm to 6.0mm. Compatible with 0.014" guide wires.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K063313 — FILTERWIRE EZ EMBOLIC PROTECTION SYSTEM (3.5 MM TO 5.5 MM) · Boston Scientific Corp · Dec 5, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 3, 2021
Cardiovascular Systems Inc. Erika Huffman Regulatory Affairs Manager 1225 Old Hwy 8 NW St. Paul, Minnesota 55112
Re: K210282
Trade/Device Name: WIRION Embolic Protection System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NTE Dated: January 29, 2021 Received: February 1, 2021
Dear Erika Huffman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or
{1}------------------------------------------------
postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Finn Donaldson Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210282
Device Name WIRION® Embolic Protection System
#### Indications for Use (Describe)
The WIRION® is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/ debris) while performing atherectomy in calcified lesions of the lower extremities. The diameter of the vessel at the site of filter basket placement should be between 3.5mm to 6.0mm. WIRION® may be used with commercially available 0.014" guide wires.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| Submitter: | Cardiovascular Systems, Inc.<br>1225 Old Highway 8 NW<br>Saint Paul, MN 55112 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Erika Huffman, MS, RAC<br>Regulatory Affairs Manager<br>Cardiovascular Systems, Inc.<br>1225 Old Highway 8 NW<br>St. Paul, MN 55112<br>Ph: 651-259-1608<br>ehuffman@csi360.com |
| Date Prepared: | January 29, 2021 |
| Trade Name: | WIRION® Embolic Protection System |
| Common Name: | Embolic Protection System |
| Regulation Number: | 870.1250, Percutaneous Catheter |
| Classification: | Class II |
| Product Code: | NTE |
| Predicate Device(s): | K200198 – WIRION™ Embolic Protection System |
| Device Description: | WIRION® is a temporary Embolic Protection System (EPS),<br>filtering distal to the intervention site. The system is a rapid<br>exchange, pre-crimped filter that can be used with any<br>commercially available 0.014" guide wires. |
| Intended Use: | The WIRION® is indicated for use as an embolic protection system<br>to contain and remove embolic material (thrombus/debris) while<br>performing atherectomy in calcified lesions of the lower<br>extremities. The diameter of the vessel at the site of filter basket<br>placement should be between 3.5mm to 6.0mm. WIRION® may be<br>used with commercially available 0.014" guide wires. |
| Comparison to<br>Predicate Device: | The modified WIRION device is identical to the predicate device as<br>follows:<br>Same regulation number, product code and classification Same intended use and indications for use Same vessel diameter range and anatomic location of use Same filter design and materials Same principles of operation |
| Same guidewire rapid exchange (RX) design Same 0.014" guidewire compatibility Same filter delivery and retrieval methods Same filter dimensions and pore size Same filter function and placement within the vessel Same filter fluoroscopy markers for visualization Same sterilization method and SAL Same number of uses per device (single use) The modified WIRION device is different from the predicate device as follows: Longer Retrieval Catheter RX guidewire lumen (RX port moved proximal) Longer shipping mandrel Retrieval Catheter tip material and color Larger sterile barrier pouch (Retrieval Catheter) Different pouch seal locations Branding New RX guidewire lumen tie-layer material Smaller nominal diameter of activating wire New syringe New tamper-evident seal on shelf box Clarified flushing step in Instructions for Use | |
| Functional and Safety Testing: | Biocompatibility testing was performed to verify the changed materials and processes are acceptable. In addition, simulated use, bond and torque strength testing, and package testing were also performed to confirm pre-determined device specifications were met. |
| Conclusion: | No new questions of safety or effectiveness were identified compared to the predicate device and the device is expected to perform as intended under the specified use conditions. |
{4}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.