← Product Code [LYY](/productcode/LYY) · K210253

# Best Glove-Latex Powder Free Examination Glove (K210253)

_Bestsafe Glove Co., Ltd. · LYY · Nov 5, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210253

## Device Facts

- **Applicant:** Bestsafe Glove Co., Ltd.
- **Product Code:** [LYY](/productcode/LYY.md)
- **Decision Date:** Nov 5, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner

## Device Story

Disposable natural rubber latex examination glove; worn on examiner's hand or finger; prevents contamination between patient and examiner. Used in clinical settings by healthcare personnel. Provides physical barrier protection. Non-sterile; powder-free. Benefits patient and examiner by reducing risk of cross-contamination during medical examinations.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM D3578-19 (dimensions, physical properties), ASTM D5151-19 (watertightness), ASTM D6124-06 (residual powder), and ASTM D5712-15 (protein content). Biocompatibility testing performed per ISO 10993-5, ISO 10993-10, and ISO 10993-11.

## Technological Characteristics

Material: Natural Rubber Latex. Standards: ASTM D3578-19, ASTM D5151-19, ASTM D6124-06, ASTM D5712-15. Form factor: Disposable, non-sterile, powder-free glove. Sizes: Small, Medium, Large, Extra Large. Thickness: 0.08 mm min (palm/finger). Tensile strength: 18 MPa min (before aging).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- MPXX™ Powder Free Natural Rubber Latex Examination Gloves ([K110250](/device/K110250.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 5, 2021

Bestsafe Glove CO., LTD. Piyawat Chirasakulkarun Chief Executive Officer 52/43 Wat Khot Hin-Khao Phai Road, T. Tubma Muang Rayong, Rayong 21000 Thailand

Re: K210253

Trade/Device Name: Best Glove-Latex Powder Free Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LYY Dated: May 25, 2021 Received: August 30, 2021

Dear Piyawat Chirasakulkarun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K210253

### Device Name

BEST GLOVE -LATEX POWDER FREE EXAMINATION GLOVE

### Indications for Use (Describe)

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

× Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRA Staff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for BESTSAFE GLOVE CO., LTD. The logo features a blue hand on the left side, followed by the word "BESTSAFE" in blue, with a plus sign in the middle of the word. Below the word "BESTSAFE" is the company name "BESTSAFE GLOVE CO., LTD" in smaller blue letters. There is a registered trademark symbol in the upper right corner of the logo.

52/43 WATKODHIN-KAOPHAI RD. T.TUBMA, A.MUANG, RAYONG, RAYONG. THAILAND 21000 TAX ID:0215563006105 TEL OFFICE : 038-949877-79 FAX : (+66)038-694493, 038-949899 TEL H/P : +66-0-86-3173770 , +66-0-81-4022099 ,+66-0-62-325-6558 EMAIL : SALES@BESTSAFE.CO.TH WEBSITE : WWW.BESTSAFE.CO.TH I WWW.BESTSAFEGLOVE.COM

# 510(k) SUMMARY K210253 Latex Powder Free Examination Glove

#### 1.0 Applicant Information

|     | Applicant:                 | BESTSAFE GLOVE CO., LTD                                                         |
|-----|----------------------------|---------------------------------------------------------------------------------|
|     | Address                    | 52/43 Watkodhin-Kaophai Rd. T. Tubma,<br>A. Muang Rayong Rayong. THAILAND 21000 |
|     | Phone Number:              | +66 (038)-949880                                                                |
|     | Fax Number:                | +66 (038)-949850                                                                |
|     | Name of Contact Person:    | Piyawat Chirasakulkarun                                                         |
|     | Contact Number:            | +66 81 4022099                                                                  |
|     | Contact Email:             | sales@bestsafe.co.th                                                            |
|     | Preparation date:          | September 6, 2021                                                               |
| 2.0 | Device Identification:     |                                                                                 |
|     | Trade/Proprietary Name(s): | BEST GLOVE -LATEX POWDER FREE EXAMINATION GLOVE                                 |
|     | Common Name:               | Latex Powder Free Examination Gloves                                            |
|     | Classification Name:       | Patient Examination Gloves                                                      |
|     | 510(k):                    | K210253                                                                         |
|     | Device Class:              | I                                                                               |
|     | Product Code:              | LYY                                                                             |
|     | Registration Number        | 21 CFR 880.6250                                                                 |
|     | Review Panel               | General Hospital                                                                |
| 3.0 | Predicate Device           |                                                                                 |
|     | Device Name:               | MPXX™ Powder Free Natural Rubber Latex Examination Gloves                       |
|     | Manufacturer:              | Total Glove Company Sdn. Bhd.                                                   |
|     | 510(k):                    | K110250                                                                         |
|     | Device Class:              | I                                                                               |
|     | Product Code:              | LYY                                                                             |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for BESTSAFE GLOVE CO., LTD. The logo features a blue hand on the left side, followed by the word "BESTSAFE" in blue with a plus sign in the middle. Below "BESTSAFE" is the text "BESTSAFE GLOVE CO., LTD" in a smaller font size. There is a registered trademark symbol in the upper right corner of the logo.

52/43 WATKODHIN-KAOPHAI RD. T.TUBMA, AMUANG, RAYONG, RAYONG. THAILAND 21000 TEL OFFICE : 038-949877-79 FAX : (+66)038-694493, 038-949899 TAX ID:0215563006105 TEL H/P : +66-0-86-3173770 , +66-0-81-4022099 ,+66-0-62-325-6558 EMAIL : SALES@BESTSAFE.CO.TH WEBSITE : WWW.BESTSAFE.CO.TH I WWW.BESTSAFEGLOVE.COM

#### 4.0 Description of the Device:

Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D3578 - 19, Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are natural in color is added) and are powder free.

#### 5.0 Indication for Use of the Device:

A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner

#### 6.0 Comparison of Technological characteristics between predicate and subject devices

| Characteristics     | References/Standards                                                         | Device performance                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                       | Comparison               |
|---------------------|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
|                     |                                                                              | Predicate                                                                                                                                                                                                                                                                     | Current                                                                                                                                                                                                               |                          |
| 510(k) Number       | -                                                                            | K110250                                                                                                                                                                                                                                                                       | K210253                                                                                                                                                                                                               | ---                      |
| Manufacturer(s)     | -                                                                            | Total Glove Company Sdn.<br>Bhd.                                                                                                                                                                                                                                              | BESTSAFE GLOVE CO.,<br>LTD                                                                                                                                                                                            | ---                      |
| Name of device      | -                                                                            | MPXX             TM Powder Free Natural<br>Rubber Latex Examination<br>Gloves                                                                                                                                                                                                 | BEST GLOVE -LATEX<br>POWDER FREE<br>EXAMINATION GLOVE                                                                                                                                                                 | ---                      |
| Indication for Use  | Medical Gloves<br>Guidance Manual,<br>Issued on January 22,<br>2008          | MPXX             TM Powder Free Natural<br>Rubber Latex Examination<br>Gloves are single use device<br>intended for medical purposes<br>that is worn on the hand of<br>healthcare and similar<br>personnel to prevent<br>contamination between<br>healthcare and the patient. | A powder-free patient<br>Examination Glove is a<br>disposable device intended<br>for medical purposes that is<br>worn on the examiner's hand<br>or finger to prevent<br>contamination between<br>patient and examiner | Same                     |
| Material            | ASTM D3578-19                                                                | Natural Rubber Latex                                                                                                                                                                                                                                                          | Natural Rubber Latex                                                                                                                                                                                                  | Identical                |
| Color               |                                                                              | Natural Colo                                                                                                                                                                                                                                                                  | Natural Colo                                                                                                                                                                                                          | Same                     |
| Size                | Medical Glove<br>Guidance Manual-<br>Labeling- Issued on<br>January 22, 2008 | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large                                                                                                                                                                                                                        | Small<br>Medium<br>Large<br>Extra Large                                                                                                                                                                               | Different                |
| Characteristics     | References/Standards                                                         | Device performance                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                       | Comparison               |
| Single Use          | Medical Gloves<br>Guidance Manual -<br>Issued on January 22,<br>2008         | Single use                                                                                                                                                                                                                                                                    | Single use                                                                                                                                                                                                            | Same                     |
| Sterile/Non-sterile | -                                                                            | Non-sterile                                                                                                                                                                                                                                                                   | Non-sterile                                                                                                                                                                                                           | Same                     |
| Dimension           | ASTM D3578-19                                                                | <b>Length</b><br>230 mm minimum                                                                                                                                                                                                                                               | <b>Length</b><br>230 mm min                                                                                                                                                                                           | Similar                  |
|                     |                                                                              | Size                                                                                                                                                                                                                                                                          | Actual value                                                                                                                                                                                                          |                          |
|                     |                                                                              | Small                                                                                                                                                                                                                                                                         | 240                                                                                                                                                                                                                   |                          |
|                     |                                                                              | Medium                                                                                                                                                                                                                                                                        | 240                                                                                                                                                                                                                   |                          |
|                     |                                                                              | Large                                                                                                                                                                                                                                                                         | 240                                                                                                                                                                                                                   |                          |
|                     |                                                                              | Extra Large                                                                                                                                                                                                                                                                   | 241                                                                                                                                                                                                                   |                          |
|                     |                                                                              | <b>Width</b><br>Medium: 95 mm ± 10 mm<br>(for Medium size)                                                                                                                                                                                                                    |                                                                                                                                                                                                                       | Different                |
|                     |                                                                              | Size                                                                                                                                                                                                                                                                          | Actual value                                                                                                                                                                                                          |                          |
|                     |                                                                              | Small                                                                                                                                                                                                                                                                         | 85                                                                                                                                                                                                                    |                          |
|                     |                                                                              | Medium                                                                                                                                                                                                                                                                        | 93                                                                                                                                                                                                                    |                          |
|                     |                                                                              | Large                                                                                                                                                                                                                                                                         | 105                                                                                                                                                                                                                   |                          |
|                     |                                                                              | Extra Large                                                                                                                                                                                                                                                                   | 115                                                                                                                                                                                                                   |                          |
| Thickness           | ASTM D3578-19                                                                | Finger = 0.08 mm. min<br>Palm = 0.08 mm. Min                                                                                                                                                                                                                                  | Finger = 0.08 mm. min<br>Palm = 0.08 mm. Min                                                                                                                                                                          | Similar                  |
|                     |                                                                              | Size                                                                                                                                                                                                                                                                          | Palm<br>(Actual value)                                                                                                                                                                                                | Finger<br>(Actual value) |
|                     |                                                                              | Small                                                                                                                                                                                                                                                                         | 0.09                                                                                                                                                                                                                  | 0.12                     |
|                     |                                                                              | Medium                                                                                                                                                                                                                                                                        | 0.09                                                                                                                                                                                                                  | 0.12                     |
|                     |                                                                              | Large                                                                                                                                                                                                                                                                         | 0.10                                                                                                                                                                                                                  | 0.13                     |
|                     |                                                                              | Extra Large                                                                                                                                                                                                                                                                   | 0.09                                                                                                                                                                                                                  | 0.12                     |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for BESTSAFE. On the left is a blue hand. To the right of the hand is the word "BESTSAFE" in blue, with a plus sign in the upper left corner of the word. Below the word "BESTSAFE" is the text "BESTSAFE GLOVE CO.,LTD" in a smaller font.

52/43 WATKODHIN-KAOPHAI RD. T.TUBMA, A.MUANG, RAYONG, RAYONG. THAILAND 21000 TEL OFFICE : 038-949877-79 FAX : (+66)038-694493, 038-949899 TAX ID:0215563006105 TEL H/P : +66-0-86-3173770 , +66-0-81-4022099 ,+66-0-62-325-6558

WEBSITE : WWW.BESTSAFE.CO.TH I WWW.BESTSAFEGLOVE.COM EMAIL : SALES@BESTSAFE.CO.TH

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for BESTSAFE GLOVE CO., LTD. The logo features a blue hand on the left side. To the right of the hand is the word "BESTSAFE" in blue, with a plus sign incorporated into the "B". Below "BESTSAFE" is the text "BESTSAFE GLOVE CO.,LTD" also in blue.

52/43 WATKODHIN-KAOPHAI RD. T.TUBMA, A.MUANG, RAYONG, RAYONG. THAILAND 21000 TEL OFFICE : 038-949877-79 FAX : (+66)038-694493, 038-949899 TAX ID:0215563006105 TEL H/P : +66-0-86-3173770 , +66-0-81-4022099 ,+66-0-62-325-6558 WEBSITE : WWW.BESTSAFE.CO.TH I WWW.BESTSAFEGLOVE.COM EMAIL : SALES@BESTSAFE.CO.TH

| Characteristics     | References/Standards                                                  | Device performance                                                                                              |                                                                                                                                                                                                                   | Comparison |              |                                                                                                                                 |              |  |      |  |  |  |  |         |  |  |  |  |         |
|---------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--------------|---------------------------------------------------------------------------------------------------------------------------------|--------------|--|------|--|--|--|--|---------|--|--|--|--|---------|
|                     |                                                                       | Predicate                                                                                                       | Current                                                                                                                                                                                                           |            |              |                                                                                                                                 |              |  |      |  |  |  |  |         |  |  |  |  |         |
| Physical Properties | ASTM D3578-19                                                         | Before Aging<br>Tensile Strength<br>18 MPa min<br>After Aging<br>Tensile Strength<br>14 MPa min                 | Before Aging<br>Tensile Strength (18 MPa min)   Size Actual value   Small 18.04   Medium 19.13   Large 18.26   Extra Large 18.64                                                                                  |            |              |                                                                                                                                 |              |  |      |  |  |  |  | Similar |  |  |  |  |         |
| Characteristics     | References/Standards                                                  | Device performance                                                                                              |                                                                                                                                                                                                                   | Comparison |              |                                                                                                                                 |              |  |      |  |  |  |  |         |  |  |  |  |         |
|                     |                                                                       | Predicate                                                                                                       | Current                                                                                                                                                                                                           |            |              |                                                                                                                                 |              |  |      |  |  |  |  |         |  |  |  |  |         |
|                     | Water Tight Test<br>(1000 ml)                                         |                                                                                                                 | ASTM D5151 – 19                                                                                                                                                                                                   |            | Pass AQL 1.5 | After Aging<br>Tensile Strength (14 MPa min)   Size Actual value   Small 15.02   Medium 18.12   Large 16.01   Extra Large 15.06 | Pass AQL 1.5 |  | Same |  |  |  |  |         |  |  |  |  | Similar |
| Powder Residual     | ASTM D6124-06<br>(Reapproved 2017)                                    | Meet $≤$ 2.0 mg/glove                                                                                           | $≤$ 2.0 mg/glove                                                                                                                                                                                                  | Similar    |              |                                                                                                                                 |              |  |      |  |  |  |  |         |  |  |  |  |         |
|                     |                                                                       |                                                                                                                 | Size Residual powder content (mg/glove)   Small 0.62   Medium 0.46   Large 0.61   Extra Large 0.61                                                                                                                |            |              |                                                                                                                                 |              |  |      |  |  |  |  |         |  |  |  |  |         |
|                     |                                                                       | Before Aging<br>Ultimate Elongation<br>650% min<br>After Aging<br>Ultimate Elongation<br>500% min (after aging) | Before Aging<br>Ultimate Elongation : 650% min   Size Actual value   Small 651   Medium 654   Large 650   Extra Large 650                                                                                         |            |              |                                                                                                                                 |              |  |      |  |  |  |  | Similar |  |  |  |  |         |
| Biocompatibility    | Primary Skin Irritation -<br>ISO 10993-10 Third<br>Edition 2010-08-01 | Not a primary skin irritant under the conditions of the study                                                   | Not a primary skin irritant under the conditions of the study                                                                                                                                                     | Same       |              |                                                                                                                                 |              |  |      |  |  |  |  |         |  |  |  |  |         |
|                     | Dermal Sensitization -<br>ISO 10993-10 Third<br>Edition 2010-08-01    | Not a contact sensitizer under the conditions of the study                                                      | Not a contact sensitizer under the conditions of the study                                                                                                                                                        | Same       |              |                                                                                                                                 |              |  |      |  |  |  |  |         |  |  |  |  |         |
|                     | In vitro cytotoxicity<br>ISO10993-5 :2009(E)                          | No data is available                                                                                            | Under the conditions of the study cytotoxic for undiluted (neat) and 1:2 dilutions but noncytotoxic for 1:4, 1:8, 1:16 and 1:32 dilutions. Moreover, under the conditions of the study, non acute systemic toxic. | Different  |              |                                                                                                                                 |              |  |      |  |  |  |  |         |  |  |  |  |         |
|                     | Acute Systemic Toxicity<br>ISO10993-11:2017(E)                        | No data is available                                                                                            | Under the conditions of the study, did not induce any systemic toxicity.                                                                                                                                          | Different  |              |                                                                                                                                 |              |  |      |  |  |  |  |         |  |  |  |  |         |
|                     |                                                                       |                                                                                                                 | After Aging<br>Ultimate Elongation : 500% min   Size Actual value   Small 501   Medium 601   Large 501   Extra Large 502                                                                                          |            |              |                                                                                                                                 |              |  |      |  |  |  |  | Similar |  |  |  |  |         |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for BESTSAFE GLOVE CO., LTD. The logo features a blue hand on the left side, followed by the word "BESTSAFE" in blue with a white cross in the middle of the "B". Below "BESTSAFE" is the text "BESTSAFE GLOVE CO., LTD" in a smaller font size. There is a registered trademark symbol in the upper right corner of the logo.

52/43 WATKODHIN-KAOPHAI RD. T.TUBMA, A.MUANG, RAYONG, RAYONG. THAILAND 21000 TAX ID:0215563006105 TEL OFFICE : 038-949877-79 FAX : (+66)038-694493, 038-949899 TEL H/P : +66-0-86-3173770 , +66-0-81-4022099 ,+66-0-62-325-6558

WEBSITE : WWW.BESTSAFE.CO.TH I WWW.BESTSAFEGLOVE.COM EMAIL : SALES@BESTSAFE.CO.TH

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for BESTSAFE GLOVE CO., LTD. The logo features a blue hand on the left side, followed by the word "BESTSAFE" in blue, with a plus sign in the middle of the word. Below "BESTSAFE" is the text "BESTSAFE GLOVE CO.,LTD" also in blue. There is a registered trademark symbol in the upper right corner of the logo.

52/43 WATKODHIN-KAOPHAI RD. T.TUBMA, A.MUANG, RAYONG, RAYONG. THAILAND 21000

TAX ID:0215563006105 TEL OFFICE : 038-949877-79 FAX : (+66)038-694493, 038-949899 TEL H/P : +66-0-86-3173770 , +66-0-81-4022099 ,+66-0-62-325-6558

WEBSITE : WWW.BESTSAFE.CO.TH | WWW.BESTSAFEGLOVE.COM EMAIL : SALES@BESTSAFE.CO.TH

## 7.0 Summary of Non-clinical performance Tests

| Test Method            | Standard                                                                 | Purpose of testing                                           | Acceptance Criteria                                                                                                                 | Result                                                                                                                                                                                                       | Status |
|------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Dimension              | ASTM D3578-19<br>Standard Specification for Rubber<br>Examination Gloves | To determine the length<br>of the gloves                     | Min 230 mm for all sizes                                                                                                            | Small: 240 mm<br>Medium: 240 mm<br>Large: 240 mm<br>Extra Large: 241 mm                                                                                                                                      | Pass   |
|                        | ASTM D3578-19<br>Standard Specification for Rubber<br>Examination Gloves | To determine the width<br>of the gloves                      | Small: 80 ± 10 mm<br>Medium: 95±10mm<br>Large: 111± 10 mm<br>Extra Large: 120 ± 10 mm                                               | Small: 85 mm<br>Medium: 93 mm<br>Large: 105 mm<br>Extra Large: 115 mm                                                                                                                                        | Pass   |
|                        | ASTM D3578-19<br>Standard Specification for Rubber<br>Examination Gloves | To determine the<br>thickness of the gloves                  | Palm 0.08 mm min<br>Finger 0.08 mm min for all sizes                                                                                | Small: Palm 0.09 mm, Finger: 0.12 mm<br>Medium: Palm: 0.09 mm, Finger: 0.12 mm<br>Large: Palm 0.10 mm, Finger: 0.13 mm<br>Extra Large: Palm 0.09 mm, Finger: 0.12 mm                                         | Pass   |
| Physical<br>Properties | ASTM D3578-19<br>Standard Specification for Rubber<br>Examination Gloves | To Determine the<br>physical properties-<br>Tensile strength | Before Ageing<br>Tensile Strength 18Mpa<br>Minimal for all sizes<br>After Ageing Tensile<br>Strength 14Mpa<br>Minimal for all sizes | Before Ageing<br>Small: 18.04 MPa<br>Medium: 19.13 MPa<br>Large: 18.26 MPa<br>Extra Large: 18.64 MPa<br>After ageing:<br>Small: 15.02 MPa<br>Medium: 18.12 MPa<br>Large: 16.01 MPa<br>Extra Large: 15.06 MPa | Pass   |

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the logo for BESTSAFE GLOVE CO., LTD. The logo features a blue hand on the left side, followed by the word "BESTSAFE" in large, bold, blue letters. Below "BESTSAFE" is the text "BESTSAFE GLOVE CO.,LTD" in smaller, blue letters. There is a registered trademark symbol in the upper right corner of the logo.

52/43 WATKODHIN-KAOPHAI RD. T.TUBMA, A.MUANG, RAYONG, THAILAND 21000

TEL OFFICE : 038-949877-79 FAX : (+66)038-694493, 038-949899 TEL H/P : +66-0-86-3173770 , +66-0-81-4022099 ,+66-0-62-325-6558 TAX ID:0215563006105

WEBSITE : WWW.BESTSAFE.CO.TH | WWW.BESTSAFEGLOVE.COM EMAIL : SALES@BESTSAFE.CO.TH

| Test Method            | Standard                                                                                         | Purpose of testing                                              | Acceptance Criteria                                                                                                             | Result                                                                                                                                                                                                                                                                                                                    | Status |
|------------------------|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Physical<br>Properties | ASTM D3578-19<br>Standard Specification for Rubber<br>Examination Gloves                         | To Determine the<br>physical properties-<br>Ultimate Elongation | Before Ageing<br>Ultimate Elongation<br>650% Min for all sizes<br>After Ageing Ultimate<br>Elongation 500% Min<br>for all sizes | Before Ageing<br>Small: 651%<br>Medium: 654 %<br>Large:650%<br>Extra Large:650%<br>After ageing:<br>Small:501%<br>Medium: 601%<br>Large:501%<br>Extra Large:502%                                                                                                                                                          | Pass   |
| Watertight test        | ASTM D5151-19<br>Standard Test Method for Detection<br>of Holes in Medical<br>Gloves             | To determine the holes in<br>the gloves                         | Sample size: 200 pcs<br>Inspection level : Gl<br>AQL 1.5<br>Acceptance Number 7<br>Rejection Number 8                           | The batch size for this sampling is 35,001-<br>150,000. Hence, according to the single<br>sampling plan GI, the sample to be drawn<br>is under code L equivalent to 200 pcs with<br>accept 7 and reject 8 to be accept under<br>AQL 1.5.<br>Small: 0 (Zero)<br>Medium: 0 (Zero)<br>Large:0 (Zero)<br>Extra Large:0 (Zero) | Pass   |
| Residual<br>powder     | ASTM D6124-06 (Reapproved 2017)<br>Standard Test Method for Residual<br>Powder on Medical Gloves | To determine the residual<br>powder in the gloves               | 2 mg per glove or less                                                                                                          | Sample size : 5 pcs<br>Requirement: 2 mg per glove or less<br>Result<br>Small:0.62 mg/glove<br>Medium: 0.46 mg/glove<br>Large: 0.61 mg/glove<br>Extra Large:0.61  mg/glove                                                                                                                                                | Pass   |

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for BESTSAFE GLOVE CO., LTD. The logo features a blue hand on the left, followed by the word "BESTSAFE" in blue, with a plus sign in the middle of the word. Below "BESTSAFE" is the text "BESTSAFE GLOVE CO., LTD" in a smaller font, also in blue. There is a registered trademark symbol in the upper right corner of the logo.

52/43 WATKODHIN-KAOPHAI RD. T.TUBMA, AMUANG, RAYONG, RAYONG. THAILAND 21000

TEL OFFICE : 038-949877-79 FAX : (+66)038-694493, 038-949899 TAX ID : 0215563006105 TEL H/P : +66-0-86-3173770 , +66-0-81-4022099 ,+66-0-62-325-6558 EMAIL : SALES@BESTSAFE.CO.TH WEBSITE : WWW.BESTSAFE.CO.TH I WWW.BESTSAFEGLOVE.COM

Status Test Method Standard Acceptance Criteria Purpose of testing Result Protein content ASTM D5712 - 15, Standard Test Less than 200 µg/dm² Pass To determine the Sample size : 3 pcs extractable protein in Method for Analysis of Aqueous Requirement: Less than 200 µg/dm2 Extractable Protein in Natural the gloves Result Small:124.36 ug /dm2 Rubber. Medium: 140.78 µg /dm² Large: 134.26 µg /dm² Extra 159.46 µg /dm²

The performance test data of the non-clinical tests meet following standards:

ASTM D 3578 – 19 Standard Specification for Rubber Examination Gloves

ASTM D 5151-19: Standard Test Method for Detection of Holes in Medical Gloves

ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves

ASTM D5712 - 15, Standard Test Method for Analysis of Aqueous Extractable Protein in Natural Rubber.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for BESTSAFE GLOVE CO., LTD. The logo features a blue hand on the left side, followed by the word "BESTSAFE" in blue, with a plus sign in the middle of the word. Below "BESTSAFE" is the text "BESTSAFE GLOVE CO., LTD" in a smaller font size.

BESTSAFE GLOVE CO.,LTD. | บริษัท เบสท์เซฟ โกลฟ จำกัด 52/43 WATKODHIN-KAOPHAI RD. T.TUBMA, AMUANG, RAYONG, RAYONG. THAILAND 21000 TEL OFFICE : 038-949877-79 FAX : (+66)038-694493, 038-949899 TAX ID:0215563006105 TEL H/P : +66-0-86-3173770 , +66-0-81-4022099 ,+66-0-62-325-6558 EMAIL : SALES@BESTSAFE.CO.TH WEBSITE : WWW.BESTSAFE.CO.TH I WWW.BESTSAFEGLOVE.COM

#### 8.0 Summary of Clinical Performance Tests:

Not Applicable

#### 9.0 Conclusion:

The conclusion drawn from the non-clinical tests demonstrates that the subject device Latex Examination Powder Free Gloves is as safe, as effective, and performs as well as or better than the legally marketed predicate device K110250

--END--

---

**Source:** [https://fda.innolitics.com/device/K210253](https://fda.innolitics.com/device/K210253)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
