Clinical case records from EU market experience; Retrospective clinical performance data
The sponsor used retrospective clinical data from patients treated in the EU to support the safety and effectiveness of the device's attachment method and overall clinical performance.
Retrospective clinical data; EU clinical experience; Post-market performance
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
IPS Preprosthetic Clinical Performance Data; Retrospective clinical case review; Follow-up/Duration: Up to 74 months post-operatively
35 patients (39 total implants) including postablative maxillary defects, severely atrophic maxilla, and cleft lip and palate (CLP) deformity; Sample Size: 35 patients / 39 implants
Not applicable for this study
Implant loosening, implant removal, post reduction, screw exposure/removal, infection/abscess, partial exposure of framework
Indications for Use
The KLS Martin IPS Preprosthetic Implant is a subperiosteal implant composed of titanium intended to construct patient specific prosthetic devices which are surgically implanted into the lower or upper jaw between the periosteum (connective tissue covering the bone) and supporting bony structures. The device is intended to provide support for multi-unit prostheses, such as dentures.
Device Story
The KLS Martin IPS Preprosthetic is a patient-specific subperiosteal implant system manufactured via additive methods (Selective Laser Melting) from Ti-6Al-4V. Input data consists of patient CT scans; physicians collaborate with KLS Martin engineers during virtual planning to design the implant. The system produces physical implants and anatomical models. The device is surgically implanted into the jaw between the periosteum and bone, fixated with bone screws, and features transgingival pillars to support multi-unit prostheses like dentures. Used in maxillofacial/mandibular reconstruction, the device restores chewing function in patients with compromised bone. Physicians use the output to plan and execute reconstructive surgeries. Benefits include patient-specific fit and stable prosthetic support.
Clinical Evidence
Clinical data from 39 implants in 35 patients (postablative maxillary defects, atrophic maxilla, cleft lip/palate) followed up to 74 months. Results: 0% implant loosening or removal. Complications included partial framework exposure (40%), infection/abscess (8.6%), exposed/removed screws (8.6%), and reduced post (2.9%).
Technological Characteristics
Material: Ti-6Al-4V (ASTM F136). Manufacturing: Additive (Selective Laser Melting). Dimensions: 1.2-10.0 mm thickness; pillars ≥4 mm diameter. Fixation: Bone screws. Sterilization: Steam (dynamic-air-removal). Connectivity: None (standalone). Software: Used for virtual planning/design.
Indications for Use
Indicated for patients requiring support for multi-unit prostheses (e.g., dentures) due to compromised maxilla or mandible bone. Contraindications include infections, hypersensitivity to foreign bodies or implant materials, circulatory/metabolic disorders, insufficient bone tissue, degenerative diseases, non-sterile surgical environments, excessive weight loads, inability to follow post-op instructions, osteoporosis/osteomalacia, bone tumors, drug/alcohol abuse, failure to locate primary nerve, uncontrolled Type II diabetes, bisphosphonate use, and bruxism.
Regulatory Classification
Identification
Subperiosteal implant material is a device composed of titanium or cobalt chrome molybdenum intended to construct custom prosthetic devices which are surgically implanted into the lower or upper jaw between the periosteum (connective tissue covering the bone) and supporting bony structures. The device is intended to provide support for prostheses, such as dentures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
KLS-Martin L.P. Susan Leander Regulatory Affairs Project Supervisor 11201 Saint Johns Industrial Parkway S Jacksonville, Florida 32246
Re: K210228
Trade/Device Name: KLS Martin IPS Preprosthetic Regulation Number: 21 CFR 872.3645 Regulation Name: Subperiosteal Implant Material Regulatory Class: Class II Product Code: ELE Dated: June 1, 2022 Received: June 7, 2022
Dear Susan Leander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210228
Device Name
KLS Martin IPS Preprosthetic
#### Indications for Use (Describe)
The KLS Martin IPS Preprosthetic Implant is a subperiosteal implant composed of titanium intended to construct patient specific prosthetic devices which are surgically implanted into the lower or upper jaw between the periosteum (connective tissue covering the bone) and supporting bony structures. The device is intended to provide support for multi-unit prostheses, such as dentures.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while the word "martin" is in gray. The letters "L.P." are in black and are slightly larger than the other letters in the logo.
11201 Saint Johns Industrial Pkwy S . Jacksonville, FL 32246 Phone: 904-641-7746 · 800-625-1557 · Fax: 904-641-7378 www.klsmartinnorthamerica.com
## Section 5
#### 510(k) Summary 21 CFR 807.92
| Submitter: | KLS-Martin L.P.<br>11201 Saint Johns Industrial Pkwy S<br>Jacksonville, FL 32246 |
|----------------------|---------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Susan Leander<br>Regulatory Affairs Project Supervisor<br>Phone: 800-625-1557<br>Email: susan.leander@klsmartin.com |
| Alternate Contact: | Melissa Bachorski<br>Assistant Director RA/QMS<br>Phone: 800-625-1557<br>Email: melissa.bachorski@klsmartin.com |
| Date Prepared: | July 7, 2022 |
| Trade Name: | KLS Martin IPS Preprosthetic |
| Common Name: | Dental implant |
| Classification Name: | Implant, subperiosteal (21 CFR 872.3645) |
| Regulatory Class: | II |
| Product Code: | ELE |
| Predicate Device: | Hollow Basket Titanium Dental Implant (K801208) |
| Reference Devices: | Pacific Implants Intraosseous Dental Implant (K813227)<br>KLS Martin Individual Patient Solutions (K191028) |
#### Device Description
The KLS Martin Individual Patient Solutions (IPS) Preprosthetic system is comprised of patient-specific models and metallic bone plates with integrated pillars used in conjunction with metallic bone screws for internal fixation of the implant to maxillofacial / midface and mandibular bones. The integrated pillars will serve as the base for temporary dentures as well as a permanent prosthesis. The devices are manufactured based on medical imaging (CT scan) of the patient's anatomy with input from the physician during virtual planning and prior to finalization and production of the device. The physician only provides input for model manipulation and interactive feedback by viewing digital models of planned outputs that are modified by trained KLS Martin engineers during the planning session. For each design iteration, verification is performed by virtually fitting the generated implant over a 3D model of the patient's anatomy to ensure its dimensional properties allow an adequate fit.
The KLS Martin IPS Planning System is utilized to plan and design the IPS Preprosthetic implant. The IPS Planning System is a collection of software and associated additive manufacturing (rapid prototyping) equipment intended to provide a variety of outputs to support reconstructive and orthognathic surgeries. The system uses electronic medical images of the patients' anatomy (CT data) with input from the physician, to manipulate original
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, "martin" is in gray, and "L.P." is in black. The logo is simple and modern, with a clean font.
patient images for planning and executing surgery. The system processes the medical images and produces a variety of patient specific physical and/or digital output devices which include anatomical models, implants, and case reports.
Implants are provided non-sterile, range in thickness from 1.2 - 10.0 mm, and are manufactured using additive methods from Ti-6Al-4V (ASTM F136). These patient-specific devices are fixated with previously cleared KLS Martin screws.
Implants have a minimum of two (2) transgingival pillars for the attachment of dental prostheses. The straight pillars (0° to the occlusal plane) have a diameter of 4 mm and are provided at lengths up to 20 mm.
#### Indications for Use
The KLS Martin IPS Preprosthetic Implant is a subperiosteal implant composed of titanium intended to construct patient specific prosthetic devices which are surgically implanted into the lower or upper jaw between the periosteum (connective tissue covering the bone) and supporting bony structures. The device is intended to provide support for of multi-unit prostheses, such as dentures.
#### Technological Characteristics & Substantial Equivalence Discussion
The intended use of the subject device, KLS Martin IPS Preprosthetic is similar to the predicate device, K801208 and the reference device, K813227. The potential impact on substantial equivalence of each technological difference was addressed through risk analysis and verification and validation testing.
#### Similarities to Predicate
The subject and predicate devices are intended to be used to provide support for prosthetic devices. Both are subperiosteal implants made of titanium that are surgically placed in the upper or lower jaw to support prostheses. Both devices have transgingival pillars for the attachment of prostheses.
The subject and predicate devices are safe and effective at restoring a patient's chewing function when the maxilla and/or mandible have been compromised.
The subject device is provided non-sterile and requires the end-user to process the implants using validated cleaning and sterilization methods prior to use as recommended in the labeling. Sterilization information is not available for the predicate device.
#### Differences from Predicate
The subject and predicate devices are attached to the bony substructures of the jaw. The predicate device is an intraosseous implant whereas the subject device is implanted using bone screws to attach the device to the bone. Although attachment to the bone is achieved differently, both are effective methods. Studies of the predicate device present evidence of osseointegration that results in stable attachment of the implant. The attachment method used by the subject device has been safely and effectively demonstrated in more than one hundred clinical cases in the EU, where the device is currently marketed. Real-world evidence of the subject device's safe and effective use is cited in peer-reviewed literature.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while the word "martin" is in gray. The letters "L.P." are in black.
#### Reference Devices
KLS Martin Individual Patient Solutions (K191028), included as a reference device, is comprised of patient-specific models and metallic bone plates used in conjunction with metallic bone screws for internal fixation of maxillofacial / midface and mandibular bones.
Both the subject and reference devices are manufactured based on medical imaging (CT scan) of the patient's anatomy with input from the physician during virtual planning and prior to finalization and production of the device. The physician only provides input for model manipulation and interactive feedback by viewing digital models of planned outputs that are modified by trained KLS Martin engineers during the planning session. For each design iteration, verification is performed by virtually fitting the generated implant over a 3D model of the patient's anatomy to ensure its dimensional properties allow an adequate fit.
Subject and reference device implants are provided non-sterile and are manufactured using additive methods from Ti-6Al-4V. These patient-specific devices are fixated with previously cleared KLS Martin screws.
The Pacific Implant, Inc. Endosseous Dental Implant (K813227) has been included as a reference device for comparison of typical loading values for dental implants. The subject and reference devices are intended to be used to provide support for prosthetic devices. Both are made of titanium; both are surgically placed in the upper or lower jaw and both are devices to support prostheses. Both devices have transgingival pillars for the attachment of prostheses. Both devices are designed for use in patients that have sub-optimal quantity or quality of bone in the upper or lower jaw.
Both the subject and reference devices are safe and effective at restoring a patient's chewing function when the jawbone has been compromised.
The subject and reference devices are provided non-sterile and require the end-user to process the implants using validated cleaning and sterilization methods prior to use as recommended in the labeling.
{6}------------------------------------------------
## kıs martı L.P.
11201 Saint Johns Industrial Pkwy S ● Jacksonville, FL 32246 Phone: 904-641-7746 • 800-625-1557 • Fax: 904-641-7378 www.klsmartinnorthamerica.com
| Device Comparison Table | | | | |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | KLS Martin IPS Preprosthetic<br>K210228 | Hollow Basket Titanium Dental<br>Implants<br>K801208 | Pacific Implant, Inc. Intraosseous Dental<br>Implant<br>K813227 | KLS Martin Individual Patient Solutions<br>K191028 |
| | (Subject Device) | (Primary Predicate) | (Reference Device) | (Reference Device) |
| Indications for Use | The KLS Martin IPS Preprosthetic Implant is a subperiosteal implant<br>composed of titanium intended to construct patient specific<br>prosthetic devices which are surgically implanted into the lower or<br>upper jaw between the periosteum (connective tissue covering the<br>bone) and supporting bony structures. The device is intended to<br>provide support for multi-unit prostheses, such as dentures. | The Hollow Basket Titanium Dental<br>Implant is a subperiosteal implant<br>composed of titanium intended to<br>construct patient specific prosthetic<br>devices which are surgically implanted<br>into the lower or upper jaw between the<br>periosteum (connective tissue covering the<br>bone) and supporting bony structures. | An endosseous dental implant is a<br>prescription device made of a material such as<br>titanium or titanium alloy that is intended to<br>be surgically placed in the bone of the upper<br>or lower jaw arches to provide support for<br>prosthetic devices, such as artificial teeth, in<br>order to restore a patient's chewing function.¹ | The KLS Martin Individual Patient Solutions<br>implant devices are intended for use in the<br>stabilization, fixation, and reconstruction of the<br>maxillofacial / midface and mandibular skeletal<br>regions. |
| Contraindications | 1. Obvious infections<br>2. Hypersensitivity to foreign bodies<br>3. Suspected sensitivity to the implant material<br>4. Circulatory problems, systemic diseases and metabolic<br>disorders<br>5. Insufficient or inadequate bone tissue<br>6. Secondary diseases such as degenerative processes that may<br>negatively influence the healing process<br>7. Interventions carried out in a non-sterile environment (e.g.<br>paranasal sinuses)<br>8. Regions exposed to inappropriate forces or excessive weight<br>loads<br>9. Patients unwilling or unable to follow instructions during the<br>postoperative phase due to their mental, neurological or<br>physical condition<br>10. Osteoporosis or osteomalacia or other severe structural bone<br>damage preventing the stable fixation of implant components<br>11. Bone tumors located in the implant base region<br>12. Obvious drug or alcohol abuse<br>13. Failure to locate a primary nerve in the lower jaw<br>14. Uncontrolled Type II diabetes<br>15. Oral or intravenous bisphosphonates<br>16. Bruxism (tooth grinding or clenching)<br>17. Smoking | Unknown – similar to K813227 | 1. Local tissue or existing dentition<br>degeneration due to²:<br>a. Excessive mobility<br>b. Loss of integration<br>c. Incompatibility of the device<br>components<br>d. Structural failure of the device<br>2. Pain<br>3. Infection<br>4. Adverse tissue reaction<br>5. Bone or nerve damage<br>a. Sinus perforation<br>b. Alveolar plate perforation<br>c. Transient or chronic pain/facial<br>paresis<br>6. Migration or thermal injury<br>a. Incompatibility with MRI<br>7. Patient Selection Criteria<br>a. Previous radiation therapy (affects<br>bone quality) | 1. Obvious infections.<br>2. Hypersensitivity to foreign bodies.<br>3. Suspected sensitivity to the implant<br>material.<br>4. Circulatory problems, systemic diseases,<br>and metabolic disorders.<br>5. Insufficient or inadequate bone tissue.<br>6. Secondary diseases such as degenerative<br>processes that may negatively influence the<br>healing process.<br>7. Interventions carried out in a non-sterile<br>environment (e.g. paranasal sinuses).<br>8. Regions exposed to inappropriate forces or<br>excessive weight loads.<br>9. Patients unwilling or unable to follow<br>instructions during the postoperative phase<br>due to their mental, neurological, or<br>physical condition.<br>10. Osteoporosis or osteomalacia or other<br>structural bone damage preventing the<br>stable fixation of implant components.<br>11. Bone tumors located in the implant base<br>region.<br>12. Obvious drug or alcohol abuse. |
| Patient-specific? | Yes. Devices are manufactured patient-specific, based on a CT scan<br>of the patient. | | No. | Yes. Devices are manufactured patient-specific,<br>based on a CT scan of the patient. |
| Classification | 21 CFR 872.3645<br>Class II | | 21 CFR 872.3640<br>Class II | 21 CFR 872.4760<br>Class II |
| Product Code | ELE | | DZE | JEY |
| Material | Anatomical Models: Epoxy/Acrylic Resins Implants: Ti-6Al-4V | | Implants: CP Titanium | Anatomical Models: Epoxy/Acrylic Resins Cutting/Marking Guides: Polyamide,<br>Ti-6Al-4V, CP Titanium Splints: acrylic/methacrylic resins |
| Manufacturing<br>Method | Epoxy/Acrylic Resins: Stereolithography (SLA) Ti-6Al-4V: 3D (Additive; Selective Laser Melting) | | Traditional milling/stamping | Epoxy/Acrylic Resins: Stereolithography<br>(SLA) CP Titanium: Traditional (Subtractive) Ti-6Al-4V: 3D (Additive; Selective Laser<br>Melting) Polyamide: 3D (Additive; Selective Laser<br>Sintering) |
| Sterilization | Unknown | | Non-sterile | Non-sterile |
| Anatomical Sites | Maxillofacial/Midface & Mandible | | Mandible | Maxillofacial / Midface & Mandible |
| Plate Specifications | | | | |
| Thickness | 1.2 mm - 10.0 mm | | Unknown | Orbital floor only: 0.3 mm – 1.0 mm Maxillofacial / midface reconstruction:<br>0.6 mm - 10.0 mm Mandibular reconstruction: 1.0 mm - 3.0 mm |
| Style | Non-locking<br>Compression<br>Threaded | | One-piece hollow, grooved, bore-holed<br>cylinders with upper portions designed to<br>support an artificial denture | Non-locking<br>Compression<br>Threaded |
| Width<br>(between screws) | Min: 4.5 mm | | No screws are used with this device | Maxillofacial / midface:<br>Min: ≥ 4.5 mm (around screw holes)<br>Min: ≥ 3 mm (not around screw hole)<br>Max: Dependent on screw-hole<br>Mandibular:<br>Min: 7 mm<br>Max: 8.5 mm |
| Length | Maxillofacial / midface:<br>Min: 29 mm<br>Max: 320 mm<br>Mandibular:<br>Min: 29 mm<br>Max: 320 mm | | Unknown | Maxillofacial / midface:<br>Min: 18 mm<br>Max: 350 mm<br>Mandibular:<br>Min: 31 mm<br>Max: 320 mm |
| Width around<br>Screws | 1.5 System: ≥ 4.5 mm<br>2.0/2.3 System: ≥ 6.5 mm<br>2.7 System: ≥ 8 mm | No screws are used with this device | No screws are used with this device | 1.5 System: ≥ 4.5 mm<br>2.0/2.3 System: ≥ 6.5 mm<br>2.7 System: ≥ 8 mm |
| Number of Screw<br>Holes | Maxillofacial / midface:<br>Min: 4 per side of defect<br>Max: Dependent on length & hole spacing<br>Mandibular:<br>Min: 4<br>Max: Dependent on length & hole spacing | Holes are present but are not screw holes.<br>These are to allow for bone ingrowth. | 14 holes are present but are not screw holes.<br>These are to allow for bone ingrowth. | Maxillofacial / midface:<br>Min: ≥ 2 per side of defect<br>Max: Dependent on length & hole spacing<br>Mandibular:<br>Min: 4<br>Max: Dependent on length & hole spacing |
| Pillar Diameter | ≥ 4 mm | Unknown | N/A | N/A |
| Angulation of<br>Pillar to the<br>Occlusal Plane | 0° | Unknown | N/A | N/A…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.