Triever Catheters

K210195 · Inari Medical · QEW · Feb 24, 2021 · Cardiovascular

Device Facts

Record IDK210195
Device NameTriever Catheters
ApplicantInari Medical
Product CodeQEW · Cardiovascular
Decision DateFeb 24, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.5150
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Triever Catheters are indicated for: - The non-surgical removal of emboli and thrombi from blood vessels. - Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. Triever Catheters are intended for use in the peripheral vasculature and for the treatment of pulmonary embolism. Triever Catheters are also intended for use in transit in the right atrium but not in conjunction with FlowTriever Catheters.

Device Story

Triever Catheters are single-use, over-the-wire aspiration catheters used for minimally invasive thromboemboli removal in peripheral vasculature and pulmonary embolism treatment. Used in conjunction with FlowTriever Catheters (except for right atrium transit), the system involves advancing the FlowTriever catheter to the thrombus, deploying self-expanding wireform disks to engage the clot, and retracting the FlowTriever into the Triever catheter for capture. Aspiration is facilitated via a 60 cc VacLok vacuum syringe. Operated by physicians in clinical settings; the device provides mechanical clot removal, reducing thrombus burden and potentially improving patient outcomes. This submission adds a warning regarding use in patients with severe pulmonary hypertension.

Clinical Evidence

No clinical data required for this submission; substantial equivalence determined via labeling modification review.

Technological Characteristics

Single-use, over-the-wire aspiration catheter system. Available in 16, 20, and 24 Fr sizes. Mechanical aspiration principle using vacuum syringe. No software or electronic components.

Indications for Use

Indicated for non-surgical removal of emboli and thrombi from blood vessels, and injection/infusion/aspiration of fluids/contrast media. Intended for peripheral vasculature, pulmonary embolism treatment, and right atrium clot transit. Contraindicated for use in right atrium in conjunction with FlowTriever Catheters.

Regulatory Classification

Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable. February 24, 2021 Inari Medical Ms. Ellen Nguyen Regulatory Affairs Associate 9 Parker. Suite 100 Irvine, California 92618 ## Re: K210195 Trade/Device Name: Triever Catheters Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW Dated: January 22, 2021 Received: January 25, 2021 ## Dear Ms. Nguyen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) #### K210195 Device Name Triever Catheters | Indications for Use ( <i>Describe</i> ) | |-----------------------------------------| | Triever Catheters are indicated for: | - · The non-surgical removal of emboli and thrombi from blood vessels. - · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. Triever Catheters are intended for use in the peripheral vasculature and for the treatment of pulmonary embolism. Triever Catheters are also intended for use in transit in the right atrium but not in conjunction with FlowTriever Catheters. Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K210195 PAGE 1 OF 2 # 510(K) SUMMARY | Date prepared | January 22, 2021 | |---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Name | Inari Medical, Inc.<br>9 Parker, Suite 100<br>Irvine, CA 92618<br>949.600.8433 | | Contact person | Ellen Nguyen<br>Regulatory Affairs Associate | | Trade name | Triever Catheters | | Common name | Embolectomy catheter | | Regulation name | Embolectomy catheter | | Classification number | 21 CFR 870.5150 | | Product code | QEW | | Regulatory class | II | | Predicate device | Triever Catheters (K202345) | | Description | The FlowTriever Retrieval/Aspiration System is a single-use over-the-wire<br>catheter-based system for the minimally invasive treatment of thromboemboli in<br>the peripheral vasculature and for the treatment of pulmonary embolism. The<br>system is comprised of two main components packaged separately:<br>Triever Catheters (available in 3 sizes: 16, 20, and 24 Fr) FlowTriever Catheters (available in 4 sizes: 6-10 mm, 11-14 mm, 15-18 mm, and 19-25 mm) The FlowTriever Catheter is inserted through the Triever Catheter and advanced<br>to the thrombus. Self-expanding wireform disks are deployed to engage thrombus<br>by retracting the outer delivery catheter. The FlowTriever Catheter is retracted<br>into the Triever Catheter to capture the targeted thrombus. Additional clot may be<br>removed by aspiration with the provided 60 cc VacLok Vacuum syringe. After the<br>procedure is complete, the Triever Catheter and FlowTriever Catheter are<br>removed from the patient. | | Indications for Use | Triever Catheters are indicated for: The non-surgical removal of emboli and thrombi from blood vessels. Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. Triever Catheters are intended for use in the peripheral vasculature and for the treatment of pulmonary embolism.<br>Triever Catheters are also intended for use in treating clot in transit in the right atrium but not in conjunction with FlowTriever Catheters. | | Device modifications | This submission added a warning to the Instruction for Use in patients with severe<br>pulmonary hypertension. | | Summary of<br>substantial equivalence | There is no change of intended use or fundamental scientific technology between<br>the proposed and predicate device. Triever Catheters have the same indication for<br>use as the predicate, K202345.<br><br>Animal testing was not required for the determination of substantial equivalence.<br><br>Clinical testing was not required for the determination of substantial equivalence.<br><br><b>Conclusion</b><br>The information provided demonstrates that including a warning for use in patients<br>with severe pulmonary hypertension in the Instructions for Use does not raise new<br>or different questions of safety and effectiveness. It can be concluded that Triever<br>Catheters with the modified labeling is substantially equivalent to the predicate<br>device. | {4}------------------------------------------------
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...