ClotTriever Thrombectomy System

K210190 · Irani Medical, Inc. · QEW · Feb 23, 2021 · Cardiovascular

Device Facts

Record IDK210190
Device NameClotTriever Thrombectomy System
ApplicantIrani Medical, Inc.
Product CodeQEW · Cardiovascular
Decision DateFeb 23, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.5150
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K210190 · Feb 23, 2021ClotTriever Thrombectomy SystemIrani Medical, Inc.CLOUT registryThe CLOUT registry clinical data was used to support the labeling change (removal of 'chronic clot' from contraindications) for the ClotTriever Thrombectomy System.Registry; Labeling change; DVT

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
CLOUT registry; RegistryPatients with deep vein thrombosis (DVT)Not applicable for this study—

Indications for Use

The ClotTriever Thrombectomy System is indicated for : - · The non-surgical removal of thrombi and emboli from blood vessels. - · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT).

Device Story

ClotTriever Thrombectomy System is a single-use, sterile device for non-surgical removal of thrombi/emboli from peripheral vasculature. System components include a reinforced polymeric sheath with self-expanding distal mesh funnel, hemostatic valve, and a catheter with reinforced coaxial shafts, expandable coring element, and thrombus collection bag. Radiopaque tips aid fluoroscopic visualization. Operated by clinicians in peripheral vascular procedures. Device functions by mechanical coring and collection of thrombus material. Output is the physical removal of clots, facilitating blood flow restoration. Benefits include minimally invasive treatment of DVT and other peripheral vascular occlusions.

Clinical Evidence

No new clinical data was required. The submission relies on clinical data from the CLOUT registry previously submitted and reviewed under K193462.

Technological Characteristics

Single-use, sterile embolectomy catheter system. Materials: reinforced polymers. Components: sheath with self-expanding distal mesh funnel, hemostatic valve, and catheter with expandable coring element and collection bag. Radiopaque markers on sheath dilator and catheter distal tips for fluoroscopic visualization. Mechanical operation; no software or energy source.

Indications for Use

Indicated for non-surgical removal of thrombi and emboli from peripheral vasculature, including deep vein thrombosis (DVT), and for injection, infusion, or aspiration of fluids/contrast media.

Regulatory Classification

Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. February 23, 2021 Inari Medical, Inc. Larry Boucher Regulatory Affairs Manager 9 Parker. Suite 100 Irvine, California 92618 ## Re: K210190 Trade/Device Name: ClotTriever Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW Dated: January 22, 2021 Received: January 25, 2021 ### Dear Larry Boucher: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210190 Device Name ClotTriever Thrombectomy System Indications for Use (Describe) The ClotTriever Thrombectomy System is indicated for : - · The non-surgical removal of thrombi and emboli from blood vessels. - · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT). Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K210190 PAGE 1 OF 2 # 510(K) SUMMARY | Date prepared | January 22, 2021 | |------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Name | Inari Medical, Inc.<br>9 Parker, Suite 100<br>Irvine, CA 92618<br>949.600.8433 | | Contact person | Larry Boucher<br>Regulatory Affairs Manager | | Trade name | ClotTriever Thrombectomy System | | Common name | Embolectomy catheter | | Regulation name | Embolectomy catheter | | Classification number | 21 CFR 870.5150 | | Product code | QEW | | Regulatory class | II | | Predicate device | ClotTriever Thrombectomy System (K193462) | | Description | The ClotTriever Thrombectomy System is a single-use, sterile medical device<br>designed for use in the peripheral vasculature. The ClotTriever Thrombectomy<br>System consists of the ClotTriever Sheath and the ClotTriever Catheter. The<br>ClotTriever Sheath comprises a reinforced polymeric sheath equipped with a self-<br>expanding distal mesh funnel, a flush/aspiration port, and a proximal hemostatic<br>valve. A dilator is provided to aid insertion. Other provided accessories include<br>the funnel loading tool and a Large Bore 60 cc syringe. The ClotTriever Catheter<br>comprises reinforced polymeric coaxial shafts terminating in an expandable<br>coring element and thrombus collection bag. Two ports terminating in stopcocks<br>are provided for de-airing the catheter shafts. To aid in fluoroscopic visualization,<br>the Sheath dilator and ClotTriever Catheter distal tips are radiopaque. | | Indications for Use | The ClotTriever Thrombectomy System is indicated for:<br>• The non-surgical removal of thrombi and emboli from blood vessels.<br>• Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.<br><br>The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT). | | Device modifications | The labeling has been changed by deleting “chronic clot” from the examples given “…for removal fibrous, adherent, or calcified material…” from the contraindications. | | Summary of substantial equivalence | The ClotTriever Thrombectomy System and predicate device has the same intended use: removal of thrombus and emboli from and infusion of fluids into the peripheral vasculature.<br><br><b>Non-Clinical Testing</b><br>Non-clinical testing was not required to support the change to the Indications for Use.<br><br><b>Clinical Testing</b><br>The CLOUT registry clinical data supporting this labeling change was previously submitted in the ClotTriever Thrombectomy System 510(k) (K193462, cleared on September 9, 2020). No additional clinical data was deemed necessary to determine substantial equivalence of the subject and predicate devices.<br><br>Removing “chronic clot” from the example “of fibrous, adherent, or calcified material” does not introduce new issues of safety or effectiveness.<br><br><b>Conclusion</b><br>Removing “chronic clot” from the example “of fibrous, adherent, or calcified material” of the ClotTriever Thrombectomy System aligns with the currently cleared Indications for Use and does not raise new or different questions of safety or and effectiveness. It can be concluded that the ClotTriever System with the modified contraindication is substantially equivalent to the predicate device. | {4}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...