K210183 · Biolase, Inc. · GEX · Apr 21, 2022 · General, Plastic Surgery
Device Facts
Record ID
K210183
Device Name
Waterlase iPLus
Applicant
Biolase, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Waterlase iPlus with the fractional handpiece is indicated for use in dermatology for skin resurfacing. Waterlase iPlus when used with the non-fractional handpieces is indicated for use in the following oral hard and soft tissue dental applications: Waterlase laser removal of porcelain and ceramic crowns and veneers General Hard Tissue Indications (for use on adult and pediatric patients) - Class I, II, III, IV and V cavity preparation - Caries removal - Hard tissue surface roughening or etching - Enameloplasty, excavation of pits and fissures for placement of sealants Root Canal Hard Tissue Indications - Tooth preparation to obtain access to root canal - Root canal preparation including enlargement - Root canal debridement and cleaning Root Canal Disinfection - Laser root canal disinfection after endodontic treatment Endodontic Surgery (Root Amputation) Indications - Flap preparation incision of soft tissue to prepare a flap and expose the bone - Cutting bone to prepare a window access to the apex (apices) of the root(s) - Apicoectomy amputation of the root end - Root end preparation for retrofill amalgam or composite - Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation. Bone Surgical Indications - Cutting, shaving, contouring and resection of oral osseous tissues (bone) - Osteotomy Soft Tissue Indications including Pulpal Tissues (for use on adult and pediatric patient) Incision, excision, vaporization, ablation and coagulation of oral soft tissues including: - Excisional and incisional biopsies - Exposure of unerupted teeth - Fibroma removal - Flap preparation incision of soft tissue to prepare a flap and expose the bone - Flap preparation incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue impactions) - Frenectomy and frenotomy - Gingival troughing for crown impressions - Gingivectomy - Gingivoplasty - Gingival incision and excision - Hemostasis - Implant recovery - Incision and drainage of abscesses - Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during apical surgery - Leukoplakia - Operculectomy - Oral papillectomies - Pulpotomy - Pulp extirpation - Pulpotomy as an adjunct to root canal therapy - Root canal debridement and cleaning - Reduction of gingival hypertrophy - Soft tissue crown lengthening - Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa - Vestibuloplasty - Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation. Laser Periodontal Procedures - Full thickness flap - Partial thickness flap - Split thickness flap - Laser soft tissue curettage - Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket - Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining junctional epithelium - Removal of granulation tissue from bony defects - Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival bleeding index, probe depth, attachment loss and tooth mobility) - Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours) - Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.) - Osseous crown lengthening - Removal of subgingival calculi in periodontal pockets with periodontitis by closed or open curettage - Waterlase Er,Cr: Y SGG assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium.
Device Story
Waterlase iPlus is an Er,Cr:YSGG (2780nm) solid-state laser system; utilizes water atomization for tissue cooling/hydration. System comprises console (laser head, power supply, cooling, microprocessor), footswitch, and fiber optic delivery cable with interchangeable handpieces. Fractional handpiece enables dermatological skin resurfacing; non-fractional handpieces support dental hard/soft tissue applications (incision, excision, ablation, vaporization, coagulation). Operated by clinicians in dental/dermatology settings. User interface via touchscreen display. Laser energy delivered to target site; water spray provides cooling. Device facilitates precise tissue removal, bone contouring, and periodontal procedures; benefits include reduced thermal damage and improved clinical indices for periodontal health.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (verification of fractional handpiece/fiber designs) and ex-vivo/in-vivo animal tissue testing to assess penetration depth/width, confirming performance within range of cleared fractional lasers.
Technological Characteristics
Er,Cr:YSGG solid-state laser; 2780nm wavelength; pulsed mode. Console-based with fiber optic delivery and water atomization. Fractional handpiece accessory. Complies with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 80601-2-60, and IEC 60601-1-2. Software conforms to IEC 62304.
Indications for Use
Indicated for adult and pediatric patients requiring oral hard/soft tissue dental procedures (cavity prep, endodontics, periodontics, bone surgery) and dermatology patients requiring skin resurfacing.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Pearl Fractional Handpiece, Cutera, Inc. (K080530)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 21, 2022
Biolase, Inc. Ed Balcos Senior Regulatory Affairs Specialist 27042 Towne Centre Drive, Suite 270 FootHill Ranch, California 92610
Re: K210183
Trade/Device Name: Waterlase iPLus Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, ONG Dated: January 22, 2021 Received: January 25, 2021
Dear Ed Balcos:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known) K210183
Device Name Waterlase iPlus Laser System
#### Indications for Use (Describe)
Waterlase iPlus with the fractional handpiece is indicated for use in dermatology for skin resurfacing.
Waterlase iPlus when used with the non-fractional handpieces is indicated for use in the following oral hard and soft tissue dental applications:
Waterlase laser removal of porcelain and ceramic crowns and veneers
General Hard Tissue Indications (for use on adult and pediatric patients)
- · Class I, II, III, IV and V cavity preparation
- · Caries removal
- Hard tissue surface roughening or etching
- · Enameloplasty, excavation of pits and fissures for placement of sealants
#### Root Canal Hard Tissue Indications
- · Tooth preparation to obtain access to root canal
- · Root canal preparation including enlargement
- Root canal debridement and cleaning
Root Canal Disinfection
- · Laser root canal disinfection after endodontic treatment
Endodontic Surgery (Root Amputation) Indications
- · Flap preparation incision of soft tissue to prepare a flap and expose the bone
- · Cutting bone to prepare a window access to the apex (apices) of the root(s)
- Apicoectomy amputation of the root end
- · Root end preparation for retrofill amalgam or composite
· Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
Bone Surgical Indications
- · Cutting, shaving, contouring and resection of oral osseous tissues (bone)
- · Osteotomy
Soft Tissue Indications including Pulpal Tissues (for use on adult and pediatric patient) Incision, excision, vaporization, ablation and coagulation of oral soft tissues including:
- · Excisional and incisional biopsies
- Exposure of unerupted teeth
- Fibroma removal
- · Flap preparation incision of soft tissue to prepare a flap and expose the bone
- · Flap preparation incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue impactions)
{4}------------------------------------------------
- · Frenectomy and frenotomy
- · Gingival troughing for crown impressions
- · Gingivectomy
- Gingivoplasty
- Gingival incision and excision
- Hemostasis
- · Implant recovery
- Incision and drainage of abscesses
- · Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during apical surgery
- Leukoplakia
- Operculectomy
- Oral papillectomies
- Pulpotomy
- Pulp extirpation
- · Pulpotomy as an adjunct to root canal therapy
- Root canal debridement and cleaning
- Reduction of gingival hypertrophy
- · Soft tissue crown lengthening
- · Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa
- · Vestibuloplasty
· Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
Laser Periodontal Procedures
- Full thickness flap
- Partial thickness flap
- · Split thickness flap
- · Laser soft tissue curettage
- · Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket
- Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining junctional epithelium
- · Removal of granulation tissue from bony defects
- · Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival bleeding index, probe depth, attachment loss and tooth mobility)
· Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours)
- · Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.)
- · Osseous crown lengthening
- · Removal of subgingival calculi in periodontal pockets with periodontitis by closed or open curettage
• Waterlase Er.Cr: Y SGG assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
K210183
# OLASF
#### 510(k) SUMMARY
#### I. SUBMITTER
Biolase,Inc. 27042 Towne Centre Drive Suite 270 Foothill Ranch, CA 92610 Tel: (949) 226-8119 Fax: (949) 273-6688 Contact Person: Ed Balcos Tel: (949) 361-1200 ebalcos@biolase.com Date: March 21, 2022
#### II. DEVICE
| Name of Device: | Waterlase iPlus Laser System, Fractional Handpiece |
|----------------------|-------------------------------------------------------------------------------------------------------|
| Common Name: | Er, Cr YSGG Laser |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery and in dermatology (21 CFR 878.4810) |
| Device Class: | II |
| Product Code: | GEX, ONG |
#### III. PREDICATE DEVICES
Pearl Fractional Handpiece, Cutera, Inc., K080530 FS-01 Handpiece, Fotona, d.d., K132806 REFERENCE DEVICE Waterlase iPlus S, Biolase, Inc., K141975
#### IV. DEVICE DESCRIPTION
Waterlase iPlus utilizes an Er, Cr YSGG (2780nm) solid-state laser and water atomization technology for hard and soft tissue incision, excision, ablation, vaporization and coagulation of soft and hard tissue.
The laser consists of a console the houses the laser head, power supply, cooling system, micro-processor and electronics, a footswitch which activates the laser, and a fiber optic cable for delivery of laser energy through a handpiece to the treatment site. A fine water spray is also emitted from the handpiece to cool and hydrate the issue. The laser is controlled through a touch screen display which serves as the User Interface.
{7}------------------------------------------------
# BOLASE
#### V. INDICATIONS FOR USE
Waterlase iPlus with the fractional handpiece is indicated for use in dermatology for skin resurfacing.
#### Waterlase iPlus when used with non-fractional handpieces is indicated for use in the following oral hard and soft tissue dental applications:
Waterlase laser removal of porcelain and ceramic crowns and veneers
General Hard Tissue Indications (for use on adult and pediatric patients)
- Class I, II, III, IV and V cavity preparation ●
- Caries removal o
- Hard tissue surface roughening or etching o
- Enameloplasty, excavation of pits and fissures for placement of sealants ●
#### Root Canal Hard Tissue Indications
- Tooth preparation to obtain access to root canal
- Root canal preparation including enlargement ●
- Root canal debridement and cleaning ●
### Root Canal Disinfection
- Laser root canal disinfection after endodontic treatment 0
#### Endodontic Surgery (Root Amputation) Indications
- Flap preparation incision of soft tissue to prepare a flap and expose the bone ●
- Cutting bone to prepare a window access to the apex (apices) of the root(s)
- Apicoectomy amputation of the root end o
- Root end preparation for retrofill amalgam or composite ●
- Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic ● tissues (i.e., granulation tissue) from around the apex
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
#### Bone Surgical Indications
- Cutting, shaving, contouring and resection of oral osseous tissues (bone) ●
- Osteotomy ●
## Soft Tissue Indications including Pulpal Tissues (for use on adult and pediatric patient)
- Incision, excision, vaporization, ablation and coagulation of oral soft tissues including:
- Excisional and incisional biopsies ●
- Exposure of unerupted teeth
- Fibroma removal
- Flap preparation incision of soft tissue to prepare a flap and expose the bone ●
{8}------------------------------------------------
## K210183
# BIOLASE
- Flap preparation incision of soft tissue to prepare a flap and expose unerupted teeth ● (hard and soft tissue impactions)
- Frenectomy and frenotomy
- Gingival troughing for crown impressions
- Gingivectomy ●
- Gingivoplasty
- Gingival incision and excision
- Hemostasis
- Implant recovery ●
- Incision and drainage of abscesses
- Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during apical surgery
- Leukoplakia
- Operculectomy ●
- Oral papillectomies
- Pulpotomy
- Pulp extirpation ●
- Pulpotomy as an adjunct to root canal therapy
- Root canal debridement and cleaning ●
- Reduction of gingival hypertrophy
- Soft tissue crown lengthening ●
- Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa ●
- Vestibuloplastv ●
- Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic . tissues (i.e., granulation tissue) from around the apex
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
### Laser Periodontal Procedures
- Full thickness flap ●
- Partial thickness flap
- Split thickness flap
- Laser soft tissue curettage ●
- Laser removal of diseased, inflamed and necrosed soft tissue within the ● periodontal pocket
- o Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining junctional epithelium
- Removal of granulation tissue from bony defects ●
- Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
- Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours)
- Ostectomy (resection of bone to restore bony architecture, resection of bone for ●
{9}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the word "BIOLASE" in large, bold, blue letters. The letters are closely spaced together, filling most of the frame. The background is white.
grafting, etc.)
- Osseous crown lengthening
- o Removal of subgingival calculi in periodontal pockets with periodontitis by closed or open curettage
- . Waterlase Er,Cr:YSGG assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
| | Subject Device | Predicate<br>(primary) | Predicate<br>(secondary) | Reference<br>Device |
|-------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------|----------------------------------------|---------------------|
| Specification | Biolase Fractional<br>Handpiece | Cutera Pearl<br>Fractional<br>Handpiece | Fotona FS01<br>Fractional<br>Handpiece | Biolase iPlus S |
| 510(k) Number | K210183 | K080530 | K132806 | K141975 |
| Laser medium | Er,Cr YSGG | Er YSGG | Er YAG | Er,Cr YSGG |
| Wavelength | 2780 nm | 2790 nm | 2940 nm | 2780 nm |
| Operating modes | pulsed | pulsed | pulsed | pulsed |
| Pulse width, us | 60 µs, 700 µs | < 1000 µs | 100 us, 300 us | 60 μs, 700 μs |
| Max Rep Rate | 15 Hz | 20 Hz | 50 Hz | 100 Hz |
| Spot size | 250 µm | 300 µm | 250 µm | 200 – 1200μm |
| Spot spacing | 900 µm in x<br>direction, 1 mm in y<br>direction (based on<br>1.5 cm/sec speed of<br>movement) | 0.6 mm (based<br>on 20%<br>density) | 1.1 mm | N/A |
| Max energy per<br>spot | 40 mJ | 200 mJ | 22 mJ | 600 mJ |
| Max fluence per<br>spot | 82 J/cm2 | 280 J/cm2 | 45 J/cm2 | 425 J/cm2 |
{10}------------------------------------------------
# BIOLASE
### VII. PERFORMANCE DATA
The following non-clinical and clinical performance information was provided in support of substantial equivalence determination:
#### Biocompatibility Testing
The biocompatibility evaluation for Waterlase iPlus Laser System was conducted in accordance with ISO 10993-1 Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing within a Risk Management Process, as recognized by the FDA. The addition of the Fraction Handpiece accessory cytotoxicity testing was completed on the only patient contacting disposable accessory applicator.
#### Software Verification and Validation
Software verification and validation were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices The results demonstrate that the Waterlase iPlus Laser System performs according to specifications and functions intended. The software design conforms with the IEC 62304 standard -Medical device software- software life cycle processes.
### Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted for the Waterlase iPlus Laser System. The device complies with the current revisions of the recognized standards IEC 60601-1, IEC 60601-2-22, IEC 60825-1, and IEC 80601-2-60 for safety and IEC 60601-1-2 for EMC.
#### Performance Testing- Bench
In an effort to ensure adding this accessory handpiece to the iPlus Laser System as well as the dermatological indication was as safe and effective as the predicate devices performance bench testing was completed. A verification of the Fractional Handpiece and trunk fiber designs was completed to gather data for peak performance settings. Verification of accessory requirements was also completed, and the accessory passed all requirement criteria.
### Performance Testing - Animal
Ex-vivo and in-vivo animal tissue testing was conducted to assess the depth and width of penetration produced with the fractional handpiece. The results support that the device can be used to produce fractional beam micro-ablation depths and widths that are within the range of other fractional laser devices cleared for dermatological skin resurfacing.
{11}------------------------------------------------
## BIOLASE
## VIII. CONCLUSION
The performance data provided in this submission supports that the device can be used safely and effectively and as intended. The proposed device's fractional handpiece's technology does not raise new types of questions regarding safety and efficacy for it's indications for use when compared to the predicate devices. Based on the above, it can be concluded that the Waterlase iPlus Laser System with Fractional Handpiece is found to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.