← Product Code [HET](/productcode/HET) · K210116

# Video Endoscopy System, 3D Video Endoscopy System (K210116)

_Surgnova Healthcare Technologies (Zhejiang) Co., Ltd. · HET · Oct 13, 2021 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210116

## Device Facts

- **Applicant:** Surgnova Healthcare Technologies (Zhejiang) Co., Ltd.
- **Product Code:** [HET](/productcode/HET.md)
- **Decision Date:** Oct 13, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1720
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Video Endoscopy System & 3D Video Endoscopy System are intended to be used to provide illumination and visualization of surgical field in a wide variety of diagnostic and therapeutic abdominal and thoracic minimally invasive procedures, including female reproductive organs (gynecology) and urological anatomy.

## Device Story

Video Endoscopy System and 3D Video Endoscopy System provide surgical field illumination and visualization for minimally invasive abdominal and thoracic procedures. The system consists of a video endoscope (with distal CMOS sensors) and a video endoscopy processor. The 2D system (EVS100 processor, LPS21000/LPS21030 endoscopes) uses one sensor; the 3D system (EVS200 processor, LPS31000/LPS31030 endoscopes) uses two sensors to output 2D/3D images. Unlike the predicate, which requires an external light source, this device features built-in white LEDs at the distal end for illumination. The processor converts electrical signals from the CMOS sensors into video signals for display on a monitor. Used in hospital operating rooms by surgeons to guide endoscopic surgery. The device benefits patients by providing high-resolution visualization of internal anatomy during minimally invasive procedures.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, including system performance verification (resolution, brightness, white balance, 3D-2D mode, color performance, field of view, geometric distortion, SNR, dynamic range, image intensity uniformity), photobiological safety (IEC 62471), biocompatibility (ISO 10993), and sterilization validation.

## Technological Characteristics

System includes video processor and rigid endoscopes (0°/30°). Illumination: built-in white LEDs. Imaging: distal CMOS sensors. Resolution: 1920x1080 HD. Connectivity: processor to monitor. Sterilization: endoscopes are terminally sterilized by users. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 10993 series, IEC 62471.

## Regulatory Identification

A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.

## Predicate Devices

- SPIES 3D System ([K150525](/device/K150525.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 13, 2021

Surgnova Healthcare Technologies (Zhejiang) Co., Ltd. Guofang MA QARA Director No.1 Xinxing Yilu Road, Emerging Industrial Cluster Area, Zonghan Subdistrict Cixi. Zhejiang 315300 China

Re: K210116

Trade/Device Name: Video Endoscopy System & 3D Video Endoscopy System Regulation Number: 21 CFR§ 884.1720 Regulation Name: Gynecologic Laparoscope and Accessories Regulatory Class: II Product Code: HET, GCJ, FGB Dated: August 31, 2021 Received: September 7, 2021

Dear Guofang MA:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210116

#### Device Name

Video Endoscopy System &3D Video Endoscopy System

Indications for Use (Describe)

The Video Endoscopy System &3D Video Endoscopy System are intended to be used to provide illumination and visualization of surgical field in a wide variety of diagnostic abdominal and thoracic minimally invasive procedures, including female reproductive organs (gynecology) and urological anatomy.

| Type of Use (Select one or both, as applicable)                                         |  |
|-----------------------------------------------------------------------------------------|--|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |  |
| <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section III 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92.

## 3.1 Submitter Information

#### 510(k) Submitter/Holder: ●

Surgnova Healthcare Technologies (Zhejiang) Co., Ltd. No.1 Xinxing Yilu Road, Emerging Industrial Cluster Area, Zonghan Subdistrict, Cixi City, Zhejiang, China

#### ● Contact

Guofang Ma QARA Director Telephone: +86-10-64116115 Fax: +86-10-64113261 Email: QARA@surgnova.com

#### ● Date Prepared: October 12, 2021

#### 3.2 Device Information

Trade/Proprietary Name: Video Endoscopy System & 3D Video Endoscopy System Common Name: Video Endoscopy System Models: EVS100, LPS21000, LPS21030, EVS200, LPS31000, LPS31030 Classification Name: Gynecologic laparoscope and accessories Classification Regulation: 21 CFR 884.1720 Product Code: HET, GCJ, FGB Device Class: Class II

#### 3.3 Predicate Device

510(k) Number: K150525 Trade/Device Name: SPIES 3D System Regulation Number: 21 CFR 884.1720 Regulation Name: Gynecologic Laparoscope and Accessories Regulatory Class: Class II Product Code: HET, GCJ, FGB Manufacturer: Karl Storz Endoscopy-America, Inc.

{4}------------------------------------------------

Premarket Notification 510(k) Submission-Section III 510(k) Summary

|                                                          | Proposed device                                              | Predicate device<br>(K150525) |
|----------------------------------------------------------|--------------------------------------------------------------|-------------------------------|
| Video Endoscopy<br>System / 3D Video<br>Endoscopy System | Video Endoscopy Processor,<br>EVS100, EVS200                 | IMAGE1 SPIES                  |
|                                                          | Video Endoscope,<br>LPS21000, LPS21030<br>LPS31000, LPS31030 | 3D TIPCAM®1                   |

The predicate device has not been subject to a design related recall.

# 3.4 Device Description

The proposed Video Endoscopy Systems include the Video Endoscopy System and 3D Video Endoscopy System, Video Endoscopy System supports 2D imaqe output, 3D Video Endoscopy System supports 2D/3D image output.

Video Endoscopy System is composed of Video Endoscope (LPS21000/LPS21030) and Video Endoscopy processor (EVS100).

| Name                      |                              | Model    |
|---------------------------|------------------------------|----------|
| Video Endoscopy<br>System | Video Endoscopy<br>Processor | EVS100   |
|                           | 0° Video Endoscope           | LPS21000 |
|                           | 30° Video Endoscope          | LPS21030 |

The Video Endoscopy Processor is a video processor, which receives the electrical signals from the Video Endoscope and process it and output the final image to the monitor. Two models 2D Video Endoscope (LPS21000/LPS21030) are available, one image sensor is located at the distal of the endoscope (LPS21000/LPS21030), they are used in conjunction with Video Endoscopy Processor (EVS100) to output 2D images.

3D Video Endoscopy System is composed of Video Endoscope (LPS31000/LPS31030) and Video Endoscopy Processor (EVS200).

| Name                         | Model                        |          |
|------------------------------|------------------------------|----------|
| 3D Video<br>Endoscopy System | Video Endoscopy<br>Processor | EVS200   |
|                              | 0° Video Endoscope           | LPS31000 |
|                              | 30° Video Endoscope          | LPS31030 |

The Video Endoscopy Processor is a video processor, which receives the electrical signals from the Video Endoscope and process it and output the final image to the monitor. Two models 3D Video Endoscope (LPS31000/LPS31030) are available, two image sensors are located at the distal of the endoscope (LPS31000/LPS31030), they are used in conjunction with Video Endoscopy Processor (EVS200) to output 2D/3D images.

{5}------------------------------------------------

Premarket Notification 510(k) Submission—Section III 510(k) Summary

Video Endoscopy processor is non-sterile device. The Video Endoscope is terminallysterilized device. The Video Endoscope must be sterilized by users before being used in surgery.

This device is intended to be used in the hospital environment, such as surgical operation room.

## 3.5 Indications for use

The Video Endoscopy System & 3D Video Endoscopy System are intended to be used to provide illumination and visualization of surgical field in a wide variety of diagnostic and therapeutic abdominal and thoracic minimally invasive procedures, including female reproductive organs (gynecology) and urological anatomy.

#### 3.6 Comparison of the technological characteristics

| Comparison<br>Items             | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Remark    |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| product Name                    | SPIES 3D System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Video Endoscopy System &<br>3D Video Endoscopy System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | /         |
| Regulation No.                  | 21 CFR 884.1720                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 21 CFR 884.1720                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same      |
| Classification                  | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same      |
| Product Code                    | HET, GCJ, FGB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | HET, GCJ, FGB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same      |
| Indications for<br>Use          | 3D TIPCAM®1: The Rigid<br>Video endoscope is intended to<br>be used together with the<br>camera control unit is for use<br>during diagnostic and/or<br>surgical procedures when<br>endoscopic video assistance is<br>required. For use in all<br>endoscopy and endoscopic<br>surgery within the peritoneal<br>and thoracic cavity, including<br>gynecological and urological<br>anatomy.<br>IMAGE1 SPIES is a camera<br>control unit (CCU) for use with<br>camera heads or video<br>endoscopes for the                                                                                                                                                                                                                                               | The Video Endoscopy System<br>& 3D Video Endoscopy System<br>are intended to be used to<br>provide illumination and<br>visualization of surgical field in<br>a wide variety of diagnostic and<br>therapeutic abdominal and<br>thoracic minimally invasive<br>procedures, including female<br>reproductive organs<br>(gynecology) and urological<br>anatomy.                                                                                                                                                                                                                                                                                                                          | Same      |
|                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |           |
| Design                          | visualization<br>and<br>documentation of endoscopic<br>and microscopic procedures.<br><br>The 3D TIPCAM®1 is connect<br>with IMAGE1 SPIES (CCU)<br>and external light source for<br>uniform illumination of the<br>surgical field, a high-resolution<br>optical system captures the<br>signals from the surgical field<br>and converged it into CMOS.<br>The CMOS converts optical<br>signals into electrical signals,<br>electrical signals are<br>transmitted through electronic<br>devices and converted into<br>video signals, after image<br>processing, video signal is<br>displayed on the monitor, and<br>endoscopic surgery is<br>performed under the guidance<br>of images displayed on a video<br>monitor.<br>Need external light source. | The distal end of the<br>endoscope is equipped with<br>white LEDs for uniform<br>illumination of the surgical field,<br>a high-resolution optical system<br>captures the signals from the<br>surgical field and converged it<br>into CMOS. The endoscopic<br>front-end CMOS converts<br>optical signals into electrical<br>signals, electrical signals are<br>transmitted through electronic<br>devices and converted into<br>video signals, after image<br>processing, video signal is<br>displayed on the monitor, and<br>endoscopic surgery is<br>performed under the guidance<br>of images displayed on a video<br>monitor.<br>Built-in LED light source and<br>heat dissipation | Different |
| Resolution                      | 1920X1080 HD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 1920X1080 HD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same      |
| Direction<br>of<br>view         | 0°, 30°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 0°, 30°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same      |
| Field angle                     | 90°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 90°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same      |
| Depth of field                  | 20 to 200mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 20 to 120mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same      |
| Outer diameter<br>of distal end | 10mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | LPS21000/LPS21030:10mm<br>LPS31000/LPS31030:10.2mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | /         |
| Working length                  | 317mm, 320mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | LPS21000/ LPS21030:<br>345.5mm, 348.2mm<br>LPS31000/ LPS31030:<br>338mm, 344mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Different |

Table 1 Comparison of subject and predicate device technological characteristics

{6}------------------------------------------------

Premarket Notification 510(k) Submission—Section III 510(k) Summary

{7}------------------------------------------------

Premarket Notification 510(k) Submission—Section III 510(k) Summary

Compared with predicate, the proposed devices use built-in LEDs light source instead of external light source and have different working length. The non-clinical testing demonstrates that the differences in technological characteristics between the subject and predicate do not raise different questions of safety and effectiveness.

## 3.7 Testing

## Non-Clinical Testing

The Video Endoscopy System and the predicate device are substantially equivalent in design concepts, technologies and materials. The Video Endoscopy System has been designed and tested and will be manufactured in accordance with both mandatory and voluntary standards, including:

Video Endoscopy System is tested according to the following standards:

- IEC 60601-1:2005, MOD ●
- IEC 60601-1-2:2014 .
- . IEC 60601-2-18:2009
- ISO 10993-5, 2009 (Cytotoxicity) ●
- . ISO 10993-11, 2010 (Acute Systemic Toxicity & Pyrogen)
- ISO 10993-10, 2010 (Sensitization and Irritation) ●

Additional bench testing for performance verification and validation purposes:

- Resolution ●
- Brightness
- . White Balance
- 3D-2D Mode
- . Color Performance
- . Field of View
- . Geometric Distortion
- . Signal-to-Noise Ratio (SNR) and Dynamic Range
- Image Intensity Uniformity (IIU)
- . Photobiological Safety per IEC 62471:2006

The Software Validation is in compliance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

Biocompatibility testing supports the device is not toxic, irritating or sensitizing. The device met all acceptance criteria for optical performance testing and was shown to have equivalent image quality to the predicate.

The list of non-clinical test performed on the proposed device.

Test Name No.

{8}------------------------------------------------

Premarket Notification 510(k) Submission-Section III 510(k) Summary

| 1 | System Performance Verification Test                             |
|---|------------------------------------------------------------------|
| 2 | Package Verification Test according to ISTA 2A-11&ASTM D 4169-16 |
| 3 | Sterilization validation according to ISO 17665-1                |

#### Clinical Testing

Clinical studies were not required to demonstrate the substantial equivalence of the Video Endoscopy System and the predicated device.

#### 3.8 Determination of substantial equivalence

The subject and predicate devices are both indicated for endoscopic observation during diagnostic and/or therapeutic procedures within the abdominal and thoracic cavities, including gynecological and urological anatomy; there is no difference in the intended use. The differences in technological characteristics between the subject and predicate do not raise different questions of safety and effectiveness. The non-clinical testing demonstrates that the subject device is as safe and effective as the predicate. Therefore, the subject device is substantially equivalent to the predicate.

---

**Source:** [https://fda.innolitics.com/device/K210116](https://fda.innolitics.com/device/K210116)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
