← Product Code [MAX](/productcode/MAX) · K210090

# Impulse AM Interbody Fusion System (K210090)

_Degen Medical · MAX · Aug 30, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K210090

## Device Facts

- **Applicant:** Degen Medical
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Aug 30, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The DeGen Medical Impulse AM™ implant is an intervertebral body fusion device intended for use in skeletally mature patients with Degenerative Disk Disease (DDD) of the lumbar spine with up to Grade 1 Spondylolisthesis at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. The DeGen Medical Impulse AM™ System is indicated to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and are intended to be used with supplemental fixation systems cleared for use in the lumbar spine. The device is to be used in patients who have had six months of nonoperative treatment.

## Device Story

Impulse AM™ Interbody Fusion System; lumbar interbody fusion device for posterior/transforaminal procedures. Implant features porous superior/inferior endplates to mitigate subsidence/expulsion; central aperture for bone graft containment; threaded holes/slots for rigid instrument connection. Available in various lengths, widths, heights, and endplate curvatures (neutral, lordotic, anatomic, hyperlordotic). Surgeon inserts device into lumbar disc space to facilitate fusion; used with autograft/allograft and supplemental fixation. Additively manufactured titanium construction. Provides structural support to disc space; promotes spinal fusion; benefits patients with discogenic back pain.

## Clinical Evidence

Bench testing only. Performed static and dynamic compression testing (ASTM F2077), static and dynamic compression shear testing (ASTM F2077), subsidence testing (ASTM F2267), and expulsion testing. No clinical data presented.

## Technological Characteristics

Additively manufactured from Puri-Ti™ titanium powder (ASTM B348). Features porous surfaces for osseointegration. Includes various endplate curvatures (neutral, lordotic, anatomic, hyperlordotic). Designed for posterior/transforaminal lumbar interbody fusion. Non-active, mechanical implant.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- DeGen Impulse™ Spacer ([K201287](/device/K201287.md))
- Alphatec Spine IdentiTi ([K183705](/device/K183705.md))
- Choice Spine TiGer Shark ([K172816](/device/K172816.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 30, 2021

DeGen Medical % Linda Braddon, Ph.D. President/CEO Secure BioMed Evaluations 7828 Hickory Flat Highway Suite 120 Woodstock, Georgia 30188

Re: K210090

Trade/Device Name: Impulse AM™ Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: July 30, 2021 Received: August 2, 2021

Dear Dr. Braddon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Brent L. Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for DeGen Medical. The logo features three curved lines emanating from three blue circles at the top, resembling a stylized representation of motion or connectivity. Below this graphic is the text "DeGen" in a bold, sans-serif font, with the word "Medical" in a smaller, lighter font underneath.

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known) K210090

Device Name

DeGen Medical Impulse AM™ Interbody Fusion System

Indications for Use (Describe)

The DeGen Medical Impulse AM™ implant is an intervertebral body fusion device intended for use in skeletally mature patients with Degenerative Disk Disease (DDD) of the lumbar spine with up to Grade 1 Spondylolisthesis at one or two contiquous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. The DeGen Medical Impulse AM™ System is indicated to be used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion and are intended to be used with supplemental fixation systems cleared for use in the lumbar spine. The device is to be used in patients who have had six months of nonoperative treatment.

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

| This section applies only to requirements of the Paperwork Reduction Act of 1995.<br>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*                                                                   |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete |
| and review the collection of information. Send comments regarding this burden estimate or any other aspect                                                                                                                     |
| of this information collection, including suggestions for reducing this burden, to:<br>Department of Health and Human Services                                                                                                 |
| Food and Drug Administration                                                                                                                                                                                                   |
| Office of Chief Information Officer                                                                                                                                                                                            |
| Paperwork Reduction Act (PRA) Staff                                                                                                                                                                                            |
| PRAStaff@fda.hhs.gov                                                                                                                                                                                                           |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of                                                                                                                             |
| information unless it displays a currently valid OMB number "                                                                                                                                                                  |

FORM FDA 3881 (6/20)

Page 1 of 1

PSC Publishing Services (301)443-6740 EP

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for DeGen Medical. The logo features three curved lines emanating from a point on the left, each ending in a blue circle. The lines are also blue and appear to be in motion. Below the graphic is the text "DeGen" in a bold, blue font, with the word "Medical" in a smaller, gray font underneath.

## 510(k) Summary K210090

In accordance with 21 CFR 807.87 (h) and 21 CRF 807.92, the 510(k) summary for the DeGen Medical Impulse AM™ System is provided below.

| Date                             | August 27, 2021                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor                          | DeGen Medical<br>1321-C North Cashua Drive<br>Florence, SC 29501<br>Phone 877-240-7838<br>Fax 843-407-0545                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| 510(k) Contact                   | Secure BioMed Evaluations<br>Linda Braddon, Ph.D.<br>7828 Hickory Flat Highway<br>Suite 120<br>Woodstock, GA 30188<br>770-837-2681<br>Regulatory@SecureBME.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Trade Name                       | Impulse AM™ Interbody Fusion System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Common Name                      | Intervertebral body fusion device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Code-                            | MAX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification                   | 21 CFR 888.3080 : Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Primary Predicate                | DeGen Impulse™ Spacer K201287                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Additional<br>Predicates         | Alphatec Spine IdentiTi K183705<br>Choice Spine TiGer Shark K172816                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Device Description               | DeGen Medical Impulse AM™ System is a lumbar interbody<br>fusion device for posterior and transforaminal lumbar fusion<br>procedures. The Impulse AM™ system includes various lengths,<br>widths, heights, and endplate shapes (curvatures include neutral,<br>lordotic, and anatomic). The DeGen Medical Impulse AM™<br>device is comprised of a single component that is additively<br>manufactured. The superior and inferior endplates feature<br>porous surfaces to mitigate subsidence and expulsion. The<br>anatomic, lordotic, and neutral configurations feature a posterior<br>face with a threaded hole and slots to rigidly connect to an<br>instrument for surgical insertion. The hyperlordotic configuration<br>features a posterior face with slots to rigidly connect an instrument<br>for surgical insertion. Superior and inferior faces feature a central<br>aperture to constrain bone graft. DeGen Medical's Impulse AM™<br>Spacers are additively manufactured from Puri-Ti™ titanium<br>powder having a chemical composition conforming to ASTM<br>B348 |
| Indications for Use              | The DeGen Medical Impulse AM™ implant is an intervertebral<br>body fusion device intended for use in skeletally mature patients<br>with Degenerative Disk Disease (DDD) of the lumbar spine with<br>up to Grade 1 Spondylolisthesis at one or two contiguous levels<br>from L2-S1. DDD is defined as discogenic back pain with<br>degeneration of the disc confirmed by history and radiographic<br>studies. The DeGen Medical Impulse AM™ System is indicated<br>to be used with autograft and/or allograft comprised of cancellous<br>and/or corticocancellous bone graft to facilitate fusion and are<br>intended to be used with supplemental fixation systems cleared<br>for use in the lumbar spine. The device is to be used in patients<br>who have had six months of nonoperative treatment.                                                                                                                                                                                                                                                                           |
| Technological<br>Characteristics | The technological design features of the subject implants were<br>compared to the predicates in intended use, indications for use,<br>design, function and technology and it was demonstrated that<br>they are substantially equivalent.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Performance<br>Testing           | Non-clinical testing was performed to demonstrate the DeGen<br>Medical Impulse AM™ System is substantially equivalent to<br>other predicate devices in accordance with “Guidance for<br>Industry and FDA Staff, Guidance for Spinal System 510(k)s",<br>May 3, 2004 and Class II Special Controls Guidance<br>Document: Intervertebral Body Fusion Device, June 12, 2007.<br>The following tests were performed:<br>• Static and dynamic compression testing per ASTM<br>F2077<br>• Static and dynamic compression shear testing per<br>ASTM F2077<br>• Subsidence testing via ASTM F2267<br>• Expulsion Testing<br>The results of these studies show the subject DeGen Medical<br>Impulse AM™ System is substantially equivalent to the other<br>predicate devices.                                                                                                                                                                                                                                                                                                            |
| Conclusions                      | Based on the indications for use, technological characteristics,<br>performance testing, and comparison to predicate devices, the<br>subject DeGen Medical Impulse AM™ System is as safe and<br>as effective as the legally marketed predicates.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for DeGen Medical. The logo features three curved lines emanating from three circles, resembling a stylized representation of motion or connectivity. Below the graphic is the text "DeGen" in a bold, blue font, with the word "Medical" in a smaller, gray font underneath.

## 510(k) Summary K210090

---

**Source:** [https://fda.innolitics.com/device/K210090](https://fda.innolitics.com/device/K210090)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
