K210084 · Hironic Co., Ltd. · GEI · Jun 3, 2022 · General, Plastic Surgery
Device Facts
Record ID
K210084
Device Name
SILKRO
Applicant
Hironic Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended for use in dermatologic and general surgical procedures for electro-coagulation.
Device Story
SILKRO is an electrosurgical system for tissue coagulation using high-frequency (RF) current. It comprises a main unit, LCD touchscreen, foot switch, and four handpieces (RM, RN, RC, RV). Operates in monopolar mode (RF energy between active electrode tip and neutral pad) or bipolar mode (RF energy between adjacent needles in electrode tip). User selects mode and adjusts parameters (intensity, duration, depth) via touchscreen. Handpieces are held at right angles to target tissue; RF energy generates heat via contact resistance, causing coagulation. Used in clinical settings by healthcare professionals. Benefits include controlled thermal coagulation for dermatologic and surgical applications.
Clinical Evidence
No clinical data. Evidence consists of bench testing (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-1-6, IEC 62304, IEC 62366, ISO 14971) and ex vivo animal testing. Animal study evaluated histological depth and zone of coagulation/thermal damage at various power levels, confirming thermal effect on tissue.
Technological Characteristics
Electrosurgical unit, 2MHz RF energy, 25W max power. Materials evaluated per ISO 10993-1. Sterilization via EO gas (ISO 11135). Shelf life (3 years) validated per ASTM F1980-16, F1929-15, F88/F88M. Software validated per IEC 62304 (moderate level of concern).
Indications for Use
Indicated for dermatologic and general surgical procedures requiring electro-coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
June 3, 2022
Hironic Co., Ltd % Sang Hwa Myung Regulatory Affair Consultant E&M D-1474, 230, Simin-daero, Dongan-gu Anyang-si, Gyeonggi-do 14067 Republic of Korea
Re: K210084
Trade/Device Name: SILKRO Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 7, 2022 Received: January 10, 2022
Dear Sang Hwa Myung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation -emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
Food and Drug Administration Indications for Use
510(k) Number (if known) K210084
Device Name SILKRO
Indications for Use (Describe)
This device is intended for use in dermatologic and general surgical procedures for electro-coagulation.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for HIRONIC. The logo consists of a red square with a white arrow pointing to the right, followed by the word "HIRONIC" in black, sans-serif letters. The logo is simple and modern, and the red and black colors give it a bold and professional look.
## 510(k) Summary
## This summary of 510(k) Safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
# K210084
1. Submitter Information - 807.92(a)(1)
| Applicant | Hironic Co., Ltd. |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | 19F, 767, Sinsu-Ro, Suji-Gu, Yongin-Si, Gyeonggi-do, 16827,<br>Republic of Korea |
| Phone Number | +82-31-525-7000 |
| Fax Number | +82-31-525-7010 |
| Contact Person | Sang Hwa, Myung |
| Contact Information | D-1474, 230, Simin-daero, Dongan-gu, Anyang-si, Gyeonggi-do,<br>14067, Republic of Korea<br>m. +82-10-4952-6638, e. mshenmc@gmail.com, f. 031-388-9263 |
| Preparation Date | Jan 07, 2022 |
Date 510(k) summary prepared: June 01, 2022
## 2. Device Name and Code - 807.92(a)(2)
| Trade/Device Name | SILKRO |
|---------------------|----------------------------------------------------------------|
| Common Name | Electrosurgical System |
| Classification Name | Electrosurgical cutting and coagulation device and accessories |
| Product Code | GEI |
| Regulation Number | 21 CFR 878.4400 |
| Regulatory Class | II |
| Review Panel | General & Plastic Surgery (ODE) |
3. Legally marketed device(s) to which equivalence is claimed - 807.92(a)(3)
| Predicate Devices | K170325 |
|---------------------|-----------------------------------------------------------------|
| | Applicant: ILOODA CO., LTD |
| | Trade/Device Name: Secret RF |
| | Regulation Number: 21 CFR 878.4400 |
| | Regulation Name: Electrosurgical Cutting and Coagulation Device |
| | and Accessories |
| | Regulatory Class: Class II |
| | Product Code: GEI, OUH |
| Reference Devices 1 | K182355 |
| | Applicant: ILOODA CO., LTD |
| | Trade/Device Name: Secret RF Smartcure Applicator |
| | Regulation Number: 21 CFR 878.4400 |
| | Regulation Name: Electrosurgical Cutting and Coagulation Device |
| | and Accessories |
| | Regulatory Class: Class II |
| | Product Code: GEI, OUH |
| Reference Device 2 | K201685 |
| | Applicant: Jeisys Medical Inc. |
| | Trade/Device Name: Potenza |
| | Regulation Number: 21 CFR 878.4400 |
| | Regulation Name: Electrosurgical Cutting and Coagulation Device |
| | and Accessories |
| | Regulatory Class: Class II |
| | Product Code: GEI |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Hironic. The logo consists of a red square with a white stylized "H" inside, followed by the word "HIRONIC" in black, sans-serif capital letters. The logo is simple and modern, with a focus on the company name.
## 4. Device Description - 807.92(a)(4)
This device is an instrument used for tissue coagulation by means of high-frequency current. It consists of the main unit, Four handpieces, GP Cable, foot switch, LCD touchscreen, ground pad and power cable.
The SILKRO has two operating modes: monopolar mode and bipolar mode. In the monopolar mode, RF energy flows from the main unit and a patient loop is formed by pairing the active electrode tip with the neutral electrode pad. Heat is not generated in the neutral electrode pad due to its low contact resistance, but heat is generated in the active electrode tip which has higher contact resistance. The higher contact resistance heats up the tissue resulting in coagulation. In the bipolar mode, RF energy is delivered between adjacent needles in the electrode tip without use of the neutral electrode pad. The user can select the mode and adjust parameters through the touch screen user interface of the electrosurgical device.
RF(HF) energy is delivered to the target tissue using a handpiece (RM, RN, RC, RV) and electrode tip, the tip being placed in light contact with the epidermis and the handpiece being held at right anqles to the tarqet tissue. As the HF energy passes through the skin, it generates an electro RF reaction, which is capable of coaqulating the tissue.
- Electrosurgical Unit Main body
- Four different handpieces (motor )
- Neutral electrode pad and neutral electrode pad cable, cleared under K092761
- Handpiece stands
- Foot switch
- Power cord
## 5. Indication for use - 807.92(a)(5)
This device is intended for use in dermatologic and general surgical procedures for electrocoagulation.
6. Summary of the Technological Characteristics of the Device Compared to the Predicate - 807.92(a)(6) (1) Predicate device
| | Proposed Device | Predicate Device |
|---------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Pending | K170325 |
| Manufacturer | Hironic Co., Ltd. | ILOODA CO., LTD |
| Trade/Device Name | SILKRO | Secret RF |
| Classification Name | Electrosurgical cutting and<br>coagulation device and<br>accessories | Electrosurgical cutting and<br>coagulation device and accessories |
| Indication for use | This device is intended for use in<br>dermatologic and general<br>surgical procedures for electro-<br>coagulation. | Secret RF is intended for use in<br>dermatologic and general surgical<br>procedures for electro-coagulation<br>and hemostasis. |
| Output energy type | Radio Frequency | Radio Frequency |
| User interface | Color Touch Panel | Color Touch Panel |
| Operating Frequency | 2MHz | 2MHz |
| Rated Input | 100-240VAC, 50/60Hz | 100-240VAC, 50/60Hz |
| Max Power | 25W ± 20%, 500Ω | 25W ± 20%, 500Ω |
| RF Intensity | 1 ~ 10 Level | 1 ~ 10 Level |
| RF Duration | 50 ms ~ 950 ms | 50 ms ~ 950 ms |
| Treatment Time | 5 ~ 15 min | 10 ~ 15 min |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Hironic. The logo consists of a red square with a white shape inside, followed by the word "HIRONIC" in black, block letters. The letters are stylized and have a modern, technological look.
| | (Recommended) | (Recommended) |
|------------------------|--------------------------------|--------------------------------|
| Needle insert depth | 0.5 ~ 3.5 mm(0.1 step) | 0.5 ~ 3.5mm (0.1 step) |
| Repetition | 0.2/0.5/0.8/1.0/2.0 sec/Single | 0.2 / 0.5 / 1 / 2 sec / Single |
| Mode of operation | Bipolar type | Bipolar type |
| Electrode(Needle) type | RF Micro needle 25pin, 49pin | RF Micro needle 25pin, 49pin |
| Sterilization | EO Gas | EO Gas |
Information provided in these 510(k) submissions shows that SILKRO are substantially equivalent to the predicate device(K170325) in terms of indication for use, performance that related with technological characteristics. SILKRO Treatment time is recommendation between 5 to 15 min and it is within range of predicate device(K17025)'s treatment time.
| | Proposed Device | Reference Device 1 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Pending | K182355 |
| Manufacturer | Hironic Co., Ltd. | ILOODA CO., LTD |
| Trade/Device Name | SILKRO | Secret RF Smartcure Applicator |
| Classification Name | Electrosurgical cutting and<br>coagulation device and<br>accessories | Electrosurgical cutting and<br>coagulation device and accessories |
| Indication for use | This device is intended for use in<br>dermatologic and general<br>surgical procedures for electro-<br>coagulation. | Secret RF Smartcure Applicator is<br>intended for use in dermatologic and<br>general surgical procedures for<br>electro-coagulation and hemostasis |
| Output energy type | Radio Frequency | Radio Frequency |
| User interface | Color Touch Panel | Color Touch Panel |
| Operating Frequency | 2MHz | 2MHz |
| Rated Input | 100-240VAC, 50/60Hz | 100-240VAC, 50/60Hz |
| Power output | 7.5 W±20%, 500Ω | 45W±20%, 500Ω |
| RF Duration | 100 ms ~ 1000 ms | 100 ms ~ 3000 ms |
| Treatment Time | 10~ 15 minutes | 10~ 15 minutes |
| Needle insert depth | 1.2, 1.5, 1.8 mm | 0.8/1.25/1.5/ 2.0 mm |
| Mode of operation | Monopolar | Monopolar |
| Electrode(Needle)<br>type | 1 pin type | 1 pin Type |
| Single Use | Single Use | Single Use |
| Sterilization | EO Gas | EO Gas |
## (2) Reference Device 1
Information provided in these 510(k) submissions shows that SILKRO electrosurgical system's RN handpiece is substantially equivalent to the Reference device 1(K182355) in terms of indication for use, and performance that related with technological characteristics. RF duration is not exact same values but our duration is within in range of reference device's duration value.
#### (3) Reference Device 2
| | Proposed Device | Reference Device 2 |
|---------------------|----------------------------------------------------------------------|-------------------------------------------------------------------|
| 510(k) Number | Pending | K201685 |
| Manufacturer | Hironic Co., Ltd. | Jeisys Medical Inc |
| Trade/Device Name | SILKRO | Potenza |
| Classification Name | Electrosurgical cutting and<br>coagulation device and<br>accessories | Electrosurgical cutting and<br>coagulation device and accessories |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "HIRONIC" in black letters with a red symbol to the left. The symbol is a stylized letter "H" with a curved top and a flat bottom. The word "HIRONIC" is in all caps and has a bold, sans-serif font. The letters are evenly spaced and the word is centered in the image.
| Indication for use | This device is intended for use in<br>dermatologic and general<br>surgical procedures for electro-<br>coagulation. | Potenza is intended for use in<br>dermatologic and general surgical<br>procedures for electro-coagulation<br>and hemostasis. |
|-------------------------|--------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Output energy type | Radio Frequency | Radio Frequency |
| Operating Frequency | 2MHz | 2MHz |
| Power output | Maximum 40 W ±20%, | Maximum 50W±20%, |
| Treatment Time | 10 ~ 15 min | 10 ~ 15 min |
| Treatment<br>Activation | Touch screen, Foot switch,<br>Neutral electrode pad and neutral<br>electrode pad cable, cleared<br>under K092761. | Touch screen, Foot switch, Neutral<br>electrode pad and neutral electrode<br>pad cable, cleared under K092761. |
| Mode of operation | Monopolar | Monopolar |
Information provide in these 510(k) submissions shows that SILKRO electrosurgical system's Electrosurgical Unit part, Bipolar and monopolar operating mode, output frequency of 2MHz, same as for the reference device. Also, treatment time, treatment activation with operation mode, same as for the reference device.
#### 7. Non-clinical tests submitted - 807.92(b)(1)
- -. Basic safety and essential performance of the SILKRO is evaluated in accordance with IEC 60601-1:2012.
- -. Effect to the device by electromagnetic disturbances is evaluated in accordance with the FDArecognized consensus standard, IEC 60601-1-2:2014.
- -. Medical electrical equipment Part 2 -2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories, IEC 60601-2-2,
- -. General Requirements for Basic Safety and Essential Performance Collateral Standard: Usability is evaluated in accordance with the FDA-recognized consensus standard, IEC 60601-1-6:2013.
- -. Risk management is recorded in the reference of ISO 14971:2007.
- -. The software for SILKRO is verified and validated in accordance with its moderate level of concern. Software life cycle processes are evaluated according to the FDA-recognized consensus standard, IEC 62304:2006.
- -. Application of usability engineering to medical devices is evaluated in accordance with IEC 62366:2007.
- ex vivo animal testing using models was also conducted to obtain histological data of values for depth and zone of coagulation and thermal damage immediately post treatment; 10 days post treatment. The treatment was performed at the intensity(power) low, mid, high. Based on this animal test, it was confirmed through mechanical and histological evaluation that RM, RV, RC handpieces can affect tissue by thermal effect of each output condition.
## 8. Biocompatibility
Hironic Co., Ltd performed biocompatibility testing for the contact part of each of handpiece's tips. FDA's "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process", June 6, 2016.
SILKRO skin contact materials (Applied part that touched patients' skin permanently within 24 hours) were evaluated according to ISO 10993-1. The results of this test confirmed that materials met the biocompatibility requirements. Biocompatibility testing including cytotoxicity, sensitization, oral
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "HIRONIC" in a stylized font. The first letter "H" is a red square with a white arrow pointing to the right. The rest of the letters are black and in a bold, sans-serif font. The letters are slightly rounded and have a modern look.
K210084
mucosal irritation was completed according to the following standards: ISO 10993-1 Biological Evaluation of Medical Devices –Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process ISO 10993-5 Biological Evaluation of Medical Devices – Part 5 Cytotoxicity ISO 10993-10 Biological Evaluation of Medical Devices Part 10: Tests for irritation and skin sensitization ISO 10993-12 Biological evaluation of medical devices -- Part 12: Sample preparation and reference materials.
Each tip are external communicating devices which come into contact with tissue for a limited period of time, i.e., less than 24 hours.
## 9. Sterilization and shelf life
## -. Sterilization
The RM Micro Needle tips (RMT-125, RMT-149, RMT-N49) and RN Needle tips are subject to Ethylene Oxide (EO) sterilization.
EO sterilization residual testing for the RM Micro Needle tips (RMT-125, RMT-149, RMT-N49) and RN Needle tips according to ISO 10993-7:2008, Biological Evaluation of Medical Devices -Part 7: Ethylene Oxide Sterilization Residuals to acceptance criteria as specified in the standard. According to ISO 11135:2014. Sterilization of health-care products - Ethylene oxide -Requirements for the development, validation and routine control of a sterilization process for medical devices, the sterility assurance level (SAL) is 10-6,
## -. Shelf life
Hironic Co., Ltd performed shelf-life testing to establish three years shelf-life based on 97-days accelerated aging in accordance with the following standards to the acceptance criteria as specified in the standard:
- ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices - Accelerated Aging
- ASTM F1929-15: Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration - Packaging Test
- ASTM F88/F88M: Standard Test Method for Seal Strength of Flexible Barrier Materials
## -. Packaging
RMT-149 is packed in PET+Steril Paper. To keeping the sterility of the product is important for this shelf life study. The test based on ASTM F1980-16, Accelerated aging theory.
## 11. Conclusions drawn from clinical and non-clinical tests submitted
SILKRO device has same indications for use and similar design and technological characteristics as the predicate device. The differences between the subject devices and the predicate device do not raise new questions of safety or efficacy. Results of performance testing demonstrated substantial equivalence of the subject device to the predicate. Therefore, the subject device is as safe and effective as previously cleared predicate device for the proposed intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.