K210071 · Surgical Information Sciences, Inc. · LLZ · Mar 31, 2021 · Radiology
Device Facts
Record ID
K210071
Device Name
SIS System (Version 5.1.0)
Applicant
Surgical Information Sciences, Inc.
Product Code
LLZ · Radiology
Decision Date
Mar 31, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Subthalamic nuclei (STN) and globus pallidus (GPe/GPi) visualization
Deep learning Convolutional Neural Network models
—
—
Ultra-high resolution 7T Magnetic Resonance images used to determine ground truth for training data.
—
—
—
Directional lead orientation detection
Deep learning Convolutional Neural Network models
Greater than 90% of orientations accurate within ± 30°
91% of trusted detections accurate within ± 30°
—
—
43 CT image series containing 55 leads
—
Indications for Use
SIS System is an application intended for use in the viewing, presentation and documentation of medical imaging, including different modules for image processing, image fusion, and intraoperative functional planning where the 3D output can be used with stereotactic image guided surgery or other devices for further processing and visualization. The device can be used in conjunction with other clinical methods as an aid in visualization of the subthalamic nuclei (STN) and globus pallidus externa and interna (GPe and GPi, respectively). Typical users of SIS System are medical professionals, including but not limited to surgeons, neurologists, and radiologists.
Device Story
Software-only device; processes patient-specific MRI and CT scans to generate 3D anatomical models of basal ganglia structures (STN, GPe, GPi). Uses pre-trained deep learning convolutional neural networks (CNNs) to segment brain structures and identify implanted leads in post-operative CT images. Determines orientation of directional leads. Used by surgeons, neurologists, and radiologists in clinical settings for stereotactic surgical planning and lead output visualization. Supplements standard clinical imaging; provides adjunctive information for surgical decision-making; aids in precise lead placement and orientation assessment.
Clinical Evidence
Bench testing only. Validation included visualization accuracy testing for STN and GPi/GPe structures, MRI-to-CT registration accuracy, and CT lead segmentation. Electrode orientation detection validated on 43 CT series (55 leads); 91% of trusted detections were accurate within ±30° of ground truth.
Technological Characteristics
Software-only device; utilizes deep learning CNNs for image segmentation. Processes MRI and CT inputs. Features include 3D anatomical modeling, image fusion (linear/non-linear registration using ANTS and ELASTIX), and directional lead orientation detection. Operates on computer platforms.
Indications for Use
Indicated for medical professionals (surgeons, neurologists, radiologists) as an aid in visualization of subthalamic nuclei (STN) and globus pallidus externa/interna (GPe/GPi) for stereotactic surgical planning and lead placement.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Surgical Information Sciences SIS Software version 3.6.0 (K192304)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Surgical Information Sciences, Inc. % Kelliann Payne Partner Hogan Lovells US LLP 1735 Market Street, 23rd Floor PHILADELPHIA, PA 19103
March 31, 2021
Re: K210071
Trade/Device Name: SIS System (Version 5.1.0) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: QIH, LLZ Dated: January 11, 2021 Received: January 11, 2021
Dear Kelliann Payne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
510(k) Number (if known)
K210071
Device Name
SIS System (version 5.1.0)
Indications for Use (Describe)
SIS System is an application intended for use in the viewing, presentation of medical imaging, including different modules for image processing, image fusion, and intraoperative functional planning where the 3D output can be used with stereotactic image quided surgery or other processing and visualization. The device can be used in conjunction with other clinical methods as an aid in visualization of the subthalamic nuclei (STN) and globus pallidus externa and interna (GPe and GPi, respectively).
Typical users of SIS System are medical professionals, including but not limited to surgeons, neurologists, and radiologists.
Type of Use (Select one or both, as applicable)
& Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) SUMMARY
### K210071
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Surgical Information Sciences, Inc. 10405 6th Avenue North, Suite 110 Plymouth, MN 55441 Contact Person: Ann Quinlan-Smith Phone: 612-325-0187 E-mail: ann.quinlan.smith@surqicalis.com
Date Prepared: March 29, 2021
## Trade Name of Device: SIS System version 5.1.0
### Common or Usual Name/Classification Name:
- Automated Radiological Image Processing Software (Product Code: QIH; 21 Primary: C.F.R 892.2050);
- Secondary: System, Image Processing, Radiological (Product Code: LLZ; 21 C.F.R 892.2050)
Regulatory Class: Class II
Predicate Device: Surgical Information Sciences SIS Software version 3.6.0 (K192304)
#### Intended Use / Indications for Use
SIS System is an application intended for use in the viewing, presentation and documentation of medical imaging, including different modules for image processing, image fusion, and intraoperative functional planning where the 3D output can be used with stereotactic image guided surgery or other devices for further processing and visualization. The device can be used in conjunction with other clinical methods as an aid in visualization of the subthalamic nuclei (STN) and globus pallidus externa and interna (GPe and GPi, respectively).
Typical users of SIS System are medical professionals, including but not limited to surgeons, neurologists, and radiologists.
{5}------------------------------------------------
# Technological Characteristics
The SIS System version 5.1.0, a software only device based on machine learning and image processing, is designed to enhance standard clinical images for the visualization of structures in the basal ganglia area of the brain, specifically the subthalamic nucleus (STN) and globus pallidus externa and interna (GPe/GPi). The output of the SIS system supplements the information available through standard clinical methods by providing additional, adjunctive information to surgeons, neurologists, and radiologists for use in viewing brain structures for planning stereotactic surgical procedures and planning of lead output.
The SIS System version 5.1.0 provides a patient-specific, 3D anatomical model of specific brain structures based on the patient's own clinical MR image using pretrained deep learning neural network models. This method incorporates ultra-high resolution 7T (7 Tesla) Magnetic Resonance images to determine ground truth for the training data set to train the deep learning models. These pre-trained deep learning neural network models are then applied to a patient's clinical image to predict the shape and position of the patient's specific brain structures of interest. The SIS System is further able to locate and identify implanted leads, where implanted, visible in post-operative CT images and place them in relation to the brain structure of interest from the preoperative processing.
The proposed device is a modification to the SIS Software version 3.6.0 that was cleared under K192304. The changes made to the SIS System include (1) an updated algorithm that is based on deep learning Convolutional Neural Network models that were architected and optimized for brain image segmentation; (2) the addition of new targets for visualization, specifically the globus pallidus externa and interna (GPe/GPi); and (3) the addition of a functionality to determine the orientation of a directional lead, following its segmentation from the post-operative CT image.
# Performance Data
Following the modifications, the software verification and validation testing was repeated to validate that the modified software functions as specified and performs similarly to the predicate device.
To validate the updated algorithm, visualization accuracy testing was conducted for the STN and GPi/GPe structures using the same test methods and acceptance criteria for the previously cleared predicate device. In addition, the company repeated the MRI to CT registration testing to ensure that 3D transformation remains accurate. The company also repeated the testing for image processing
{6}------------------------------------------------
of CT images to validate the lead segmentation. Finally, the electrode orientation detection software was validated on 43 CT image series that contained 55 leads. The software was characterized by two probabilities: the probability of a trusted detection being accurate (within ± 30° of the ground truth) and the probability of an untrusted detection being accurate. When the software trusted the lead detection, it was correct in 91% of cases. This testing demonstrated that greater than 90% of orientations presented to the user are accurate within ± 30°. The results of this testing demonstrated that the SIS System version 5.1.0 has been fully verified and validated and the updated device performs as intended and is as safe and effective compared to the predicate.
## Substantial Equivalence
Both the subject and predicate devices are applications intended for use in the viewing, presentation and documentation of medical imaging, including different modules for image processing, image fusion, and intraoperative functional planning where the 3D output can be used with stereotactic image guided surgery or other devices for further processing and visualization. Both devices can be used in conjunction with other clinical methods as an aid in visualization of the target brain structures. In addition, typical users of both devices are medical professionals, including, but not limited to surgeons, neurologists and radiologists.
The subject device, like the predicate, operates on other computer platforms and uses a proprietary algorithm to generate 3D segmented anatomical models from patients' MRI scans. The subject device employs an updated version of the algorithm based on deep learning Convolutional Network Models, which were trained to identify the region of interest and individually predict the location and size of the anatomical structures of interest. Furthermore, the addition of the globus pallidus externa and interna (GPe/GPi) structures as well as the functionality to detect the orientation of the implanted directional lead, further facilitate the fundamental clinical purpose for which the predicate was cleared, namely assistance with visualization, surgical planning, image review and analysis. Validation testing demonstrated that the subject device is as safe and effective as the predicate device. The table below provides a summary comparison between the subject and predicate devices.
{7}------------------------------------------------
| | SIS System version 5.1.0<br>(subject device) | SIS Software version 3.6.0<br>(predicate device) |
|-------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for Use | SIS System is an application intended for use in the<br>viewing, presentation and<br>documentation of medical<br>imaging, including different<br>modules for image<br>processing, image fusion,<br>and intraoperative functional<br>planning where the 3D<br>output can be used with<br>stereotactic image guided<br>surgery or other devices for<br>further processing and<br>visualization. The device<br>can be used in conjunction<br>with other clinical methods<br>as an aid in visualization of<br>the subthalamic nuclei<br>(STN) and globus pallidus<br>externa and interna (GPe<br>and GPi, respectively).<br><br>Typical users of the SIS<br>System are medical<br>professionals, including but<br>not limited to surgeons,<br>neurologists and<br>radiologists. | SIS Software is an application intended for use in the<br>viewing, presentation and<br>documentation of medical<br>imaging, including different<br>modules for image<br>processing, image fusion, and<br>intraoperative functional<br>planning where the 3D output<br>can be used with stereotactic<br>image guided surgery or<br>other devices for further<br>processing and visualization.<br>The device can be used in<br>conjunction with other clinical<br>methods as an aid in<br>visualization of the<br>subthalamic nuclei (STN).<br><br>Typical users of SIS Software<br>are medical professionals,<br>including but not limited to<br>surgeons, neurologists, and<br>radiologists. |
| User Population | Medical professionals,<br>including but not limited to<br>surgeons, neurologists and<br>radiologists. | Medical professionals,<br>including but not limited to<br>surgeons, neurologists, and<br>radiologists. |
| Allows for importing of<br>digital imaging sets | Yes | Yes |
| Uses proprietary<br>software algorithm to<br>generate 3D segmented<br>anatomical models from<br>patient's MR scans | Yes | Yes |
| | SIS System version 5.1.0<br>(subject device) | SIS Software version 3.6.0<br>(predicate device) |
| Allows for review and<br>analysis of data in 2D<br>and 3D formats | Yes | Yes |
| Performs image fusion<br>of datasets using<br>automated or manual<br>image matching<br>technique | Yes | Yes |
| Segments structures in<br>images with manual and<br>automated tools and<br>converts them into 3D<br>objects for display | Yes | Yes |
| Creates hybrid datasets<br>by filing in segmented<br>regions slice-by-slice<br>on anatomical datasets | Yes | Yes |
| Can be downloaded to<br>planning system | Yes | Yes |
| Segmentation of CT<br>scan to identify<br>structures in relation to<br>those visualized on MR | Yes | Yes |
| Feature to Account for<br>CT images with gantry<br>tilt | Yes | Yes |
| Cross-registers images<br>and creates 3D (fused)<br>model | Yes | Yes |
| Uses registration<br>methods (linear and<br>non-linear) by multiple<br>registration tools (ANTS<br>and ELASTIX) | Yes | Yes |
{8}------------------------------------------------
{9}------------------------------------------------
# Conclusion
The updated SIS System version 5.1.0 is as safe and effective as the predicate version previously cleared in K192304. The subject device has the same intended use and similar technological characteristics and principles of operation, with minor differences supported by performance validation testing demonstrating that the subject device is as safe and effective as the predicate device. Thus, the minor technological differences between SIS System version 5.1.0 and its predicate device raise no new issues of safety or effectiveness, and the updated SIS System version 5.1.0 is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.