K203845 · Philips Medical Systems Nederland B.V. · PSY · Sep 17, 2021 · Pathology
Device Facts
Record ID
K203845
Device Name
Philips IntelliSite Pathology Solution
Applicant
Philips Medical Systems Nederland B.V.
Product Code
PSY · Pathology
Decision Date
Sep 17, 2021
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 864.3700
Device Class
Class 2
Indications for Use
The Philips IntelliSite Pathology Solution (PIPS) is an automated digital slide creation, viewing, and management system. The PIPS is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The PIPS is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. The PIPS comprises the Image Management System (IMS), the Ultra Fast Scanner (UFS) and display. The PIPS is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using PIPS.
Device Story
System creates, manages, and displays digital images of glass pathology slides; intended for use by pathologists in clinical settings. Input: glass slides (FFPE tissue) scanned by Ultra Fast Scanner (UFS). Processing: Image Management System (IMS) handles image storage and rendering; subject device introduces client-side rendering (JavaScript) for decompression, sharpening, and contrast enhancement to reduce network bandwidth and improve scalability. Output: high-resolution digital images viewed on a display. Pathologists use these images to perform diagnostic interpretations equivalent to conventional light microscopy. Benefits: enables digital workflow, remote viewing, and efficient slide management for pathology departments.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical verification testing, including pixel comparison tests and regression testing, demonstrating that the client-side rendering modification maintains image quality equivalence to the predicate.
Technological Characteristics
System includes UFS (optical/mechanical/electronic), IMS (software), and PP27QHD display. Client-side rendering utilizes WebGL JavaScript API to offload processing to client GPU (min 4GB VRAM, 112 GB/s bandwidth). Connectivity via intranet/internet. Standards: ISO 14971, ISO 15223-1, IEC 62304, IEC 62366-1.
Indications for Use
Indicated for pathologists to review and interpret digital images of FFPE surgical pathology slides. Not for use with frozen section, cytology, or non-FFPE hematopathology specimens.
Regulatory Classification
Identification
The whole slide imaging system is an automated digital slide creation, viewing, and management system intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides. The system generates digital images that would otherwise be appropriate for manual visualization by conventional light microscopy.
Special Controls
A whole slide imaging system must comply with the following special controls: (1) Premarket notification submissions must include the following information: (i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system. (ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate: (A) Slide feeder; (B) Light source; (C) Imaging optics: (D)Mechanical scanner movement; (E) Digital imaging sensor; (F) Image processing software; (G)Image composition techniques; (H)Image file formats; (I) Image review manipulation software; (J) Computer environment; (K)Display system. (iii)Detailed bench testing and results at the system level, including for the following, as appropriate: (A)Color reproducibility; (B) Spatial resolution; (C) Focusing test; (D) Whole slide tissue coverage; (E) Stitching error: (F) Turnaround time. (iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate: (A)Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included. (C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (e.g., main sign-out diagnosis). (D) A detailed human factors engineering process must be used to evaluate the whole slide imaging system user interface(s). (2) Labeling compliant with 21 CFR 809.10(b) must include the following: The intended use statement must include the information described in paragraph (i) (1)(i) of this section, as applicable, and a statement that reads, "It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device." (ii) A description of the technical studies and the summary of results, including those that relate to paragraph (1)(ii) and (1)(iii) of this section, as appropriate. (iii) A description of the performance studies and the summary of results, including those that relate to paragraph (1)(iv) of this section, as appropriate. (iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system.
(ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate:
(A) Slide feeder;
(B) Light source;
(C) Imaging optics;
(D) Mechanical scanner movement;
(E) Digital imaging sensor;
(F) Image processing software;
(G) Image composition techniques;
(H) Image file formats;
(I) Image review manipulation software;
(J) Computer environment; and
(K) Display system.
(iii) Detailed bench testing and results at the system level, including for the following, as appropriate:
(A) Color reproducibility;
(B) Spatial resolution;
(C) Focusing test;
(D) Whole slide tissue coverage;
(E) Stitching error; and
(F) Turnaround time.
(iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate:
(A) Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (
*e.g.,* main sign-out diagnosis).(D) A detailed human factor engineering process must be used to evaluate the whole slide imaging system user interface(s).
(2) Labeling compliant with 21 CFR 809.10(b) must include the following:
(i) The intended use statement must include the information described in paragraph (b)(1)(i) of this section, as applicable, and a statement that reads, “It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device.”
(ii) A description of the technical studies and the summary of results, including those that relate to paragraphs (b)(1)(ii) and (iii) of this section, as appropriate.
(iii) A description of the performance studies and the summary of results, including those that relate to paragraph (b)(1)(iv) of this section, as appropriate.
(iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
INSTRUMENT ONLY
## I Background Information:
A 510(k) Number
K203845
B Applicant
Philips Medical Systems Nederland B.V.
C Proprietary and Established Names
Philips IntelliSite Pathology Solution
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PSY | Class II | 21 CFR 864.3700 - Whole Slide Imaging System | PA - Pathology |
## II Submission/Device Overview:
A Purpose for Submission:
Philips Medical Systems Nederland B.V., submitted K203845 due to a significant change to the Image Management System (IMS) subsystem of their Philips IntelliSite Pathology Solution (PIPS, K192259). The change is as follows: client-side rendering.
B Type of Test:
Not applicable.
## III Intended Use/Indications for Use:
A Intended Use(s):
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
See Indications for Use below.
## B Indication(s) for Use:
The Philips IntelliSite Pathology Solution (PIPS) is an automated digital slide creation, viewing, and management system. The PIPS is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The PIPS is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.
The PIPS comprises the Image Management System (IMS), the Ultra Fast Scanner (UFS) and display. The PIPS is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using PIPS.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
## IV Device/System Characteristics:
### A Device Description:
Philips IntelliSite Pathology Solution (PIPS) is an automated digital slide creation, viewing, and management system. PIPS consists of two subsystems and a display component:
**Ultra-Fast Scanner (UFS)**: There are no major changes to the UFS cleared under K172174. In summary, the UFS subsystem scans cover-slipped glass slides at 40x magnification to produce high-quality Whole Slide Images (WSIs) of the tissue material mounted on the glass slides. The slides are placed in racks which are then placed in rack slots in the UFS. The scanning procedure requires minimal user interaction. The UFS has a loading capacity for large batches (up to 300 slides). On the touchscreen the scanning progress is shown. Proprietary software is used for image processing during acquisition. Philips' proprietary format, iSyntax, is used to store and transmit the images between the UFS and the IMS. The UFS consists of optical, mechanical and electronic elements as well as software components. The UFS is located and operated on site in healthcare institutions such as clinical pathology laboratories, etc.
**Image Management System (IMS)**: The IMS is a software only subsystem and consists of the IMS Application Server and Storage software and IMS Viewer software installed on compatible server hardware. Functionality of the IMS includes the ability to view WSIs, organize workload, and annotate and bookmark scanned WSIs. The IMS Viewer will operate in the user's work environment using the display (component of the PIPS) to display WSIs. The IMS Application Server and Storage software will operate in a controlled environment, typically a server room separate from the pathology departments. One or more client computers with intranet/internet connection to the IMS, meeting the minimum requirements (see Table 16-3 and Table 16-4 of Section 16.3.1), and display are required for viewing images from the IMS. Client computers are
K203845 - Page 2 of 7
{2}
not part of the PIPS and will either be located in healthcare institutions such as clinical pathology laboratories, etc. or in a pathology office.
**Client-Side Rendering**: Image processing is performed at the client computer (i.e., client-side rendering) instead of at the server side. Client-side rendering utilizes Web Graphics Library (WebGL), which is a JavaScript Application Programming Interface (API), for image processing within a browser. WebGL allows using the computational power of a Graphics Processing Unit (GPU) on the client computer. Offloading image processing to the GPU speeds up rendering. A Philips service engineer can only configure the server or client-side rendering option. This option is called "Use my computer for image processing". When the option "Use my computer for image processing" is enabled, additional requirements with respect to the Video Card are applicable as follows: Video card minimum specifications - GPU Memory: 4GB; GeForce GTX 1050 Ti or similar; Memory bandwidth: 112 GB/s. The recommended browser for client-side rendering is Chrome.
**Philips PP27QHD Display**: There are no changes introduced for the display. In summary, the Philips PP27QHD display is connected to a client computer with access to the IMS Viewer. The display is calibrated using the built-in calibration sensor and software tool, which perform automatically a periodic calibration. The pathologist uses the display for viewing WSIs using the IMS Viewer software.
**B Instrument Description Information**:
1. **Instrument Name**:
Philips IntelliSite Pathology Solution (PIPS)
2. **Specimen Identification**:
Not applicable.
3. **Specimen Sampling and Handling**:
Not applicable.
4. **Calibration**:
Not applicable.
5. **Quality Control**:
Not applicable.
**V Substantial Equivalence Information**:
**A Predicate Device Name(s)**:
K203845 - Page 3 of 7
{3}
Philips IntelliSite Pathology Solution (PIPS)
## B Predicate 510(k) Number(s):
K192259
## C Comparison with Predicate:
| Device & Predicate Device(s): | K203845 | K192259 |
| --- | --- | --- |
| Device Trade Name | Philips Intellisite Pathology Solution (PIPS) | Philips Intellisite Pathology Solution (PIPS) |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The Philips IntelliSite Pathology Solution (PIPS) is an automated digital slide creation, viewing, and management system. The PIPS is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The PIPS is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. The PIPS comprises the Image Management System (IMS), the Ultra Fast Scanner (UFS) and display. The PIPS is for creation and viewing of digital images of | Same |
K203845 - Page 4 of 7
{4}
K203845 - Page 5 of 7
| | scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using PIPS. | |
| --- | --- | --- |
| General Device Characteristic Differences | | |
| Image Review and Manipulation Software – Computer Environment | To reduce the network bandwidth requirements and improve scalability, an option is added (configurable by a Philips service engineer) that moves the image processing logic (i.e. decompression, sharpening and contrast enhancement) from the server side to the client side. The programming language is JavaScript. | Most image processing logic (e.g. decompression, sharpening and contrast enhancement) are performed on the server side. The programming language is C#. |
VI Standards/Guidance Documents Referenced:
- ISO 14971: 2007: Medical devices – Application of risk management to medical devices
- ISO 15223-1: 2016: Medical devices: Symbols to be used with medical device labels, labelling and information to be supplied – Part 1: General requirements
- IEC 62304:2006 Amd 1: 2015: Medical device software – Software life-cycle processes
- IEC 62366-1: 2015: Medical devices – Application of usability engineering to medical devices
VII Performance Characteristics (if/when applicable):
{5}
K203845 - Page 6 of 7
# A Analytical Performance:
1. Precision/Reproducibility:
Not applicable.
2. Linearity:
Not applicable.
3. Analytical Specificity/Interference:
Not applicable.
4. Accuracy (Instrument):
Not applicable.
5. Carry-Over:
Not applicable.
# B Other Supportive Instrument Performance Characteristics Data:
The K203845 was submitted to seek clearance for one significant change, the client-side rendering to the Image Management System (IMS), the software subsystem of the Philips IntelliSite Pathology Solution (PIPS, K192259).
Verification testing activities have been performed to support the substantial equivalence determination for the modified IMS containing the client-side rendering change which included:
1. A pixel comparison test to determine pixel equivalence using the 'Regression SingleTest' methodology that loads a reference image and reads the displayed pixel from the user interface into a test image to verify that the displayed RGB output values of each pixel are identical with the reference image input.
Reference Data Set: A Reference Data Set is created by importing a set of 85 Whole Slide Images (WSIs) in a clinically validated product and storing from the pixels from 3377 selected regions of these WSIs in png images (i.e. reference png image) with $1.2*10^{9}$ pixels.
Test acceptance criteria: The displayed RGB output values of each pixel in the test image of the proposed product need to be identical with the reference bitmap image from the clinically validated product. When all test images are identical with the related reference bitmap images then the test results meet the acceptance criteria and the conclusion is that the test 'passed'.
Test execution: During the test execution, the same WSIs used to create the Reference Data Set are imported in the proposed product. For this RegressionSingleTest a version of the IMS HTML viewer is used that does not contain users controls but allows for retrieval of the pixel data by a test framework.
{6}
IMS HTML viewer test method: The RegressionSingleTest executes the following steps for each browser (i.e., Chrome):
1. Import 85 WSIs from the Reference Data Set in the proposed IMS.
2. For each of the 3377 selected regions:
- Open the region in the IMS HTML viewer,
- Read the displayed pixels from the User Interface (HTMLCanvasElement class) into a test image,
- Load the pixels from the reference bitmap image,
- Compare the displayed pixels (RGB) of the test image with the pixels in the related reference bitmap image for equivalence.
3. When all 3377 test images are identical with the reference bitmap images then report ‘passed’, otherwise report ‘fail’.
Test Results: All test results passed.
2. Testing of requirements, risk control measures, integration and regression
The verification testing activities performed, support the substantial equivalence determination of the subject Philips Intellisite Pathology Solution (PIPS) (K203845) with the modified Image Management System to the predicate Philips Intellisite Pathology Solution (PIPS) (K192259).
**Cybersecurity**
Security Threat analysis was performed for IMS version 4.1.1 and compared to previous version, IMS 3.2.1. No additional security risks were identified.
**VIII Proposed Labeling:**
The labeling supports the finding of substantial equivalence for this device.
**IX Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K203845 - Page 7 of 7
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.