Ponto Bone Anchored Hearing System, MONO Surgery Kit
K203807 · Oticon Medical AB · MAH · Mar 3, 2021 · Ear, Nose, Throat
Device Facts
Record ID
K203807
Device Name
Ponto Bone Anchored Hearing System, MONO Surgery Kit
Applicant
Oticon Medical AB
Product Code
MAH · Ear, Nose, Throat
Decision Date
Mar 3, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.3302
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Ponto Bone Anchored Hearing System's intended use is for improvement of hearing for patients with conductive or mixed hearing losses, whether unilaterally or bilaterally fitted, or for those with single-sided deafness.
Device Story
Ponto Bone Anchored Hearing System consists of external sound processor and surgically implanted titanium abutment. Processor captures sound; converts to vibrations; transmits through skull bone to cochlea via bone conduction. Used in clinical settings by ENT surgeons for implant installation; patients use processor daily. MONO Surgery Kit modifies osteotomy preparation from two-step to single-step drilling. Output is amplified sound for patients with conductive/mixed hearing loss or SSD. Benefits include improved hearing and alternative to AC CROS aids.
Clinical Evidence
Bench testing only. Comparative testing performed on artificial bone to evaluate MONO Surgery Kit against predicate MIPS kit. Metrics included heat generation, insertion torque, installation turns, seating, and ISQ-value measurements. Results confirmed equivalence to predicate performance.
Technological Characteristics
Implant material: Grade 4 titanium. Surface: Smooth with laser ablation in thread valleys. Connection: Hexagon fit with internal screw. Surgical instruments: Single-use drills. Sterilization: E-beam radiation. MONO kit uses single-step drilling vs. two-step MIPS. No software or electronic components described for the surgical kit.
Indications for Use
Indicated for patients with conductive or mixed hearing losses (PTA BC threshold <= 45 dB HL for Ponto 4, <= 55 dB HL for Ponto 3 Power, <= 65 dB HL for Ponto 3 SuperPower); bilateral fitting for symmetric conductive/mixed loss; patients with profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (SSD, PTA AC threshold <= 20 dB HL); and patients indicated for AC CROS hearing aids who cannot/will not use them. Contraindicated for patients below age 5.
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 4, 2021
Oticon Medical AB Carolina Wessling RA Manager Datavagen 37B Askim, SE-436 32 Sweden
Re: K203807
Trade/Device Name: Ponto Bone Anchored Hearing System, MONO Surgery Kit Regulation Number: 21 CFR 874.3300 Regulation Name: Hearing Aid Regulatory Class: Class II Product Code: MAH
Dear Carolina Wessling:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated March 3, 2021. Specifically, FDA is updating this SE Letter (e.g., typo in the trade/device name).
Please note that the 510(k) submission was not re-reviewed. For questions this letter please contact Shu-Chen Peng, Ph.D., OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices, 301-796-6481, Shu-Chen.Peng@fda.hhs.gov.
Sincerely,
Shuchen Peng -S
Shu-Chen Peng, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 3, 2021
Oticon Medical AB Carolina Wessling RA Manager Datavagen 37B Askim, SE-436 32 Sweden
Re: K203807
Trade/Device Name: Pronto Bone Anchored Hearing System, MONO Surgery Kit Regulation Number: 21 CFR 874.3300 Regulation Name: Hearing Aid Regulatory Class: Class II Product Code: MAH Dated: February 2, 2021 Received: February 5, 2021
#### Dear Carolina Wessling:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shu-Chen Peng
Shu-Chen Peng, Ph.D. Acting Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203807
Device Name
Ponto Bone Anchored Hearing System
#### Indications for Use (Describe)
The Ponto Bone Anchored Hearing System (Ponto sound processors and implant system) is intended for the following patients and indications:
· Patients with conductive or mixed hearing losses, who can still benefit from amplification of the sound. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2 and 3 kHz) of the indicated ear should be better than or equal to 45 dB HL for use with the Ponto 4 sound processors, 55 dB HL for use with the Ponto 3 Power sound processors and 65 dB HL for use with the Ponto 3 SuperPower sound processor.
· Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 4 kHz, or less than 15 dB at individual frequencies.
· Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e. single sided deafness or "SSD"). The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3 kHz).
· Also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
The placement of a bone anchored implant is contraindicated for patient below the age of 5.
The Ponto sound processors are intended to be used with either the Ponto implant system or with specific compatible BAHA abutments/implants from Cochlear Bone Anchored Solutions (BAS) (refer to the Ponto labeling for specific compatible Cochlear models). In addition, selected Cochlear Baha sound processors can be used with the Ponto implant/ abutment system (refer to the Ponto labeling for compatible BAHA sound processor models).
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt"> <span style="font-family:Arial"> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt"> <span style="font-family:Arial"> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
# Oticon Medical AB's Ponto Bone Anchored Hearing System
## Ponto Implant System, MONO Surgery Kit
#### Submitter: Oticon Medical AB Datavägen 37B SE-436 32 Askim Sweden
Phone: +46 31 748 61 00 Facsimile: +46 31 687 756
Contact Person: Carolina Anker Wessling Date Prepared: December 22, 2020
510(k) number: K203807
Name of Device: Ponto Bone Anchored Hearing System, Ponto Implant System, MONO Surgery Kit,
Common or Usual Name: Hearing aid, bone conduction
Classification Name: Hearing aid, bone conduction, implanted
Regulatory Class: 21 CFR §874.3300, Class II
Product Code: MAH
## Predicate Devices
| Predicate Device | 510(k) no. | Manufacturer |
|---------------------------------------|------------|-------------------|
| Ponto Bone Anchored Hearing<br>System | K152067 | Oticon Medical AB |
# Device Description and purpose of the 510(k) notice
The Ponto Bone Anchored Hearing System consists of an external sound processor unit and an implant with a skin penetrating abutment. The implant with the abutment is
{5}------------------------------------------------
surqically anchored in the bone behind the ear. Vibrations generated by the sound processor are transmitted directly through the skull bone to the cochlea as bone conduction sound. The sound processor can be connected and disconnected by the user by the snap coupling. The technological characteristics of the Ponto Bone Anchored Hearing System, Ponto Implant System, remain unchanged from the original design (most recently cleared in K152067).
The Ponto Bone Anchored Hearing System also include accessories and instruments for installation of the implantable components, and the Ponto Surgical Manual and Surgical Manual Addenda include step by step instructions for a number of safe alternative surgical approaches for implant installation.
The main purpose of this 510(k) notification is a modification to the drilling to prepare the osteotomy for installation of a Ponto bone anchored implant, from two-step drilling (MIPS previously cleared in K152067) to a single drill step (MONO). The MONO Surgery Kit is a further development of the MIPS Surgery Kit previously cleared in K152067.
# Intended Use / Indications for Use
## Intended use
The Ponto Bone Anchored Hearing System's intended use is for improvement of hearing for patients with conductive or mixed hearing losses, whether unilaterally or bilaterally fitted, or for those with single-sided deafness.
# Indications for use
The Ponto Bone Anchored Hearing System (Ponto sound processors and implant system) is intended for the following patients and indications:
- . Patients with conductive or mixed hearing losses, who can still benefit from amplification of the sound. The pure tone average (PTA) bone conduction (BC) threshold (measured at 0.5, 1, 2 and 3 kHz) of the indicated ear should be better than or equal to 45 dB HL for use with the Ponto 3 and Ponto 4 sound processors, 55 dB HL for use with the Ponto 3 Power sound processors and 65 dB HL for use with the Ponto 3 SuperPower sound processor.
- Bilateral fitting is applicable for most patients having a symmetrically conductive . or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 4 kHz, or less than 15 dB at individual frequencies.
- . Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e. single sided deafness or "SSD"). The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3 kHz).
- Also indicated for any patient who is indicated for an air-conduction contralateral ● routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
{6}------------------------------------------------
The placement of a bone anchored implant is contraindicated for patient below the age of 5.
The Ponto sound processors are intended to be used with either the Ponto implant system or with specific compatible BAHA abutments/implants from Cochlear Bone Anchored Solutions (BAS) (refer to the Ponto labeling for specific compatible Cochlear models). In addition, selected Cochlear Baha sound processors can be used with the Ponto implant/abutment system (refer to the Ponto labeling for compatible BAHA sound processor models).
# Summary of Technological Characteristics
Minor modifications have been made to the drills previously cleared in 510(k) K152067. In the MONO Surgery Kit, the guide drill is removed, and the widening drill is modified to enable a single drill step to prepare the osteotomy for Ponto implant installation.
The MONO drill is provided in a kit packaging together with Cannula, Insertion indicator and Soft healing cap. The Cannula, Insertion indicator and Soft healing cap components included in the MONO Surgery Kit are identical to those included in the previously cleared MIPS Surgery Kit (K152067).
The technological principle and characteristics of the Ponto Bone Anchored Hearing System, Ponto Implant System, remain unchanged from the original design (most recently cleared in K152067). No changes have been made to the implant components within the Ponto Bone Anchored Hearing System compared to previously cleared K152067.
## Performance data
Pre-clinical bench tests were performed to compare the technical performance of the new drill against its precursors, that have proven to perform well in a clinical setting.
Comparative testing to install 4 mm Ponto implant (Ø4.5 mm) in artificial bone using the MONO Surgery Kit and the predicate devices were performed. The testing included heat generation, insertion torque, installation turns, seating and measurements of ISQ-value at implant level. The results were according to the requirements and confirmed equivalence with the predicate drills/drill protocol.
The above-mentioned tests confirm that the MONO procedure and MONO Surgery Kit are as safe and efficient as the predicate MIPS procedure and MIPS Surgery Kit (K152067). In all instances, the modified instruments functioned as intended and the performance was as expected. Hence, it has been concluded that further testing will not raise new issues of safety and efficacy.
{7}------------------------------------------------
## Substantial Equivalence Conclusions
The Ponto Bone Anchored Hearing System, including MONO Surgery Kit, is as safe and effective as the predicate Ponto Bone Anchored Hearing System (K152067). The Ponto Bone Anchored Hearing System has the same intended use and indications as previously cleared in K152067. Further, no changes have been made to the overall characteristics or principles of operation of the Ponto Bone Anchored Hearing System compared to the predicate device (K152067).
The minor technological differences between the MONO Surgery Kit and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the installation of Ponto Bone Anchored Hearing System implants using the MONO procedure is as safe and effective as the predicate installation procedures (K152067).
Accompanying documentation provide information on selection of surgical technique suited for the individual patient. The MONO Surgery Kit and MONO procedure do not alter the intended use of the Ponto Bone Anchored Hearing System and do not affect the safety and effectiveness of the Ponto Bone Anchored Hearing System when used as labelled.
Thus, the Ponto Bone Anchored Hearing System, including MONO Surgery Kit, is substantially equivalent to the previously cleared Ponto Bone Anchored Hearing System (K152067).
{8}------------------------------------------------
#### Substantial Equivalence Chart
| | Predicate 510(k) K152067 | This 510(k) K203807 |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Oticon Medical AB | Oticon Medical AB |
| Intended Use | The Ponto system is intended for improvement of<br>hearing for patients with conductive and mixed hearing<br>losses, bilateral fitting and single sided deafness. | The Ponto bone anchored hearing system's intended<br>use is for improvement of hearing for patients with<br>conductive or mixed hearing losses, whether<br>unilaterally or bilaterally fitted, or for those with single-<br>sided deafness. |
| Indications for Use | The Ponto bone anchored hearing system (Ponto sound<br>processors and implant system) is intended for the following<br>patients and indications:<br>• Patient with conductive or mixed hearing losses, who can<br>still benefit from amplification of the sound. The pure tone<br>average (PTA) bone conduction (BC) threshold (measured<br>at 0.5, 1, 2 and 3 kHz) of the indicated ear should be better<br>than or equal to 45 dB HL for use with the Ponto, Ponto<br>Pro and Ponto Plus sound processors, 55 dB HL for use<br>with Ponto Pro Power and Ponto Plus Power sound<br>processors.<br>• Bilateral fitting is applicable for most patients having a<br>symmetrically conductive or mixed hearing loss. The<br>difference between the left and right sides' BC thresholds<br>should be less than 10 dB on average measured at 0.5, 1,<br>2 and 4 kHz, or less than 15 dB at individual frequencies.<br>• Patients who have a profound sensorineural hearing loss in<br>one ear and normal hearing in the opposite ear (i.e. single<br>sided deafness or "SSD"). The pure tone average (PTA) air<br>conduction (AC) threshold of the hearing ear should then<br>be better than or equal to 20 dB HL (measured at 0.5, 1, 2<br>and 3 kHz).<br>• Also indicated for any patient who is indicated for an air-<br>conduction contralateral routing of signals (AC CROS)<br>hearing aid, but who for some reason cannot or will not use<br>an AC CROS.<br>The placement of a bone anchored implant is contraindicated<br>for patient below the age of 5.<br>The Ponto sound processors are intended to be used with<br>either the Ponto implant system or with specific compatible<br>Baha abutments/implants from Cochlear Bone Anchored<br>Solutions (BAS) (refer to the Ponto labeling for specific<br>compatible Cochlear models). In addition, selected Cochlear<br>Baha sound processors can be used with the Ponto<br>implant/abutment system (refer to the Ponto labeling for<br>compatible Baha sound processor models). | The Ponto bone anchored hearing system (Ponto sound<br>processors and implant system) is intended for the following<br>patients and indications:<br>• Patients with conductive or mixed hearing losses, who can<br>still benefit from amplification of the sound. The pure tone<br>average (PTA) bone conduction (BC) threshold (measured<br>at 0.5, 1, 2 and 3 kHz) of the indicated ear should be better<br>than or equal to 45 dB HL for use with the Ponto 3 and<br>Ponto 4 sound processors, 55 dB HL for use with the<br>Ponto 3 Power sound processors and 65 dB HL for use<br>with the Ponto 3 SuperPower sound processor.<br>• Bilateral fitting is applicable for most patients having a<br>symmetrically conductive or mixed hearing loss. The<br>difference between the left and right sides' BC thresholds<br>should be less than 10 dB on average measured at 0.5, 1,<br>2 and 4 kHz, or less than 15 dB at individual frequencies.<br>• Patients who have a profound sensorineural hearing loss in<br>one ear and normal hearing in the opposite ear (i.e. single<br>sided deafness or "SSD"). The pure tone average (PTA) air<br>conduction (AC) threshold of the hearing ear should then<br>be better than or equal to 20 dB HL (measured at 0.5, 1, 2<br>and 3 kHz).<br>• Also indicated for any patient who is indicated for an air-<br>conduction contralateral routing of signals (AC CROS)<br>hearing aid, but who for some reason cannot or will not use<br>an AC CROS.<br>The placement of a bone anchored implant is contraindicated<br>for patient below the age of 5.<br>The Ponto sound processors are intended to be used with<br>either the Ponto implant system or with specific compatible<br>Baha abutments/implants from Cochlear Bone Anchored<br>Solutions (BAS) (refer to the Ponto labeling for specific<br>compatible Cochlear models). In addition, selected Cochlear<br>Baha sound processors can be used with the Ponto<br>implant/abutment system (refer to the Ponto labeling for<br>compatible Baha sound processor models) |
| Technological<br>characteristics /<br>anatomical sites | The Ponto bone anchored hearing system consists of<br>an external sound processor unit and an implant with a<br>skin penetrating abutment. The implant is placed in or<br>around the mastoid region of the skull on either one or<br>both sides. The implant is then used as anchorage for<br>the skin penetrating abutment that connects to the<br>external sound processor. Vibrations generated by the<br>sound processor are transmitted directly through the<br>skull bone to the cochlea as bone conduction sound.<br>The sound processor has a coupling so that it can<br>easily be connected to and disconnected from the<br>abutment by the user. | Same / No change |
| | Predicate 510(k) K152067 | This 510(k) K203807 |
| Implant system<br>components | Implants<br>Abutments<br>Connection screw<br>Healing cap<br>Cover screw | Same / No change |
| Intended use; Implant | The implants are intended to be placed in bone tissue<br>for long term use (more than 25 years). The implants<br>are intended to integrate with the bone tissue and<br>provide a reliable anchorage for a sound processor. | Same / No change |
| Compatible sound<br>processors | All Oticon Medical sound processors and selected<br>sound processors from Cochlear Bone Anchored<br>Solutions.<br>See Ponto labeling for compatible Baha sound<br>processor models. | Same / No change |
| Surgical instruments | Single use drills:<br>Linear incision drill (Guide drill / Wide Countersink)<br>MIPS Surgery Kit<br>Reusable instruments | Single use drills:<br>Linear incision drill (Guide drill / Wide Countersink)<br>MIPS Surgery Kit<br>MONO Surgery Kit<br>Reusable instruments |
| Implant system<br>installation | Surgical procedure performed by an ENT surgeon. | Same / No change |
| Procedures for Ponto<br>implant installation | Three (3) alternative procedures available:<br>Linear incision, one-stage<br>Linear incision, two-stage<br>MIPS | Same, with addition of one (1) alternative procedure:<br>MONO |
| Selecting procedure | Linear incision, single-stage surgery is indicated for: Adult<br>patients with normal bone quality and thickness above 3 mm,<br>where no complications during surgery are expected.<br>Children with normal bone quality and a bone thickness above<br>4 mm (typically 12 years or older) provided that age,<br>development status and other known factors have been<br>considered and found suitable for single-stage surgery.<br>Linear incision, two-stage surgery is recommended<br>for/when: Adult patients with an expected bone depth below 3<br>mm or expected poor bone quality. (Reasons for expecting<br>poor bone quality or thin bone may for example include disease<br>or history of irradiation.)<br>Children with a bone thickness below 4 mm, or where age<br>development status or other factors make single-stage surgery<br>unsuitable.<br>An implant is placed in association with the removal of an<br>acoustic neuroma.<br>Contact with the dura mater or the wall of the sigmoid sinus is<br>expected, or if there is any risk of complications.<br>MIPS surgical procedure is indicated for: Adult patients with<br>normal bone quality and bone thickness above 3 mm, where no<br>complications during surgery are expected.<br>Children with normal bone quality and a bone thickness above<br>4 mm (typically 12 years or older) provided that age,<br>development status and other known factors have been<br>considered and found suitable for single-stage surgery.<br>Patients, as per above, with a skin thickness of 12 mm or less. | Same, with addition of:<br>MONO surgical procedure is indicated for: Adult patients (18<br>years and above) with normal anatomy and expected bone<br>thickness of at least 5mm, where no complications during<br>surgery are expected.<br>Patients, as per above, with a skin thickness of 12 mm or less.<br>Use of MONO surgical technique is contraindicated for children<br>and patients with expected bone thickness below 5 mm. |
| Implant material | Titanium, grade 4 | Same / No change |
| Implant surface | Smooth titanium surface<br>Laser ablation applied to parts of the thread valleys.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.