Eustachi Ear Pressure Relief Device
K203754 · Exercore, LLC · MJV · Feb 19, 2021 · EN
Device Facts
| Record ID | K203754 |
| Device Name | Eustachi Ear Pressure Relief Device |
| Applicant | Exercore, LLC |
| Product Code | MJV · EN |
| Decision Date | Feb 19, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic, Pediatric |
Indications for Use
The Eustachi is indicated for the treatment of negative middle ear pressure. Negative middle ear pressure can lead to fluid accumulation in the middle ear, impaired hearing, and hearing loss. The Eustachi provides a method of ventilating the middle ear by momentarily increasing the air pressure in the nose and the eustachian tube. Equalizing the pressure can prevent the accumulation of fluid and prevent hearing loss.
Device Story
Battery-operated handheld device; delivers controlled airflow to nasal cavity to facilitate eustachian tube opening. User places nosepiece against one nostril, pinches other nostril shut, activates device via power button, and swallows during operation. Airflow equalizes pressure in middle ear; prevents fluid accumulation and hearing loss. Used in home setting by patients or under adult supervision. Output is mechanical air pressure; informs clinical decision-making by providing non-invasive pressure relief. Benefits patient by restoring middle ear ventilation.
Clinical Evidence
No clinical data. Evidence consists of bench testing (dimensional, pressure, flow rate, drop tests), biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2), and usability testing (IEC 62366-1, IEC 60601-1-6) to validate OTC home use.
Technological Characteristics
Handheld, battery-powered (2 AA) device. Nosepiece constructed of ABS plastic. Delivers air pressure 31–41 kPa at 1.7–2.1 LPM. Surface-contact device for limited duration. Complies with IEC 60601-1, IEC 60601-1-11, and IEC 60601-1-2. Non-sterile.
Indications for Use
Indicated for treatment of negative middle ear pressure in patients experiencing fluid accumulation, impaired hearing, or hearing loss. Suitable for OTC home use by patients or under adult supervision.
Predicate Devices
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 19, 2021
Exercore, LLC % Lisa Pritchard Regulatory, Quality & Compliance Consultant DuVal & Associates. P.A. 825 Nicollet Mall, Suite 1840 Minneapolis, Minnesota 55402
Re: K203754
Trade/Device Name: Eustachi Ear Pressure Relief Device Regulatory Class: Unclassified Product Code: MJV Dated: December 22, 2020 Received: December 23, 2020
Dear Lisa Pritchard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Shu-Chen Peng Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203754
Device Name Eustachi Ear Pressure Relief Device
### Indications for Use (Describe)
The Eustachi is indicated for the treatment of negative middle ear pressure.
Negative middle ear pressure can lead to fluid accumulation in the middle ear, impaired hearing, loss. The Eustachi provides a method of ventilating the middle ear by momentarily increasing the air pressure in the nose and the eustachian tube. Equalizing the pressure can prevent the accumulation of fluid and prevent hearing loss.
| Type of Use (Select one or both, as applicable) | <table><tr><td><input checked="true" type="checkbox"/></td><td>One-Time Contribution (26 CFR 301.6114-2)</td></tr><tr><td><input type="checkbox"/></td><td>Recurring Use (26 CFR 301.6114-2)</td></tr></table> | <input checked="true" type="checkbox"/> | One-Time Contribution (26 CFR 301.6114-2) | <input type="checkbox"/> | Recurring Use (26 CFR 301.6114-2) |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|-------------------------------------------|--------------------------|-----------------------------------|
| <input checked="true" type="checkbox"/> | One-Time Contribution (26 CFR 301.6114-2) | | | | |
| <input type="checkbox"/> | Recurring Use (26 CFR 301.6114-2) | | | | |
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the ExerCore logo. The logo consists of a stylized letter "e" formed by two curved lines, one green and one blue, encircling each other. To the right of the symbol is the word "exercore" in a simple, sans-serif font, with all letters in lowercase and black.
# 510(k) Summary - K203754
#### I. SUBMITTER
Exercore, LLC 8170 Old Carriage Court North, Suite 200 Minneapolis, MN 55379
Phone: 888-406-0668
Contact Person: Kevin Connelly Date Prepared: February 16, 2021
- II. DEVICE
Name of Device: Eustachi® Ear Pressure Relief Device Common or Usual Name: Eustachi Classification Name: Middle Ear Inflation Device Regulatory Class: Unclassified Product Code: MJV
#### III. PREDICATE DEVICE
EarPopper, K073401 This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The Eustachi device is a battery-operated device designed to blow a controlled flow of air into the nose to facilitate opening the eustachian tubes. For use, the nosepiece is fitted tightly against one nostril to make an airtight seal. While pinching the other nostril shut, the Eustachi device is activated by pressing and holding the power button. The patient swallows while the device is running, allowing regulated air to move from the nose to the eustachian tube to help open the eustachian tube and equalize pressure in the middle ear.
{5}------------------------------------------------
Eustachi is provided non-sterile, and is designed for reuse with a single patient.
#### V. INDICATIONS FOR USE
The Eustachi is indicated for the treatment of negative middle ear pressure.
Negative middle ear pressure can lead to fluid accumulation in the middle ear, impaired hearing, and hearing loss. The Eustachi provides a method of ventilating the middle ear by momentarily increasing the air pressure in the nose and the eustachian tube. Equalizing the pressure can prevent the accumulation of fluid and prevent hearing loss.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Both the subject and predicate devices are designed for user by patients to relieve negative middle ear pressure. Both devices create a flow of air into the nose to eustachian tube, delivered at a specified pressure and volume. Both require the patient to swallow during use to facilitate movement of air into the eustachian tube. At a high level, the subject and predicate devices are based on equivalent technological characteristics as shown in Table 1. Validation from non-clinical testing and usability evaluation demonstrated that minor differences in technological characteristics do not raise any new questions of safety or effectiveness.
| Item | Subject Device<br>Eustachi | Primary Predicate Device<br>K073401<br>EarPopper |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Treatment of negative middle ear pressure | Treatment of negative middle ear pressure |
| Indications for Use | The Eustachi is indicated for the treatment of negative middle ear pressure.<br><br>Negative middle ear pressure can lead to fluid accumulation in the middle ear, impaired hearing, and hearing loss. The Eustachi provides a method of ventilating the middle ear by momentarily increasing the air pressure in the nose and the eustachian tube. Equalizing the pressure can | The EarPopper is indicated for the treatment of negative middle ear pressure.<br><br>Negative middle ear pressure can lead to fluid accumulation in the middle ear, impaired hearing, and hearing loss. The EarPopper provides a method of ventilating the middle ear by momentarily increasing the air pressure in the nose and the eustachian tube. Equalizing the pressure can prevent the |
Table 1: Comparison of Technological Characteristics
{6}------------------------------------------------
| Item | Subject Device<br>Eustachi | Primary Predicate Device<br>K073401<br>EarPopper |
|--------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| | prevent the accumulation of fluid<br>and prevent hearing loss. | accumulation of fluid and<br>prevent hearing loss. |
| Type of Use | Over-the-Counter | Prescription Use |
| Patient Use | Performed at home by patient on<br>their own or under adult<br>supervision | Performed at home by patient on<br>their own or under adult<br>supervision |
| Intended<br>Use<br>Environment | Home Use | Home Use |
| Components | Handheld unit | Handheld unit |
| Mechanism<br>of Action | Provides flow of air to open<br>eustachian tube | Provides flow of air to open<br>eustachian tube |
| Patient<br>Contact Type | Surface-contact device in contact<br>with mucosal tissue for limited<br>duration | Surface-contact device in contact<br>with mucosal tissue for limited<br>duration |
| User<br>Interface | Single button press on handheld<br>unit | Single button press on handheld<br>unit |
| Air Pressure | 31 – 41 kPa (4.5 to 6 PSI) | 3 – 6 PSI |
| Air Flow<br>Rate | 1.7 – 2.1 LPM | 0.6 – 1.3 LPM |
| Power<br>Supply | Alkaline Battery-powered (2 AA) | Alkaline Battery-powered (9V) |
| Sterility | Non-sterile | Non-sterile |
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination:
# Biocompatibility Testing
The biocompatibility evaluation of the Eustachi device was conducted in accordance with ISO 10993-1: 2018 (FDA recognition # 2-258) and FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process," issued June 16, 2018. The battery of testing included the following tests:
- Cytotoxicity
- Sensitization ●
- Irritation / Intracutaneous reactivity .
The Eustachi nosepiece (ABS plastic) is considered to be a surface-contact device in contact with mucosal tissue for limited cumulative duration (less than 24 hours).
{7}------------------------------------------------
# Cleaning Validation
Cleaning validation was conducted in accordance with AAMI TIR30:2011/(R)2016 to validate effectiveness of the Eustachi cleaning instructions.
### Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the Eustachi device. The device complies with applicable sections of IEC 60601-1: 2012 and IEC 60601-1-11: 2015 standards for electrical safety, and IEC 60601-1-2 edition 4 for EMC.
### Bench Testing
Bench testing was conducted to validate intended performance of the Eustachi device including dimensional verification, pressure value verification, flow rate verification, and drop test verification.
## Usability Testing
A study was conducted in accordance with IEC 62366-1: 2016 and IEC 60601-1-6: 2013 to verify the ability of potential users to complete the following tasks when no training is provided to simulate over-the-counter home use:
- Identify whether the device would be appropriate for them to use;
- Understand and complete the tasks required for use of the device; ●
- Understand when a healthcare professional should be consulted ●
### Software Verification and Validation
The Eustachi does not contain software; no software verification or validation was required.
### Animal Performance Study
No animal performance studies were required for the Eustachi device.
### Clinical Study
No clinical studies were required for the Eustachi device.
#### CONCLUSIONS VIII.
The biocompatibility, electrical safety, EMC, and bench testing conducted provide evidence that the Eustachi performs comparable to the predicate device. The Eustachi has the same intended use as the predicate device. The usability testing conducted provides evidence that the Eustachi labeling is sufficient to support over-the-counter (OTC) use by ensuring the device performs safely when used by lay users with no previous training. The data
{8}------------------------------------------------
provided supports substantial equivalence of the Eustachi to the predicate device.