← Product Code [MSC](/productcode/MSC) · K203727

# Harmony Latex Dam, Harmony Non-Latex Dam (K203727)

_Pamco Distributing, Inc. · MSC · Aug 5, 2021 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K203727

## Device Facts

- **Applicant:** Pamco Distributing, Inc.
- **Product Code:** [MSC](/productcode/MSC.md)
- **Decision Date:** Aug 5, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Harmony® Latex & the Harmony® Polyisoprene Dams are used as a barrier when engaging in oral/vaginal and oral/anal sex to help reduce the transmission of bodily fluids, harmful pathogens and sexually transmitted infections.

## Device Story

Harmony® Latex and Polyisoprene Dams are rectangular, non-porous polymer films; placed over vagina or anus during oral sexual contact; act as physical barrier to infectious agents; used by individuals during sexual activity; provide protection against transmission of bodily fluids and STIs; non-sterile; single-use; vanilla-scented; natural color.

## Clinical Evidence

Bench testing only. Testing included dimensional analysis, tensile strength, elongation at break, tearing force, tear resistance, and freedom from holes per ISO 29942:2011. Viral barrier properties and packaging integrity were verified. Biocompatibility testing performed per ISO 10993-1 included cytotoxicity (ISO 10993-5), irritation and sensitization (ISO 10993-10), and acute systemic toxicity (ISO 10993-11). All results were acceptable.

## Technological Characteristics

Rectangular polymer films (152.4 mm x 254 mm x 0.12 mm). Materials: Natural rubber latex or synthetic polyisoprene. Non-sterile. Vanilla fragrance. Mechanical testing per ISO 29942:2011. Biocompatibility per ISO 10993-1, 10993-5, 10993-10, and 10993-11.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- Trust Dam ([K091769](/device/K091769.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 5, 2021

Pamco Distributing, Inc. % Catriona Boyd Senior Medical Device Consultant Ironstone Product Development 250 Carlaw Avenue, Suite 108 Toronto. Ontario M4M 3L1 Canada

Re: K203727

> Trade/Device Name: Harmony® Latex Dams, Harmony® Polyisoprene Dams Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: MSC Dated: July 2, 2021 Received: July 6, 2021

Dear Catriona Boyd:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203727

Device Name Harmony® Latex Dams Harmony® Polyisoprene Dams

Indications for Use (Describe)

The Harmony® Latex & the Harmony® Polyisoprene Dams are used as a barrier when engaging in oral/vaginal and oral/ anal sex to help reduce the transmission of bodily fluids, harmful pathogens and sexually transmitted infections.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

# Harmony® Latex and Harmony® Polyisoprene Dams

# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Pamco Distributing Inc. 231 Arnold Street Kitchener, ON, Canada N2H 6E8 Phone: +1-519-648-2200 Contact Person: Annette Maclean

Date Prepared: August 3, 2021

#### Name of Device

Harmonv® Latex Dams Harmony® Polyisoprene Dams

#### Device Classification and Product Code

Regulation Number: 21 CFR 884.5300 Requlation Name: Condom Requlatory Class: Class II Product Code: MSC (Barrier, Std, Oral Sex)

## Predicate Device

Trust Dam (K091769)

The predicate has not been subject to a design-related recall.

#### Indications for Use

The Harmony® Latex & the Harmony® Polyisoprene Dams are used as a barrier when engaging in oral/vaginal and oral/anal sex to help reduce the transmission of bodily fluids, harmful pathogens and sexually transmitted infections.

## Device Description

The Harmony® Latex and Harmony® Polyisoprene Dams are rectangular shaped non-porous, intact polymer film that are placed over the vagina or anus during oral sexual contact. The dam provides coverage to the external female genitalia or to the anal acts as a physical barrier to infectious agents responsible for the transmission of sexually transmitted infections (STIs). Both Harmony® Latex and Harmony® Polyisoprene Dams are 152.4 mm (6") x 254 mm (10") and 0.12 mm thick. The Harmony® Latex Dam contains natural polyisoprene rubber and the Harmon® Polyisoprene Dam contains synthetic polyisoprene rubber. The dams are slightly scented with vanilla fragrance and are provided in natural color.

{4}------------------------------------------------

# Comparison of Indications for Use and Technological Characteristics with the Predicate Device

The intended use of the subject devices, Harmony® Latex & the Harmony® Polyisoprene Dams, and the predicate device, Trust Dams (K091769), are the same. Both the subject devices and the predicate device are used as a barrier when engaging in oral/vaginal and oral/anal sex to help reduce the transmission of bodily fluids, harmful pathogens and sexually transmitted infections. Table 1 below compares the technological characteristics of the subject devices and the predicate device.

|                             | Harmony® Dams<br>(Subject Devices)                                                                                                                                                                                                                   | Trust Dam<br>(Predicate Device)                                                                                                                                                                                    |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                 | Harmony® Latex Dams<br>Harmony® Polyisoprene Dams                                                                                                                                                                                                    | Trust Dams                                                                                                                                                                                                         |
| Manufacturer                | Pamco Distributing Inc.                                                                                                                                                                                                                              | Line One Laboratories Inc                                                                                                                                                                                          |
| 510(k) Number               | K203727                                                                                                                                                                                                                                              | K091769                                                                                                                                                                                                            |
| Regulatory Class            | II                                                                                                                                                                                                                                                   | II                                                                                                                                                                                                                 |
| Regulation Number           | 21 CFR 884.5300                                                                                                                                                                                                                                      | 21 CFR 884.5300                                                                                                                                                                                                    |
| Product Code                | MSC                                                                                                                                                                                                                                                  | MSC                                                                                                                                                                                                                |
| Classification Name         | Barrier, Std, Oral Sex                                                                                                                                                                                                                               | Barrier, Std, Oral Sex                                                                                                                                                                                             |
| Intended Use                | The Harmony® Latex & the<br>Harmony® Polyisoprene Dams are<br>used as a barrier when engaging in<br>oral/vaginal and oral/anal sex to help<br>reduce the transmission of bodily<br>fluids, harmful pathogens and<br>sexually transmitted infections. | The Line One Trust Dam is used as a<br>barrier when engaging in oral/vaginal<br>sex and oral/anal sex to help reduce<br>the transmission of bodily fluids,<br>harmful germs, and sexually<br>transmitted diseases. |
| Nominal Dimensions          | 152.4 mm X 254 mm, Thickness-<br>0.12 mm                                                                                                                                                                                                             | 154 mm X 250 mm, Thickness- 0.07<br>mm                                                                                                                                                                             |
| Material                    | Harmony® Latex Dams:<br>Natural rubber latex<br><br>Harmony® Polyisoprene Dams:<br>Synthetic polyisoprene                                                                                                                                            | Natural rubber latex                                                                                                                                                                                               |
| Lubricated                  | No                                                                                                                                                                                                                                                   | No                                                                                                                                                                                                                 |
| Color                       | Natural Color / No Dye Added                                                                                                                                                                                                                         | Yellow, Red, Green                                                                                                                                                                                                 |
| Fragrance/Flavour           | Vanilla                                                                                                                                                                                                                                              | Banana, strawberry, mint                                                                                                                                                                                           |
| Mechanical<br>Properties    | Tested in accordance with ISO<br>29942:2011, including the following<br>tests:<br>Tensile strength<br>Elongation at break<br>Tearing force<br>Freedom from holes<br>Tear resistance                                                                  | Tested in accordance with ASTM<br>D3492:2008                                                                                                                                                                       |
| Sterility                   | Non-sterile                                                                                                                                                                                                                                          | Non-sterile                                                                                                                                                                                                        |
| Shelf-life                  | 3 years                                                                                                                                                                                                                                              | 5 years                                                                                                                                                                                                            |
| Biocompatibility<br>Tested? | Yes                                                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                |

|  |  |  | Table 1 - Comparison of Technological Characteristics of the Subject and Predicate Devices |
|--|--|--|--------------------------------------------------------------------------------------------|
|--|--|--|--------------------------------------------------------------------------------------------|

{5}------------------------------------------------

The intended use of the Harmony® Latex & the Harmony® Polyisoprene Dams is the same as the Trust Dam. The differences between the subject and predicate device include the materials (polyisoprene), colors, flavors, and shelf life. However, these differences in technological characteristics do not raise different questions of safety or effectiveness.

# Summary Discussion of Non-Clinical Data

The following non-clinical testing was completed on the Harmony® Latex and Harmony® Polyisoprene Dams:

- . Per ISO 29942:2011 Prophylactic dams -- Requirements and test methods
	- o Dimensional Testing
	- Tensile Testing and Elongation at Break O
	- Tear Resistance and Tearing Force o
	- o Freedom from Holes and Visual Defects
- Viral Barrier Properties ●
- . Packaging Integrity Testing
- Per the 2020 FDA guidance Use of International Standard ISO 10993-1 "Biological evaluation of ● medical devices - Part 1: Evaluation and testing within a risk management process":
	- Cytotoxicity (ISO 10993-5:2009) o
	- Irritation Testing (ISO 10993-10:2010) o
	- Sensitization (ISO 10993-10:2010) o
	- Acute Systemic Toxicity (ISO 10993-11:2017) o

All test results were acceptable.

## Conclusions

The results from the non-clinical bench testing above demonstrate that the Harmony® Latex Dams and Harmony® Polyisoprene Dams are as safe and effective as the predicate device and support a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/device/K203727](https://fda.innolitics.com/device/K203727)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
