Solevy Co. LLC Water-Based Personal Lubricant

K203654 · Solevy Co., LLC · NUC · Apr 19, 2021 · Obstetrics/Gynecology

Device Facts

Record IDK203654
Device NameSolevy Co. LLC Water-Based Personal Lubricant
ApplicantSolevy Co., LLC
Product CodeNUC · Obstetrics/Gynecology
Decision DateApr 19, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2

Intended Use

Solevy Co. LLC Water-Based Lubricant is a water-based personal lubricant, for penile and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

Device Story

Solevy Co. LLC Water-Based Lubricant is a clear, colorless, semi-viscous personal lubricant. It is applied topically to the penis or vagina to provide moisture and lubrication, enhancing comfort during intimate sexual activity. The device is intended for over-the-counter use by consumers. It is compatible with natural rubber latex, polyurethane, and polyisoprene condoms. The product is supplied in polyethylene bottles (1-16 fl. oz.) with polypropylene caps or lotion pumps, sealed with induction seals or tamper-resistant shrink wrap. It functions as a physical lubricant to supplement natural body secretions.

Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-5, -10, and -11 (cytotoxicity, sensitization, vaginal irritation, acute systemic toxicity). Condom compatibility tested per ASTM D7661-10. Shelf-life validated via real-time aging study (3 years). Microbiological testing performed per USP <61>, <1111>, <51>, and <62>.

Technological Characteristics

Water-based personal lubricant. Ingredients: Glycerin, Water, Cellulose Gum, Methylparaben, Propylparaben. Physical properties: Viscosity 1,200-1,800 cps (USP <912>), pH 6.0-7.0, Osmolality 750-950 mOsm/kg (USP <785>), Specific Gravity 1.12-1.20 (USP <841>). Non-sterile. Packaging: Polyethylene bottles with polypropylene caps/pumps. Condom compatibility tested per ASTM D7661-10.

Indications for Use

Indicated for individuals requiring personal lubrication for penile and/or vaginal application to enhance comfort during sexual activity; compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. April 19, 2021 Solevy Co. LLC Joe Mendoza Ouality/Regulatory Manager 29017 Avenue Penn Santa Clarita, CA 91355 Re: K203654 > Trade/Device Name: Solevy Co. LLC Water-Based Lubricant Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: December 11, 2020 Received: December 15, 2020 Dear Joe Mendoza: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K203654 Device Name Solevy Co. LLC Water-Based Lubricant #### Indications for Use (Describe) Solevy Co. LLC Water-Based Lubricant is a water-based personal lubricant, for penile and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyurethane, and polyisoprene condoms. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary - K203654 | 510(k) Owner: | Solevy Co. LLC | | |---------------------------|----------------------------------------------|---------------------------------------| | Street Address: | 29017 Avenue Penn<br>Santa Clarita, CA 91355 | | | Contact Person: | Joe Mendoza<br>Quality/Regulatory Manager | | | Contact Number: | Phone: (661) 622-4880 | | | Summary Preparation Date: | April 16, 2021 | | | Trade Name: | Solevy Co. LLC Water-Based Lubricant | | | Common Name: | Personal Lubricant | | | Device Classification: | Regulation Name: | Condom | | | Regulation Number: | 21 CFR 884.5300 | | | Product Code: | NUC (Lubricant, Personal) | | | Device Class: | Class II | | Predicate Device: | Product Name: | JO H2O Water Based Personal Lubricant | | | 510(k) Number: | K150480 | | | Manufacturer: | United Consortium, Inc. | | | Product Code: | NUC | | | Device Class: | Class II | The predicate device has not been subject to a design-related recall. #### Device Description Solevy Co. LLC Water-Based Lubricant is a clear, colorless, semi-viscous personal lubricant that is compatible with condoms made of natural rubber latex, polyurethane, and polyisoprene. The device is a non-sterile lubricant, for penile and/or vaginal application, to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication The product is provided in volumes of 1, 2.5, 4.5, 8, and 16 fl. oz in polyethylene bottles. The 1-8 fl. oz. bottles are closed with polypropylene (PP) caps, while the 16 fl. oz. bottle includes a polypropylene lotion pump. The 16 fl. oz. bottle is sealed using tamper-resistant shrink wrap band. The 1-8 fl. oz. bottles are sealed using an induction seal constructed of aluminized mylar. {4}------------------------------------------------ The device specifications are listed in Table 1 below: #### Table 1: Device Specifications | Property | Specification | | |------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | Appearance | Clear, semi-viscous liquid | | | Color | Colorless | | | Odor | Odorless | | | Viscosity per USP<912> | 1,200 - 1,800 cps | | | pH | 6.0 - 7.0 | | | Osmolality per USP<785> | 750 - 950 mOsm/kg | | | Specific Gravity per USP<841> | 1.12 - 1.20 | | | Total aerobic microbial count<br>(TAMC) per USP <61> and <1111> | <100 cfu/g | | | Total yeast and mold count<br>(TYMC) per USP <61> and <1111> | <10 cfu/g | | | Antimicrobial effectiveness per<br>USP <51> | Meets USP <51> acceptance criteria for Category 2<br>products. Category 2, bacteria should show not less<br>than 2.0 log reduction at 14 days and no increase from<br>14-day count at the 28-day count. Yeast and molds<br>should show no increase from the initial calculated<br>count at 14 and 28 days | | | Presence of Pathogenic<br>organisms ( <i>Staphylococcus aureus</i> , <i>Pseudomonas aeruginosa</i> , and <i>Candida albicans</i> per USP <62>) | Absent | | #### Indications for Use Solevy Co. LLC Water-Based Lubricant is a water-based personal lubricant, for penile and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyurethane, and polyisoprene condoms. #### Predicate Device Comparison A comparison of the indications for use and technological characteristics of the subject and predicate device is summarized in Table 2 below: {5}------------------------------------------------ Table 2: Predicate Device Comparison | Feature | Solevy Co. LLC Water-Based<br>Lubricant (K203654) | JO H2O Water-Based Personal<br>Lubricant (K150480) | |----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | Solevy Co. LLC Water-Based<br>Lubricant is a water-based<br>personal lubricant, for penile<br>and/or vaginal application,<br>intended to lubricate and<br>moisturize, to enhance the ease<br>and comfort of intimate sexual<br>activity and supplement the<br>body's natural lubrication. This<br>product is compatible with<br>natural rubber latex,<br>polyurethane, and polyisoprene<br>condoms. | JO H20 Personal Lubricant is a<br>water-based personal lubricant,<br>for penile and/or vaginal<br>application, intended to<br>lubricate and moisturize, to<br>enhance the ease and comfort<br>of intimate sexual activity and<br>supplement the body's natural<br>lubrication. This product is<br>compatible with natural rubber<br>latex, polyurethane, and<br>polyisoprene condoms. | | Product Code | NUC | NUC | | Ingredients | Glycerin, Water, Cellulose Gum,<br>Methylparaben, Propylparaben | Glycerin, Water, Cellulose Gum,<br>Methylparaben, Propylparaben | | Appearance | Clear, colorless, semi-viscous<br>liquid | Clear, colorless, semi-viscous<br>liquid | | Water-Based | Yes | Yes | | Over-the-Counter Use | Yes | Yes | | Condom<br>Compatibility | Natural Rubber Latex,<br>Polyisoprene, Polyurethane | Natural Rubber Latex,<br>Polyisoprene, Polyurethane | | Biocompatibility<br>Tested | Yes | Yes | | Non-Sterile | Yes | Yes | | Shelf Life | 3 years | 3 years | The indications for use for the subject and predicate devices are the same; therefore, they have the same intended use. The subject and predicate devices have the same technological characteristics. As the technological characteristics of the subject and predicate device are the same, there are no differences raising different questions of safety and effectiveness. {6}------------------------------------------------ # Summary of Non-Clinical Performance Data ## Biocompatibility Solevy Co. LLC Water-Based Lubricant has undergone biocompatibility testing in accordance with the 2020 FDA guidance document "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process." Testing included: - . Cytotoxicity (ISO 10993-5:2009) - . Sensitization (ISO 10993-10:2010) - . Vaginal Irritation (ISO 10993-10:2010) - Acute Systemic Toxicity (ISO 10993-11:2006) ● The results of testing demonstrated the subject device is non-cytotoxic, non-sensitizing, nonirritating, and non-systemically toxic. # Shelf-Life The Solevy Co. LLC Water-Based Lubricant has a three-year shelf-life based on the results of a real-time aging study. The shelf-life study evaluated all device specifications listed above in Table 1, Device Specifications. The subject device met all device specifications over the stated shelf-life duration. #### Condom Compatibility The compatibility of the Solevy Co. LLC Water-Based Lubricant was evaluated with natural rubber latex, polyisoprene, and polyurethane condoms per ASTM D7661-10 Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms. The results of testing demonstrated that the Solevy Co. LLC Water-Based Lubricant is compatible with natural rubber latex, polyisoprene, and polyurethane condoms. #### Conclusion The results of performance testing described above demonstrate that the Solevy Water Based Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...