K203614 · Auris Health, Inc., A Johnson and Johnson Family Company · QNW · Mar 11, 2021 · Ear, Nose, Throat
Device Facts
Record ID
K203614
Device Name
Monarch Bronchoscope
Applicant
Auris Health, Inc., A Johnson and Johnson Family Company
Product Code
QNW · Ear, Nose, Throat
Decision Date
Mar 11, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The reprocessed Monarch Bronchoscope, used in conjunction with the Monarch Platform, is intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
Device Story
Reprocessed version of Monarch Bronchoscope; used with Monarch Platform robotic system. Device comprises inner scope and outer sheath with 4-way articulation; distal tip contains CMOS imager and LED illumination. Physician controls device via Monarch Platform interface; live video transmitted to display. Working channel allows introduction of biopsy or other instruments. Used in clinical settings for airway access; enables diagnostic/therapeutic interventions. Reprocessing involves soil removal, decontamination, and functional testing; device can be reprocessed once. Output is visual feedback for physician to guide navigation and procedures. Benefits include safe, effective airway access for diagnostic/therapeutic tasks.
Clinical Evidence
Bench testing only. Testing included cleaning validation, functional performance, system compatibility, electrical safety (AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-18), biocompatibility (ISO 10993-1:2018), and sterilization validation (ISO 11135:2018, ISO 10993-7:2008). All tests passed.
Technological Characteristics
Single-use flexible bronchoscope; 4-way articulation; CMOS imager; LED illumination; aluminosilicate glass lens. Materials identical to predicate. Sterilization: Ethylene oxide (ISO 11135:2018). Connectivity: Wired to Monarch Platform. Standards: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 10993-1:2018, ISO 11135:2018.
Indications for Use
Indicated for patients requiring bronchoscopic visualization and access to airways for diagnostic and therapeutic procedures.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 11, 2021
Auris Health Inc., a Johnson and Johnson Family Company Somi Ekwealor Staff Regulatory Affairs Analyst 150 Shoreline Drive Redwood City, California 94065
Re: K203614
Trade/Device Name: Monarch Bronchoscope Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: QNW Dated: December 8, 2020 Received: December 14, 2020
Dear Somi Ekwealor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shu-Chen Peng, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203614
Device Name Monarch Bronchoscope
Indications for Use (Describe)
The reprocessed Monarch Bronchoscope, used in conjunction with the Monarch Platform, is intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
| Type of Use (Select one or both, as applicable) | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|
| <table style="border: none;"><tr><td style="border: none;">☑ Remediation Use (Part 31, GER 321, Corrective Action)</td><td style="border: none;">☐ On-site Corrective Use (31, GER 321, Corrective Action)</td></tr></table> | ☑ Remediation Use (Part 31, GER 321, Corrective Action) | ☐ On-site Corrective Use (31, GER 321, Corrective Action) |
| ☑ Remediation Use (Part 31, GER 321, Corrective Action) | ☐ On-site Corrective Use (31, GER 321, Corrective Action) | |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K203614 510(k) Summary
### General Information
| 510(k) Submitter | Auris Health, Inc., a Johnson and Johnson Family<br>Company<br>150 Shoreline Drive<br>Redwood City, CA 94065 |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Original Device<br>Manufacturer | Auris Health, Inc., a Johnson and Johnson Family<br>Company<br>150 Shoreline Drive<br>Redwood City, CA 94065 |
| Reprocessed Device<br>Manufacturer | Auris Health, Inc., a Johnson and Johnson Family<br>Company<br>2383 Bering Dr.<br>San Jose, CA 95131 |
| FDA Registration Number | 3014447948 |
| Primary Correspondent | Somi Ekwealor, MSRS, RAC<br>Staff Regulatory Affairs Specialist<br>Johnson and Johnson, Robotics & Digital Solutions |
| Contact Information | Email: sekwealo@its.jnj.com<br>Phone: (408) 320-5385 |
| Date Prepared | 08 December 2020 |
# Device Identification
### Proposed Reprocessed Device:
| Proprietary Name | Monarch Bronchoscope |
|---------------------|--------------------------------------------------|
| Common Name | Reprocessed Bronchoscope (Flexible Or Rigid) |
| Classification Name | Bronchoscope (flexible or rigid) and accessories |
| Regulation Number | 21 CFR 874.4680 |
| Product Code | QNW |
| Regulatory Class | II |
| Model Number | MBR-000211-B |
### Predicate OEM Device:
| Proprietary Name | Monarch Platform |
|------------------------|--------------------------------------------------|
| Common Name | Bronchoscope (Flexible Or Rigid) |
| Premarket Notification | K193534 |
| Classification Name | Bronchoscope (flexible or rigid) and accessories |
| Regulation Number | 21 CFR 874.4680 |
| Product Code | EOQ |
| Regulatory Class | II |
| Model Number | MBR-000211-A |
{4}------------------------------------------------
### Device Description
The reprocessed Monarch Bronchoscope, MBR-000211-B, (hereafter referred to as "Proposed Device") is identical to the Original Monarch Bronchoscope, MBR-000211-A, (hereafter referred to as "Predicate Device"), which is a component of, and must be used with, the Monarch Platform, cleared under K193534. The Monarch Bronchoscope is connected to the robotic arms of the Monarch Platform to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The Proposed Device is a reprocessed single-use device that can be reprocessed up to one (1) time.
The Monarch Bronchoscope, consisting of the Inner Scope ("scope") and the Outer Sheath ("sheath"), has 4-way articulation controlled by continuous, direct, visual control of the physician using the Monarch Platform. The Proposed Device contains a working channel through which biopsy devices, or other working channel instruments, may be introduced. The distal tip of the Proposed Device has a camera control unit (CCU) that collects live images that are then transmitted to the physician's display interface of the Monarch Platform. The camera transmits vision data to the Monarch Tower through the camera cable. The single-use, manually controlled, working channel instruments compatible with the Proposed Device are identical to the working channel instruments compatible with the predicate Monarch Platform bronchoscope.
### Intended Use/Indications for Use
The reprocessed Monarch Bronchoscope, used in conjunction with the Monarch Platform, is intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
### Summary of Technological Characteristics
The Proposed Device (MBR-000211-B) is a single-use device identical to the Predicate Device (MBR-000211-A) in design, patient-contacting materials, clinical applications, patient population, performance specifications, and principles of operation. All technological characteristics including articulation, vision, and compatibility with working channel instruments are identical. Key technological characteristics are listed in the table below.
| Key Attributes | Proposed Device (MBR-0000211-B) |
|----------------------------|---------------------------------|
| Product Code | QNW |
| Regulation Number | 21 CFR 874.4680 |
| Classification | II |
| Intended for Single Use | Yes |
| Field of View (FOV) in air | 90 degrees |
| Direction of view | 0 degrees |
| FOV depth | 3-30mm |
| Imaging type | CMOS Imager |
| Illumination type | LED |
{5}------------------------------------------------
| Key Attributes | Proposed Device (MBR-0000211-B) |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Active angulation degrees up/down or<br>4 directions | 180/180/180/180 |
| Pixel resolution | 200 x 200 |
| Camera lens | Aluminosilicate glass |
| Light source | LED (covered in<br>cyanoacrylate adhesive) |
| Working Channel Instruments<br>Compatibility | Auris working channel instruments and third party<br>instruments that meet working channel length and<br>diameter requirements (e.g., REBUS probe) |
This premarket notification is submitted to demonstrate the ability to reprocess the Monarch Bronchoscope one (1) time without impacting safety or effectiveness. Auris Health's reprocessing of the device includes removal of adherent soil and decontamination. Each individual device is tested for appropriate function of its components prior to packaging, labeling, and sterilization operations.
### Performance Testing
The Reprocessed Monarch Bronchoscope was tested for performance in accordance with internal design specification and with the applicable performance standards to demonstrate safety and effectiveness. This includes the following tests:
| Test Name | Description | Results |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Cleaning Validation | Worst case device conditioning was taken into<br>consideration while conducting this testing. | Pass |
| Functional<br>Performance and<br>System<br>Compatibility | Functional performance and system compatibility were<br>performed to verify the performance of the Proposed Device<br>was not negatively impacted by reprocessing. Worst case<br>device conditioning was taken into consideration while<br>conducting this testing. | Pass |
| Electromagnetic<br>Compatibility and<br>Electrical Safety | The Proposed Device has been fully evaluated for electrical<br>safety and EMC compliance to the following standards:<br>AAMI/ANSI ES60601-1, IEC 60601-1-2, and IEC 60601-<br>2-18. Worst case device conditioning was taken into<br>consideration while conducting this testing. | Pass |
| Biocompatibility | Evaluates the end of life biocompatibility of the limited<br>contact device in accordance with ISO 10993-1:2018. Worst<br>case device conditioning was taken into consideration while<br>conducting this testing. | Pass |
| Sterilization | Sterilization was assessed in accordance with ISO 11135:<br>2018. Sterilization residuals were assessed in accordance<br>with ISO 10993-7:2008. The Proposed Device uses a<br>validated Ethylene oxide sterilization process using a half-<br>cycle overkill approach to achieve a minimum sterility<br>assurance level of 10-6. | Pass |
{6}------------------------------------------------
The performance testing demonstrates that the reprocessed Monarch Bronchoscope is as safe and effective as the legally marketed Original Monarch Bronchoscope and operates as originally intended.
#### Conclusion
The indications for use/intended use, patient-contacting materials, clinical applications, patient population, performance specifications, technological characteristics, and principles of operation of the Proposed Device are identical to the Predicate Device.
Performance testing and process validation results demonstrated substantial equivalence of the Proposed Device (reprocessed Monarch Bronchoscope, MBR-000211-B) to the Predicate Device (Original Monarch Bronchoscope, MBR-000211-A) with respect to safety and and effectiveness. therefore conclude that the reprocessed Monarch We Bronchoscope is as safe, effective, and substantially equivalent to the Predicate Device, Monarch Bronchoscope.
{7}------------------------------------------------
The following device is included in this submission:
| Description | OEM Model Number | Reprocessed Model Number |
|----------------------|------------------|--------------------------|
| Monarch Bronchoscope | MBR-000211-A | MBR-000211-B |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.