DynaFuse Fixation System

K203595 · MedShape, Inc. · HWC · Jan 4, 2022 · Orthopedic

Device Facts

Record IDK203595
Device NameDynaFuse Fixation System
ApplicantMedShape, Inc.
Product CodeHWC · Orthopedic
Decision DateJan 4, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The MedShape DynaFuse Fixation System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation of bones appropriate for the size of the device.

Device Story

DynaFuse Fixation System is a sterile, single-use orthopedic implant consisting of a threaded bone fastener and washer. Designed to apply compression across fracture or fusion sites; used for bone reconstruction, osteotomy, arthrodesis, and fracture fixation. Operated by surgeons in clinical/OR settings. Provides mechanical stabilization of bone segments to facilitate healing. Benefits patient by maintaining compression at the surgical site.

Clinical Evidence

Bench testing only. Testing included static torsion (ASTM F543), static strain (ASTM F2516), fatigue strain (ASTM E606), and bacterial endotoxin testing (ANSI/AAMI ST72:2011).

Technological Characteristics

System consists of threaded bone fastener and washer. Materials: titanium alloy and nickel-titanium. Mechanical fixation principle. Sterile, single-use. Testing performed per ASTM F543, ASTM F2516, ASTM E606, and ANSI/AAMI ST72:2011.

Indications for Use

Indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation in patients requiring bone fixation appropriate for the device size.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ January 4, 2022 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the full name. The full name of the agency is "U.S. Food & Drug Administration". MedShape, Inc. Ryan O'flaherty Product Manager 1575 Northside Drive NW, Suite 440 Atlanta, Georgia 30318 Re: K203595 Trade/Device Name: DynaFuse Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC. HTN Dated: November 29, 2021 Received: December 1, 2021 Dear Ryan O'flaherty: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K203595 Device Name DynaFuse Fixation System Indications for Use (Describe) The MedShape DynaFuse Fixation System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation of bones appropriate for the size of the device. Type of Use (Select one or both, as applicable) | Registration Use (Part 81 CFR 801.2 Subpart D) | <span></span> | |------------------------------------------------|---------------| | Over-The-Counter Use (81 CFR 801.3 Subpart E) | <span></span> | > Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the MedShape logo. The logo consists of a blue abstract shape on the left and the word "MEDSHAPE" in gray on the right. The abstract shape is a combination of a square and a circle. The word "MEDSHAPE" is written in a sans-serif font, and there is a trademark symbol after the word. # 510(k) Summary 510(k) Number: K203595 D. December 7th, 2020 Date Submitted: This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92. - A. Submitter: MedShape, Inc. 1575 Northside Drive, Suite 440 Atlanta, Georgia 30318 - B. Company Contact: Ryan O'Flaherty Product Manager (404) 249-9155 (404) 249-9158 (fax) ryan.oflaherty@medshape.com - C. Device Information: DynaFuse™ Fixation System Trade Name: Common Name: Smooth or threaded metallic bone fixation fastener - Device Classification: Class II 21 CFR 888.3040 (Primary) Smooth or threaded metallic bone fixation fastener Panel Code: 87, Orthopedic Panel Product Code: HWC Class II 21 CFR 888.3030 Single/multiple component metallic bone fixation appliances and accessories Panel Code: 87, Orthopedic Panel Product Code: HTN - E. Predicate Devices: Trilliant Surgical Tiger Cannulated Screw, K153338 (Primary) Tyber Medical Trauma Screw, K133842 and K153575 - F. Referenced Predicate Devices: MedShape DynaNail Mini, K182677 {4}------------------------------------------------ #### G. Physical Description: The proposed MedShape DynaFuse Fixation System is a sterile, single use orthopedic implant system consisting of a threaded bone fastener and washer. The DynaFuse Fixation System is designed to apply compression across a target fracture or fusion site and is intended to be used for fracture and osteotomy fixation, including joint arthrodesis. The devices provided sterile, intended for single use, and are comprised of titanium alloy and nickel-titanium. #### H. Indications for Use: The MedShape DynaFuse Fixation System is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation of bones appropriate for the size of the device. - l. Comparison of Characteristics / Performance Testing / Substantial Equivalence: The MedShape DynaFuse Fixation system and its predicate devices have similar indications for use, design, materials, and technological principles of operation. The following non-clinical tests were performed to demonstrate substantial equivalence to the predicate devices: - Static Torsion Test per ASTM F543 - Static Strain Test per ASTM F2516 - Fatigue Strain Test per ASTM E606 - Bacterial endotoxin testing based on an endotoxin limit of 20EU/device per ANSI/AAMI ST72:2011 Additional engineering analysis of the MedShape DynaFuse Fixation System and the predicate devices was also performed to demonstrate substantial equivalence. No new questions of safety or effectiveness were identified during device testing and analysis; therefore, the DynaFuse Fixation System is considered substantially equivalent to the predicate devices. RET Ryan O'Flaherty Product Manager (404) 249-9155 (404) 249-9158 (fax) ryan.oflaherty@medshape.com Image /page/4/Picture/15 description: The image shows the logo for MedShape. The logo consists of a blue abstract shape on the left and the word "MEDSHAPE" in gray on the right. The abstract shape appears to be a combination of a square and a circle, with the square positioned above the circle.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...