ArtiSential Laparoscopic Instruments-Electrodes

K203580 · Livsmed, Inc. · GEI · Apr 30, 2021 · General, Plastic Surgery

Device Facts

Record IDK203580
Device NameArtiSential Laparoscopic Instruments-Electrodes
ApplicantLivsmed, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateApr 30, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

ArtiSential Laparoscopic Instruments-Electrodes, Monopolar Series are indicated for cutting and coagulation in endoscopic, gynecological, and general abdominal and thoracic and general laparoscopic procedures.

Device Story

Sterile, single-use, invasive laparoscopic instrument; used for monopolar electrosurgical cutting and coagulation. Device connects to standard electrosurgical power source via cable. Operator (surgeon) uses device to manipulate tissue (dissection/scissors) while applying radiofrequency energy. Output is thermal effect on tissue for cutting/coagulation. Used in OR setting. Benefits include precise tissue management in minimally invasive procedures.

Clinical Evidence

Bench testing only. Includes biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-18, 60601-2-2), sterilization (ISO 11135), shelf-life/packaging (ASTM F88, ASTM F1929-15, ISO 11737-2), and thermal effect testing on liver, kidney, and muscle tissue at 10W, 40W, and 300W.

Technological Characteristics

Monopolar electrosurgical instrument; radiofrequency energy; shaft diameters 8mm; lengths 250mm, 380mm, 450mm; jaw types: dissector, scissors; EO sterilization; single-use; non-software device.

Indications for Use

Indicated for cutting and coagulation in endoscopic, gynecological, and general abdominal and thoracic and general laparoscopic procedures. Prescription use only.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The word "FDA" is in a larger, bolder font than the rest of the text. The overall design is clean and professional, reflecting the FDA's role in regulating food and drugs in the United States. April 30, 2021 LivsMed Inc. Dong Wook Lee QMR (Quality Management Representative) #304, D-dong, 700, Pangyo-ro, Bundang-gu 13516 Seongnam-si, Gyeonggi-do Republic of Korea Re: K203580 Trade/Device Name: ArtiSential Laparoscopic Instruments-Electrodes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: March 5, 2021 Received: March 8, 2021 Dear Dong Wook Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdth/cfdocs/cfpmn/pmn.cfm identifies cocombination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-DICE website the assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K203580 Device Name ArtiSential Laparoscopic Instruments-Electrodes Indications for Use (Describe) ArtiSential Laparoscopic Instruments-Electrodes, Monopolar Series are indicated for cutting and coagulation in endoscopic, gynecological, and general abdominal and thoracic and general laparoscopic procedures. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary ## 1. General Information | Applicant/Submitter: | LivsMed Inc. | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address: | #304, D-dong, 700, Pangyo-ro, Bundang-gu, 13516<br>Seongnam-si, Gyeonggi-do, Republic of Korea<br>Tel) +82-70-4282-7652<br>Fax) +82-31-706-3211 | | Contact Person: | Dong Wook Lee / QMR<br>(Quality Management Representative) | | Address: | #304, D-dong, 700, Pangyo-ro, Bundang-gu, 13516<br>Seongnam-si, Gyeonggi-do, Republic of Korea<br>Tel) +82-70-7709-4993<br>Fax) +82-31-706-3211<br>Email) dongwook.livsmed@gmail.com | | Preparation Date: | 01-27-2021 | ## 2. Device Name and Code | Device Trade Name | ArtiSential Laparoscopic Instruments-Electrodes | |---------------------|------------------------------------------------------| | Common Name | Electrosurgical Instruments | | Classification Name | Electrosurgical, cutting & coagulation & accessories | | Product Code | GEI | | Regulation Number | 21 CFR 878.4400 | | Classification | Class II | | Review Panel | General & Plastic Surgery | ## 3. Predicate Device and Reference Device ArtiSential Laparoscopic Instruments-Electrodes, Monopolar Series are substantially equivalent to the following devices. Table 3.1 Predicate device | Applicant | Device Name | 510(k) Number | |--------------|----------------------------------------------------|---------------| | LivsMed Inc. | ArtiSential Laparoscopic<br>Instruments-Electrodes | K200501 | {5}------------------------------------------------ Table 3.2 Reference Device | Applicant | Device Name | 510(k) Number | |-----------|------------------------------------------------------------|---------------| | DANNIK | DANNIK Disposable<br>Monopolar Laparoscopic<br>Instruments | K201063 | ## 4. Device Description This product is sterile, single-use, invasive instrument that used in laparoscopic surgery. There are two Version, one for Dissector and Scissors. The two models are same except for jaw. This product is a specific component, but not the entire electrosurgical device. The device is not intended to be marketed with multiple components, accessories, and as part of a system. ### 5. Indications for Use - 5.1 Indications for use ArtiSential Laparoscopic Instruments-Electrodes, Monopolar Series are indicated for cutting and coagulation in endoscopic, gynecological, and general abdominal and thoracic and general laparoscopic procedures. ### 6. Technical Characteristics in Comparison to Predicate Device and Reference Device Table 6.1 Substantial Equivalence Assessment of Proposed device and predicate device and reference device | | Proposed<br>device | Predicate<br>Device | Reference<br>device | Substantial<br>Equivalence | |--------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|------------------------------------------------------------------|----------------------------| | 510(K) Number | K203580 | K200501 | K201063 | N/A | | Manufacture | LivsMed Inc. | LivsMed Inc. | DANNIK | N/A | | Device Name | ArtiSential<br>Laparoscopic<br>Instruments-<br>Electrodes | ArtiSential<br>Laparoscopic<br>Instruments-<br>Electrodes | DANNIK<br>Disposable<br>Monopolar<br>Laparoscopic<br>Instruments | N/A | | Clearance Date | N/A | 05-11-2020 | 06-09-2020 | N/A | | Classification /<br>Regulation | Class 2 /<br>878.4400 | Class 2 /<br>878.4400 | Class 2 /<br>878.4400 | N/A | | Product Code | GEI | GEI | GEI | Same | | Intended for | Prescription<br>Use | Prescription<br>Use | Prescription<br>Use | Same | {6}------------------------------------------------ | Indications for Use | ArtiSential<br>Laparoscopic<br>Instruments-<br>Electrodes,<br>Monopolar<br>Series have<br>application in<br>endoscopic,<br>gynecological,<br>and general<br>abdominal and<br>thoracic and<br>general<br>laparoscopic<br>procedures.<br>When<br>connected by a<br>standard cable<br>to an<br>electrosurgical<br>power source,<br>the device may<br>be utilized for<br>monopolar<br>cautery. | ArtiSential<br>Laparoscopic<br>Instruments-<br>Electrodes,<br>Monopolar<br>Series have<br>application in<br>endoscopic,<br>gynecological,<br>and general<br>abdominal and<br>thoracic and<br>general<br>laparoscopic<br>procedures.<br>When<br>connected by a<br>standard cable<br>to an<br>electrosurgical<br>power source,<br>the device may<br>be utilized for<br>monopolar<br>cautery. | The DANNIK<br>Disposable<br>Monopolar<br>Laparoscopic<br>Instrument<br>have<br>application in a<br>variety of<br>minimally<br>invasive<br>procedures to<br>facilitate<br>grasping,<br>mobilization,<br>dissection and<br>transection of<br>tissue. | Same | |--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------| | Energy Type | Radiofrequency | Radiofrequency | Radiofrequency | Same | | Electrode type<br>(monopolar or<br>bipolar) | Monopolar | Monopolar | Monopolar | Same | | Jaw type | Dissector,<br>Scissors | Hook, Spatula | Dissector,<br>Scissors | Same | | Physical<br>dimensions and<br>design (size,<br>length) | - Shaft<br>diameter: 8mm<br><br>- Shaft Length:<br>250mm,<br>380mm,<br>450mm | - Shaft<br>diameter: 8mm<br><br>- Shaft Length:<br>250mm,<br>380mm,<br>450mm | - Shaft<br>diameter: 3mm,<br>5mm<br><br>- Shaft Length:<br>330mm | Same | | Sterilization | EO | EO | EO | Same | # 7. Performance Data - 7.1 Biocompatibility The device has been evaluated for its biological safety according to ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process". Following endpoints have been assessed during the evaluation: - Cytotoxicity = Page 3 / 5 {7}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for LIV'SMED. The logo is in a dark blue color. The apostrophe in the logo is a lighter blue color. - Sensitization = - ் Intracutaneous reactivity - Acute systemic toxicity - ### 7.2 Electrical Safety The ArtiSential Laparoscopic Instruments-Electrodes, Monopolar Series have been tested according to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-18 and IEC 60601-2-2. The device had passed all performed tests. ### 7.3 Sterilization ArtiSential Laparoscopic Instruments-Electrodes, Monopolar Series are provided sterile, intended to be single-use. This product is EO-Sterilization in accordance with ISO-11135. ### 7.4 Shelf life The proposed expiration date is 2 years from the manufacturing date. Following tests were conducted both zero time point and 2 years accelerated aging point and the device had passed all tests as below. - 1) Packaging and sterility test - . Packaging integrity test - . Sealing strength test of packaging (per ASTM F 88) - . Dye penetration test (per ASTM F1929-15) - . Sterility test (per ISO 11737- 2:2009) - 2) Performance test - . Appearance - Dimension . - . Operational test - . Tensile Strength - . Feedthrough test ### 7.5 Performance test {8}------------------------------------------------ The device had passed all performed tests as below. - . Appearance - . Dimension - Operational test . - Tensile Strength . - Feedthrough test . ### 7.6 Thermal effect Thermal effects testing was conducted on liver, kidney, abdominal muscle at three different power settings (10, 40, and 300W). Based on these performance characteristics, the results demonstrate that the performance requirements were met, the device performs as intended and that the subject device has substantially equivalent performance characteristics to the predicate device. ### 8. Substantial Equivalence Indication for use, principles of operation, energy type, electrode type, jaw type, physical dimension, rated voltage, materials (electrode, shaft), articulating feature, tip rotation, sterilization method of the proposed ArtiSential Laparoscopic Instruments-Electrodes, Monopolar Series are as same as the predicate device. Although there is a minor difference of raw material (insulation) between the ArtiSential Monopolar Series and the predicate device, the difference does not raise new or different questions of safety and efficacy. There is no new technology and no difference that would raise new or different questions of safety or efficacy. ### 9. Conclusions In conclusion, the comparison carried out covers all products, models, sizes, and the entire intended purpose of the device under evaluation. The subject device which is the ArtiSential Laparoscopic Instruments-Electrodes, Monopolar Series are considerably as same as the predicate device in principles of operation, technological characteristics, as well as performance characteristics. The testing was conducted to evaluate the performance of subject device in comparison to the predicate device. Results of validation and verification activities in design control that include testing/certification to designated standards and performance testing of the device has demonstrated substantial equivalence of the subject device to the predicate in terms of safety and effectiveness for requested intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...