K203535 · Shimadzu Corporation Medical Systems · OWB · Apr 28, 2021 · Radiology
Device Facts
Record ID
K203535
Device Name
Trinias
Applicant
Shimadzu Corporation Medical Systems
Product Code
OWB · Radiology
Decision Date
Apr 28, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Trinias is an angiographic X-ray system, which is used for diagnostic imaging and interventional procedures. The Trinias is intended to be used for cardiac angiography, neurovascular angiography, abdominal angiography, peripheral angiography, rotational angiography, multi-purpose angiography and whole body radiographic/fluoroscopic procedures
Device Story
Trinias is an angiographic X-ray system for diagnostic and interventional procedures. It captures X-ray images using a digital flat panel detector and high-voltage generator; images are processed by the DAR-9500f digital angiography system. The system includes a patient table (KS-100), C-arm (MH-500/MH-600), and collimator (F-100). Operated by clinicians in clinical settings, the system provides real-time fluoroscopic and radiographic visualization to guide interventional procedures. Output is displayed on monitors for physician review, facilitating clinical decision-making during minimally invasive interventions. Benefits include high-resolution imaging for complex vascular and cardiac procedures. Modifications include updated user interfaces, new table/C-arm models, and expanded flat panel detector options.
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through bench testing, standards compliance (electrical safety, electromagnetic compatibility, radiation protection, and usability), and software validation.
Indicated for patients requiring diagnostic imaging and interventional procedures, including cardiac, neurovascular, abdominal, peripheral, and rotational angiography, as well as whole-body radiographic/fluoroscopic procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shimadzu Corporation Medical Systems % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
Re: K203535
Trade/Device Name: Trinias Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB Dated: March 29, 2021 Received: March 30, 2021
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
April 28, 2021
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-leam). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K203535
Device Name Trinias
Indications for Use (Describe)
The Trinias is an angiographic X-ray system, which is used for diagnostic imaging and interventional procedures. The Trinias is intended to be used for cardiac angiography, neurovascular angiography, abdominal angiography, peripheral angiography, rotational angiography, multi-purpose angiography and whole body radiographic/fluoroscopic procedures
| Type of Use (Select one or both, as applicable) | | | | |
|-------------------------------------------------|---------------------------------------------|--|--|--|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary: 510(k) Number K203535
# 4 SHIMADZU
Shimadzu Corporation 1, Nishinokyo-Kuwabaracho, Nakagyo-ku, Kyoto, 604-8511, Japan Registration Number: 8030233 Phone: +81-75-823-1305 Fax : +81-75-823-1377 Date Prepared: March 29, 2021 Contact: Shigeru Naganishi; Phone: +81-75-823-1920 E-mail: shige@shimadzu.co.jpshige@shimadzu.co.jp
## 1. Identification of the Device:
Trade/Device Names: Trinias Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: II Product Code: OWB Common/Usual Name: Interventional Fluoroscopic X-Ray System
# 2. Equivalent legally marketed device: K123508
Trade/Device Name: Trinias Manufacturer: Shimadzu. Regulation Number: 21CFR892.1650 Regulation Name: Interventional Fluoroscopic X-Ray System Regulatory Class: II Product Code: OWB Common/Usual Name: Interventional Fluoroscopic X-Ray System
- 3. Indications for Use (intended use): The Trinias is an angiographic X-ray system, which is used for diagnostic imaging and interventional procedures. The Trinias is intended to be used for cardiac angiography, neurovascular angiography, abdominal angiography, peripheral angiography, rotational angiography, multi-purpose angiography and whole body radiographic/fluoroscopic procedures.
# 4. Description of the Device:
This notification is for a modified device. The modifications are: Updated user interfaces (wireless mouse, keyboard) A new model of catheterization table A new type of digital system console Additional x-ray tube choices Add alternate choices for the same sizes of digital flat panel detectors An additional size of available flat panel detector (12" x 16") An additional type of control cabinet. Model identifications: KS-70 Patient Table KS-100 New Patient Table F-100 New Collimator LX-6308, LX-6311, E79039X: X-ray Tubes DAR-9500f, Digital Angiography System; Flat panel. New MH-500 Suspended C-Arm
{4}------------------------------------------------
New MH-600 Lateral C-Arm D150GC-40 High Voltage Generator
- 5. Safety and Effectiveness, comparison to predicate device. The results of bench and standards testing indicates that the new device is as safe and effective as the predicate device. Proper system operation is fully verified upon installation. Most of the components employed are identical to the predicate device.
| | Trinias K123508 | Modified Trinias |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use: | The Trinias is an angiographic X-ray system,<br>which is used for diagnostic imaging and<br>interventional procedures as described in 21<br>CFR 892.1650. The Trinias is intended to be<br>used for cardiac angiography, neurovascular<br>angiography, abdominal angiography,<br>peripheral angiography, rotational<br>angiography, multi-purpose angiography and<br>whole body radiographic/fluoroscopic<br>procedures | The Trinias is an angiographic X-ray system, which<br>is used for diagnostic imaging and interventional<br>procedures. The Trinias is intended to be used for<br>cardiac angiography, neurovascular angiography,<br>abdominal angiography, peripheral angiography,<br>rotational angiography, multi-purpose<br>angiography and whole body<br>radiographic/fluoroscopic procedures. (SAME<br>except CFR reference removed for international<br>consistency) |
| Patient Table | Image: KS-70 patient table | Newly available: KS-100, table top can tilt.<br>Image: KS-100 patient table |
| | C-arms | |
| Ceiling | Image: MH-200S C-arm<br>MH-200S | Image: MH-500 C-arm<br>MH-500 |
| | Trinias K123508 | Modified Trinias |
| Floor | MH-300<br>Image: MH-300 | MH-600<br>Image: MH-600 |
| | MH-300/400 Biplane<br>Image: MH-300/400 Biplane | MH-300/400 Biplane is still available under<br>K123508. |
| Digital X-Ray Receptor Panels Comparison | | |
| SIZE: | Shimadzu Reference/<br>Manufacturer Model | Shimadzu Reference/<br>Manufacturer Model |
| 9" x 9" | PANEL MAIN ASSY, A-9 | (No longer available) |
| 17" x 17" | PANEL MAIN ASSY, S-17 | (No longer available) |
| 8" x 8" | SFD-0808<br>Varex PaxScan 2020+ (194μm) | SFD-0808AF<br>Varex PaxScan 2020X (194μm) (Updated version<br>of the predicate) |
| 12" x 12" | SFD-1212<br>Varex PaxScan 3030+ 194μm | SFD-1212AF<br>Varex PaxScan 3030X (194μm) (Updated version<br>of the predicate) |
| 16" x 12" | This size was not available | SFD-1612AF<br>Varex PaxScan 4030X (194μm) (Updated version<br>of the predicate) |
| X-Ray Generation | | |
| | Trinias K123508 | Modified Trinias |
| Generator | 100 kW | SAME |
| Model # | D150GC-40 | Same as predicate |
| Control Method | 50kHz Inverter | Same as predicate |
| Rated output | 100kW | Same as predicate |
| Radiography tube Voltage | 40kV - 150kV | Same as predicate |
| Radiography tube current | 10 to 1250mA | Same as predicate |
| Radiography mAs | 0.5 to 800mAs | Same as predicate |
| Radiography time | 0.001 to 10sec | Same as predicate |
| Fluoroscopy tube voltage | 50kV to 125kV | Same as predicate |
| Fluoroscopy tube current | 0.3 to 30mA | 0.3 to 38mA |
| Short time rating | 100kv - 1000mA | Same as predicate |
| Long time rating | 100kv - 30mA | 100kV - 38mA |
| Collimator | F-50: Has 4 BH (beam hardening) filters with Irradiation area of 400 x 400 mm max | F-100: Has 16 BH filters with irradiation area of 421 x 421 mm max. New collimator offers more filtration options. |
| Safety Standards | | |
| US Performance Standard | 21CFR1020.30, 21CFR1020.31 and 21CFR2020.32 | SAME |
| IEC Safety Standards | See list below | SAME |
# 6. Substantial Equivalence Chart: Below.
{5}------------------------------------------------
{6}------------------------------------------------
- 7. Summary of non-clinical testing: Software was validated according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005. Because the system uses Wi-Fi and Ethernet, we observed the recommendations contained in the FDA Guidance Document: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014. The digital panel software employed was already reviewed by FDA. Labeling was developed and information provided in accordance with this FDA Guidance Document: Pediatric Information for X-ray Imaging Device Premarket Notifications, Guidance for Industry and Food and Drug Administration Staff, November 2017. Labeling also includes reference to the Image Gently website (http://www.imagegently.org/). Because the device contains wireless technology, we consulted Radio Frequency Wireless Technology in Medical Devices - Guidance for Industry and FDA Staff, AUGUST 2013 and we incorporated those recommendations into our labeling.
{7}------------------------------------------------
This device has been tested and is certified to comply with the US Radiation Safety Performance Standards as listed in the table above. Performance and safety testing was conducted by third party NRTL certified testing laboratories and the device was found to comply with the following FDA recognized standards:
- US Performance Standard 21CFR1020.30, 21CFR1020.31 and 21CFR2020.32 .
- IEC 60601-1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012 Medical electrical equipment Part . 1: General requirements for basic safety and essential performance FDA 19-4
- 0 IEC 60601-1-2:2014 Collateral Standard: Electromagnetic disturbances - Requirements and tests FDA 19-8
- 9 IEC 60601-1-3:2008 + A1:2013 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic Xray equipment FDA 12-269
- . IEC 60601-1-6:2010 + A1:2013 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability FDA 5-89
- IEC 60601-2-43:2010 + A1:2017 Particular requirements for the basic safety and essential . performance of X-ray equipment for interventional procedures FDA 12-308
- Additionally evaluated in accordance with .
- IEC 62366: 2007 + A1: 2014 Medical devices Part 1: Application of usability engineering to medical 0 devices FDA 5-114
- IEC 62304: 2006 Medical device software Software life cycle processes FDA 13-79 .
- . EN 60601-1:2006 + A11:2011 + A1:2013 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . EN 60601-1-3:2008 + A1:2013 + A11:2016 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- . EN 60601-1-6:2010 + A1:2015 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- EN 60601-2-43:2010 + A1:2018 Medical electrical equipment Part 1-6: General requirements for . basic safety and essential performance - Collateral standard: Usability
- . EN 62304:2006 Medical device software - Software life cycle processes
- ANSI/AAMI ES60601-1:2005 + A2:2010 + A1:2012 .
We consulted the following FDA guidance documents in the development of the Trinias: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, Guidance for Industry and Food and Drug Administration Staff; and Pediatric Information for X-ray Imaging Device Premarket Notifications.
- 8. Summary of clinical testing: Not applicable. Clinical testing was not deemed to show substantial equivalence. We relied on non-clinical testing and compliance with standards.
- 9. Conclusion: After analyzing standards compliance results and bench tests, it is the conclusion of Shimadzu Corporation that the MODIFIED Trinias is as safe and effective as the predicate device, has few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.