The Albumin BCP2 assay is used for the quantitation of albumin in human serum or plasma on the ARCHITECT c System. The Albumin BCP2 assay is to be used as an aid in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Device Story
Automated clinical chemistry assay; utilizes bromocresol purple (BCP) reagent; binds specifically to human albumin in serum or plasma samples; forms colored complex; absorbance measured at 604 nm; absorbance directly proportional to albumin concentration. Used on ARCHITECT c8000 System in clinical laboratory settings; operated by trained laboratory personnel. Provides quantitative albumin results to clinicians; assists in diagnosis and management of liver and kidney disorders.
Clinical Evidence
No clinical data. Performance established via bench testing: precision (CLSI EP05-A3), linearity (CLSI EP06-A), analytical specificity/interference, and method comparison (n=127 samples, R=1.00, slope=1.00).
Technological Characteristics
Colorimetric assay; active ingredient Bromocresol Purple (0.154 g/L); pH 5.4; automated on ARCHITECT c8000 System; analytical measuring interval 0.3–9.0 g/dL; extended measuring interval 9.0–22.4 g/dL; traceable to ERM-DA470/IFCC standard.
Indications for Use
Indicated for the quantitation of albumin in human serum or plasma to aid in the diagnosis and treatment of liver or kidney diseases. No specific age or gender restrictions provided.
Regulatory Classification
Identification
An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Predicate Devices
Albumin BCP (k981814)
Related Devices
K203248 — Albumin BCG2 · Abbott Ireland Diagnostics Division · Nov 23, 2021
K193001 — Albumin BCP · SENTINEL CH. SpA · Dec 19, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo features the Department of Health & Human Services seal on the left. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION".
March 18, 2022
Abbott Ireland Diagnostics Division Suzanne Cheang Regulatory Affairs Manager Lisnamuck Longford, Ireland
Re: K203530
Trade/Device Name: Albumin BCP2 Regulation Number: 21 CFR 862.1035 Regulation Name: Albumin Test System Regulatory Class: Class II Product Code: CJW Dated: December 17, 2021 Received: December 20, 2021
Dear Suzanne Cheang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) k203530
Device Name Albumin BCP2
Indications for Use (Describe)
The Albumin BCP2 assay is used for the quantitation of albumin in human serum or plasma on the ARCHITECT c System.
The Albumin BCP2 assay is to be used as an aid in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5: 510(k) Summary (Summary of Safety and Effectiveness)
This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### I. 510(k) Number
k203530
### II. Applicant Name
Abbott Ireland Diagnostics Division Lisnamuck, Longford Longford, IE
Primary contact person for all communications:
Suzanne Cheang, Manager, Regulatory Affairs Abbott Diagnostics Division Phone (972) 518-6617 Fax (972) 518-7498
Secondary contact person for all communications:
Magdalena Suszko, Manager, Regulatory Affairs Abbott Diagnostics Division Phone (224) 667-9025 Fax (224) 667-4836
Date Summary Prepared: December 16, 2021
{4}------------------------------------------------
#### III. Device Name
Albumin BCP2
### Reagents
Trade Name: Albumin BCP2 Device Classification: Class II Classification Name: Albumin test system Governing Regulation Number: 21 CFR 862.1035 Product Code: CJW
# IV. Predicate Device
Albumin BCP (k981814)
# V. Description of Device
# A. Principles of the Procedure
The Albumin BCP2 assay is an automated clinical chemistry assay. The Albumin BCP2 procedure is based on the binding of bromocresol purple specifically with human albumin to produce a colored complex. The absorbance of the complex at 604 nm is directly proportional to the albumin concentration in the sample.
Methodology: Colorimetric (Bromocresol Purple)
{5}------------------------------------------------
### B. Reagents
The various kit configurations of the Albumin BCP2 reagent kits are described below.
| | List Number | |
|------------------------------|-------------|---------|
| | 04U4520 | 04U4530 |
| Tests per cartridge | 150 | 550 |
| Number of cartridges per kit | 4 | 4 |
| Tests per kit | 600 | 2200 |
| Reagent 1 (R1) | 15.8 mL | 49.9 mL |
Active ingredient: Bromocresol Purple 0.154 g/L. Inactive ingredients: Reagent 1 sodium hydroxide/acetic acid (pH 5.4) and detergent/surfactant (0.8%). Preservatives: ProClin 300 and ProClin 950.
#### VI. Intended Use of the Device
The Albumin BCP2 assay is used for the quantitation of albumin in human serum or plasma on the ARCHITECT c System.
The Albumin BCP2 assay is to be used as an aid in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
# VII. Comparison of Technological Characteristics
The Albumin BCP2 assay (subject device) is an automated clinical chemistry assay for the quantitation of albumin in human serum or plasma on the ARCHITECT c System.
The similarities and differences between the subject device and the predicate device are presented in the following table.
{6}------------------------------------------------
| Characteristics | Subject Device<br>Albumin BCP2 (List No. 04U45) | Predicate Device<br>Albumin BCP (k981814; List No. 7D54) |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>and Indications<br>for Use | The Albumin BCP2 assay is used<br>for the quantitation of albumin in<br>human serum or plasma on the<br>ARCHITECT c System.<br>The Albumin BCP2 assay is to be<br>used as an aid in the diagnosis andinvolving primarily the liver or<br>kidneys. | The Albumin BCP assay is used for the<br>quantitation of albumin in human serum or<br>plasma. |
| Platform | ARCHITECT c System | Samea |
| Methodology | Colorimetric (Bromocresol Purple) | Same |
| Specimen Type | Human serum or plasma | Same |
| Assay Principle<br>/ Principle of<br>Procedure | The Albumin BCP2 procedure is<br>based on the binding of<br>bromocresol purple specifically<br>with human albumin to produce a<br>colored complex. The absorbance<br>of the complex at 604 nm is<br>directly proportional to the<br>albumin concentration in the<br>sample. | Same |
| Standardization | ERM-DA470/IFCC | Same |
| Use of<br>Calibrators | Yes | Same |
| Use of Controls | Yes | Same |
| Assay Range | Analytical Measuring Interval:<br>0.3 – 9.0 g/dL<br>Extended Measuring Interval:<br>9.0 – 22.4 g/dL<br>Reportable Interval:<br>0.3 – 22.4 g/dL | Measuring Interval: 0.31 to 11.0 g/dL |
| Characteristics | Subject Device<br>Albumin BCP2 (List No. 04U45) | Predicate Device<br>Albumin BCP (k981814; List No. 7D54) |
| Precision | Samples with albumin<br>concentrations between 0.4 and<br>8.2 g/dL were evaluated. The<br>samples demonstrated<br>% coefficients of variation<br>(%CV) ≤ 1.7% and standard<br>deviations (SD) ≤ 0.04 g/dL. | Samples with albumin concentrations<br>between 2.4 and 3.7 g/dL demonstrated<br>%CVs ranging from 1.2% to 1.4%. |
| Lower Limits of<br>Measurement | Limit of Blank: 0.0 g/dL<br>Limit of Detection: 0.3 g/dL<br>Limit of Quantitation: 0.3 g/dL | Limit of Detection: 0.3 g/dL<br>Limit of Quantitation: 0.31 g/dL |
| Tube Types | Serum:<br>- Serum tubes<br>- Serum separator tubes<br><br>Plasma:<br>- Dipotassium EDTA tubes<br>- Lithium heparin tubes<br>- Lithium heparin separator tubes<br>- Sodium heparin tubes | Serum:<br>- Glass or plastic tubes with or without gel<br>barrier<br><br>Plasma:<br>- Glass or plastic lithium heparin tubes<br>(with or without gel barrier)<br>- Glass or plastic sodium heparin tubes |
#### Comparison of Subject Device (Albumin BCP2) to Predicate Device (Albumin BCP)
ª In accordance with FDA Guidance Document "Data for Commercialization of Original Equipment Manufacturer, Secondary and Generic Reagent for Automated Analyzers", issued June 10, 1996, the assay equivalency study on ARCHITECT c System vs. the original platform, AEROSET, was performed and submitted under k980367/A005 in May 2002.
ERM = European Reference Materials Standard Reference Material
IFCC - International Federation of Clinical Chemistry and Laboratory Medicine
{7}------------------------------------------------
{8}------------------------------------------------
### VIII. Summary of Nonclinical Performance
### A. Reportable Interval
Based on the limit of detection (LoD), limit of quantitation (LoQ), precision, and linearity, the ranges over which results can be reported are provided below according to the definitions from Clinical and Laboratory Standards Institute (CLSI) EP34, 1st ed. *
| | g/dL |
|--------------------------------------|------------|
| Analytical Measuring Interval (AMI)a | 0.3 - 9.0 |
| Extended Measuring Interval (EMI)b | 9.0 - 22.4 |
| Reportable Intervalc | 0.3 - 22.4 |
a AMI: The AMI extends from the LoQ to the upper limit of quantitation (ULoQ). This is determined by the range of values in g/dL that demonstrated acceptable performance for linearity, imprecision, and bias.
b The EMI extends from the ULoQ to the ULoQ × dilution factor.
6 The reportable interval extends from the LoD to the upper limit of the EMI.
NOTE: The default Low Linearity value of the assay file corresponds to the lower limit of the Analytical Measuring Interval (AMI), and result values below 0.3 g/dL are reported as "<0.3 g/dL".
Clinical and Laboratory Standards Institute (CLS). Establishing and Verifying an Extended Measuring Interval Through Speciment Dilution and Spiking. 1st ed. CLSI Document EP34. Wayne, PA: CLSI; 2018.
{9}------------------------------------------------
#### B. Within-Laboratory Precision
A study was performed based on guidance from CLSI EP05-A3. * Testing was conducted using 3 lots of the Albumin BCP2 reagent, 3 lots of the Consolidated Chemistry Calibrator, 1 lot of commercially available controls, and 3 instruments. Two controls and 3 human serum panels were tested in duplicate, twice per day on 20 days on 3 reagent lot/calibrator lot/instrument combinations, where a unique reagent lot and a unique calibrator lot is paired with 1 instrument. The performance from a representative combination is shown in the following table.
| | | | Within-Run<br>(Repeatability) | | Within-Laboratorya | |
|-----------------|----|----------------|-------------------------------|-----|-----------------------|----------------------|
| Sample | n | Mean<br>(g/dL) | SD | %CV | SD<br>(Rangeb) | %CV<br>(Rangeb) |
| Control Level 1 | 80 | 3.7 | 0.04 | 1.1 | 0.05<br>(0.05 - 0.05) | 1.4<br>(1.3 - 1.4) |
| Control Level 2 | 80 | 2.5 | 0.04 | 1.6 | 0.04<br>(0.04 - 0.04) | 1.6<br>(1.4-1.7) |
| Panel 1 | 80 | 0.4 | 0.02 | 5.3 | 0.04<br>(0.02 - 0.04) | 10.4<br>(5.5 - 10.4) |
| Panel 2 | 80 | 5.3 | 0.05 | 1.0 | 0.05<br>(0.03 - 0.06) | 1.0<br>(0.6-1.0) |
| Panel 3 | 80 | 8.2 | 0.03 | 0.3 | 0.05<br>(0.05 - 0.07) | 0.7<br>(0.7 - 0.9) |
a Includes within-run, between-run, and between-day variability.
b Minimum and maximum SD or %CV across all reagent lot and instrument combinations.
# C. Accuracy
A study was performed to estimate the bias of the Albumin BCP2 assay relative to a standard reference material (ERM - DA470k/IFCC). Testing was conducted using 2 lots of the Albumin BCP2 reagent, 2 lots of the Consolidated Chemistry Calibrator, and 1 instrument. The bias was within ± 2.8%.
<sup>&</sup>quot; Clinical and Laboratory Standards Institute (CLSI). Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline-Third Edition. CLSI Document EP05-A3. Wayne, PA: CLSI; 2014.
{10}------------------------------------------------
# D. Lower Limits of Measurement
A study was performed based on guidance from CLSI EP17-A2. * Testing was conducted using 3 lots of the Albumin BCP2 reagent kit on each of 2 instruments over a minimum of 3 days. The maximum observed limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) values are summarized below.
| | g/dL |
|------|------|
| LoBa | 0.0 |
| LoDb | 0.3 |
| LoQc | 0.3 |
a The LoB represents the 95th percentile from n ≥ 60 replicates of zero-analyte samples.
b The LoD represents the lowest concentration at which the analyte can be detected with 95% probability based on n ≥ 60 replicates of low-analyte level samples.
6 The LoQ is defined as the lowest concentration at which a maximum allowable precision of 20% CV was met and was determined from n ≥ 60 replicates of low-analyte level samples.
#### E. Linearity
A study was performed based on guidance from CLSI EP06-A. \$ The assay was demonstrated to be linear across the analytical measuring interval of 0.3 to 9.0 g/dL.
<sup>+</sup> Clinical and Laboratory Standards Institute (CLS). Evaluation of Detection Capability for Clinical Laboratory Measurent Procedures; Approved Guideline-Second Edition. CLSI Document EP17-A2. Wayne, PA: CLSI; 2012.
S Clinical and Laboratory Standards Institute (CLS). Evaluation of the Linearity of Quanitiative Measurement Procedures: A Statistical Approach; Approved Guideline. CLSI Document EP06-A. Wayne, PA: CLSI; 2003.
{11}------------------------------------------------
# F. Potentially Interfering Endogenous and Exogenous Substances
A study was performed based on guidance from CLSI EP07-A2. ** Each substance was tested at 2 levels of the analyte (approximately 3.5 g/dL and 5.0 g/dL).
No significant interference (interference within ± 10%) was observed at the following concentrations.
Potentially Interfering Endogenous Substances
| | Interferent Level | |
|-------------------------------------|-------------------|--|
| Potentially Interfering Substance | Default Units | |
| Conjugated Bilirubin | 60 mg/dL | |
| Unconjugated Bilirubin | 60 mg/dL | |
| Hemoglobin | 2000 mg/dL | |
| Triglycerides | 3025 mg/dL | |
| | Interferent Level | |
| Potentially Interfering Substance | Default Units | |
| Acetaminophen | 250 mg/L | |
| Acetylcysteine | 1663 mg/L | |
| Acetylsalicylic Acid | 1000 mg/L | |
| Aminosalicylic Acid | 80 mg/dL | |
| Ampicillin-Na | 1000 mg/L | |
| Ascorbic Acid | 300 mg/L | |
| Ca-dobesilate | 200 mg/L | |
| Cefotaxime | 31 mg/dL | |
| Cefoxitin | 2500 mg/L | |
| Cyclosporine | 5 mg/L | |
| Desacetylcefotaxime | 6 mg/dL | |
| Doxycycline | 50 mg/L | |
| Ibuprofen | 500 mg/L | |
| Levodopa | 20 mg/L | |
| Methyldopa | 20 mg/L | |
| Metronidazole | 200 mg/L | |
| Phenylbutazone | 400 mg/L | |
| Rifampicin | 60 mg/L | |
| Sodium Heparin | 10 U/mL | |
| Theophylline (1,3-dimethylxanthine) | 100 mg/L | |
<sup>**</sup> Clinical and Laboratory Standards Institute (CLS). Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition. CLSI Document EP07-A2. Wayne, PA: CLSI; 2005.
{12}------------------------------------------------
# Potentially Interfering Exogenous Substances
{13}------------------------------------------------
# G. Method Comparison
A study was performed based on guidance from CLSI EP09-A3** using the Passing-Bablok regression method.
| Albumin BCP2 vs Albumin BCP on the ARCHITECT c System | | | | | | |
|-------------------------------------------------------|-----|-------|----------------------------|-----------|-------|------------------------|
| | n | Units | Correlation<br>Coefficient | Intercept | Slope | Concentration<br>Range |
| Serum | 127 | g/dL | 1.00 | -0.20 | 1.00 | 0.6 - 9.6 |
# H. Tube Type
A study was performed to evaluate the suitability of specific blood collection tube types for use with the Albumin BCP2 assay. Samples were collected from a minimum of 40 donors and evaluated across tube types. The following blood collection tube types were determined to be acceptable for use with the Albumin BCP2 assay:
- Serum tubes .
- Serum separator tubes .
- Dipotassium EDTA tubes •
- Lithium heparin tubes .
- Lithium heparin separator tubes .
- Sodium heparin tubes .
<sup>**</sup> Clinical and Laboratory Standards Institute (CLS). Measurement Procedure Comparison and Bias Patient Samples; Approved Guideline-Third Edition. CLSI Document EP09-A3. Wayne, PA: CLSI; 2013.
{14}------------------------------------------------
# I. Dilution Verification
A study was performed to evaluate the performance of the Albumin BCP2 automated dilution protocol relative to the manual dilution procedure on the ARCHITECT c System. Five human serum samples were prepared by spiking human serum albumin (HSA) stock solution into reagent grade water to obtain the target concentration values of 10.0 g/dL, 12.5 g/dL, 17.5 g/dL, and 20.0 g/dL. Each sample was divided into multiple aliquots. An aliquot of each sample was tested using the 1:2.5 automated dilution protocol on the ARCHITECT c System. The additional aliquots were used to prepare 6 sets of manual dilutions (1:2.5 dilution with saline) of each sample and the manually diluted samples were tested.
The % difference values for the automated dilution protocol versus the manual dilution procedure ranged from of -2.9% to -1.5% and therefore, demonstrated acceptable performance.
{15}------------------------------------------------
# IX. Summary of Clinical Performance
This section does not apply.
# X. Conclusion Drawn from Nonclinical Laboratory Studies
The results presented in this 510(k) pre-market notification demonstrate that the performance of the subject device, Albumin BCP2 (List No. 04U45), is substantially equivalent to the predicate device, Albumin BCP (List No. 7D54, k981814).
The similarities and differences between the subject device and predicate device are presented in Section 5-VII.
There is no known potential adverse effect to the operator when using this in vitro device according to the Albumin BCP2 reagent package insert instructions.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.