← Product Code [NKB](/productcode/NKB) · K203482

# M.U.S.T. Pedicle Screw Extension and Long Tab Implants (K203482)

_Medacta International S.A. · NKB · Jan 26, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K203482

## Device Facts

- **Applicant:** Medacta International S.A.
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Jan 26, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The M.U.S.T. Pedicle screws system is intended for posterior non-cervical pedicle fixation (T1-S2/ilium) and non-pedicle fixation, or anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

## Device Story

M.U.S.T. Pedicle Screw Extension and Long Tab implants are spinal fixation components used in thoracolumbar surgery. The system includes cannulated/non-cannulated pedicle screws, rods, lateral connectors, and set screws. The Long Tab (LT) screws feature two tabs on the screw head to facilitate rod reduction during surgery; tabs are broken off post-reduction, leaving a standard tulip profile. The system is used by surgeons in clinical settings for spinal stabilization and fusion. The device provides mechanical support to the spine, aiding in the correction of deformities or stabilization of injured segments. The output is the physical fixation of the spinal construct, which promotes bone fusion and patient recovery.

## Clinical Evidence

No clinical studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including geometrical analysis, cadaveric surgical simulation, and mechanical performance testing (static/dynamic axial compression, torsion, tensile force resistance) per ASTM F1717, F543, and F2193 standards. Pyrogenicity assessments were also performed.

## Technological Characteristics

Materials include Titanium and Cobalt-Chrome (CoCr). Components include cannulated/non-cannulated pedicle screws, straight/pre-contoured rods, lateral connectors, and set screws. Mechanical testing performed per ASTM F1717-18, ASTM F543-17, and ASTM F2193-18a. Devices are sterile, implantable, and intended for posterior/anterolateral spinal fixation.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- M.U.S.T. Pedicle Screw System ([K153664](/device/K153664.md))
- M.U.S.T. Pedicle Screw System ([K121115](/device/K121115.md))
- M.U.S.T. Pedicle Screw System ([K132878](/device/K132878.md))
- M.U.S.T. Pedicle Screw System ([K141988](/device/K141988.md))
- M.U.S.T. Pedicle Screw System ([K162061](/device/K162061.md))
- M.U.S.T. Pedicle Screw System ([K171170](/device/K171170.md))
- M.U.S.T. Combined Set Screws ([K171758](/device/K171758.md))
- M.U.S.T. Pedicle Screw System ([K193365](/device/K193365.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date January 26, 2021. The text is in a simple, sans-serif font. The date is written out in full, with the month, day, and year clearly visible. The text is centered and takes up most of the image.

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Medacta International SA % Chris Lussier Senior Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis, Tennessee 38118

Re: K203482

Trade/Device Name: M.U.S.T. Pedicle Screw Extension and Long Tab Implants Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWQ, KWP Dated: November 24, 2020 Received: November 27, 2020

Dear Chris Lussier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K203482

### Device Name

M.U.S.T. Pedicle Screw Extension and Long Tab implants

### Indications for Use (Describe)

The M.U.S.T. Pedicle screws system is intended for posterior non-cervical pedicle fixation (T1-S2/ilium) and non-pedicle fixation, or anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

#### Submitter I.

Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66 Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director of Quality and Regulatory, Medacta USA Date Prepared: November 24, 2020

#### II. Device

| Device Proprietary Name: | M.U.S.T. Pedicle Screw Extension and Long Tab Implants |
|--------------------------|--------------------------------------------------------|
| Common or Usual Name:    | Pedicle screw spinal system                            |
| Classification Name:     | Thoracolumbosacral Pedicle Screw System                |
| Primary Product Code:    | NKB                                                    |
| Secondary Product Code:  | KWQ, KWP                                               |
| Regulation Number:       | 21 CFR 888.3070, 21 CFR 888.3060, 21 CFR 888.3050      |
| Device Classification    | II                                                     |

#### III. Predicate Device

Substantial equivalence is claimed to the following devices:

Primary predicate device:

- M.U.S.T. Pedicle Screw System, K153664, Medacta International SA ●
Reference predicate devices:

- M.U.S.T. Pedicle Screw System, K121115, Medacta International SA
- M.U.S.T. Pedicle Screw System, K132878, Medacta International SA ●
- M.U.S.T. Pedicle Screw System, K141988, Medacta International SA
- M.U.S.T. Pedicle Screw System, K162061, Medacta International SA
- M.U.S.T. Pedicle Screw System, K171170, Medacta International SA
- M.U.S.T. Combined Set Screws, K171758, Medacta International SA ●
- M.U.S.T. Pedicle Screw System, K193365, Medacta International SA

#### IV. Device Description

The M.U.S.T. Extension and M.U.S.T. Long Tab is intended to be used as part of the M.U.S.T. Pedicle Screw system (K121115, K132878, K141988, K153664, K162061, K171170, K171758, K193365) for 

{4}------------------------------------------------

the stabilization and the fusion of the lumbar and thoracic spine. The M.U.S.T. pedicle screw system includes cannulated or non-cannulated poly-axial pedicle screws (K121115, K132878, K153664), cannulated or non-cannulated mono-axial pedicle screws (K132878), set screws (K121115, K171758), straight and pre-bent rods (K121115, K141988, K162061), lateral connectors (K162061) and cross connectors (K132878, K193365). The M.U.S.T. pedicle screw system also includes the enhanced screws and rods designed for percutaneous surgery (K141988).

The M.U.S.T. Extension and M.U.S.T. Long Tab introduce new sizes of sterile implantable devices intended to be used for posterior non-cervical pedicle fixation (T1-S2/ilium) or anterolateral fixation (T8-L5). The devices subject of this submission are:

- New size of Straight Rods (both Titanium and CoCr version): length 600mm Hex interface at . the end
- New size of Straight Rods (Anodized Titanium version): length 600mm ●
- Pre-contoured Rods (both Titanium and CoCr version) ●
- New sizes of Lateral Connectors: length 80; 100; 125 & 150mm ●
- New sizes of Enhanced Cannulated Pedicle Screws: Ø8; Ø9; Ø10 length from 30 to 100mm ●
- New cannulated pedicle screws with thread 4 leads (15 and 25 mm) ●
- Setscrews in Titanium ●

Intended purpose and the performance specification of the devices are equivalent to the ones of the current US cleared portfolio: screws, rods and lateral connectors are already used in the MUST implant construct.

#### V. Indications for Use

The M.U.S.T. Pedicle screws system is intended for posterior non-cervical pedicle fixation (T1-S2/ilum) and non-pedicle fixation, or anterolateral fixation (T8-L5). These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

#### VI. Comparison of Technological Characteristics

#### M.U.S.T. Extension .

The new sizes of Straight Rods (Titanium and CoCr version) included in the M.U.S.T. Extension and the predicate devices (K121115) share the following characteristics:

- indication for use; ●
- design;
- diameters;
- materials; ●

{5}------------------------------------------------

- biocompatibility; ●
- device usage;
- sterility;
- shelf life;
- packaging. ●

The only difference between the Straight Rods included in the M.U.S.T. Extension and the predicate devices (K121115) is the rods length.

#### M.U.S.T. Extension ●

The new size of Straight Rod and Enhanced Straight Rod (Anodized Titanium version) included in the M.U.S.T. Extension and the predicate devices (K162061) share the following characteristics:

- indication for use;
- geometry;
- diameters; ●
- materials; ●
- biocompatibility;
- device usage;
- sterility;
- shelf life;
- packaging.

The only difference between the Straight Rods included in the M.U.S.T. Extension and the predicate devices (K162061) is the rods length.

#### M.U.S.T. Extension ●

The subject Pre-contoured Rods and Straight Enhanced Rods (both Titanium and CoCr version) included in the M.U.S.T. Extension and the predicate devices (K141988) share the following characteristics:

- indication for use; ●
- design; ●
- diameters; ●
- materials;
- biocompatibility; ●
- device usage; ●
- sterility;
- shelf life;
- packaging.

The only difference between the Pre-contoured Rods and Straight Enhanced Rods included in the M.U.S.T. Extension and the predicate devices (K141988) is the rods length.

{6}------------------------------------------------

#### M.U.S.T. Extension ●

The new sizes of Lateral connectors included in the M.U.S.T. Extension and the predicate devices (K162061) share the following characteristics:

- indication for use; ●
- geometry;
- diameters; ●
- materials; ●
- interconnection mechanism;
- biocompatibility;
- device usage;
- sterility;
- shelf life;
- packaging. ●

The only difference between the Lateral connectors included in the M.U.S.T. Extension and the predicate devices (K162061) is the connector length.

#### M.U.S.T. Extension ●

The new sizes of Enhanced Cannulated Pedicle Screws included in the M.U.S.T. Extension and the predicate devices (K171170 and K141988) share the following characteristics:

- indication for use;
- design; ●
- materials; ●
- biocompatibility; ●
- device usage;
- sterility;
- shelf life; ●
- packaging. ●

The only difference between the Enhanced Cannulated Pedicle Screws included in the M.U.S.T. Extension and the predicate devices (K141988) is the screws diameters and lengths; the thread of the new sizes is substantially equivalent to the thread shape of the predicate device.

#### M.U.S.T. Long Tab (LT) ●

The new M.U.S.T. LT Cannulated pedicle screws included in the M.U.S.T. LT and the predicate devices (K153664 and K171170) share the following characteristics:

- indication for use;
- design; ●
- materials;

{7}------------------------------------------------

- diameters;
- screw lengths; ●
- biocompatibility;
- device usage;
- sterility; ●
- shelf life;
- packaging. ●

The only difference between the M.U.S.T. LT Cannulated pedicle screws included in the M.U.S.T. LT and the predicate devices (K153664 and K171170) is the pedicle screw head that has two long tabs, that can easily allow reduction operation during the surgery (two different lengths of reduction; 15mm and 25mm). The MUST LT tulip is equivalent to the predicate device in terms of material and geometry. In fact, the two tabs are intended to the broken, during surgery and after rod reduction. The resulting tulip has a similar profile and dimensions reduced as the predicate device.

The new M.U.S.T. LT Combined setscrew in the M.U.S.T. LT and the predicate devices (K171758) share the following characteristics:

- indication for use; ●
- diameters; ●
- biocompatibility; ●
- device usage;
- sterility;
- shelf life; ●
- packaging.

The differences between the M.U.S.T. LT Combined setscrews in the M.U.S.T. LT and the predicate devices (K171758) are material and dimensions.

#### VII. Performance Data

The addition of the M.U.S.T. Extension and M.U.S.T. LT components to M.U.S.T. - Pedicle Screw System portfolio was evaluated by risk analysis to identify any new risks associated with the change. Based on the risk analysis, design verification was conducted to written protocols with pre-defined acceptance criteria. The protocols and pre-defined acceptance criteria were based on the standards, FDA guidance, and comparison to the predicate device system.

The following validation and tests are being provided in support of the substantial equivalence determination:

## Non-Clinical Studies

- DESIGN VALIDATION ●

{8}------------------------------------------------

- Geometrical Analysis MUST LT Thread Geometry: To ensure that the implant has o sufficient fixation in the bone, a geometrical analysis was performed in comparison to approved and marketed implants, related to the most important dimensions of the threads.
- Cadaver surgery performed according to the well-known MIS percutaneous (Minimally o invasive system) technique (99.PERC46.12 rev.04) by experienced surgeons and according to the M07.138.001 prepared for the Chicago Wetlab.
- Tab welding MUST LT Flexion strength test according to IL07.09.598 to characterize the o specific design of the tulip in the dedicated breaking area
- PERFORMANCE TESTING ●
	- Worst Case definition for ASTM F1717 testing: discussion of implant types and sizes, chosen o as the worst case for mechanical testing of the spinal construct in a vertebrectomy model.
	- Mechanical evaluation under static and dynamic axial compression according to ASTM о F1717-18 Standard test methods for Spinal Implants Construct in a Vertebrectomy model: Endolab 970 200220 30 1413 part 1 rev.0 and 970 200220 30 1414 part 1 rev.0
	- Mechanical evaluation under static torsion according to ASTM F1717-18 Standard test o methods for Spinal Implants Construct in a Vertebrectomy model: Endolab 970 200220 30 1413 part 2 rev.0 and 970 200220 30 1414 part 2 rev.0
	- Mechanical evaluation of axial tensile force resistance according to ASTM F543-17 Standard o specification and test methods for metallic medical bone screws and ASTM F2193-18a Standard Specification and test methods for components used in the surgical fixation of the spinal skeletal and 970.200227.120.868-part1 rev. 0
	- Mechanical evaluation of torsion force resistance according to ASTM F543-17 Standard o specification and test methods for metallic medical bone screws and ASTM F2193-18a Standard Specification and test methods for components used in the surgical fixation of the spinal skeletal system: Endolab 970.200227.120.867-part2 rev. 0 and 970.200227.120.868part2 rev. 0
	- o Mechanical evaluation of torsional yield strength, maximum torque and braking angle according to ASTM F543-17 Standard specification and test methods for metallic medical bone screws - A1: Endolab 970.200227.120.867-part3 rev. 0 and 970.200227.120.868-part3 rev. O
	- Characterization and evaluation of risk of breakage during the insertion in the bone of MUST o screws according to ASTM F543-17 Standard specification and test methods for metallic medical bone screw: Medacta test report ASTM F543 Summary Report

{9}------------------------------------------------

- . PYROGENICITY
	- 0 Pyrogenicity Assessment, ENDOTOXINS-MEDIATED PYROGENICITY ASSESSMENT REPORT FOR: "MUST PEDICLE SCREWS IMPLANTS SURGICAL KITS" TF VI-PS-01, RAS-01.008.141 Rev. 5, Dated: 17 October 2020
	- 0 Pyrogenicity Assessment, ENDOTOXINS-MEDIATED PYROGENICITY ASSESSMENT REPORT FOR: "MUST LT IMPLANTS (VI-PS-01)", RAS-01.008.277, Rev. 0, Dated: 16 September 2020

Clinical Studies:

- No clinical studies were conducted.
## VIII. Conclusion

Based on the above information, the M.U.S.T. Extension and M.U.S.T. LT components are substantially equivalent to the identified predicate devices.

Substantial equivalence has been demonstrated through a comparison of intended use, design and technological characteristics, as well as performance evaluations.

---

**Source:** [https://fda.innolitics.com/device/K203482](https://fda.innolitics.com/device/K203482)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
