← Product Code [GEX](/productcode/GEX) · K203441

# The Alma Hybrid Laser System (K203441)

_Alma Lasers , Ltd. · GEX · Jul 2, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K203441

## Device Facts

- **Applicant:** Alma Lasers , Ltd.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Jul 2, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Alma Hybrid Laser System, Delivery Devices, Applicators and Accessories are intended for use in surgical applications requiring the ablation, vaporization, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery).

## Device Story

The Alma Hybrid Laser System is a surgical laser console featuring an optical bench, microcontroller, and touch-screen interface. It utilizes two wavelengths (10600 nm CO2 and 1570 nm) delivered via articulated arm and various applicators (HyLight, Pixel, ProScan). The device is operated by clinicians in aesthetic surgery settings. It transforms electrical energy into laser output for tissue ablation, vaporization, incision, and coagulation. The system supports monochromatic or sequential radiation modes and fractional patterns (grid, hygrid). Clinicians use the touch-screen to set parameters; the laser output is applied to the patient's skin to achieve resurfacing or lesion removal. The device benefits patients by providing precise tissue management for dermatological conditions. It includes a cooling system, air-purge system, and footswitch control.

## Clinical Evidence

No human clinical studies were performed. Evidence consists of bench testing (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 10993-1, ISO 17665-1, IEC 62304, ISO 14971) and a porcine animal study (n=3) for the ProScan applicator. Histology results at 0, 3, 7, and 14 days showed re-epithelialization and no adverse events, supporting safety and effectiveness.

## Technological Characteristics

Laser system with 10600 nm CO2 and 1570 nm wavelengths. Features articulated arm, dual-axis scanner, and cooled tips. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 10993-1, ISO 17665-1, and IEC 62304. Software-controlled via microcontroller. Sterilization via moist heat (ISO 17665-1).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Alma Lasers Pixel CO2 Laser System ([K103501](/device/K103501.md))
- M22 And ResurFx Systems ([K170060](/device/K170060.md))

## Reference Devices

- Alma Lasers System ([K080463](/device/K080463.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 2, 2021

Alma Lasers, Ltd. Avi Hirshnzon Evp RAqa 18 Haharash Street, North Industrial Park Caesarea Ha Zafon, 3079895 Israel

Re: K203441

Trade/Device Name: The Alma Hybrid Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, ONG Dated: November 20, 2020 Received: November 23, 2020

Dear Avi Hirshnzon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203441

Device Name The Alma Hybrid Laser System

### Indications for Use (Describe)

The Alma Hybrid Laser System, Delivery Devices, Applicators and Accessories are intended for use in surgical applications requiring the ablation, vaporization, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery).

HyLight-CO2

The Alma Hybrid CO2 non-fractional applicator, with wavelength of 10600 nm is cleared for use for the particular indications as follows:

### Dermatology & Plastic Surgery

The ablation, vaporization, excision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:

- * laser skin resurfacing
- * laser derm-abrasion
- * laser burn debridement.

Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of: warts, acne scars, nevi epidermal, syringoma. Vaporization/coagulation of warts.

### Pixel

The Alma Hybrid Pixel CO2 fractional applicator, with wavelength of 10600 nm is indicated for: The ablation, vaporization, and coagulation of soft tissue in dermatology and plastic surgery in the performance of skin resurfacing.

ProScan

The Alma Hybrid ProScan CO2 fractional applicator, with wavelength of 10600 nm is indicated for:

· Laser skin resurfacing (ablation and/or vaporization) of soft tissue.

The Alma Hybrid ProScan 1570nm fractional applicator, with wavelength of 1570 nm, is indicated for:

· Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue

The Alma Hybrid ProScan CO2 &1570nm fractional applicator, with wavelengths of 10600 nm & indicated for laser skin resurfacing (ablation and/or vaporization) of soft tissue.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

#### K203441 510(k) Summary

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92

| Owner Name               | Alma Lasers Inc.                                                                                                                                                                                                                                                                                            |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address                  | 18 Haharash Street<br>NORTH INDUSTRIAL PARK<br>CAESAREA, IL 3079895                                                                                                                                                                                                                                         |
| Contact Person           | Kathy Maynor<br>Regulatory Consultant<br>Email : regulatory@almalasers.com<br>Phone: 352-586-3113 (cell)<br>Facsimile: 646-805-1305<br><br>Secondary contact: Jessica Rivera-Montejo<br>Assoc. Director of Regulatory<br>Email: regulatory@almalasers.com<br>Phone: 224-377-2019<br>Facsimile: 646-805-1305 |
| Summary Preparation Date | July 2, 2021                                                                                                                                                                                                                                                                                                |

#### Submitter Information [21 CFR 807.92(a) (1)] I.

#### II. Name of device [21 CFR 807.92 (a) (2)]

| Trade or Proprietary Name                                                     | The Alma Hybrid Laser System |                                                |                   |
|-------------------------------------------------------------------------------|------------------------------|------------------------------------------------|-------------------|
| Common Device Name(s) and<br>Regulatory Class                                 | Product<br>Code(s)           | Classification Panel                           | Regulation        |
| Electrosurgical cutting and<br>coagulation device and<br>accessories Class II | GEX                          | General & Plastic<br>Surgery Panel, 79<br>(SU) | § 21 CFR 878.4810 |
| Powered Laser Surgical<br>Instrument With<br>Microbeam/Fractional Output      | ONG                          | General & Plastic<br>Surgery Panel, 79 (SU)    | § 21 CFR 878.4810 |

#### III. Predicate Devices [21 CFR 807.92(a) (3)]

| Type      | 510(k) # | Trade Name                         | Product Code |
|-----------|----------|------------------------------------|--------------|
| Primary   | K103501  | Alma Lasers Pixel CO2 Laser System | GEX          |
| Reference | K080463  | Alma Lasers System                 | GEX          |
| Secondary | K170060  | M22 And ResurFx Systems            | GEX          |

{5}------------------------------------------------

#### IV. Device Description [21 CFR 807.92(a) (4)]

The Alma Hybrid Laser System consists of:

- Laser system console (containing the optical bench assembly and laser, the ● microcontroller control electronics and system software, the high voltage power supply, the laser cooling system, the compressed air-purge system, and the service panel)
- LCD control panel with touch-screen technology: the LCD display provides information . on the status and settings of the The Alma Hybrid Laser System with touch-screen technology provided to input commands into the system.
- Two wavelengths – CO2 and 1570 nm
- . Articulated arm
- Footswitch
- Delivery devices (Pixel applicators, Hylight applicators, Proscan applicator) ●

#### V. Intended use of device and Indications for Use [21 CFR 807.92(a) (5)]

The Alma Hybrid Laser System, Delivery Devices, Applicators and Accessories are intended for use in surgical applications requiring the ablation, vaporization, incision, and coagulation of soft tissue in medical specialties including: aesthetic surgery (dermatology and plastic surgery).

## HyLight-CO2

The Alma Hybrid CO2 non-fractional applicator, with wavelength of 10600 mm is cleared for use for the particular indications as follows:

### Dermatology & Plastic Surgery

The ablation, vaporization, excision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:

- * laser skin resurfacing
- * laser derm-abrasion
- * laser burn debridement.

Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of: warts, acne scars, nevi epidermal, syringoma. Vaporization/coagulation of warts.

### Pixel

The Alma Hybrid Pixel CO2 fractional applicator, with wavelength of 10600 nm is indicated for: The ablation, vaporization, and coagulation of soft tissue in dermatology and plastic surgery in the performance of skin resurfacing.

## ProScan

The Alma Hybrid ProScan CO2 fractional applicator, with wavelength of 10600 nm is indicated for: ·Laser skin resurfacing (ablation and/or vaporization) of soft tissue

The Alma Hybrid ProScan 1570nm fractional applicator, with wavelength of 1570 mm, is indicated for: · Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue The Alma Hybrid ProScan CO2 &1570nm fractional applicator, with wavelengths of 10600 nm & 1570nm is indicated for laser skin resurfacing (ablation and/or vaporization) of soft tissue.

{6}------------------------------------------------

| Feature                              | Device<br>Alma Pixel CO2 (K103501)<br>Primary Predicate                                                                                                            | Hybrid System (K203441)<br>(Subject Device)                                        |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Hylight CO2                          |                                                                                                                                                                    |                                                                                    |
| Wavelength                           | 10600 nm                                                                                                                                                           | Same                                                                               |
| Pilot beam                           | 650 nm, 3mW                                                                                                                                                        | Same                                                                               |
| Spot size (mm)                       | 0.15 dia at 50 mm working distance; 0.2 mm dia. At 100 mm working distance                                                                                         | 0.15 - 3.0 at 50 mm working distance<br>0.25- 3.1 (Dia.) at200 mm working distance |
| Repetition Rate                      | 1-5 Hz                                                                                                                                                             | Same                                                                               |
| Pulse Duration                       | 1-1000 ms                                                                                                                                                          | Same                                                                               |
| Repeat Delay                         | 0.01 - 1sec                                                                                                                                                        | Same                                                                               |
| Min output energy                    | 5 mJ                                                                                                                                                               | Same                                                                               |
| Output Power                         | 30W, 70W                                                                                                                                                           | Same                                                                               |
| Fluence                              | Max 1746 J/mm² for 30W system<br>Max 3530 J/mm² for the 70W system                                                                                                 | Same                                                                               |
| CO2 Pixel (7x7 pattern; 9x9 pattern) |                                                                                                                                                                    |                                                                                    |
| Spot Size                            | 0.125 to 0.20 mm dia. pixels in a 10 x10 mm²area (49 pixels) @ 50 mm distance<br>- 0.125 to 0.20 mm dia. pixels in a 11 x 11 mm²area (81 pixels) @ 100 mm distance | Same                                                                               |
| Repetition Rate                      | 1-5 Hz                                                                                                                                                             | 0.5-2Hz                                                                            |
| Pulse Duration                       | 100-300 ms for pulsed                                                                                                                                              | 1- 405 ms                                                                          |
| Energy                               | 5mJ/Pixel - 150 mJ/Pixel                                                                                                                                           | Same                                                                               |
| CO2 ProScan                          |                                                                                                                                                                    |                                                                                    |
| Scanner                              | Dual axis scanner                                                                                                                                                  | Same                                                                               |
| Spot size (mm)                       | 0.35 mm@ 100 mm distance                                                                                                                                           | Same                                                                               |
| Output energy                        | Up to 1000 mJ                                                                                                                                                      | 120 (30W model)/240 mJ (70W model)                                                 |
| Beam density                         | ~5% - 95% untreated tissue between spots                                                                                                                           | ~63% - 97% untreated tissue between spots                                          |

### VI. Summary of technological characteristics of the device compared to the predicates [21 CFR 807.92(a)(6)]

| Device         | Feature | Lumenis ReSurfx<br>K170060 - Secondary Predicate | Hybrid System (K203441)<br>(Subject Device) |
|----------------|---------|--------------------------------------------------|---------------------------------------------|
| ProScan 1570   |         |                                                  |                                             |
| Wavelength     |         | 1565 nm                                          | 1570 nm                                     |
| Pilot beam     |         | 650 nm, <5mW                                     | 650 nm, 3mW                                 |
| Scanner        |         | Dual axis scanner                                | Same                                        |
| Tip            |         | cooled                                           | Same                                        |
| Scan size (mm) |         | 5-18 mm                                          | Up to 30 mm diameter                        |
| Output energy  |         | 10 mJ per beam – 70 mJ per beam                  | 24mJ-144 mJ/pixel                           |
| Beam density   |         | Up to 500 microbeams/cm2                         | Up to 390 pixels/cm2                        |

The technical characteristics for the Hybrid Laser System are substantially equivalent to the predicate devices. The Alma Hybrid 10600nm non-fractional module, with

{7}------------------------------------------------

wavelength of 10600 nm have differences in spot size, pulse duration and output energy, but the fluence is within the range of the predicate devices.

The Alma Hybrid Pixel CO2 fractional applicator, with wavelength of 10600 nm have differences in pulse duration, but the max energy output is within the range of the predicate device.

The Alma Hybrid ProScan fractional applicator, with wavelengths of 10600 nm and/or 1570nm wavelengths offers two main treatment modes Monochromatic or Sequential Radiation. the Alma Hybrid ProScan fractional applicator, with wavelengths of 10600 nm, and/or 1570nm Treatment Modes include three fractional patterns - grid, hygrid and grid sequential- with either single or dual wavelengths. The fractional grid patterns are substantially equivalent to their predicate devices with differences in spot size, beam density and pulse duration. Alma did perform histology testing to justify the technical differences.

#### VII. Performance Testing [21 CFR 807.92(b)(1)]

IEC 60601-1 Medical Electrical Equipment-Part 1: General Requirements for safety IEC 60601-1-2 Medical Electrical Equipment 1-2 General Requirements for basic safety and essential performance - Electromagnetic Compatibility IEC 60601-2-22 Medical Electrical Equipment-Part 2-22: Particular requirements for

basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment

IEC 60825-1 Safety of laser products-Part 1: Equipment Classification, requirements and user's guide

ISO 10993-1 Biological evaluation of medical devices -- Part 1

ISO 17665-1 Sterilization of health care products - moist heat

IEC 62304 Medical Device Software

ISO 14971 Risk Analysis

In addition, software verification and performance validation testing were performed.

Alma also performed histology testing on porcine animals for the ProScan Applicator. Testing was performed safely on the test animals, and the histology results complied with the FDA requirements at 0, 3, 7 and 14 days. Three [3] Domestic female (Mixed Landrace & Large White) crossbred swine were used in this study. During the in-life stage, vital signs, ECG and % saturation were monitored, clinical observations and body weights were monitored and recorded. Re-epithelialization was observed three days after radiation in all specimens. No adverse events or unexpected complications have been detected in the swines. On the last day of trial, biopsies were taken from the center of each radiated point by punch biopsy and were sent to H&E. At the end of the procedure at first, third, and seventh days, the animal was awakened and transferred to the recovery room. Euthanizing of the pigs were done at the end of the study on day 14.

In all instances, the Alma Hybrid Laser System, Delivery Devices and Accessories

{8}------------------------------------------------

functioned as intended and the results observed were as expected.

#### VIII. Clinical Data [21 CFR 807.92(b) (2)]

Based on the similarities in the safety and effectiveness profiles of the subject and the predicate, no human clinical studies were deemed needed to support this submission.

#### IX. Conclusions Safety and Effectiveness SE [21 CFR 807.92(b) (3)]

The Alma Hybrid Laser System is as safe and effective as the predicate devices. The proposed Alma Hybrid Laser System has the same intended use and indications, similar technological characteristics, and the same principles of operation as its predicate devices. Thus, the Alma Hybrid Laser System is substantially equivalent to its predicate

---

**Source:** [https://fda.innolitics.com/device/K203441](https://fda.innolitics.com/device/K203441)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).