Disposable Ear Thermometer Probe Cover

K203431 · Huaian Zhongxin Packing Material Co., Ltd. · FLL · Jul 28, 2021 · General Hospital

Device Facts

Record IDK203431
Device NameDisposable Ear Thermometer Probe Cover
ApplicantHuaian Zhongxin Packing Material Co., Ltd.
Product CodeFLL · General Hospital
Decision DateJul 28, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2

Indications for Use

The probe cover is used as a sanitary barrier between the infra red thermometer and the ear canal. The probe cover is provided non-sterile and for single use only.

Device Story

Disposable polypropylene probe cover; functions as sanitary barrier between infrared ear thermometer and patient ear canal; prevents cross-contamination of ear secretions/particulates during temperature measurement. Used in home or clinical settings by patients or healthcare providers. Device is non-sterile, single-use, and transparent. Fits onto thermometer probe tip to maintain hygiene during intermittent temperature monitoring.

Clinical Evidence

No clinical data; bench testing only. Compliance verified via ASTM E1104-98, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation/sensitization).

Technological Characteristics

Material: Biocompatible polypropylene. Form factor: Disposable, transparent, colorless, odorless, single-size probe cover. Standards: ASTM E1104-98, ISO 10993-5, ISO 10993-10. Non-sterile. No electronic or software components.

Indications for Use

Indicated for use as a sanitary barrier between an infrared thermometer and the ear canal for patients of all ages in home or professional settings.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration". July 28, 2021 Huaian Zhongxin Packing Material Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM. 608, No. 738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China Re: K203431 Trade/Device Name: Disposable Ear Thermometer Probe Cover Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: July 1, 2021 Received: July 9, 2021 Dear Boyle Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K203431 Device Name Disposable Ear Thermometer Probe Cover Indications for Use (Describe) The probe cover is used as a sanitary barrier between the infra red thermometer and the ear canal. The probe cover is provided non-sterile and for single use only. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary (K203431) This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92. ### 1.0 Submitter's Information Name: Huaian Zhongxin Packing Material Co.,Ltd. Address: No.128-5,West Meigao Road,Qingpu Industrial Park, Huaian, Jiangsu Province,223022 China Tel: +86 -15751384913 Fax: +86- 517-83852820 Contact: Fenq Zuo Date of Preparation: July.24,2021 #### Designated Submission Correspondent Mr. Boyle Wanq Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn #### 2.0 Device Information Trade name: Disposable Ear Thermometer Probe Cover Common name: Thermometer Probe Covers and Sheaths Classification name: Clinical electronic thermometer Model(s): 4#.6# #### 3.0 Classification Production code: FLL Requlation number: 21CFR 880.2910 Classification: Class II Panel: General Hospital #### 4.0 Predicate Device Information Manufacturer: Kaz, USA Inc. Device: Braun Thermoscan® Pro 4000 Series/IRT 4000 Series {4}------------------------------------------------ Thermometers 510(k) number: K101747 #### 5.0 Device Description The Disposable Ear Thermometer Probe Cover is a disposable plastic cover made of a biocompatible polypropylene material that is dimensionally manufactured to set tolerances so that it can be fitted on the probe tip of the thermometer. The probe cover is used as a sanitary barrier between the infra red thermometer and the ear canal to prevent any ear secretions or particulates from being transferred between different people. The subject device is used as barriers between Braun Thermoscan® PRO3000 series/PRO4000 series/PRO6000 series ear thermometer, and also compatible with Braun Thermoscan® IRT1020/IRT2020/IRT2520/IRT3020/ IRT3520/IRT4020/IRT4520/IRT6020/IRT6520 ear thermometer and users' auditory canal measuring sites to avoid possible contamination and infection during temperature measuring. The probe covers are transparent, colorless and odorless, non-sterile and intended for single use only. #### 6.0 Indication for Use Statement The probe cover is used as a sanitary barrier between the infra red thermometer and the ear canal. The probe cover is provided non-sterile and for single use only. | Item | Subject Device<br>K203431 | Predicate Device<br>K101747 | Remark | |------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|--------| | Type of<br>Thermometer | Disposable Ear Thermometer<br>Probe Cover | Probe Cover | -- | | Product Code | FLL | FLL | Same | | Regulation No. | 21 CFR 880.2910 | 21 CFR 880.2910 | Same | | Class | II | II | Same | | Indications for<br>use | The probe cover is used as a<br>sanitary barrier between the<br>infra red thermometer and the<br>ear canal. The probe cover is<br>provided non-sterile and for<br>single use only. | The Braun<br>Thermoscan®IRT 4000<br>series and Braun<br>Thermoscan® PRO 4000<br>series Clinical Infrared Ear<br>Thermometers is indicated | Same* | #### 7.0 Comparison to the Predicate Device {5}------------------------------------------------ | | | for the intermittent<br>measurement and<br>monitoring of human body<br>temperature by consumers<br>of all ages in a home<br>use/professional use<br>environment.<br>The probe cover is used as<br>a sanitary barrier between<br>the infra red thermometer<br>and the ear canal. | | |------------------------------------------|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------| | Prescription/ove<br>r-the-counter<br>use | Over-the-counter use | Over-the-counter use | Same | | Design Features | Conforms to ASTM Standard<br>E1104 | Conforms to ASTM<br>Standard E1104 | Same | | Design<br>Configurations | One size | One size | Same | | Performance<br>Specifications | Conforms to ASTM Standard<br>E1104 | Conforms to ASTM<br>Standard E1104 | Same | | Prescription vs.<br>OTC | OTC | OTC | Same | | Sterile vs.<br>Non-Sterile | Non-sterile | Non-sterile | Same | | Patient contact<br>materials | Polypropylene | Polypropylene | Same | | Biocompatibility | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | Same | *The predicate's thermometer probe covers are accessories to the thermometer itself, and were included in its 510(k) clearance under K101747.Only the probe covers are included in this current 510(k),as there is no thermometer or any accessory marked with the current subject device. The subject device's indications are only applicable to the probe covers. The proposed device's Indications for Use are, therefore, more narrow than the Indications for Use stated in K101747, which applied to both the thermometer and the probe covers, as the subject device's indications are only applicable to the probe covers. ## 8.0 Summary of Non-Clinical Testing Non clinical tests were conducted to verify that the subject devices met all design specifications as were Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: {6}------------------------------------------------ ● ASTM E1104-98 (Reapproved 2016) Standard Specification for Clinical Thermometer Probe Covers and Sheaths. ● ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity ● ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization #### 9.0 Summary of Clinical Testing Clinical testing was not required for this submission. #### 10.0 Conclusion Performance testing contained in this submission demonstrates the minor differences in technological characteristics between the subject device and the predicate do not raise different questions of safety and effectiveness. Based on the performance testing and compliance with acceptable voluntary standards, the subject device is substantially equivalent to its predicate device in K101747.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...