The Triathlon AS-1 patient specific cutting guides are disposable, single-use surgical instruments intended to assist orthopedic surgeons in the positioning of femoral and tibial total knee arthroplasty components intraoperatively, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. They are intended for use with the Cruciate Retaining (CR), Condylar Stabilizing (CS), Posterior Stabilized (PS) and Tritanium® components of the Triathlon® Total Knee System and the Total Stabilizer (TS) and Posterior Stabilized Rotation (PSR) Triathlon® tibial inserts.
Device Story
Triathlon AS-1 consists of patient-specific, single-use disposable cutting guides (jigs) and associated surgical documentation (plan, protocol). Input: patient CT scans. Process: software uses patient anatomy to size and position off-the-shelf Stryker Triathlon implants; 3D planning optimizes implant fit; guides are designed to match patient femoral/tibial contours to facilitate surgical technique. Output: physical cutting guides used by orthopedic surgeons in the OR to prepare bone for implant placement. Benefit: simplified surgical technique and optimized implant positioning. No implants or reusable instruments are included.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing: cadaver testing for surgeon evaluation and usability, implant sizing and positioning testing, and software verification and validation.
Technological Characteristics
Patient-specific, single-use cutting guides manufactured from biocompatible nylon via selective laser sintering. Supplied sterile. Designed for compatibility with Stryker Triathlon manual surgical instrumentation. Software-driven design process based on patient CT imaging.
Indications for Use
Indicated for patients undergoing total knee arthroplasty requiring femoral and tibial component positioning, where anatomic landmarks are identifiable on patient imaging scans. Compatible with Triathlon Total Knee System components (CR, CS, PS, Tritanium) and Triathlon tibial inserts (TS, PSR).
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
iTotal® Cruciate Retaining (CR) Knee Replacement System (K180906)
iTotal® Identity Posterior Stabilized Knee Replacement System (K201023)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 19, 2021
Conformis Inc. Mary Kruitwagen Sr. Regulatory Affairs Specialist 600 Technology Park Drive, 4th Floor Billerica, Massachusetts 01821
Re: K203421
Trade/Device Name: Triathlon AS-1 Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Codes: OOG, JWH Dated: March 30, 2021 Received: April 1, 2021
Dear Mary Kruitwagen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
valence determination does not mean that FDA
Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Peter G. Allen Digitally signed by Peter G. Date: 2021.04.19 13:33:19 -5 -04'00'
For: Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203421
Device Name Triathlon AS-1
## Indications for Use (Describe)
The Triathlon AS-1 patient specific cutting guides are disposable, single-use surgical instruments intended to assist orthopedic surgeons in the positioning of femoral and tibial total knee arthroplasty components intraoperatively, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. They are intended for use with the Cruciate Retaining (CR), Condylar Stabilizing (CS), Posterior Stabilized (PS) and Tritanium® components of the Triathlon® Total Knee System and the Total Stabilizer (TS) and Posterior Stabilized Rotation (PSR) Triathlon® tibial inserts.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Conformis. The logo features the word "CONFORMIS" in a stylized font, with the letters "C," "O," "N," "F," and "M" in black and the letters "O," "R," "I," and "S" in a light brown color. Above the word is a curved, orange shape that resembles a "C."
| 510(k) Number: | K203421 |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name and Address: | Conformis, Inc.<br>600 Technology Park Drive, Fourth Floor<br>Billerica, Massachusetts 01821<br>USA |
| Establishment Registration Numbers(s): | 3009844603 and 3004153240 |
| Date Summary Prepared: | November 16, 2020 |
| Contact Person: | Mary Kruitwagen<br>Title: Senior Regulatory Affairs Specialist<br>Telephone: 781-345-9038<br>Fax: 781-345-0147<br>Email: Mary.Kruitwagen@conformis.com |
| Secondary Contact: | Kara Johnson<br>Title: Director Regulatory Affairs<br>Telephone: 781-832-5402<br>Fax: 781-345-0147<br>Email: Kara.Johnson@conformis.com |
| Trade Name/Device Name: | Triathlon AS-1 |
| Common/ Usual Name: | Knee Arthroplasty Implantation System |
| Device Class: | II |
| Regulation Number: | 21 CFR 888.3560 |
| Regulation Description: | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis. |
| Product Code: | OOG, JWH |
| Product Code Regulation : | 21 CFR 888.3560 |
| Product Code Device Name: | Knee Arthroplasty Implantation System |
| Legally Marketed Predicate Device:<br>Primary Predicate: | 510(k): K180906<br>Device: iTotal® Cruciate Retaining (CR) Knee Replacement System<br>SE Date: May 16, 2018<br>Product codes: JWH, OIY, OOG<br>Class: II |
{4}------------------------------------------------
| Legally Marketed Predicate Device: | K201023 |
|------------------------------------|---------|
| Secondary Predicate: | |
| 510(k): | K201023 |
|----------------|--------------------------------------------------------|
| Device: | iTotal® Identity Posterior Stabilized Knee Replacement |
| System | |
| SE Date: | June 16, 2020 |
| Product codes: | JWH, OIY, OOG |
| Class: | II |
**Device Description:** The subject device, Triathlon AS-1, is a new Conformis Inc. device offering. The Triathlon AS-1 is comprised of patient matched single-use disposable cutting guides (also referred to as instruments or jigs) with corresponding surgical documentation which includes the Instructions for Use, Surgical Protocol and Surgical Plan. The single-use patient-matched instruments (including documentation) are similar to those of the legally marketed predicate device Knee Replacement Systems by Conformis Inc. (predicate devices K180906, K201023).
The subject device, Triathlon AS-1, does not include an implant or reusable instrumentation associated with knee replacement systems. The subject device, however, is designed to be compatible with selected legally marketed Triathlon Total Knee implants and reusable instrumentation from Stryker Orthopaedics. While no implant is part of this subject device, the subject device utilizes software to determine the size and position of the compatible implant for an individual patient and then design the subject device around the patient's anatomy to prepare for the identified implant.
For the subject device, the predicate Conformis design software, surgical plan, and single-use cutting guides (instrumentation) are being modified to accurately size, position, and prepare the bone for off-the-shelf Triathlon femoral and tibial implants. Using patient imaging (CT scans), a Triathlon Total Knee implant set is sized and positioned to best meet the unique anatomic requirements of the specific patient. The Triathlon AS-1 planning process allows for efficient 3D planning, providing optimized fit of an off-the-shelf Triathlon implant and providing the design of the patient-matched, single-use guides to prepare for the planned implant in the planned position.
Jigs are designed to fit the contours of the patient's femoral and tibial anatomies and to facilitate a simpler surgical technique. The design of the instrumentation will be modified as needed to be compatible with the Stryker Orthopaedics Triathlon manual surgical instrumentation.
Each set of instruments is designed for single use, specifically for one patient. The disposable single-use instrument set is manufactured from biocompatible nylon material and supplied sterile.
{5}------------------------------------------------
| Indications For Use: | The Triathlon AS-1 patient-specific cutting guides are disposable, single-use surgical instruments intended to assist orthopedic surgeons in the positioning of femoral and tibial total knee arthroplasty components intraoperatively, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. They are intended for use with the Cruciate Retaining (CR), Condylar Stabilizing (CS), Posterior Stabilized (PS) and Tritanium® components of the Triathlon® Total Knee System and the Total Stabilizer (TS) and Posterior Stabilized Rotation (PSR) Triathlon® tibial inserts. |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Intended Use/Indications for Use Differences | The subject device, single-use patient-matched instruments and the corresponding surgical documentation, is similar to the predicate iJigs provided with the Conformis Total Knee Replacement Systems. The predicate devices describe these similar instruments (iJigs) in their premarket submissions. The subject device performs the same function and has the same intended use as the iJigs in the predicate devices. |
| | The modifications to the single-use patient matched instruments do not raise any new issues with the safety or effectiveness as compared to the predicate single-use patient matched instruments. The cemented or uncemented implant fixation method has no bearing on the subject device. |
| | The differences in the wording of the subject device's Intended Use/Indications For Use statements do not represent a change from the Indications For Use or Intended Use of the predicate devices for the equivalent single-use patient matched instruments. |
| Summary of Technological Characteristics | There is no new technology employed as part of this subject device. |
| | The CAD design processes for the subject device are the similar to those employed for the predicate devices. The instrumentation has the same function and technological characteristics as those of the predicate devices. The subject device is composed of the same material that uses the same selective laser sintering manufacturing process as the predicate devices. The subject device uses the same packaging materials and sterilization methods as those of the predicate devices. |
| Substantial Equivalence: | The subject device is substantially equivalent to the predicate devices Conformis Total Knee Replacement System as described in their premarket submissions (K180906 and 201023). The following testing was conducted and used to establish substantial equivalence: <ul><li>Cadaver testing (surgeon evaluation and usability)</li><li>Implant sizing and positioning testing</li><li>Software verification and validation</li></ul> |
| Safety and Effectiveness Data | The assessment of the non-clinical data provided in this submission supports that the subject device is safe, effective and performs as well |
{6}------------------------------------------------
as the predicate device. No new issues of safety or effectiveness were raised.
Conclusion:
In conclusion, the information contained within this submission supports safety, effectiveness of this device and that it performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.