← Product Code [HRS](/productcode/HRS) · K203414

# DePuy Synthes 2.7mm VA LCP Clavicle Plate System, DePuy Synthes 2.7mm VA LCP Clavicle Hook Plate System (K203414)

_DePuy Synthes · HRS · Dec 16, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K203414

## Device Facts

- **Applicant:** DePuy Synthes
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Dec 16, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The DePuy Synthes 2.7mm VA LCP Clavicle Plate System is indicated for fixation of fractures, and nonunions of the clavicle in adults, and in both adolescents (12-18 years) and transitional adolescents (18-21 years), in which the clavicular growth plates have fused or in which the growth plates will not be crossed by the plate system. The DePuy Synthes 2.7mm VA LCP Clavicle Hook Plate System is indicated for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.

## Device Story

Metallic bone fixation system for clavicle fractures; includes lateral, shaft, and medial plates, plus hook plates (Long, Short, Button types). Used for temporary fixation, correction, or stabilization during open reduction internal fixation (ORIF). Operated by surgeons in healthcare facilities. Implants available in stainless steel or titanium alloy; non-implantable templates used for size selection. Plates provide mechanical stabilization of bone segments to facilitate healing. No electronic components or software.

## Clinical Evidence

No clinical data; clinical testing was not necessary for the determination of substantial equivalence.

## Technological Characteristics

Metallic bone fixation plates (stainless steel or titanium alloy). Includes non-implantable templates. Sterile plates sterilized via gamma irradiation. No software, electronics, or energy sources.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- DePuy Synthes 2.7mm VA LCP Clavicle Plate System ([K201321](/device/K201321.md))
- DePuy Synthes 2.7mm VA LCP Clavicle Hook Plate System ([K201959](/device/K201959.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 16, 2020

DePuy Synthes % Georgina Mueller Regulatory Affairs Specialist II Synthes GmbH Luzernstrasse 21 Zuchwil. SO 4528 Switzerland

Re: K203414

Trade/Device Name: DePuy Synthes 2.7mm VA LCP Clavicle Plate System, DePuy Synthes 2.7mm VA LCP Clavicle Hook Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: November 16, 2020 Received: November 19, 2020

Dear Georgina Mueller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Stereotaxic, Trauma and Restorative Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K203414

Device Name

DePuy Synthes 2.7mm VA LCP Clavicle Plate System

Indications for Use (Describe)

The DePuy Synthes 2.7mm VA LCP Clavicle Plate System is indicated for fixation of fractures, and nonunions of the clavicle in adults, and in both adolescents (12-18 years) and transitional adolescents (18-21 years), in which the clavicular growth plates have fused or in which the growth plates will not be crossed by the plate system.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K203414

Device Name

DePuy Synthes 2.7mm VA LCP Clavicle Hook Plate System

Indications for Use (Describe)

The DePuy Synthes 2.7mm VA LCP Clavicle Hook Plate System is indicated for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.

Type of Use (Select one or both, as applicable)

| <span style="font-size: 10pt;"> <span style="font-family: Wingdings;">ü</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|

| X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

## DePuy Synthes 2.7mm VA LCP Clavicle Plate System

| Sponsor                          | DePuy Synthes<br>Georgina Mueller<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>Phone: +41 61 965 63 14                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared                    | November 16, 2020                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Proprietary Name                 | DePuy Synthes 2.7mm VA LCP Clavicle Plate System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification Name              | Single/multiple component metallic bone fixation appliances and<br>accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification                   | Class II<br>Regulation Number: 21 CFR 888.3030<br>Product Code: HRS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Predicate device                 | DePuy Synthes 2.7mm VA LCP Clavicle Plate System (K201321)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Description               | The DePuy Synthes 2.7mm VA LCP Clavicle Plate System consists of<br>lateral, shaft and medial plates designed for temporary fixation,<br>correction or stabilization of clavicle bones. The subject plates are<br>available in various sizes in both sterile and non-sterile configurations<br>and are available in stainless steel and titanium alloy. The system also<br>consists of non-implantable templates that correspond to the implants.<br>Templates are intended for implant size selection and are available in<br>non-sterile.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Indications for use              | Fixation of fractures, osteotomies, and non-unions of the clavicle in<br>adults, and in both adolescents (12-18 years) and transitional<br>adolescents (18-21 years), in which the clavicular growth plates have<br>fused or in which the growth plates will not be crossed by the plate<br>system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Technological Characteristics    | The design, features, and specifications of the subject device remain<br>unchanged compared to the previously cleared version of this device<br>cleared via K201321 (predicate device).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Non-clinical Performance<br>Data | Non-clinical performance data was not necessary for the determination<br>of substantial equivalence.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Clinical Performance Data        | Clinical testing was not necessary for the determination of substantial<br>equivalence.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Substantial Equivalence          | The following subject device characteristics remain unchanged<br>compared to the previously cleared version of these devices<br>(predicate):<br>- Intended for the use in temporary fixation, correction or stabilization<br>of clavicle bones during open reduction internal fixation (ORIF)<br>performed by surgeons within a health care facility. Plates are<br>available sterile and non-sterile. Sterile plates are sterilized by<br>gamma irradiation.<br>- Indicated for fixation of fractures, osteotomies, and non-unions of<br>the clavicle in adults, and in both adolescents (12-18 years) and<br>transitional adolescents (18-21 years), in which the clavicular<br>growth plates have fused or in which the growth plates will not be<br>crossed by the plate system.<br>- Design, features, and specifications (technological characteristics)<br>The subject devices have added contraindications compared to the<br>predicate devices which has no contraindications. The differences in<br>contraindications of the subject devices compared to the predicate<br>devices do not raise any new questions of safety and effectiveness.<br>It is concluded that the information provided herein supports substantial<br>equivalence of the subject devices. |

{5}------------------------------------------------

# DePuy Synthes 2.7mm VA LCP Clavicle Hook Plate System

| Sponsor                          | DePuy Synthes<br>Georgina Mueller<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>Phone: +41 61 965 63 14                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared                    | November 16, 2020                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Proprietary Name                 | DePuy Synthes 2.7mm VA LCP Clavicle Hook Plate System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification Name              | Single/multiple component metallic bone fixation appliances and<br>accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification                   | Class II<br>Regulation Number: 21 CFR 888.3030<br>Product Code: HRS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Predicate device                 | DePuy Synthes 2.7mm VA LCP Clavicle Hook Plate System (K201959)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device Description               | The DePuy Synthes 2.7mm VA LCP Clavicle Hook Plate System<br>consists of the plate types Long, Short and Button which can be used                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                  | to treat simple and complex clavicle fractures including malunions, non- unions and isolated ligamentous injuries of the AC joint. The subject plates are available in three hook depths, in left- and right-side versions, in both sterile and non-sterile configurations and are available in stainless steel and titanium. The system also consists of non- implantable templates that correspond to the implants. Templates are intended for implant size selection and are available non-sterile.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Indications for use              | The DePuy Synthes 2.7mm VA LCP Clavicle Hook Plate System is indicated for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Technological Characteristics    | The design, features, and specifications of the subject device remain unchanged compared to the previously cleared version of this device cleared via K201959 (predicate device).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Non-clinical Performance<br>Data | Non-clinical performance data was not necessary for the determination of substantial equivalence.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Clinical Performance Data        | Clinical testing was not necessary for the determination of substantial equivalence.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Substantial Equivalence          | The following subject device characteristics remain unchanged compared to the previously cleared version of these devices (predicate):  Intended for the use in fixation of the clavicle and the acromioclavicular joint during open reduction internal fixation (ORIF) performed by surgeons within a health care facility. Subject plates are available sterile and nonsterile, whereas the predicate plates are available sterile only. Sterile plates are sterilized by gamma irradiation. Indicated for fixation of lateral clavicle fractures and dislocations of the acromioclavicular joint. Design, features, and specifications (technological characteristics)  The subject devices have revised contraindications compared to the compared to the predicate device. The differences in contraindications of the subject devices compared to the predicate devices do not raise any new questions of safety and effectiveness. It is concluded that the information provided herein supports substantial equivalence of the subject devices. |

{6}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K203414](https://fda.innolitics.com/device/K203414)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
