K203385 · Paragon 28, Inc. · MBI · Feb 11, 2021 · Orthopedic
Device Facts
Record ID
K203385
Device Name
TenoTac Soft Tissue Fixation System
Applicant
Paragon 28, Inc.
Product Code
MBI · Orthopedic
Decision Date
Feb 11, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The TenoTac® Soft Tissue Fixation System is intended to be used for soft tissue to bone fixation. Specific indications for the TenoTac® device include: Foot & Ankle: Medial/lateral repair and reconstruction, mid and forefoot repair, hallux malleus repair and reconstruction, metatarsal ligament/tendon repair and reconstruction including plantar plate attenuation and tear, and the correction of hammer toe, claw toe, mallet toe, crossover toe, floating toe, and any other lesser toe deformities, correction of metatarsophalangeal joint instability due to shortening from interphalangeal fusion, correction of metatarsophalangeal joint instability due to shortening from Weil osteotomy, Achilles tendon repair. Hand & Wrist: Collateral ligament repair, Scapholunate ligament reconstruction, tendon transfers in phalanx, Volar plate reconstruction.
Device Story
TenoTac® Soft Tissue Fixation System; two-piece titanium alloy (Ti-6Al-4V ELI) construct consisting of threaded male tack and threaded female sleeve; used for soft tissue-to-bone fixation in foot, ankle, hand, and wrist; inserted into bone canal to secure soft tissue; operated by orthopedic surgeons in clinical/surgical settings; provides mechanical fixation to facilitate tissue healing; benefits patient through stabilization of ligaments and tendons during reconstruction or deformity correction.
Clinical Evidence
No clinical or non-clinical testing was necessary to support substantial equivalence.
Technological Characteristics
Construct: two-piece threaded titanium alloy (Ti-6Al-4V ELI per ASTM F136) tack and sleeve system. Available in various sizes/lengths. Mechanical fixation principle. No software or energy source.
Indications for Use
Indicated for soft tissue to bone fixation in foot, ankle, hand, and wrist procedures, including ligament/tendon repair, reconstruction, and correction of toe deformities (hammer, claw, mallet, crossover, floating) and joint instability.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
February 11, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Paragon 28, Inc Haylie Hertz Regulatory Affairs Specialist 14445 Grasslands Dr. Englewood, Colorado 80112
Re: K203385
Trade/Device Name: TenoTac® Soft Tissue Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: November 13, 2020 Received: November 18, 2020
Dear Haylie Hertz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203385
Device Name TenoTac® Soft Tissue Fixation System
Indications for Use (Describe)
The TenoTac® Soft Tissue Fixation System is intended to be used for soft tissue to bone fixation. Specific indications for the TenoTac® device include:
Foot & Ankle: Medial/lateral repair and reconstruction, mid and forefoot repair, hallux malleus repair and reconstruction, metatarsal ligament/tendon repair and reconstruction including plantar plate attenuation and tear, and the correction of hammer toe, claw toe, mallet toe, crossover toe, floating toe, and any other lesser toe deformities, correction of metatarsophalangeal joint instability due to shortening from interphalangeal fusion, correction of metatarsophalangeal joint instability due to shortening from Weil osteotomy, Achilles tendon repair.
Hand & Wrist: Collateral ligament repair, Scapholunate ligament reconstruction, tendon transfers in phalanx, Volar plate reconstruction.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| 510(k) Number: | K203385 |
|--------------------|------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | TenoTac® Soft Tissue Fixation System |
| Manufacturer: | Paragon 28, Inc.<br>14445 Grasslands Dr.<br>Englewood, CO 80112 |
| Contact: | Haylie Hertz<br>Regulatory Affairs Specialist<br>Paragon 28, Inc.<br>Phone: 303-720-0017<br>hhertz@paragon28.com |
| Date Prepared: | February 10, 2021 |
| Common Name: | Soft Tissue Fixation Device |
| Classification: | 21 CFR 888.3040 |
| Class: | II |
| Product Code: | MBI |
### Predicate Device:
The TenoTac® Soft Tissue Fixation System (K182898) serves as the primary predicate device.
#### Reference Device:
The Zimmer Biomet (formerly Biomet Sports Medicine) JuggerKnot™ Soft Anchors (K110145) serve as a reference device.
#### Indications for Use:
The TenoTac® Soft Tissue Fixation System is intended to be used for soft tissue to bone fixation.
Specific indications for the TenoTac® include:
Foot & Ankle: Medial/lateral repair and reconstruction, mid and forefoot repair, hallux valgus repair, hallux malleus repair and reconstruction, metatarsal ligament/tendon repair and reconstruction including plantar plate attenuation and tear, and the correction of hammer toe, claw toe, mallet toe, crossover toe, floating toe, and any other lesser toe deformities, correction of metatarsophalangeal joint instability due to shortening from interphalangeal fusion, correction of metatarsophalangeal joint instability due to shortening from Weil osteotomy, Achilles tendon repair.
{4}------------------------------------------------
Hand & Wrist: Collateral ligament repair, Scapholunate ligament reconstruction, tendon transfers in phalanx, Volar plate reconstruction.
## Device Description:
The TenoTac® Soft Tissue Fixation System is manufactured from titanium alloy (Ti-6Al4V ELI per ASTM F136) and is comprised of specialized threaded tacks/male implants and associated threaded sleeves/female implants for attaching soft tissue to bone. The tacks and sleeves are available in various sizes and lengths to accommodate different bone sizes.
## Substantial Equivalence:
The TenoTac Soft Tissue Fixation system is intended for soft tissue to bone fixation. Compared to the predicate TenoTac device, the subject device is a two-piece titanium tack and sleeve construct inserted on two ends of a bone canal. There are no differences between these devices. Compared to the JuggerKnot device, the subject device achieves the same intended effect by fixing soft tissue through a bone canal. Differences include the shape of the anchors and the material of the anchors. All indications for the subject device are within the indications of the predicate devices. In addition, the subject device possesses the same technological characteristics as the predicate device, including performance, basic design, material, manufacturing, and sizes. Therefore, the TenoTac Soft Tissue Fixation System is substantially equivalent to the predicate devices with respect to indications, design, function, and performance.
## Performance Testing:
Neither clinical or non-clinical testing were necessary to support substantial equivalence of the subject device.
## Conclusion:
The TenoTac Soft Tissue Fixation System possesses the same intended use and technological characteristics as the predicate and reference devices. Therefore, the TenoTac Soft Tissue Fixation System is substantially equivalent for its intended use.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.