ASTM Level 1/EN14683 Type IIR 3-Ply disposable Surgical Mask, Model number: FM-1400G
K203376 · Medtecs (Taiwan) Corp. · FXX · Mar 2, 2022 · General, Plastic Surgery
Device Facts
| Record ID | K203376 |
| Device Name | ASTM Level 1/EN14683 Type IIR 3-Ply disposable Surgical Mask, Model number: FM-1400G |
| Applicant | Medtecs (Taiwan) Corp. |
| Product Code | FXX · General, Plastic Surgery |
| Decision Date | Mar 2, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4040 |
| Device Class | Class 2 |
Intended Use
ASTM Level 1/EN14683 Type IIR 3-Ply disposable Surgical Mask, Model number: FM-140G are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids and particulate material. The face masks are single use, disposable device, provided non-sterile.
Device Story
Single-use, 3-ply flat-folded surgical mask; inner/outer layers spun-bond polypropylene; middle layer melt-blown polypropylene filter. Features elastic ear loops (latex-free) and malleable polyethylene nose piece with galvanized iron wire. Used in operating rooms by surgical personnel to provide barrier protection against microorganisms, body fluids, and particulates. Device is non-sterile, disposable, and intended for single use.
Clinical Evidence
Bench testing only. Testing performed on 3 non-consecutive lots (96 samples total) per ASTM F2100-19. Results: Fluid resistance (ASTM F1862) met criteria at 80 mmHg; Bacterial Filtration Efficiency (ASTM F2101) >99%; Differential Pressure (EN14683 Annex C) <5.0 mmH2O/cm2; Particulate Filtration Efficiency (ASTM F2299) >95%; Flammability (16 CFR 1610) Class I. Biocompatibility testing (ISO 10993-5, -10) confirmed non-cytotoxic, non-irritating, and non-sensitizing.
Technological Characteristics
3-ply construction: spun-bond polypropylene (inner/outer), melt-blown polypropylene (middle). Malleable polyethylene nose piece with galvanized iron wire. Dimensions: 17.5 cm x 9.3 cm. Non-sterile. Latex-free. Complies with ASTM F2100-19 and EN14683 standards.
Indications for Use
Indicated for operating room personnel during surgical procedures to protect the patient and personnel from transfer of microorganisms, body fluids, and particulate material.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
- Surgical Face Mask with Ear-Loop, YN-50 JAG (K123115)
Reference Devices
- Avianz® Surgical Face Mask (K200847)
Related Devices
- K123115 — ACME SURGICAL FACE MASK WITH EAR-LOOP MODEL YN-501 AB/W/G/P · Acme Filter Mask, Inc. · May 20, 2013
- K043199 — SURGICAL FACE MASK · Hong YA Non-Woven Products Co., Ltd. · Feb 16, 2005
- K202491 — Disposable Surgical Face Mask · Jiangsu Newvalue Medical Products Co., Ltd. · Jan 6, 2021
- K210433 — Surgical Face Mask (Ear loops and Tie-on) · Wuhan Dymex Healthcare Co., Ltd. · Jun 7, 2021
- K210042 — Disposable Surgical Mask · Suzhou Letian Protective Products Co., Ltd. · Nov 17, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" below it in a smaller font.
March 2, 2022
Medtecs (Taiwan) Corp. % Sandy Liu Consultant Jin Services Co. 9F-1, No13, Lane41, Zhangrong Rd, Sec. 5, North District Tainan City, 70447 Taiwan
Re: K203376
Trade/Device Name: ASTM Level 1/EN14683 Type IIR 3-Ply Disposable Surgical Mask, Model number: FM-1400G Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: February 7, 2022 Received: February 18, 2022
Dear Sandy Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203376
Device Name
ASTM Level 1/EN14683 Type IIR 3-Ply disposable Surgical Mask, Model number: FM-140G
Indications for Use (Describe)
ASTM Level 1/EN14683 Type IIR 3-Ply disposable Surgical Mask, Model number: FM-140G are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids and particulate material. The face masks are single use, disposable device, provided non-sterile.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MEDTECS International Corporation Limited. The logo features a stylized human figure in blue and green, with a blue circle representing the head and green shapes forming the body. To the right of the figure, the word "MEDTECS" is written in blue, with the "T" in green. Below "MEDTECS", the words "INTERNATIONAL CORPORATION LIMITED" are written in smaller blue font.
# K203376 510(k) Summary As required by 21CFR 807.92 Date of Preparation: 2020.10.30
#### Applicant Information
| Company Name: | MEDTECS (TAIWAN) CORP. |
|---------------------------|----------------------------|
| Company Address: | 11F., No. 9, Songgao Road |
| | Xinyi Dist., 11073, Taiwan |
| Telephone: | +886-2-27392222 |
| Fax: | +886-2-27297896 |
| Contact Person: | William Yang |
| Summary Preparation Date: | 2020.10.30 |
## Official Correspondent
| Company Name: | Jin Services Co. |
|------------------|-------------------------------------------------------------------------------------------|
| Company Address: | 9F-1, No13, Lane41, Zhangrong Rd, Sec. 5, North Distrit, Tainan City, Taiwan70447, Taiwan |
| Telephone: | +886-917535026 |
| Email: | contact@fdaclass.com |
| Contact Person: | Sandy Liu, Consultant |
#### Device Name:
| Trade Name: | ASTM Level 1/EN14683 Type IIR 3-Ply disposable Surgical Mask, |
|----------------------|---------------------------------------------------------------|
| Model number: | FM-140G |
| Classification Name: | Surgical Mask |
| Regulation Number: | 878.4040 |
| Product Code: | FXX |
| Device Class: | Class II |
| Panel: | General Hospital |
## PREDICATE DEVICE:
K123115, Surgical Face Mask with Ear-Loop, YN-50 JAG Acme Filter Mask Inc.
# REFERENCE DEVICE:
K200847, Avianz® Surgical Face Mask
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for MEDTECS International Corporation Limited. The logo features a green and blue figure on the left, followed by the company name in blue with a green "T". Underneath the company name is the text "INTERNATIONAL CORPORATION LIMITED" in a smaller font. The logo is clean and professional, and the colors are eye-catching.
## MEXPO INTERNATIONAL INC. Device Description
The ASTM Level 1/EN14683 Type IIR 3-Ply disposable Surgical Mask, Model number: FM-140G is a single-use, three layer, flat-folded mask with ear loops and nose piece. The inner and outer layers are constructed of spun-bond non-woven polypropylene and the middle layer is constructed of melt blown non-woven polypropylene. The mask is held in place over the mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece is made of malleable polyethylene with Galvanized
iron wire and allows the user to fit the facemask around their nose. The ASTM Level 1/EN14683 Type IIR 3-Ply disposable Surgical Mask, Model number: FM-140G is sold non-sterile and is intended to be a single use, disposable device.
#### Intended Use:
ASTM Level 1/EN14683 Type IIR 3-Ply disposable Surgical Mask, Model number: FM-140G are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids and particulate material. The face masks are single use, disposable device, provided non-sterile.
#### Technological Characteristics Comparison
The following is a summary of the technological characteristics of the ASTM Level 1/EN14683 Type IIR 3-Ply disposable Surgical Mask, Model number: FM-140G as compared to the predicate device.
| Items | Subject Device<br>ASTM Level 1/EN14683 Type<br>IIR 3-Ply disposable Surgical<br>Mask, Model number: FM-<br>140G | Predicate Device<br>Surgical Face Mask with Ear-Loop,<br>YN-50 JAG | Comparison<br>Result | |
|--------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|----------------------|
| Submitter | MEDTECS (TAIWAN) CORP. | Acme Filter Mask Inc. | N/A | |
| 510(k)<br>Number | N/A | K123115 | N/A | |
| Device<br>Regulation<br>number | 878.4040 | 878.4040 | same | |
| Classification | II | II | same | |
| | Subject Device | Predicate Device | | |
| Items | ASTM Level 1/EN14683 Type<br>IIR 3-Ply disposable Surgical<br>Mask, Model number: FM-140G | Surgical Face Mask with Ear-Loop,<br>YN-50 JAG | Comparison<br>Result | |
| FDA Product<br>Code | FXX | FXX | same | |
| Indications<br>for Use | ASTM Level 1/EN14683 Type<br>IIR 3-Ply disposable Surgical<br>Mask, Model number: FM-<br>140G are intended to be worn<br>by operating room personnel<br>during surgical procedures to<br>protect both the surgical patient<br>and the operating room<br>personnel from transfer of<br>microorganisms, body fluids<br>and particulate material. The<br>face masks are single use,<br>disposable device, provided<br>non-sterile. | Surgical Face Mask is Device that is<br>intended to be worn by operating room<br>personnel during surgical procedures to<br>protect both the surgical patient and the<br>operation room personnel from transfer<br>of microorganisms, body fluids and<br>particulate material. | Identical | |
| Prescription<br>for use | No | No | same | |
| Over the<br>Counter | Yes | Yes | same | |
| | Design | | | |
| Inner and<br>Outer Layers | Spun-bond polypropylene | Spun-bond polypropylene | same | |
| Middle Layer | Melt blown polypropylene filter | Melt blown polypropylene filter<br>Identical | same | |
| Ear loops | not made with natural rubber latex | not made with natural rubber latex | same | |
| Nose Piece | Malleable polyethylene with<br>Galvanized iron wire | Malleable polyethylene with<br>aluminum wire | Different | |
| Dimensions | $17.5$ cmL x $9.3$ cm $\pm 0.5$ cm H<br>( $6.89$ x $3.66\pm 0.2$ inches) | $17.5$ cm L x $9.5$ cm H | Identical | |
| | | Subject Device<br>ASTM Level 1/EN14683 Type<br>IIR 3-Ply disposable Surgical<br>Mask, Model number: FM-<br>140G | Predicate Device<br>Surgical Face Mask with Ear-Loop,<br>YN-50 JAG | Comparison<br>Result |
| Items | Mask Style | 3 flats pleated | 3 flats pleated | same |
| | Design<br>Features | Malleable nosepiece, flat<br>pleated, elastic ear loops | Malleable nosepiece, flat pleated,<br>elastic ear loops | same |
| | Model size | One-Size (regular) fits all | One-Size (regular) fits all | same |
| | Sterility | Non-sterile | Non-sterile | same |
| | Use | Single Use, Disposable | Single Use, Disposable | same |
| | Color | Green | Green | same |
| | Contain any<br>drugs or<br>biologics | No | No | same |
| | face shield<br>attached | no | no | same |
| | Foam strip<br>attached | no | no | same |
| | LATEX-<br>FREE | Yes | Yes | same |
| | | Performance | | |
| | ASTM F2100<br>Level | Level 1 | Level 1 | same |
| | Fluid<br>Resistance<br>Performance | Fluid Resistance | Fluid Resistance | identical |
| | Bacterial<br>Filtration<br>Efficiency | Higher than 99% | Higher than 99% | identical |
| | Differential<br>Pressure<br>(Delta P) | Avg of 3.1 $mmH_2O/cm^2$ | Average 3.33 ( $mmH_2O/cm^2$ ) | Different |
| | Particulate<br>Filtration | Avg of 96.14% for 0.1 Sub-micron<br>Particulate Filtration | Average 94.79% for Solid Aerosol<br>Filtration Efficiency | Different* |
| Items | Subject Device<br>ASTM Level 1/EN14683 Type<br>IIR 3-Ply disposable Surgical Mask, Model number: FM-140G | Predicate Device<br>Surgical Face Mask with Ear-Loop, YN-50 JAG | Comparison Result | |
| Efficiency | | | | |
| Flammability | Class I (No Flame Spread) | Class I (No Flame Spread) | Identical | |
| Biocompatibility | | | | |
| Cytotoxicity,<br>ISO10993-5 | Non-cytotoxic | Non-cytotoxic | Identical | |
| Irritation,<br>ISO10993-10 | Non-irritating | Non-irritating | Identical | |
| Sensitization,<br>ISO10993-10 | Non-sensitizing | Non-sensitizing | Identical | |
| Labeling, Package and Storage | | | | |
| Storage indication | Store in a dry and well-ventilated environment. Avoid high temperature and keep away from fire and flammable materials. | Store in a dry and well-ventilated environment. Avoid high temperature and keep away from fire and flammable materials. | Identical | |
| Product labeling | All information showing on the Gift box | All information showing on the Gift box | Same | |
| Package materials | Paper un-seal gift box | Paper un-seal gift box | Same | |
| Product package | Paper Gift box (50 pcs/box)<br>Paper Carton box (40 gift boxes/carton) | Paper Gift box (50 pcs/box)<br>No information for Carton box | Identical | |
| UDI included on the box | Yes (both gift box and Carton) | No information for UDI | Different | |
| Shelf Life Claim | 5 Years | 5 Years | Same | |
| Item | Standard | Acceptance Criteria<br>(for Level 1 barrier) | Results | |
| Fluid Resistance<br>Performance | F2100-19 clause 9.4/ASTM F1862-17:<br>Standard Test Method for<br>Resistance of Medical<br>Face Masks to Penetration<br>by Synthetic Blood | At least 29 out of 32 specimens show passing<br>results at 80 mmHg | All samples met the<br>predetermined acceptance<br>criteria. | |
| Bacterial<br>Filtration<br>Efficiency | F2100-19 clause 9.1/ASTM F2101-19:<br>Standard Test Method for<br>Evaluating the Bacterial<br>Filtration Efficiency<br>(BFE) of Medical Face | ≥ 95% | All samples met the<br>predetermined acceptance<br>criteria. | |
| Item | Standard | Acceptance Criteria<br>(for Level 1 barrier) | Results | |
| Differential<br>Pressure (Delta P) | Mask Materials, Using a<br>Biological Aerosol of<br><i>Staphylococcus aureus</i> | | | |
| | F2100-19 clause<br>9.2/EN14683:2019<br>Medical Face Masks—<br>Requirements and Test<br>Methods Annex C | < 5.0 mm H2O/cm² | All samples met the<br>predetermined acceptance<br>criteria. | |
| | | | | |
| | | | | |
| Particulate<br>Filtration<br>Efficiency | F2100-19 clause<br>9.3/ASTM F2299-17: Test<br>Method for Determining<br>the Initial Efficiency of<br>Materials Used in Medical<br>Face Masks to Penetration<br>by Particulates Using<br>Latex Spheres | ≥ 95% | All samples met the<br>predetermined acceptance<br>criteria. | |
| | Particulates Using Latex | | | |
| | Spheres | | | |
| | | | | |
| | | | | |
| | | | | |
| | | | | |
| Flammability | F2100-19 clause 9.5/16<br>CFR 1610-2008: Standard<br>for the Flammability of<br>Clothing Textiles | Class I, does not Ignite | All samples met the<br>predetermined acceptance<br>criteria. | |
| | | | | |
| | | | | |
| Cytotoxicity | ISO10993-5 Third edition:<br>Biological evaluation of<br>medical devices-Part 5:<br>Tests for in vitro<br>cytotoxicity | Non-cytotoxic | All samples met the<br>predetermined acceptance<br>criteria. | |
| | | | | |
| | | | | |
| Irritation | ISO10993-10 Third<br>Edition: Biological<br>evaluation of medical<br>devices-Part 10: Tests for<br>irritation and skin<br>sensitization | Non-irritating | All samples met the<br>predetermined acceptance<br>criteria. | |
| | | | | |
| | | | | |
| Item | Standard | Acceptance Criteria<br>(for Level 1 barrier) | Results | |
| Sensitization | ISO10993-10 Third<br>Edition: Biological<br>evaluation of medical<br>devices-Part 10: Tests for<br>irritation and skin<br>sensitization | Non-sensitizing | All samples met the<br>predetermined acceptance<br>criteria. | |
| Performance<br>Testing of<br>Shipping<br>Containers and<br>Systems | ASTM D4169: Standard<br>Practice for Performance<br>Testing of Shipping<br>Containers and Systems | No visible damage was<br>found on sample<br>appearance after the test<br>(Drop, Compression,<br>Fixed vibration,<br>Altitude, Vibration,<br>Concentrated Impact) | All samples met the<br>predetermined acceptance<br>criteria. | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MEDTECS International Corporation Limited. The logo features a green and blue figure on the left, followed by the company name in bold, blue letters. The "T" in MEDTECS is green, and the words "INTERNATIONAL CORPORATION LIMITED" are printed in smaller letters below the company name.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for MEDTECS International Corporation Limited. The logo features a stylized figure in blue and green, with a blue circle representing the head and green shapes forming the body. To the right of the figure is the company name, "MEDTECS," with the "T" in green and the rest of the letters in blue. Below the company name is the text "INTERNATIONAL CORPORATION LIMITED" in a smaller font size.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for MEDTECS International Corporation Limited. The logo features a blue and green figure of a person on the left side. To the right of the figure is the company name, "MEDTECS INTERNATIONAL CORPORATION LIMITED", with the "T" in "MEDTECS" being green.
## Summary of Non-Clinical Testing
Per FDA document Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submissions, the below testing has been completed on 3 nonconsecutive lots of 32 samples for a total of 96 samples of the ASTM Level 1/EN14683 Type IIR 3-Ply disposable Surgical Mask, Model number: FM-140G ,
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for MEDTECS International Corporation Limited. The logo features a stylized human figure in blue and green, with the head in blue and the body in green. To the right of the figure is the company name, "MEDTECS," with the "T" in green and the rest of the letters in blue. Below the company name is the text "INTERNATIONAL CORPORATION LIMITED" in a smaller font size.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for MEDTECS International Corporation Limited. The logo features a green and blue figure on the left, with the word "MEDTECS" in blue and green on the right. The words "INTERNATIONAL CORPORATION LIMITED" are in smaller blue font below the word "MEDTECS".
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for MEDTECS International Corporation Limited. The logo features a stylized human figure in blue and green, with the head in blue and the body in green. To the right of the figure is the company name, "MEDTECS," with the "T" in green and the rest of the letters in blue. Below the company name is the text "INTERNATIONAL CORPORATION LIMITED" in a smaller font size.
## Conclusions:
The conclusion drawn from the non-clinical tests demonstrates that the subject device, the ASTM Level 1/EN14683 Type IIR 3-Ply disposable Surgical Mask, Model number: FM-140G, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K123115.