ENDOtube

K203350 · Nissha Medical Technologies Sas · OCX · Sep 29, 2021 · Gastroenterology, Urology

Device Facts

Record IDK203350
Device NameENDOtube
ApplicantNissha Medical Technologies Sas
Product CodeOCX · Gastroenterology, Urology
Decision DateSep 29, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The ENDOtube devices are a combination of tubes, fittings and valves intended to provide sterile water and/or CO2 and/ or air to an endoscope during gastrointestinal endoscopic procedure. These devices are intended to be used by healtheare professional only. In procedure room, all ENDOtube devices are 24 hours use disposable products except the one-way valves which are single use products and the insufflation connectors which are multiple patient use.

Device Story

ENDOtube kits consist of tubes, fittings, and valves; provide sterile water, air, or CO2 to endoscopes during GI procedures. Components include wash bottle caps, irrigation tubing for peristaltic pumps, and insufflation tubing with filters. Used in procedure rooms by healthcare professionals. Connects to endoscope air/water channels or water-jet channels; interfaces with various endoscope brands (Olympus, Pentax, Fujifilm) and irrigation pumps (Olympus, Erbe, Emed). Provides sterile fluid/gas delivery; supports visualization and irrigation during endoscopy. Benefits include maintaining sterile supply lines and preventing cross-contamination via single-use valves and disposable tubing.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing only, including connection/usability testing, leak testing, durability testing, sterilization validation, and biocompatibility testing per AAMI/ISO 10993.

Technological Characteristics

Materials: PVC, silicone, SAN, ABS, PC, TPU. Biocompatible, DEHP-free. Sensing/Actuation: Passive fluid/gas delivery via peristaltic pump or bottle pressure. Connectivity: Mechanical interface to endoscopes/pumps. Sterilization: EO for tubes/valves; insufflation connectors are reprocessable. Form factor: Tubing kits with fittings/valves.

Indications for Use

Indicated for use by healthcare professionals during gastrointestinal endoscopic procedures to provide sterile water, air, or CO2 to an endoscope.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. September 29, 2021 Nissha Medical Technologies SAS Roch Mignot Ouality & Regulatory Affairs Manager ZA Des Boutries 12 Rue Des Cayennes Conflans STE Honorine, Yvelines 78700 FRANCE Re: K203350 Trade/Device Name: ENDOtube Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OCX Dated: August 12, 2021 Received: August 19, 2021 Dear Roch Mignot: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {2}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K203350 Device Name ENDOtube ### Indications for Use (Describe) The ENDOtube devices are a combination of tubes, fittings and valves intended to provide sterile water and/or CO2 and/ or air to an endoscope during gastrointestinal endoscopic procedure. These devices are intended to be used by healtheare professional only. In procedure room, all ENDOtube devices are 24 hours use disposable products except the one-way valves which are single use products and the insufflation connectors which are multiple patient use. Type of Use (Select one or both, as applicable) | <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |----------------------------------------------------------------------------| | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for "NISSHA MEDICAL TECHNOLOGIES". The word "NISSHA" is written in large, bold, blue letters. Below the word "NISSHA" is the phrase "MEDICAL TECHNOLOGIES" written in smaller, light gray letters. The logo is simple and modern. #### SUBMITTER l. ## NISSHA MEDICAL TECHNOLOGIES SAS Z.A. DES BOUTRIES 12, RUE DES CAYENNES B.P. 310 78703 CONFLANS SAINTE HONORINE CEDEX France Phone: +33 1 39 72 66 66 Fax: +33 1 39 72 61 61 Under registration number: 3004762958 ## A division of NISSHA MEDICAL TECHNOLOGIES 400 EXCHANGE STREET BUFFALO, NY 14204 United States of America Tel: 716.853.7500 ## Contact Person: Roch MIGNOT, Quality and Requlatory Affairs Manager - France Tel: +33 1 39 72 77 - Cellphone: +33 7 85 19 39 07 Email: rmignot@nisshamedical.com Date Prepared: October 30, 2020 Date Revised: December 30, 2020 Date Latest Reviewed: August 03, 2021 #### II. DEVICE | Device / Property Name | ENDOtube | |------------------------------------------|--------------------------------------| | Common / Usual Name | Endoscopic tubing kit | | Classification / Regulation Number 21CFR | 876.1500 | | Classification Name | Endoscope and accessories | | Product Code | OCX | | Product Code Name | Endoscopic Irrigation/Suction System | | Regulatory Class | II | | Device Panel | Gastroenterology/Urology | The ENDOtube products include several families as detailed bellow in VOL_002_Submission Context in the file 005_Product List and bellow. {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Nissha Medical Technologies. The word "NISSHA" is written in large, bold, blue letters. Below the word "NISSHA" is the phrase "MEDICAL TECHNOLOGIES" in smaller, gray letters. The logo is simple and modern, and the colors are clean and professional. | Family | Reference | Description | |----------------------------------------------|---------------|---------------------------------------------------------------------------------------------------------| | ENDOtube for Wash Bottle | S-FLO000001 | Sterile disposable tubing - Olympus Gl endoscopes | | | S-FLP000002 | Sterile disposable tubing - Pentax Gl endoscopes | | | S-FLF000003 | Sterile disposable tubing - Fujifilm Gl endoscopes 5-600 series | | | S-FLF000004 | Sterile disposable tubing - Fujifilm Gl endoscopes 700 series | | ENDOtube for Wash Bottle | S-FLO001CO2 | Sterile disposable tubing + CO2 option - Olympus Gl endoscopes | | ENDOtube for Wash Bottle | S-FLP002CO2 | Sterile disposable tubing + CO2 option - Pentax Gl endoscopes | | ENDOtube for Wash Bottle | S-FLF003CO2 | Sterile disposable tubing + CO2 option - Fujifilm Gl endoscopes 5-600 series | | ENDOtube for Wash Bottle | S-FLF004CO2 | Sterile disposable tubing + CO2 option - Fujifilm Gl endoscopes 700 series | | ENDOtube for Insufflation | S-TUBCO2LUER | CO2 insufflation sterile tubing with hydrophobic filter - Luer connection | | | S-TUBCO2FUJI | CO2 insufflation sterile tubing with hydrophobic filter - Adapter connection | | | 04628 | Connection adapter for insufflator FUJIFILM GW-1 | | ENDOtube for Insufflation | 04700 | Connection adapter for insufflator OLYMPUS UCR | | ENDOtube for Insufflation | 04712 | Connection adapter for insufflator FUJIFILM GW-100 | | ENDOtube for Wash Pump | S-PLO000100 | Sterile disposable irrigation tubing - Emed pump/Olympus Gl endoscopes | | | S-PLO000300 | Sterile disposable irrigation tubing - Olympus pump/Olympus Gl endoscopes | | | S-PLO000400 | Sterile disposable irrigation tubing - Endogator pump/Olympus Gl endoscopes | | | S-PLO000600 | Sterile disposable irrigation tubing - Erbe pump/Olympus Gl endoscopes | | | S-PLPPLF200 | Sterile disposable irrigation tubing - Emed pump/Pentax-Fujifilm Gl endoscopes | | | S-PLPPLF500 | Sterile disposable irrigation tubing - Endogator pump/Pentax-Fujifilm Gl<br>endoscopes | | | S-PLPPLF700 | Sterile disposable irrigation tubing - Erbe pump/Pentax-Fujifilm Gl endoscopes | | ENDOtube for Wash Pump | S-100100 | Sterile single use one-way water jet connector for Olympus | | ENDOtube for Wash Pump | S-100300 | Sterile single use one-way water jet connector for Fujifilm | | ENDOtube for Wash Bottle<br>CO2 Insufflation | S-FLOCO2LUER | Sterile ENDOtube for Wash Bottle with CO2 tubing - for Olympus/Luer Insufflator<br>Connection | | | S-FLOCO2FUJI | Sterile ENDOtube for Wash Bottle with CO2 tubing - for Olympus/Adapter<br>Insufflator Connection | | | S-FLPCO2LUER | Sterile ENDOtube for Wash Bottle with CO2 tubing - for Pentax/Luer Insufflator<br>Connection | | | S-FLPCO2FUJI | Sterile ENDOtube for Wash Bottle with CO2 tubing - for Pentax/Adapter<br>Insufflator Connection | | | S-FLFCO2LUER | Sterile ENDOtube for Wash Bottle with CO2 tubing - for Fujifilm 5-600/Luer<br>Insufflator Connection | | | S-FLFCO2FUJI | Sterile ENDOtube for Wash Bottle with CO2 tubing - for Fujifilm 5-600/Adapter<br>Insufflator Connection | | | S-FLF7CO2LUER | Sterile ENDOtube for Wash Bottle with CO2 tubing - for Fujifilm 700/Luer<br>Insufflator Connection | | | S-FLF7CO2FUJI | Sterile ENDOtube for Wash Bottle with CO2 tubing - for Fujifilm 700/Adapter<br>Insufflator Connection | #### PREDICATE DEVICE lll. In order to properly cover the entire product range of the application, two predicates have been selected. The primary predicate is ENDOGATOR™ (K092429). | Device Name | ENDOGATOR™ | ENDO SMART CAP™ | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | K092429 | K093665 | | Class | II | II | | Indications For Use | The ENDOGATOR™ system (tubing<br>and accessories to accommodate<br>various endoscopes and irrigation<br>pumps) is intended to provide<br>irrigation via sterile water supply<br>during GI endoscopic procedures<br>when used in conjunction with an<br>irrigation pump (or cautery unit). | The ENDO SMARTCAP™ is<br>intended to be used with an<br>air or CO2 source and/or a<br>pump along with a sterile<br>water source to supply air or<br>CO2 and sterile water to an<br>endoscope during endoscopic<br>procedures. | {6}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Nissha Medical Technologies. The word "NISSHA" is written in large, bold, blue letters. Below the word "NISSHA", the words "MEDICAL TECHNOLOGIES" are written in smaller, gray letters. The logo is simple and modern. #### IV. Device Description #### IV.1. General The ENDOtube kits are a combination of tubes, fittings and valves intended to provide sterile water and/or CO2 and/or air to an endoscope during a GI endoscopic procedure. Visual description of ENDOtube products in the endoscopic column can be found in the Figure 1. Image /page/5/Figure/7 description: The image shows a medical device with different tubes and components. There are three tubes labeled "Endo tube for insufflation", "Endo tube for wash bottle", and "Endo tube for wash pump". The tubes are connected to the medical device, which includes a monitor, a control panel, and a pump. Figure 1: Visual description of the ENDOtube The ENDOtube products are compatible with various endoscopes (FujiFilm® 500/600 series and 700 series, OLYMPUS® 140, 160, 180 and 190 and PENTAX®), and irrigation pumps (OLYMPUS®, Erbe® and EMED®). #### IV.2. ENDOtube for Wash Bottle The ENDOtube for Wash Bottle kits are a combination of tubes, fittings and valves intended to provide sterile water, air or CO2 to an endoscope during a GI endoscopic procedure. The product is composed of at least two tubes bounded to a hermetic universal cap. The cap is screwed on a sterile water bottle. One tube is connected to the air output of the endoscopic column, and drive air into the bottle to put the sterile water bottle under is therefore forced into a second tube that is connected to the air/water channel of the endoscope. The device can also provide directly air or CO2 to the endoscope through another channel. #### IV.3. ENDOtube for Wash Pump The ENDOtube for Wash Pump kits are a combination of tubes, fittings and valves intended to provide sterile water to an endoscope through a peristaltic pump during a GI endoscopic procedure. The product is composed of one or more tubes bounded one after another. On one extremity, the tube is inserted into sterile water bottle. On the other extremity the tube is connected to the water-jet channel of the endoscope. The central part is inserted into the pumping part of a peristaltic pump. {7}------------------------------------------------ The ENDOtube for Wash Pump family contains also two independent single use one-way valves, used as an interface to connect the ENDOtube for Wash Pump tube to the endoscope waterjet channel. #### ENDOtube for Insufflation IV.4. The ENDOtube for Insufflation kits are a combination of tubes, fittings and filters intended to provide CO2 or air to an endoscope during a GI endoscopic procedure. Form on side, the tube is connected to an independent CO2 or air supply. On the other side, it's connected to an ENDOtube for Wash Bottle. The product drive air or CO2 to put the sterile water bottle under pressure, to power it through the ENDOtube for Wash Bottle to the endoscope. The ENDOtube for Insufflation family contains also three independent insufflation connectors used to adapt the ENDOtube for Insufflation tube to common independent CO2 or air supply devices. #### IV.5. ENDOtube for Wash Bottle CO2 Insufflation The ENDOtube for Wash Bottle CO2 Insufflation kits are a combination of tubes, fittings, valves and filters intended to provide sterile water and/or CO2 and/or air to an endoscope during a GI endoscopic procedure. It's the combination of an ENDOtube for Wash Pump and an ENDOtube for Insufflation, bounded through a valve. #### Indications for Use V. The ENDOtube devices are a combination of tubes, fittings and valves intended to provide sterile water and/or CO2 and/or air to an endoscope during gastrointestinal endoscopic procedure. These devices are intended to be used by healthcare professional only. In procedure room, all ENDOtube devices are 24 hours use disposable products except the one-way valves which are single use products and the insufflation connectors which are multiple patient use. {8}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Nissha Medical Technologies. The word "NISSHA" is in large, blue, sans-serif font. Below that, in a smaller, gray, sans-serif font, are the words "MEDICAL TECHNOLOGIES." #### VI. Technological characteristics comparison | Trade Name | ENDOtube (subject<br>device) | ENDOGATOR™<br>(primary predicate) | ENDO SMARTCAP™ | |-------------------------------------------------------------|--------------------------------------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------| | 510(k) number | K203350 | K092429 | K093665 | | Intended Use | Endoscopic Accessories | Endoscopic Accessories | Endoscopic Accessories | | Endoscope column<br>compatibility | OLYMPUS® GI<br>endoscopes 140, 160,<br>180 and 190 | OLYMPUS® GI<br>endoscopes 140, 160,<br>180 and 190 | OLYMPUS® GI<br>endoscopes 140, 160<br>and 180 | | | PENTAX® GI endoscopes | PENTAX® GI endoscopes | PENTAX® GI endoscopes | | | FUJIFILM® GI 500/600<br>series and 700 series<br>endoscopes | N/A | FUJIFILM® GI 500/600<br>series and 700 series<br>endoscopes | | | OLYMPUS®, Erbe®,<br>EMED® | OLYMPUS®, Erbe®,<br>EMED® | OLYMPUS®, Erbe®,<br>EMED® | | Irrigation peristatic<br>pump compatibility | OLYMPUS®, Erbe®,<br>EMED® | OLYMPUS®, Erbe®,<br>EMED® | OLYMPUS®, Erbe®,<br>EMED® | | Insufflator compatibility | OLYMPUS® UCR,<br>FUJIFILM® GW-1 and<br>GW-100 | Not compatible with<br>OLYMPUS® or FUJIFILM®<br>GW-1 and GW-100 | N/A | | Universal water bottle<br>cap | Common sterile water<br>bottle brands | Common sterile water<br>bottle brands | Common sterile water<br>bottle brands | | Fluid supply | Sterile water; Air; CO2 | Sterile water | Sterile water; Air; CO2 | | One-way valve (for<br>irrigation pump<br>compatible device) | Present | Present | Present | | Tube use period | Multi-patient use,<br>without reprocessing<br>Up to 24 hours | Multi-patient use,<br>without reprocessing<br>Up to 24 hours | Multi-patient use,<br>without reprocessing<br>Up to 24 hours | | One-way valve use<br>period | Single use<br>Up to 2 hours | Single use<br>Not communicated but<br>expected to be up to 2<br>hours | Single use<br>Not communicated but<br>expected to be up to 2<br>hours | | Insufflation connector<br>use period | Multi-patient use | N/A | N/A | | Insufflation connector<br>process | Reprocessable | N/A | N/A | | Device shelf-life | 2 years | 2 years | 2 years | | Tube and valve<br>sterilization | EO sterilization | EO sterilization | EO sterilization | | Insufflation connector<br>sterilization | Non-sterile | N/A | N/A | | Main materials | Biocompatible | Biocompatible | Biocompatible | | | Inert metallic<br>materials, | Inert metallic<br>materials, | Inert metallic materials, | | | elastomers and | elastomers and | elastomers and | | | thermoplastic | thermoplastic | thermoplastic polymer | | | polymer | polymer | | | | Not made with DEHP | Contain DEHP | Contain DEHP | | Device tissue contact | Intestinal mucosa<br>indirect contact for the<br>patient | Intestinal mucosa<br>indirect contact for the<br>patient | Intestinal mucosa<br>indirect contact for the<br>patient | | | Skin direct contact for<br>the end-user | Skin direct contact for<br>the end-user | Skin direct contact for<br>the end-user | | | PVC/Silicone | PVC/Silicone | PVC/Silicone | | Tube materials | PVC/Silicone | PVC/Silicone | PVC/Silicone | | Connector and valve<br>materials | SAN/ABS | SAN/ABS | SAN/ABS | | Trade Name | ENDOtube (subject<br>device) | ENDOGATOR™<br>(primary predicate) | ENDO SMARTCAP™ | | 510(k) number | K203350 | K092429 | K093665 | | Bottle cap material | PC | PE/SAN | PE/SAN | | Bottle cap seal material | TPU | Silicone | Silicone | | Testing | Fluid and Pressure<br>Testing, flow rated data | Not provided in 510(k)<br>summary | Not provided in 510(k)<br>summary | {9}------------------------------------------------ Image /page/8/Picture/2 description: The image shows the logo for Nissha Medical Technologies. The word "NISSHA" is written in large, bold, blue letters. Below the word "NISSHA" is the phrase "MEDICAL TECHNOLOGIES" written in smaller, gray letters. The logo is simple and modern. #### Substantial Equivalence Discussion VII. Based on the above comparison of intended use, endoscope compatibilities, testing methods and construction materials employed, the subject device ENDOtube (multiple models) are substantially equivalent to the two named predicates. The main asset of a device that is intended to provide sterile water, air or CO2 to an endoscope are the safety of use and the compatibility of the product with the environment of a medical theater. The comparison between ENDOtube devices and the predicates are mainly based on these attributes. - Compatibility with the main endoscope, pump and sterile water bottle available on the market - Ability to provide sterile water, air or CO2 - Sterilized, cross-contamination proof devices - Main materials used #### VIII. Non-clinical Testing Summary | Description | Parameter | Test description | Conclusion | |-------------------------------------------------------------|--------------------------------------------------------------|----------------------------------|------------| | Endoscope<br>column<br>compatibility | OLYMPUS® GI endoscopes<br>140, 160, 180 and 190 | Connection and usability testing | Pass | | | PENTAX® GI endoscopes | Connection and usability testing | Pass | | | FUJIFILM® GI 500/600<br>series and 700 series<br>endoscopes | Connection and usability testing | Pass | | Irrigation peristaltic<br>pump compatibility | Olympus®, Erbe®, EMED® | Connection and usability testing | Pass | | Insufflator<br>compatibility | OLYMPUS® UCR,<br>FUJIFILM® GW-1 and GW-<br>100 | Connection and usability testing | Pass | | Universal water<br>bottle cap | Presence and effectiveness | Connection and usability testing | Pass | | Fluid supply | Sterile water; Air; CO2 | Leak test and usability testing | Pass | | One-way valve (for<br>irrigation pump<br>compatible device) | Presence and effectiveness | Cross-contamination test | Pass | | Tube use period | Multi-patient use, without<br>reprocessing<br>Up to 24 hours | Durability test | Pass | | One-way valve use | Single use | Operational test | Pass | - CONFIDENTIAL AND PROPRIETARY -ENDOtube 510(k) Traditional Submission {10}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for Nissha Medical Technologies. The word "NISSHA" is written in large, bold, blue letters. Below the word "NISSHA" is the phrase "MEDICAL TECHNOLOGIES" written in smaller, light gray letters. The logo is simple and modern. | Description | Parameter | Test description | Conclusion | |--------------------------------------|----------------------------------------------------------------------|-------------------------------------------|------------| | period | Up to 2 hours | | | | Device shelf-life | 2 years | Performance test | Pass | | Insufflation connector<br>use period | Multi-patient use | Operational test | Pass | | Insufflation connector<br>processing | Reprocessable | Performance test | Pass | | Tube and valve<br>sterilization | EO sterilization | Sterilization validation | Pass | | Main materials | Biocompatible | Biocompatibility testing (AAMI/ISO 10993) | Pass | | | Inert metallic materials,<br>elastomers and<br>thermoplastic polymer | Biocompatibility testing (AAMI/ISO 10993) | Pass | | | No DEHP | Chemical characterization | Pass | #### Clinical Testing IX. Not required for a finding of substantial equivalence. #### X. Conclusion Based on the above comparisons of intended use, endoscope, testing methods and construction materials employed, the subject device ENDOtube (multiple models) are substantially equivalent to the two named predicates. The main asset of a device that is intended to provide sterile water, air or CO2 to an endoscope are the safety of use and the compatibility of the product with the environment of a medical theater. The comparison between ENDOtube devices and the predicates are mainly based on these following attributes: - Compatibility with the main endoscope, pump and sterile water bottle available on the market; - Ability to provide sterile water, air or CO2; - Sterilized, cross-contamination proof devices; - Main materials used. Therefore, the ENDOtube is substantially equivalent to the named predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...