← Product Code [ODP](/productcode/ODP) · K203342

# CONSTRUX Mini Ti Spacer System (K203342)

_Orthofix, Inc. · ODP · Mar 18, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K203342

## Device Facts

- **Applicant:** Orthofix, Inc.
- **Product Code:** [ODP](/productcode/ODP.md)
- **Decision Date:** Mar 18, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The CONSTRUX Mini Ti Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. The CONSTRUX Mini Ti Spacer System is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation system; the hyperlordotic implants (≥10°) are required to be used with an anterior cervical plate. Patients must have undergone a regimen of at least six weeks of non-operative treatment prior to being treated with the CONSTRUX Mini Ti Spacer System in the cervical spine.

## Device Story

CONSTRUX Mini Ti Spacer System consists of 3D-printed porous titanium interbody implants; designed for cervical intervertebral disc space insertion via anterior approach. Implants stabilize adjacent vertebrae, maintain disc height, and prevent vertebral collapse; must be used with autograft/allograft and supplemental fixation (anterior cervical plate required for hyperlordotic implants ≥10°). Used by surgeons in clinical settings to facilitate spinal fusion. Benefits include structural support for bone growth in DDD patients.

## Clinical Evidence

Bench testing only. Evaluated per ASTM F2077 (static/dynamic axial compression, compression shear, torsion), ASTM F2267 (subsidence), and ASTM F1877 (particle characterization). Results demonstrate substantial equivalence to predicate devices.

## Technological Characteristics

3D-printed porous titanium implants with functional gradient porous structure. Mechanical testing per ASTM F2077, ASTM F2267, and ASTM F1877. Provided sterile. No software or energy source.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- CONSTRUX Mini PTC Spacer System ([K202666](/device/K202666.md))
- SureMAX Family of Cervical Spacers ([K193359](/device/K193359.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 18, 2021

Orthofix Inc. Natalia Volosen Regulatory Affairs Principal 3451 Plano Parkway Lewisville, Texas 75056

Re: K203342

Trade/Device Name: CONSTRUX Mini Ti Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: March 8, 2021 Received: March 9, 2021

Dear Ms. Volosen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent L. Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203342

Device Name CONSTRUX Mini Ti Spacer System

#### Indications for Use (Describe)

The CONSTRUX Mini Ti Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.

The CONSTRUX Mini Ti Spacer System is intended for use with autograft comprised of cancellous and/ or corticocancellous bone graft and supplemental fixation system; the hyperlordotic implants (≥10°) are required to be used with an anterior cervical plate.

Patients must have undergone a regimen of at least six weeks of non-operative treatment prior to being treated with the CONSTRUX Mini Ti Spacer System in the cervical spine.

| Type of Use (Select one or both, as applicable)                               |                                                                              |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> | <div> <span> </span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </div> |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Device Trade Name:                          | CONSTRUX Mini Ti Spacer System                                                                                                |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:                               | Orthofix Inc.<br>3451 Plano Parkway<br>Lewisville, TX 75056, USA<br>Phone: (214) 937-2145<br>Fax: (214) 937-3322              |
| Contact Person:                             | Ms. Natalia Volosen<br>Regulatory Affairs Principal<br>nataliavolosen@orthofix.com                                            |
| Date Prepared:                              | January 20, 2021                                                                                                              |
| Registration Number:                        | 2183449                                                                                                                       |
| Product Code:                               | ODP                                                                                                                           |
| Classifications:                            | Class II - 21 CFR §888.3080, Intervertebral body fusion<br>device                                                             |
| Primary Predicate:<br>Additional Predicate: | K202666 – CONSTRUX™ Mini PTC Spacer System,<br>SE 10/14/2020<br>K193359 – SureMAX™ Family of Cervical Spacers,<br>SE 1/3/2020 |

Reason for the 510(k) Submission: Addition of 3D printed interbody devices

# Device Description:

The CONSTRUX Mini Ti Spacer System is comprised of a variety of 3D printed implants that have porous titanium end plates and a functional gradient porous structure. CONSTRUX Mini Ti spacers are implanted in the cervical intervertebral disc space and are intended to facilitate vertebral fusion by stabilizing adjacent vertebrae, maintaining disc height and preventing the collapse of one vertebra onto another.

The CONSTRUX Mini Ti Spacer System is not intended to be used as a stand-alone device and must be used with supplemental fixation. The implants are used singly and are implanted using an anterior approach. The CONSTRUX Mini Ti implants are provided sterile.

# Indications for Use and Intended Use:

The CONSTRUX Mini Ti Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.

The CONSTRUX Mini Ti Spacer System is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental 

{4}------------------------------------------------

fixation system; the hyperlordotic implants (≥10°) are required to be used with an anterior cervical plate.

Patients must have undergone a regimen of at least six weeks of non-operative treatment prior to being treated with the CONSTRUX Mini Ti Spacer System in the cervical spine.

## Performance Testing Summary:

CONSTRUX Mini Ti spacer was evaluated via mechanical testing per ASTM F2077 (including static and dynamic axial compression, static and dynamic compression shear and static and dynamic torsion), ASTM F2267 (subsidence) and characterization of particles per ASTM F1877. The results demonstrated the performance of CONSTRUX Mini Ti spacer is substantially equivalent to the predicate devices.

## Comparison of Technology:

CONSTRUX Mini Ti Spacer System have the same intended use, indications for use, technological characteristics, materials, the same principles of operation and same design as the predicate device CONSTRUX Mini PTC Spacer System (K202666) and SureMAXTM Family of Cervical Spacers (K193359).

## Conclusion:

The subject device is substantially equivalent to the CONSTRUX Mini PTC Spacer System (K202666) primary predicate and SureMAX™ Family of Cervical Spacers (K193359) additional predicate with respect to indications, design, materials, function, and performance.

---

**Source:** [https://fda.innolitics.com/device/K203342](https://fda.innolitics.com/device/K203342)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
