K203321 · Urogen Pharma, Ltd. · QBL · Dec 9, 2020 · General Hospital
Device Facts
Record ID
K203321
Device Name
Cystoject Syringe Lever
Applicant
Urogen Pharma, Ltd.
Product Code
QBL · General Hospital
Decision Date
Dec 9, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The Cystoject Syringe Lever Device is intended for use in the administration of sterile materials under aseptic conditions, in a clinical urology setting, by a clinician and in accordance with the best judgment of a physician.
Device Story
Cystoject Syringe Lever is a reusable, non-sterile mechanical device used in clinical urology settings. It houses a standard 30 mL syringe; features a lever mechanism to push the syringe plunger via a rotating knob; includes a clutch for rapid lever positioning. Device does not contact patients or fluids. Operated by clinicians to facilitate controlled delivery of sterile materials. Benefits include improved mechanical control during administration. Device requires cleaning and sterilization or high-level disinfection between uses.
Clinical Evidence
Bench testing only. Validation testing included residual fluid volume post-dispensing and load/unload testing. Results demonstrated the device met all predetermined acceptance criteria.
Technological Characteristics
Mechanical, non-powered, reusable syringe lever. Features a housing for a 30 mL syringe, a manual rotating knob, and a clutch mechanism. Provided non-sterile; requires cleaning and sterilization or high-level disinfection. No electrical components, software, or patient-contacting materials.
Indications for Use
Indicated for use by clinicians in a urology setting for the administration of sterile materials using a 30 mL syringe.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 9, 2020
UroGen Pharma Ltd. % James G. Ottinger, R.Ph. Executive Vice President, Regulatory and Quality UroGen Pharma, Inc. 400 Alexander Park Drive Princeton, NJ 08540
Re: K203321
Trade/Device Name: Cystoject Syringe Lever Regulation Number: 21 CFR§ 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: QBL Dated: November 11, 2020 Received: November 12, 2020
Dear James G. Ottinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark J. Antonino, M.S. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203321
Device Name Cystoject Syringe Lever
Indications for Use (Describe)
The Cystoject Syringe Lever Device is intended for use in the administration of sterile materials under aseptic conditions, in a clinical urology setting, by a clinician and in accordance with the best judgment of a physician.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for UroGen Pharma. The logo consists of a stylized blue circle with a branching design inside, resembling a molecule or cell structure. To the right of the circle are the words "UroGen" in a bold, blue font, with the trademark symbol "®" next to it. Below "UroGen" is the word "Pharma" in a smaller, lighter blue font.
9 Ha'Ta'asiya St. POB 2397 Ra'anana 4365405, Israel Tel.: +972 - 9 - 770 - 7600 Fax:+972 - 77 - 417 – 1410 e-mail: info@urogen.com
# 510(K) SUMMARY [SPECIAL 510(K)] Cystoject Syringe Lever
## 510(k) Number: K203321
### SUBMITTER I.
# Applicant's Name and address:
UroGen Pharma Ltd. 9 Ha`Ta`asiya Street Ra'anana, Israel, 4365405 Tel: +972-9-770-7600 Fax: +972-77-417-1410
### Primary Contact Person:
James G. Ottinger, R.Ph. Executive Vice President , Regulatory and Quality UroGen Pharma, Inc. Telephone: +1-646-768-9780 Mobile: +1-610-745-1230 Fax: +1-646-918-1262 e-mail: jim.ottinger@urogen.com
Date Prepared: December 1, 2020
#### II. DEVICE
| Trade Name: | Cystoject Syringe Lever |
|-----------------------|----------------------------------|
| Common or Usual Name: | Syringe Lever |
| Classification Name: | |
| Regulation: | Piston Syringe (21 CFR 880.5860) |
| Class: | II |
| Product Code: | QBL (Piston Syringe Lever) |
| Panel: | General Hospital |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for UroGen Pharma. The logo consists of a blue abstract graphic on the left, resembling a cell or branching structure within a partial circle. To the right of the graphic is the text "UroGen" in a bold, blue sans-serif font. Below "UroGen" is the word "Pharma" in a smaller, lighter blue font.
K203321 Page 2 of 3
### III. PREDICATE DEVICES
Predicate device Uroject12 Syringe Lever, by UroGen Pharma, Ltd., Product code QBL, cleared under: K190987.
This predicate has not been subject to a design-related recall.
### 510(k) Summary
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Act of 1990 and 21 CFR 807.92.
#### IV. DEVICE DESCRIPTION
The Cystoject Syringe Lever is a reusable device that is intended for use in the delivery of sterile materials under aseptic conditions, in a clinical urology setting, by a clinician and in accordance with the best judgment of the physician. The Cystoject Syringe Lever body contains a housing for holding a standard syringe and a lever that pushes the syringe's plunger when the user rotates its knob. A clutch mechanism allows the user to quickly move the lever to the syringe plunger's position. The Cystoject Syringe Lever does not require an energy source, does not utilize electrical components, and does not utilize software or hardware. The Cystoject Syringe Lever does not come into direct or indirect contact with the patient nor with fluids intended for instillation.
The Cystoject Syringe Lever is provided non-sterile as a reusable, reprocessable device to be cleaned and sterilized or high-level disinfected before each use.
The Cystoject Syringe Lever indicated for use with a 30 mL syringe.
#### V. INDICATIONS FOR USE
The Cystoject Syringe Lever Device is intended for use in the administration of sterile materials under aseptic conditions, in a clinical urology setting, by a clinician and in accordance with the best judgment of a physician.
#### VI. SUBSTANTIAL EQUIVALENCE TO PREDICATE DEVICE
The Cystoject Syringe Lever is substantially equivalent to the predicate device based on the following:
### Intended Use
The intended use of the subject and predicate devices is identical.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for UroGen Pharma. The logo consists of a blue circular design on the left, resembling a stylized molecule or cell. To the right of the design is the text "UroGen" in a larger, bolder blue font, with the word "Pharma" underneath in a smaller font. The overall design is clean and professional, conveying a sense of innovation and scientific expertise.
Ha'Ta'asiva St. POB : a'anana 4365405 Israe 972 - 77 - 417 e-mail: info@urogen.com
## Technology
Both the Cystoject Syringe Lever and the Uroject12 Syringe Lever have the same principle and mode of operation. The Cystoject Syringe Lever has the same general design as the Uroject12 Syringe Lever, except for modifications to enable a different compatible syringe size (i.e., 30 mL with the Uroject12 Syringe Lever) and change in the exterior device color. These differences in technological characteristics do not raise different questions of safety and effectiveness.
#### VII. PERFORMANCE DATA
The summary of design control activities identified the following non-clinical performance tests to support the substantial equivalence determination.
### Validation Testing
The only testing included in this submission that pertains to the validation activities for the additional device are the Residual Fluid Volume post dispensing and Load/Unload Testing.
The results of the validation testing demonstrated that the Cystoject Syringe Lever is considered safe and effective for its intended use.
### Summary
The results of the validation testing demonstrate that the Cystoject Syringe Lever met all predetermined acceptance criteria.
#### VIII. CONCLUSION
The performance testing demonstrates that the Cystoject Syringe Lever is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.