Emprint SX Ablation Platform with Thermosphere Technology
K203303 · Covidien, LLC · NEY · Dec 17, 2020 · General, Plastic Surgery
Device Facts
Record ID
K203303
Device Name
Emprint SX Ablation Platform with Thermosphere Technology
Applicant
Covidien, LLC
Product Code
NEY · General, Plastic Surgery
Decision Date
Dec 17, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Emprint™ SX Ablation Platform with Thermosphere™ Technology is intended for use in percutaneous, laparoscopic, and intraoperative coagulation) of soft tissue, including partial or complete ablation of nonresectable liver tumors. The Emprint™ SX Ablation Platform is not intended for use in cardiac procedures. The system's optional 3D navigation feature assists in the placement of the Emprint™ SX Navigation Antenna with Thermosphere™ Technology using real-time image guidance during intraoperative and laparoscopic ablation procedures. The navigation feature enhances the output of a compatible medical ultrasound imaging system and displays an image of the antenna and its predicted trajectory on a computer monitor. The size and shape of the predicted zone relative to the position of the antenna are displayed on the enhanced ultrasound image.
Device Story
Microwave-based ablation system; delivers 2450 MHz, 100W energy to soft tissue via antenna for coagulation. Optional 3D navigation feature uses electromagnetic field sensors on ultrasound probes and antenna; tracks position relative to Table Top Field Generator (TTFG) placed under patient. System overlays antenna trajectory and predicted ablation zone onto real-time ultrasound images on auxiliary monitor. Operated by physicians in OR/clinic settings via touchscreen interface. Software controls power (5-100W) and time (up to 10 min); optional temperature probe enables automatic generator shut-off. Closed-loop saline cooling system maintains antenna shaft temperature. Software update (v1.1.1) allows clearing general system errors without power-cycling, reducing procedural delays. Benefits include precise antenna placement and controlled ablation of nonresectable liver tumors.
Clinical Evidence
Bench testing only. Comprehensive software verification testing performed to validate the error-handling update. No clinical data required.
Technological Characteristics
Microwave ablation system; 2450 MHz, 100W generator. Closed-loop room-temperature saline cooling. Electromagnetic tracking system (TTFG). Touchscreen GUI. Software-controlled power/time. Compatible with open/laparoscopic ultrasound probes. Software version 1.1.1.
Indications for Use
Indicated for percutaneous, laparoscopic, and intraoperative soft tissue coagulation, including partial or complete ablation of nonresectable liver tumors. Contraindicated for cardiac procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Emprint™ SX Ablation Platform with Thermosphere™ (K171358)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency responsible for regulating food and drug products in the United States.
December 17, 2020
Medtronic Inc. Liron Bar-Yaakov Senior Manager of Regulatory Affairs 5920 Longbow Dr. Boulder, Colorado 80301
Re: K203303
Trade/Device Name: Emprint SX Ablation Platform with Thermosphere Technology Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: NEY Dated: November 4, 2020 Received: November 9, 2020
Dear Liron Bar-Yaakov:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K203303
Device Name
The Emprint™ SX Ablation Platform with Thermosphere™ Technology
### Indications for Use (Describe)
The Emprint™ SX Ablation Platform with Thermosphere™ Technology is intended for use in percutaneous, laparoscopic, and intraoperative coagulation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.
The Emprint™ SX Ablation Platform is not intended for use in cardiac procedures.
The system's optional 3D navigation feature assists in the placement of the Emprint™ SX Navigation Antenna with Thermosphere™ Technology using real-time image guidance during intraoperative and laparoscopic ablation procedures. The navigation feature enhances the output of a compatible medical ultrasound imaging system and displays an image of the antenna and its predicted trajectory on a computer monitor. The size and shape of the predicted zone relative to the position of the antenna are displayed on the enhanced ultrasound image.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt;"> <svg height="10" width="10"> <rect height="10" style="fill:lightgrey;stroke-width:1;stroke:rgb(0,0,0)" width="10"></rect> <line style="stroke:rgb(0,0,0);stroke-width:1" x1="0" x2="10" y1="0" y2="10"></line> <line style="stroke:rgb(0,0,0);stroke-width:1" x1="0" x2="10" y1="10" y2="0"></line> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt;"> <svg height="10" width="10"> <rect height="10" style="fill:white;stroke-width:1;stroke:rgb(0,0,0)" width="10"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K203303
Covidien LLC Special 510(k) Submission Emprint™ SX Ablation Platform with Thermosphere™ Technology Emprint™ SX 1.1.1 Software Update
## 510(k) Summary
#### l. Submitter
Medtronic Inc 5920 Longbow Dr. Boulder, CO 80301
Contact: Liron Bar-Yaakov Senior Manager of Regulatory Affairs 5920 Longbow Dr Boulder, CO 80301 Telephone: (508) 261-8000 Email: Liron.Baryaakov@medtronic.com
Date of Summary Preparation: November 05, 2020
#### ll. Devices
### Table 5.1 Devices:
| Device Trade Name | Emprint™ SX Ablation Platform with Thermosphere™ Technology |
|---------------------|-------------------------------------------------------------|
| Device Common Name | Microwave Ablation System |
| Classification Name | System, Ablation, Microwave and Accessories |
| Regulatory Class | II |
| Product Code | NEY |
#### III. Predicate Devices
This Special 510(k) submission describes enhancements to the software of the previously cleared Emprint™ SX Ablation Platform with Thermosphere™ Technology.
### Table 5.2 Predicate Devices
| Device Name | Classification Name and Class | 510(k) Clearance No |
|-----------------------------------------------------|----------------------------------------------------------|---------------------|
| Emprint™ SX Ablation Platform with<br>Thermosphere™ | System, Ablation, Microwave<br>and Accessories; Class II | K171358 |
#### IV. Device Description
The Emprint™ SX Ablation Platform with Thermosphere™ Technology is microwave-based system intended to deliver energy through an antenna inserted into soft tissue for the purpose of coagulating (ablating) a defined tissue volume. The Emprint™ SX Ablation Platform utilizes a 2450 MHz 100W generator to deliver power to a single microwave ablation antenna. The platform is software-controlled and a touchscreen Graphical User Interface is used to select the desired ablation time (up to 10 minutes) and power (5 to 100W) settings. Using an optional temperature probe, the ablation platform can be set to monitor the temperature of a desired target and to automatically turn the generator off, when the target reaches a pre-set temperature. The ablation platform uses circulating, room-temperature, normal saline to cool the non-radiating portion of the antenna shaft and to provide a consistent ablation zone. The normal saline is pumped from an IV bag through the antenna shaft and back to the IV bag in a closed system.
{4}------------------------------------------------
Covidien LLC Special 510(k) Submission
Emprint™ SX Ablation Platform with Thermosphere™ Technology
Emprint™ SX 1.1.1 Software Update
Using an optional/selectable navigation feature, the Emprint™ SX Ablation Platform assists physicians in the accurate placement of the trackable Emprint™ SX Navigation Antenna into a target anatomical structure by overlaying the image of the antenna and its trajectory onto an ultrasound image in real time.
Sensors are attached to platform-compatible, open and laparoscopic ultrasound probes and to the platform's trackable antenna. These sensors detect fluctuations in an electromagnetic field emitted from the system's Table Top Field Generator (TTFG). The TTFG is placed underneath the patient and transmits positional information that is processed by the system's navigation hardware and interpreted by its software. The platform's software allows the surgeon to see a representation of the ablation zone on an auxiliary display and the system's operator to control the ablation system via the touchscreen interface.
| Table 5.3 EmprintTM SX Ablation Platform components: | |
|-------------------------------------------------------------------------------|-------------|
| Component Name | Catalog No. |
| EmprintTM SX Ablation Platform with ThermosphereTM Technology | CASYS100 |
| EmprintTM SX Ultrasound Surgical Navigation (software) Application | CAUS1SN |
| EmprintTM Ablation Reusable Cable | CA190RC1 |
| EmprintTM SX Table Top Field Generator | CAFG1 |
| EmprintTM SX Table Top Field Generator Cart | CAFGCART1 |
| EmprintTM SX Field Generator Spacers (Bed Pads) | CAFGSP1 |
| EmprintTM SX Short Navigation Antenna with ThermosphereTM Technology, 15cm | CA15L2N |
| EmprintTM SX Standard Navigation Antenna with ThermosphereTM Technology, 20cm | CA20L2N |
| EmprintTM SX Long Navigation Antenna with ThermosphereTM Technology, 30cm | CA30L3N |
| EmprintTM SX Open Ultrasound Tracking Sensor | CABK8816 |
| EmprintTM SX Open Ultrasound Tracking Sensor | CAAL9132 |
| EmprintTM SX Laparoscopic Ultrasound Tracking Sensor | CAAL9150L |
| EmprintTM SX Laparoscopic Ultrasound Tracking Sensor | CABK8666L |
| Remote Temperature Probe | RTP20 |
| Ablation Footswitch | RFASW |
### Table 5.3 Emprint™ SX Ablation Platform components:
#### V. Description of Changes
Modifications have been made only to the software of the Emprint™ SX Ablation Platform with Thermosphere™ Technology. Enhancements were made to the Software to allow the user to clear a general system error without power-cycling the system in order to deliver ablation therapy, mitigating any risk of delay in procedure.
These software changes, do not affect the Emprint™ SX Ablation Platform with Thermosphere™ Technology indication for use, essential function, performance, or underlying principles of operation.
{5}------------------------------------------------
Covidien LLC Special 510(k) Submission Emprint™ SX Ablation Platform with Thermosphere™ Technology Emprint™ SX 1.1.1 Software Update
#### VI. Indications for use
The Emprint™ SX Ablation Platform with Thermosphere™ Technology is intended for use in percutaneous, laparoscopic, and intraoperative coagulation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.
The Emprint™ SX Ablation Platform is not intended for use in cardiac procedures.
The system's optional, 3-D navigation feature assists in the placement of the Emprint™ SX Navigation Antenna with Thermosphere™ Technology using real-time image guidance during intraoperative and laparoscopic ablation procedures. The navigation feature enhances the output of a compatible medical ultrasound imaging system and displays an image of the antenna and its predicted trajectory on a computer monitor. The size and shape of the predicted ablation zone relative to the position of the antenna is displayed on the enhanced ultrasound image.
#### VII. Comparison of technological characteristics with the predicate device
Emprint™ SX Ablation Platform with Thermosphere™ Technology current software (predicate device) and the updated software update share the same underlying technology. Both devices are used to deliver microwave energy to soft tissue in a highly controlled way. Signal frequency, power and exposure time are controlled by the system's 2450 MHz generator. The design of the radiating portion of the accessory and continuous cooling of the accessory during the ablation procedure determines the shape (near spherical) and size of the resulting ablations.
The enhancement to CAUS1SN software allows the user to avoid power cycling the system removing a delay to the procedure due to the receipt of a general system error. This is completed through a feature update to the Emprint API code, responsible for processing messages from the generator. There are no effects on any other portion of the software architecture, or on the systems underlying technological characteristics.
#### VIII. Performance Data
Comprehensive SW Verification testing was performed by Medtronic in support of the proposed change. No additional pre-clinical or clinical data is required to support the substantial equivalence determination for this software update.
#### IX. Conclusions
The Modified Emprint™ SX Ablation Platform with Thermosphere™ Technology software (CAUS1SN) update is nearly identical to the predicate Emprint™ SX Ablation Platform with Thermosphere™ Technology software. The subject and predicate devices share intended use, technological characteristics, and meet identical performance requirements. The results of comprehensive verification testing raised no new questions of safety or effectiveness. The current Emprint™ SX Ablation Platform with Thermosphere™ Technology software version, and the Emprint™ SX Ablation Platform with Thermosphere™ Technology new software version are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.