3M Attest Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator
K203284 · 3M Company · QKM · Dec 7, 2020 · General Hospital
Device Facts
Record ID
K203284
Device Name
3M Attest Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator
Applicant
3M Company
Product Code
QKM · General Hospital
Decision Date
Dec 7, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
Use the 3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E for pack control monitoring of the following hydrogen peroxide sterilization sterilizers and cycles: STERRAD® 100S System, STERRAD® NX® System (Standard and Advanced cycles), STERRAD® NX® System with AllClear™ Technology (Standard and Advanced cycles), STERRAD® 100NX® System (Standard, Flex, Express and Duo cycles) STERRAD® 100NX® System with AllClear™ Technology (Standard, Flex, Express and Duo cycles) vaporized hydrogen peroxide sterilizers and STERIS® V-PRO® 1 (Lumen cycle), STERIS ® V-PRO® 1 Plus (Lumen and Non-Lumen cycles) and STERIS ® V-PRO® maX Low Temperature Sterilization System (Lumen and Flexible cycles), STERIS® V-PRO® 60 Temperature Sterilization System (Lumen and Flexible cycles), STERIS® V-PRO® maX 2 Temperature Sterilization System (Lumen, Flexible, and Fast Non-Lumen cycles), and the STERIS® V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles).
Device Story
Chemical indicator strip for vaporized hydrogen peroxide sterilization monitoring; non-cellulose coated strip in film laminate. Input: exposure to vaporized hydrogen peroxide (time, temperature, concentration). Operation: color changes from blue to pink along strip; progression visible through window with 'REJECT' and 'ACCEPT' zones. Used in healthcare facilities for pack control monitoring within wrapped trays, rigid containers, or pouches. Output: visual colorimetric indicator of sterilization cycle completion. Healthcare providers use output to verify sterilization parameters were achieved; ensures sterility assurance of processed items; benefits patient by confirming effective sterilization of surgical instruments.
Clinical Evidence
Bench testing only. Simulated use testing performed in the STERIS V-PRO s2 sterilizer across Lumen, Non-Lumen, Flexible, and Fast cycles. Results confirmed the indicator correctly transitions to 'ACCEPT' in complete cycles and remains in 'REJECT' for incomplete cycles.
Technological Characteristics
Non-cellulose based coated indicator strip in film laminate. Colorimetric sensing principle (blue to pink). Shelf life: 15 months. Endpoint specs: 5.1 mg/L concentration, 1 minute exposure, 50°C temperature.
Indications for Use
Indicated for pack control monitoring of hydrogen peroxide sterilization cycles in specified STERRAD and STERIS V-PRO systems. No specific patient population or contraindications defined.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E (K193110)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 7, 2020
3M Company Andrew Wingen Regulatory Affairs Commercialization Strategy Lead 2510 Conway Ave, Bldg 275-5W-06 St. Paul, Minnesota 55144-1000
Re: K203284
Trade/Device Name: 3M Attest Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: QKM Dated: November 6, 2020 Received: November 9, 2020
Dear Andrew Wingen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203284
#### Device Name
3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator (1348/1348E)
#### Indications for Use (Describe)
Use the 3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E for pack control monitoring of the following hydrogen peroxide sterilization sterilizers and cycles: STERRAD® 100S System, STERRAD® NX® System (Standard and Advanced cycles), STERRAD® NX® System with AllClear™ Technology (Standard and Advanced cycles), STERRAD® 100NX® System (Standard, Flex, Express and Duo cycles) STERRAD® 100NX® System with AllClear™ Technology (Standard, Flex, Express and Duo cycles) vaporized hydrogen peroxide sterilizers and STERIS® V-PRO® 1 (Lumen cycle), STERIS ® V-PRO® 1 Plus (Lumen and Non-Lumen cycles) and STERIS ® V-PRO® maX Low Temperature Sterilization System (Lumen and Flexible cycles), STERIS® V-PRO® 60 Temperature Sterilization System (Lumen and Flexible cycles), STERIS® V-PRO® maX 2 Temperature Sterilization System (Lumen, Flexible, and Fast Non-Lumen cycles), and the STERIS® V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles).
Type of Use (Select one or both, as applicable)| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the 3M logo. The logo is red and consists of the number "3" and the letter "M". The number and letter are in a bold, sans-serif font. The logo is simple and recognizable.
# Special 510(k) Summary for 3MTM Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E K203284
Sponsor Information:
3M Company 3M Health Care 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000
Contact: Andrew Wingen Regulatory Affairs Commercialization Strategy Lead Phone Number: (651) 733-0929 Fax Number: (651) 737-5320
Date of Summary: 4 December 2020
{4}------------------------------------------------
#### 1. Device Name and Classification:
| Common or Usual Name: | Chemical Indicator |
|------------------------|---------------------------------------------------------------------------------------|
| Proprietary Name: | 3MTM AttestTM Vaporized Hydrogen Peroxide Tri-Metric<br>Chemical Indicator 1348/1348E |
| Classification Name: | Indicator, physical/chemical sterilization process |
| Device Classification: | Class II, 21 CFR § 880.2800 |
| Product Code: | QKM |
#### 2. Predicate Device:
K193110, 3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E
## 3. Description of Device:
The 3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E is a chemical indicator consisting of a non-cellulose based coated indicator strip sensitive to vaporized hydrogen peroxide, contained in a film laminate.
The 3M™ Attest™ CI 1348/1348E verifies that the stated values for the three critical parameters of exposure time, temperature, and amount of vaporized hydrogen peroxide have been achieved within a package or containment device (i.e. wrapped trays, rigid containers, sterilization pouches, and other types of packs) and/or at a specific location within the load or empty chamber.
Upon exposure to vaporized hydrogen peroxide, the color of the coated indicator strip progressively changes from blue toward pink along the strip. The progression of the blue to pink color change along the strip is visible through a window with marked "REJECT" and "ACCEPT" zones. The extent of the progression depends on exposure time, temperature, and amount of vaporized hydrogen peroxide.
## 4. Indications for Use
Use the 3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E for pack control monitoring of the following hydrogen peroxide sterilization sterilizers and cycles: STERRAD® 100S System, STERRAD® NX® System (Standard and Advanced cycles), STERRAD® NX® System with AllClear™ Technology (Standard and Advanced cycles), STERRAD® 100NX® System (Standard, Flex, Express and Duo cycles) STERRAD® 100NX® System with AllClear™ Technology
{5}------------------------------------------------
#### PREMARKET NOTIFICATION [Special 510(k)] 3MTM Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E
(Standard, Flex, Express and Duo cycles) vaporized hydrogen peroxide sterilizers and STERIS® V-PRO® 1 (Lumen cycle), STERIS ® V-PRO® 1 Plus (Lumen and Non-Lumen cycles) and STERIS ® V-PRO® maX Low Temperature Sterilization System (Lumen, Non-Lumen and Flexible cycles), STERIS® V-PRO® 60 Temperature Sterilization System (Lumen, Non-Lumen and Flexible cycles), STERIS® V-PRO® maX 2 Temperature Sterilization System (Lumen, Non-Lumen, Flexible, and Fast Non-Lumen cycles), and the STERIS® V-PRO® s2 Low Temperature Sterilization System (Lumen, Non-Lumen, Flexible, and Fast cycles).
| Feature | Submission Device: 3MTM AttestTM<br>Vaporized Hydrogen Peroxide Tri-<br>Metric Chemical Indicator | Predicate Device:<br>3MTM Attest™ Vaporized Hydrogen<br>Peroxide Tri-Metric Chemical<br>Indicator (K193110) | Comparison |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Models | 1348, 1348E | 1348, 1348E | Identical |
| Device<br>Design | The 3MTM Attest™ Vaporized<br>Hydrogen Peroxide Tri-Metric<br>Chemical Indicator 1348/1348E is a<br>chemical indicator consisting of a non-<br>cellulose based coated indicator strip<br>sensitive to vaporized hydrogen<br>peroxide, contained in a film laminate.<br>The 3M™ Attest™ CI 1348/1348E<br>verifies that the stated values for the<br>three critical parameters of exposure<br>time, temperature, and concentration of<br>vaporized hydrogen peroxide have been<br>achieved within a package or<br>containment device (i.e. wrapped trays,<br>rigid containers, sterilization pouches,<br>and other types of packs) and/or at a<br>specific location within the load or<br>empty chamber.<br>Upon exposure to vaporized hydrogen<br>peroxide, the color of the coated<br>indicator strip progressively changes<br>from blue toward pink along the strip.<br>The progression of the blue to pink<br>color change along the strip is visible<br>through a window with marked<br>"REJECT" and "ACCEPT" zones. The<br>extent of the progression depends on<br>exposure time, temperature, and<br>concentration of vaporized hydrogen<br>peroxide. | The 3MTM Attest™ Vaporized<br>Hydrogen Peroxide Tri-Metric<br>Chemical Indicator 1348/1348E is a<br>chemical indicator consisting of a non-<br>cellulose based coated indicator strip<br>sensitive to vaporized hydrogen<br>peroxide, contained in a film laminate.<br>The 3MTM Attest™ CI 1348/1348E<br>verifies that the stated values for the<br>three critical parameters of exposure<br>time, temperature, and concentration of<br>vaporized hydrogen peroxide have been<br>achieved within a package or<br>containment device (i.e. wrapped trays,<br>rigid containers, sterilization pouches,<br>and other types of packs) and/or at a<br>specific location within the load or<br>empty chamber.<br>Upon exposure to vaporized hydrogen<br>peroxide, the color of the coated<br>indicator strip progressively changes<br>from blue toward pink along the strip.<br>The progression of the blue to pink<br>color change along the strip is visible<br>through a window with marked<br>"REJECT" and "ACCEPT" zones. The<br>extent of the progression depends on<br>exposure time, temperature, and<br>concentration of vaporized hydrogen<br>peroxide. | Identical |
| Sterilizers in<br>which Color | STERRAD® 100S<br>STERRAD® NX (Standard and | STERRAD® 100S<br>STERRAD® NX (Standard and | Similar. The<br>submission |
| Change | Advanced cycles) | Advanced cycles) | device adds |
| | | | |
| Performance was<br>Demonstrated<br>(Blue toward Pink) | STERRAD® NX with ALLClear™<br>Technology (Standard and Advanced<br>cycles)<br>STERRAD® 100NX (Standard, Flex,<br>Express, and Duo cycles)<br>STERRAD® 100NX with ALLClear™<br>Technology (Standard, Flex, Express,<br>and Duo cycles)<br>STERIS® V-PRO® maX (Lumen,<br>Non-Lumen, and Flexible cycles)<br>STERIS® V-PRO® maX2 (Lumen,<br>Non-Lumen, Flexible, and Fast Non-<br>Lumen cycles)<br>STERIS® V-PRO® 60 (Lumen, Non-<br>Lumen, and Flexible cycles)<br>STERIS® V-PRO® s2 (Lumen, Non-<br>Lumen, Flexible, and Fast cycles)<br>Use the 3MTM Attest™ Vaporized<br>Hydrogen Peroxide Tri-Metric<br>Chemical Indicator 1348/1348E for<br>pack control monitoring of the<br>following hydrogen peroxide<br>sterilization sterilizers and cycles:<br>STERRAD® 100S System,<br>STERRAD® NX® System (Standard<br>and Advanced cycles), STERRAD®<br>NX® System with AllClear™<br>Technology (Standard and Advanced<br>cycles), STERRAD® 100NX® System<br>(Standard, Flex, Express and Duo<br>cycles) STERRAD® 100NX® System<br>with AllClear™ Technology (Standard,<br>Flex, Express and Duo cycles) | STERRAD® NX with ALLClear™<br>Technology (Standard and Advanced<br>cycles)<br>STERRAD® 100NX (Standard, Flex,<br>Express, and Duo cycles)<br>STERRAD® 100NX with ALLClear™<br>Technology (Standard, Flex, Express,<br>and Duo cycles)<br>STERIS® V-PRO® maX (Lumen,<br>Non-Lumen, and Flexible cycles)<br>STERIS® V-PRO® maX2 (Lumen,<br>Non-Lumen, Flexible, and Fast Non-<br>Lumen cycles)<br>STERIS® V-PRO® 60 (Lumen, Non-<br>Lumen, and Flexible cycles)<br>Use the 3MTM Attest™ Vaporized<br>Hydrogen Peroxide Tri-Metric<br>Chemical Indicator 1348/1348E for<br>pack control monitoring of the<br>following hydrogen peroxide<br>sterilization sterilizers and cycles:<br>STERRAD® 100S System,<br>STERRAD® NX® System (Standard<br>and Advanced cycles), STERRAD®<br>NX® System with AllClear™<br>Technology (Standard and Advanced<br>cycles), STERRAD® 100NX® System<br>(Standard, Flex, Express and Duo<br>cycles) STERRAD® 100NX® System<br>with AllClear™ Technology (Standard,<br>Flex, Express and Duo cycles) | the<br>STERIS®<br>V-PRO® s2<br>sterilizer to<br>the<br>indications<br>for use and<br>testing was<br>conducted to<br>demonstrate<br>color change<br>performance. |
| Indications<br>for Use | vaporized hydrogen peroxide sterilizers<br>and STERIS® V-PRO® 1 (Lumen<br>cycle), STERIS ® V-PRO® 1 Plus<br>(Lumen and Non-Lumen cycles) and<br>STERIS ® V-PRO® maX Low<br>Temperature Sterilization System<br>(Lumen, Non-Lumen and Flexible<br>cycles), STERIS® V-PRO® 60<br>Temperature Sterilization System<br>(Lumen, Non-Lumen and Flexible<br>cycles), STERIS® V-PRO® maX 2<br>Temperature Sterilization System<br>(Lumen, Non-Lumen, Flexible, and<br>Fast Non-lumen cycles), STERIS® V-<br>PRO® s2 Low Temperature<br>Sterilization System (Lumen, Non-<br>Lumen, Flexible, and Fast cycles). | vaporized hydrogen peroxide sterilizers<br>and STERIS® V-PRO® 1 (Lumen<br>cycle), STERIS ® V-PRO® 1 Plus<br>(Lumen and Non-Lumen cycles) and<br>STERIS ® V-PRO® maX Low<br>Temperature Sterilization System<br>(Lumen, Non-Lumen and Flexible<br>cycles), STERIS® V-PRO® 60<br>Temperature Sterilization System<br>(Lumen, Non-Lumen and Flexible<br>cycles), and STERIS® V-PRO® maX 2<br>Temperature Sterilization System<br>(Lumen, Non-Lumen, Flexible, and<br>Fast Non-lumen cycles). | Similar. The<br>submission<br>device adds<br>the<br>STERIS®<br>V-PRO® s2<br>sterilizer to<br>the<br>indications<br>for use. |
| Indicator<br>Agent | Proprietary | Proprietary | Identical |
| Stability of<br>the endpoint<br>reaction | At least one month (4 weeks) | At least one month (4 weeks) | Identical |
## 5. Summary of Technological Characteristics compared to Predicate Device
{6}------------------------------------------------
## PREMARKET NOTIFICATION [Special 510(k)] 3M™ Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E
{7}------------------------------------------------
#### PREMARKET NOTIFICATION [Special 510(k)] 3MTM Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E
| Shelf life | 15 months | | | 6 months | | | Similar. 15<br>month shelf<br>life data has<br>been<br>generated for<br>the<br>submission<br>device using<br>the same<br>protocol<br>used in the<br>predicate<br>submission. |
|-------------------------------------------------------------|------------------------|------------------|-----------------|------------------------|------------------|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Endpoint<br>Specifications<br>(Minimum<br>Stated<br>Values) | VH2O2<br>Concentration | Exposure<br>Time | Temperature | VH2O2<br>Concentration | Exposure<br>Time | Temperature | Identical |
| | 5.1 mg/L | 1<br>minute | 50 degrees<br>C | 5.1 mg/L | 1<br>minute | 50 degrees<br>C | |
## 6. Summary of Nonclinical Testing
Provided below are the non-clinical test methodologies performed to demonstrate the Tri-Metric CI met the acceptance criteria of the standard.
| Test Method | Purpose | Acceptance Criteria | Results |
|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Health Care<br>Facility<br>Simulated<br>Use Testing<br>on STERIS<br>V-PRO s2<br>sterilizer | Assess color<br>change of<br>indicator after<br>exposure to<br>representative<br>complete and<br>incomplete<br>cycles for the<br>STERIS V-<br>PRO s2<br>sterilizer | The 3M Tri-Metric CI turns from blue<br>toward pink in the ACCEPT region of<br>the indicator window to indicate a<br>"PASS" (reaches endpoint) when<br>exposed to a complete Lumen, Non-<br>Lumen, Flexible, or Fast Cycle in the<br>STERIS V-PRO s2 Low Temperature<br>Sterilization System.<br>The 3M Tri-Metric CI turns from blue<br>toward pink only within the REJECT<br>region of the indicator window to<br>indicate a "FAIL" (does not reach<br>endpoint) when exposed to an<br>incomplete Lumen, Non-Lumen,<br>Flexible, or Fast Cycle in the STERIS<br>V-PRO s2 Low Temperature<br>Sterilization System. | All Tri-<br>Metric CIs<br>tested met<br>acceptance<br>criteria.<br>All Tri-<br>Metric CIs<br>tested met<br>acceptance<br>criteria. |
## 7. Conclusion
Based on the non-clinical performance data, the submission device 3MTM Attest™ Vaporized Hydrogen Peroxide Tri-Metric Chemical Indicator 1348/1348E is as safe, as effective, and performs as well as or better than the predicate of the same name (K193110), Class II (21 CFR 880.2800), product code QKM.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.