← Product Code [NFJ](/productcode/NFJ) · K203244

# CellChek 20 rc (K203244)

_Konan Medical, Inc. · NFJ · Jun 15, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K203244

## Device Facts

- **Applicant:** Konan Medical, Inc.
- **Product Code:** [NFJ](/productcode/NFJ.md)
- **Decision Date:** Jun 15, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The CellChek 20 rc is a software program intended to analyze ophthalmic images captured by the Konan Specular Microscope XVII for examination of corneal endothelium.

## Device Story

CellChek 20 rc is a software program for analyzing corneal endothelial images captured by the Konan Specular Microscope XVII. It is installed on a general-use computer and used in research and learning centers. The device calculates cell density, coefficient of variation of cell area, and percent hexagonality. It provides both manual and automatic analysis methods. In manual mode, users identify cells and boundaries by placing dots or tracing; in automatic mode, the software detects cells and boundaries, allowing user modification of results. The output includes analysis results and images, which can be displayed, stored, printed, or exported. The device aids healthcare providers in examining corneal endothelium by automating cell distribution analysis.

## Clinical Evidence

No clinical data. Performance was established via non-clinical software testing in accordance with IEC 62304 standards, confirming the software functions are equivalent to the reference device (CellChek 20).

## Technological Characteristics

Software-only device for general-use computers. Implements image analysis algorithms (auto-trace, auto-center, flex-center methods) for corneal endothelium. Developed per IEC 62304. Connectivity includes image import/export capabilities. No physical materials or energy sources.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Image Storage System, Model KSS-400 ([K081797](/device/K081797.md))

## Reference Devices

- Konan Specular Microscopy XVII, CellChek 20 ([K191558](/device/K191558.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

June 15, 2021

Konan Medical, Inc. % Alan Donald President Matrix Medical Consulting, INC. 8880 Rio San Diego Drive Suite 800 San Diego, California 92108

Re: K203244

Trade/Device Name: CellChek 20 rc Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: NFJ, NQE Dated: April 20, 2021 Received: April 21, 2021

Dear Alan Donald:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203244

Device Name CellChek 20 rc Software

Indications for Use (Describe)

The CellChek 20 rc is a software program intended to analyze ophthalmic images captured by the Konan Specular Microscope XVII for examination of corneal endothelium.

| Type of Use ( <i>Select one or both, as applicable</i> ) |
|----------------------------------------------------------|
|----------------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Konan Medical. The word "KONAN" is in large, bold, sans-serif font, with the "O" in blue and the rest of the letters in black. Below "KONAN" is the word "MEDICAL" in a smaller, sans-serif font, also in black. A blue line is below the word "MEDICAL".

10-29. Mivanisl

The following 510(k) Summary is being submitted in accordance with the Requirements of MDA 1990 and CFR 807.92.

# 510(k) Summary

### Submitter Information

| A. Company Name:          | Konan Medical, Inc.              |
|---------------------------|----------------------------------|
| B. Company Address:       | 10-29, Miyanishicho, Nishinomiya |
|                           | Hyogo 662-0976, Japan            |
| C. Company Phone:         | +81-798-36-3456                  |
| D. Company Facsimile:     | +81-798-26-1028                  |
| E. Contact Person:        | Runa Emura                       |
| F. Date Summary Prepared: | June 10, 2021                    |

### Device Identification

| A. Device Trade Name:            | CellChek 20 rc                                                |
|----------------------------------|---------------------------------------------------------------|
| B. 510(k) number:                | K203244                                                       |
| C. Common Name:                  | Computer system for analysis of specular<br>microscope images |
| D. Classification Name(s):       | System, Image Management, Ophthalmic                          |
| E. Classification Regulation(s): | 892.2050                                                      |
| F. Device Class:                 | Class 2                                                       |
| G. Product Codes:                | NFJ                                                           |
| H. Prescription (Rx) or OTC?     | Rx                                                            |
| I. Advisory Panel:               | Ophthalmic                                                    |

### Identification of Predicate Device

The predicate device is the Image Storage System, Model KSS-400, which was cleared by FDA under 510(k) number K081797 on Jan 22, 2009.

{4}------------------------------------------------

-29, Miyanishicho, Nishinomiya, Hyogo 662-0976, JAPAN
TEL +81-798-36-3456 FAX +81-798-26-1028

# ldentification for Reference Device

The reference device is the Konan Specular Microscopy XVII, CellChek 20, which was cleared by FDA under 510(k) number K191558 on March 26, 2020.

| A. Common Name:                  | Specular Microscope               |
|----------------------------------|-----------------------------------|
| B. Classification Name(s):       | AC-powered slitlamp biomicroscope |
| C. Classification Regulation(s): | 886.1850                          |
| D. Device Class:                 | Class 2                           |
| E. Product Codes:                | NQE                               |
| F. Prescription (Rx) or OTC?     | Rx                                |
| G. Advisory Panel:               | Ophthalmic                        |

### Device Description

Konan Medical has developed the CellChek 20 rc to provide photographic data taken exclusively by the Konan Specular Microscope XVII, CellChek 20, which was cleared by FDA under 510(k) number K191558 on Mar 26, 2020, to research and learning centers for the advancement of ophthalmic sciences and practice. The CellChek 20 rc was developed based on the software program of CellChek 20.

CellChek 20 rc is a software program to analyze ophthalmic images for examination of corneal endothelium. This has the cell counting analysis program, and allows for analysis of the images of the cell distribution of the eye.

The software program is installed on a general-use computer to analyze corneal endothelial images photographed exclusively by the Konan Specular Microscope XVII, CellChek 20. The analysis function is to calculate mainly the cell density, the coefficient of variation of cell area, and the percent hexagonality. In the manual methods, cornea endothelial cells and cell boundaries are actually identified by users. In the automatic methods, this software detects cells and cell boundaries, however, users can modify the detection results. During operating, the users interact with the software by visually placing dots in the center of each of cells and/or by tracing cell boundaries displayed on a computer screen, or use the automatic algorithm.

| A. Common Name:            | Computer system for analysis of specular<br>microscope images |
|----------------------------|---------------------------------------------------------------|
| B. Classification Name(s): | System, Image Management, Ophthalmic                          |

{5}------------------------------------------------

-29, Miyanishicho, Nishinomiya, Hyogo 662-0976, JAPAN
TEL +81-798-36-3456 FAX +81-798-26-1028

Image /page/5/Picture/3 description: The image shows the logo for Konan Medical. The word "KONAN" is in bold, with the "O" in blue. Below that, the word "MEDICAL" is in a smaller, thinner font. There is a blue line underneath the word "MEDICAL".

| C. Classification Regulation(s): | 892.2050   |
|----------------------------------|------------|
| D. Device Class:                 | Class 2    |
| E. Product Codes:                | NFJ        |
| F. Prescription (Rx) or OTC?     | Rx         |
| G. Advisory Panel:               | Ophthalmic |

CellChek 20 rc was developed according to the harmonized standard for software, IEC 62304, and FDA requirements for software and cybersecurity for the 510(k) clearance. Additionally, risk analysis studies revealed no new risks for the CellChek 20 rc when compared to the predicate device.

As shown in the above, it is concluded that the safety and the performance of CellChek 20 rc is substantially assured.

#### Indication for Use

The CellChek 20 rc is a software program intended to analyze ophthalmic images captured by the Konan Specular Microscope XVII for examination of corneal endothelium.

#### Technological Characteristics

The CellChek 20 rc is substantially equivalent to the predicate device, the Image Storage System, Model KSS-400 as shown the below comparison table. Since the analytical functions of the CellChek 20 rc are the same as in the CellChek 20 (K191558), it is shown below as a reference device.

| Comparative Table of Predicate and Reference Devices for the CellChek 20 rc |                                                                                                           |                                                                                                              |                                                                                                                  |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Name                                                                        | Model KSS-400                                                                                             | CellChek 20                                                                                                  | CellChek 20 rc                                                                                                   |
| 510(k) No.<br>(ID No.)                                                      | K081797                                                                                                   | K191558                                                                                                      | K203244                                                                                                          |
| Device name<br>registered                                                   | Image Storage<br>System, Model<br>KSS-400                                                                 | Konan Specular<br>Microscope XVII                                                                            | CellChek 20 rc                                                                                                   |
| Indications for<br>Use                                                      | The KSS-400 Image<br>Storage System is<br>intended to be used for<br>automated image<br>analysis and data | The Konan Specular<br>Microscope XVII,<br>CellChek 20, is a<br>non-contact ophthalmic<br>microscope, optical | The CellChek 20 rc is a<br>software program<br>intended to analyze<br>ophthalmic images<br>captured by the Konan |

Konan Medical, Inc.

{6}------------------------------------------------

10-29, Miyanishicho, TEL +81-798

| Comparative Table of Predicate and Reference Devices for the CellChek 20 rc          |                                                                            |                                                                                                                                                  |                                                                        |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Name                                                                                 | Model KSS-400                                                              | CellChek 20                                                                                                                                      | CellChek 20 rc                                                         |
|                                                                                      | storage for corneal<br>images taken with<br>Konan specular<br>microscopes. | pachymeter, and<br>camera intended for<br>examination of the<br>corneal endothelium<br>and for measurement<br>of the thickness of the<br>cornea. | Specular Microscope<br>XVII for examination of<br>corneal endothelium. |
| Computer used                                                                        | General-use computer<br>by User.                                           | This device has a<br>built-in computer.                                                                                                          | General-use computer<br>by User.                                       |
| Analysis                                                                             | · Auto trace method                                                        | · Auto trace method                                                                                                                              | · Auto trace method                                                    |
|                                                                                      | · Manual trace method                                                      | · Trace method                                                                                                                                   | · Trace method                                                         |
|                                                                                      |                                                                            | · Auto center method                                                                                                                             | · Auto center method                                                   |
|                                                                                      | · Center method                                                            | · Center method                                                                                                                                  | · Center method                                                        |
|                                                                                      |                                                                            | · Auto F Center method                                                                                                                           | · Auto F Center method                                                 |
|                                                                                      | · Flex Center method                                                       | · F Center method                                                                                                                                | · F Center method                                                      |
| Display and<br>Storage of<br>Analysis results<br>and Images                          | Yes                                                                        | Yes                                                                                                                                              | Yes                                                                    |
| Print of Analysis<br>results and<br>Images                                           | Yes                                                                        | Yes                                                                                                                                              | Yes                                                                    |
| Load (Import) of<br>Images                                                           | Yes (Load)                                                                 | Yes (Import)                                                                                                                                     | Yes (Import)                                                           |
| Outboard (Export)<br>of Analysis<br>results and<br>Images                            | Yes (Outboard)                                                             | Yes (Export)                                                                                                                                     | Yes (Export)                                                           |
| Note                                                                                 |                                                                            |                                                                                                                                                  |                                                                        |
| CellChek 20 rc was developed based on the software program of CellChek 20 (K191558). |                                                                            |                                                                                                                                                  |                                                                        |

## Summary of Testing

### Non-Clinical Performance Testing

The following testing was performed on the CellChek 20 rc which was the same software function standard as those for CellChek 20.

- CellChek 20 rc device was subjected to software testing in accordance with IEC62304.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Konan Medical. The word "KONAN" is in bold, with the "KO" in blue and the "NAN" in black. The letters "MEDICAL" are in a smaller font size and are located below the word "KONAN". There is a blue line below the word "MEDICAL".

10-29, Miyanishicho, Nish TEL +81-798-36-3456

### Conclusion

From the above, it is concluded that the CellChek 20 rc software is substantially equivalent in the intended use, Indications for Use, technology, and functionality of the predicate device, the Image Storage System, Model KSS-400.

---

**Source:** [https://fda.innolitics.com/device/K203244](https://fda.innolitics.com/device/K203244)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
