The retrospective study was used to evaluate the performance of the VBrain AI software in contouring brain metastases, acoustic neuromas, and meningiomas on axial T1 contrast-enhanced MRI images.
Retrospective study; Multicenter; Multinational; AI performance validation; Brain tumor segmentation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective, blinded, multicenter, multinational study; Retrospective, blinded, multicenter, multinational study
Adult patients with brain metastases, acoustic neuromas, or meningiomas; Sample Size: 116 cases (238 tumors); Number of Sites: 4 institutions (3 US and 1 non-US)
Not applicable for this study
Lesion-wise sensitivity, false-positive rate, lesion-wise Dice coefficient, average Hausdorff distance, and average centroid distance
Independent dataset consisting of 116 cases with 238 tumors acquired from 4 different institutions (3 US and 1 non-US).
3 (board-certified radiation oncologists)
Indications for Use
VBrain is a software device intended to assist trained medical professionals, during their clinical workflows of radiation therapy treatment planning, by providing initial object contours of known (diagnosed) brain tumors (i.e., the region of interest, ROI) on axial T1 contrast-enhanced brain MRI images. VBrain uses an artificial intelligence algorithm (i.e., deep learning neural networks) to contour (segment) brain tumor on MRI images for trained medical professionals' attention, which is meant for informational purposes only and not intended for replacing their current standard practice of manual contouring process. VBrain does not alter the original MRI image, nor does it intend to be used to detect tumors for diagnosis. VBrain is intended only for generating Gross Tumor Volume (GTV) contours of brain metastases, meningiomas, and acoustic neuromas on axial T1 contrast-enhanced MRI images; It is not intended to be used with images of other brain tumors. The user must know the tumor type when they use VBrain. VBrain is intended to be used on adult patients only. Medical professionals must finalize (confirm or modify) the contours generated by VBrain, as necessary, using an external platform available at the facility that supports DICOM-RT viewing/editing functions, such as image visualization software and treatment planning system.
Device Story
VBrain is a software device for radiation therapy treatment planning. It takes axial T1 contrast-enhanced brain MRI images as input. Using deep learning neural networks, it automatically segments (contours) known brain metastases, meningiomas, and acoustic neuromas. The device operates on a PACS network; it receives patient scans and exports results as DICOM-RT objects (RTSTRUCT). It is used by trained medical professionals (e.g., radiation oncologists, dosimetrists) in clinical settings. The output is an initial contour set for professional review; clinicians must finalize, confirm, or modify these contours using external DICOM-RT compatible software (e.g., treatment planning systems). The device does not alter original images or perform diagnostic detection. By providing an automated starting point for contouring, it aims to assist clinical workflows, potentially improving efficiency in radiation therapy planning while maintaining the standard of care through mandatory human oversight.
Software-based radiological image processing device. Operates on Linux OS. Utilizes deep learning neural networks for automated segmentation. Inputs: axial T1 contrast-enhanced MRI. Outputs: DICOM-RT (RTSTRUCT). Configured for PACS network integration. Software classified as Major Level of Concern.
Indications for Use
Indicated for adult patients with known (diagnosed) brain metastases, meningiomas, or acoustic neuromas. Used by trained medical professionals (radiologists, oncologists, physicians, technologists, dosimetrists, physicists) to assist in radiation therapy treatment planning by providing initial GTV contours on axial T1 contrast-enhanced MRI images. Not for tumor detection/diagnosis. Not for use with other tumor types.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
March 19, 2021.
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
Vysioneer Inc % Chiu S. Lin Consultant 33 Rogers Street, # 308 CAMBRIDGE MA 02142
Re: K203235
Trade/Device Name: VBrain Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: QKB Dated: February 9, 2021 Received: February 10, 2021
Dear Chiu Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{2}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203235
Device Name V Brain
## Indications for Use (Describe)
V Brain is a software device intended to assist trained medical professionals, during their clinical workflows of radiation therapy treatment planning, by providing initial object contours of known (diagnosed) brain tumors (i.e., region of interest, ROI) on axial T1 contrast-enhanced brain MRI images.
V Brain uses an artificial intelligence algorithm (i.e., deep learning neural networks) to contour (segment) brain tumor on MRI images for trained medical professionals' attention, which is meant for informational purposes only and not intended for replacing their current standard practice of manual contouring process. VBrain does not alter the original MRI image, nor does it intend to be used to detect tumors for diagnosis. VBrain is intended only for generating Gross Tumor Volume (GTV) contours of brain metastases, meningiomas, and acoustic neuromas on axial T1 contrast-enhanced MRI images; It is not intended to be used with images of other brain tumors. The user must know the tumor type when they use VBrain. VBrain is intended to be used on adult patients only.
Medical professionals must finalize (confirm or modify) the contours generated by VBrain, as necessary, using an external platform available at the facility that supports DICOM-RT viewing/editing functions, such as image visualization software and treatment planning system.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image features the logo for Vysioneer, a company name displayed in a bold, sans-serif font. Above the name is a stylized 'V' shape, rendered in a shade of purple. Below the name is a horizontal purple line, and underneath that is the alphanumeric code 'K203235'.
# Section 5 510(k) Summary
#### 5.1 Submitter
Vysioneer Inc.
33 Rogers St. #308, Cambridge, MA 02142
| Contact Person: | Jen-Tang Lu, PhD (Chief Executive Officer) |
|------------------------|--------------------------------------------|
| Phone: | 609-865-8659 |
| Email: | jt@vysioneer.com |
| Date Summary Prepared: | February 09, 2021 |
#### 5.2 Device Name
| Trade Name: | VBrain |
|-----------------------------------|-----------------------------------------------------------------|
| Common Name: | Radiological Image Processing Software for<br>Radiation Therapy |
| Regulation Number / Product Code: | 21 CFR 892.2050 / QKB |
### 5.3 PREDICATE DEVICE
Primary Predicate #1: AccuContour™, K191928, Xiamen Manteia Technology LTD Primary Predicate #2: MIM - MRT Dosimetry, K182624, MIM Software Inc.
#### Intended Use / Indications for Use 5.4
VBrain is a software device intended to assist trained medical professionals, during their clinical workflows of radiation therapy treatment planning, by providing initial object contours of known (diagnosed) brain tumors (i.e., the region of interest, ROI) on axial T1 contrast-enhanced brain MRI images.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Vysioneer. The logo consists of a stylized "V" shape in purple, with the company name "VYSIONEER" written in a sans-serif font below it, also in purple. The logo is simple and modern.
VBrain uses an artificial intelligence algorithm (i.e., deep learning neural networks) to contour (segment) brain tumor on MRI images for trained medical professionals' attention, which is meant for informational purposes only and not intended for replacing their current standard practice of manual contouring process. VBrain does not alter the original MRI image, nor does it intend to be used to detect turnors for diagnosis. VBrain is intended only for generating Gross Tumor Volume (GTV) contours of brain metastases, meningiomas, and acoustic neuromas on axial T1 contrast-enhanced MRI images; It is not intended to be used with images of other brain tumors. The user must know the tumor type when they use VBrain is intended to be used on adult patients only.
Medical professionals must finalize (confirm or modify) the contours generated by VBrain, as necessary, using an external platform available at the facility that supports DICOM-RT viewing/editing functions, such as image visualization software and treatment planning system.
#### Device Description ર્ રંડ
VBrain is a software device indicated for use in the analysis of brain MRI images. The device consists of image preprocessing, deep learning neural networks, and postprocessing components, and is intended to assist trained medical professionals, during clinical workflows of radiation therapy treatment planning, by highlighting and contouring known (diagnosed) brain tumors on the axial T1 contrast-enhanced MRI images. The software is configured to work on a PACS network. Upon user's request, it will patient scans or users can send corresponding MR images, and the device will utilize deep learning neural networks to generate contours for the detected/diagnosed brain tumors and export the results as DICOM-RT objects (using the RT Structure Set ROI Contour attribute, RTSTRUCT) back to the network. The medical professionals must finalize (confirm and modify) the contours produced by VBrain as necessary using an external platform that supports RT DICOM viewing/editing, such as a treatment planning system.
#### Comparison with Predicate Devices 5.6
VBrain is substantially equivalent to a combination of the primary predicate devices AccuContour™ (K191928) and MIM - MRT Dosimetry (K182624).
The proposed device, VBrain, and the primary predicates, AccuContour™ (K191928) and K182624 (MIM - MRT Dosimetry), are all software devices intended to be used in the workflow of radiation therapy by providing tools of segmenting (contouring) of tumors and/or organs on MRI and/or CT images. Both the proposed device and AccuContour™ (K191928) are AI-based (deep learning) software regulated under the Product Code QKB (Radiological Image Processing Software For Radiation Therapy). On the other hand, both the proposed device and K182624 (MIM - MRT Dosimetry) provide tools for tumor contouring. The only difference is that VBrain uses deep learning (neural networks) to automatically generate tumor contours as a starting point for user's review and edit, while K182624 (MIM - MRT Dosimetry) provides a semi-automatic
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Vysioneer. The logo consists of a stylized letter "V" formed by two diagonal lines in a purple color. Below the symbol, the word "VYSIONEER" is written in a sans-serif font, also in purple. A thin purple line is located below the word.
tool (propagation tool) that requires user's input to start the image segmentation (contouring) process.
Although the proposed new device. VBrain, uses a data-driven deep learning-based algorithm for contouring of known brain tumors, the primary predicate MIM - MRT Dosimetry (K182624) uses a semi-automatic algorithm that requires user's input to start the contouring process. The specific design of the proposed device does not raise different questions of safety and effectiveness, because the new device only provides initial object contours of known (diagnosed) brain tumors for the medical professionals' attention, which are meant for informational purposes only and not intended for replacing their current standard practice of manual contouring process. Medical professionals must use VBrain in conjunction with appropriate software to review and edit results generated automatically by VBrain. In addition, VBrain does not alter the original MRI image, nor does it intend to be used to detect tumors for diagnosis. The medical professionals must know the tumor type when they use VBrain. Consequently, the new device does not change any medical professionals' workflow planning procedure and therefore does not raise different questions of safety and effectiveness.
Please see Table 5-1 comparing the intended use and key technological characteristics of the proposed device and the predicate devices.
| Company | Vysioneer Inc. | Xiamen Manteia Technology LTD. (Primary) | MIM Software Inc. (Primary) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | VBrain | AccuContour™ | MIM - MRT Dosimetry |
| 510k Number | Pending | K191928 | K182624 |
| Regulation No. | 21CFR 892.2050 | 21CFR 892.2050 | 21CFR 892.2050 |
| Classification | II | II | II |
| Product Code | QKB | QKB | LLZ |
| Intended Use/Indication for Use | VBrain is a software device intended to assist trained medical professionals, during their clinical workflows of radiation therapy treatment planning, by providing initial object contours of known (diagnosed) brain tumors (i.e., the region of interest, ROI) on axial T1 contrast-enhanced brain MRI images. | It is used by radiation oncology department to register multimodality images and segment (non-contrast) CT images, to generate needed information for treatment planning, treatment evaluation and treatment adaptation. The product has two image process functions: (1) Deep learning contouring: it can automatically contour | MIM software is used by trained medical professionals as a tool to aid in evaluation and information management of digital medical images. The medical image modalities include, but are not limited to, CT, MRI, CR, DX, MG, US, SPECT, PET and XA as supported by ACR/NEMA DICOM 3.0. |
| VBrain uses an artificial<br>intelligence algorithm<br>(i.e., deep learning<br>neural networks) to<br>contour (segment) brain<br>tumor on MRI images<br>for trained medical<br>professionals' attention,<br>which is meant for<br>informational purposes<br>only and not intended<br>for replacing their<br>current standard practice<br>of manual contouring<br>process. VBrain does not<br>alter the original MRI<br>image, nor does it intend<br>to be used to detect<br>tumors for diagnosis.<br>VBrain is intended only<br>for generating Gross<br>Tumor Volume (GTV)<br>contours of brain<br>metastases,<br>meningiomas, and<br>acoustic neuromas on<br>axial T1 contrast-<br>enhanced MRI images; it<br>is not intended to be<br>used with images of<br>other brain tumors. The<br>user must know the<br>tumor type when they<br>use VBrain. VBrain is<br>intended to be used on<br>adult patients only.<br>Medical professionals<br>must finalize (confirm or<br>modify) the contours<br>generated by VBrain, as<br>necessary, using an<br>external platform<br>available at the facility<br>that supports DICOM-<br>RT viewing/editing<br>functions, such as image<br>visualization software<br>and treatment planning<br>system | the organ-at-risk,<br>including head and neck,<br>thorax, abdomen and<br>pelvis (for both male and<br>female),<br>(2) Automatic<br>Registration, and<br>(3) Manual Contour.<br><br>It also has the following<br>general functions:<br>(1) Receive,<br>add/edit/delete, transmit,<br>input/export, medical<br>images and DICOM<br>data;<br>(2) Patient management;<br>(3) Review of processed<br>images;<br><br>(4) Open and Save of<br>files. | MIM assists in the<br>following indications:<br>• Receive, transmit,<br>store, retrieve,<br>display, print, and<br>process medical<br>images and DICOM<br>objects.<br>• Create, display and<br>print reports from<br>medical images.<br>• Registration, fusion<br>display, and<br>review of medical<br>images for diagnosis,<br>treatment evaluation,<br>and treatment planning.<br>• Evaluation of cardiac<br>left ventricular end-<br>diastolic volume,<br>end-systolic volume, and<br>ejection<br>fraction.<br>• Localization and<br>definition of objects,<br>such as tumors and<br>normal tissues in<br>medical images.<br>• Creation,<br>transformation, and<br>modification of contours<br>for applications<br>including, but not<br>limited to, quantitative<br>analysis, aiding adaptive<br>therapy, transferring<br>contours to radiation<br>therapy treatment<br>planning systems, and<br>archiving contours for<br>patient follow-up and<br>management.<br>• Quantitative and<br>statistical analysis of<br>PET/SPECT brain scans<br>by comparing to other<br>registered PET/SPECT<br>brain scans | |
| | | | • Planning and<br>evaluation of permanent<br>implant brachytherapy<br>procedures (not<br>including<br>radioactive<br>microspheres).<br>• Calculating absorbed<br>radiation dose as a result<br>of administering a<br>radionuclide. |
| | | | When using device<br>clinically, the user<br>should only use FDA<br>approved<br>radiopharmaceuticals. If<br>using with unapproved<br>ones, this device should<br>only be used for research<br>purposes. |
| | | | Lossy compressed<br>mammographic images<br>and digitized film screen<br>images must not be<br>reviewed for<br>primary image<br>interpretations.<br>Images that are printed<br>to film must be printed<br>using an FDA-approved<br>printer for the diagnosis<br>of digital mammography<br>images. |
| | | | Mammographic images<br>must be viewed on a<br>display system that has<br>been cleared by the FDA<br>for the diagnosis of<br>digital mammography<br>images. The software is<br>not to be used for<br>mammography CAD. |
| Segmentation<br>(Contouring)<br>Technology | Deep learning | Deep learning | Atlas-based algorithm<br>and propagation tools<br>(requiring user's input to |
| | | | start the image<br>segmentation process) |
| Operating System | Linux operating system | Microsoft Windows | Microsoft Windows and<br>Apple macOS operating<br>systems…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.