K203159 · Carestream Health · MQB · Dec 2, 2020 · Radiology
Device Facts
Record ID
K203159
Device Name
Lux 35 Detector
Applicant
Carestream Health
Product Code
MQB · Radiology
Decision Date
Dec 2, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is intended to capture for display radiographic images of human anatomy including both pediatric and adult patients. The device is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. Excluded from the indications for use are mammography, fluoroscopy, and angiography applications.
Device Story
Lux 35 Detector is a digital flat panel X-ray imager; captures incident X-rays via cesium iodide scintillator; converts energy to visible light then electrical charge in amorphous silicon pixel array. Operates wirelessly (battery-powered) or tethered; transmits data to Carestream DRX-1 System console. Used in clinical settings by radiologic technologists; replaces conventional screen-film or CR systems. Image acquisition software performs defect correction, gain/offset calibration, and sub-sampling. ImageView software on console renders images for physician review; DICOM output. Improved DQE and MTF compared to predicate; reduced weight and ergonomic design (rounded edges, finger grips) facilitate clinical workflow; enables high-quality diagnostic imaging.
Clinical Evidence
No clinical data. Bench testing only. Phantom image study compared image quality attributes (detail, sharpness, noise, artifacts) between subject and predicate. >84% of responses rated subject device 0 or higher (equivalent or better) than predicate. All attribute ratings were significantly greater than 0.
Technological Characteristics
Amorphous silicon flat panel image sensor; cesium iodide (CsI) scintillator; glass-reinforced epoxy laminate substrate. Wireless (battery) or tethered connectivity. Dimensions: 35cm x 42.1cm active area; 139-micron pixel pitch. Complies with ANSI AAMI ES60601-1, IEC 60601-1-2, ISO 14971, ISO 10993-1.
Indications for Use
Indicated for capturing radiographic images of human anatomy in pediatric and adult patients for general projection radiography. Excluded: mammography, fluoroscopy, and angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
December 2, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Carestream Health, Inc. % Gina Maiolo Regulatory Affairs Manager 150 Verona Street ROCHESTER NY 14608
Re: K203159
Trade/Device Name: Lux 35 Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: October 30, 2020 Received: November 2, 2020
Dear Gina Maiolo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203159
Device Name Lux 35 Detector
Indications for Use (Describe)
"The device is intended to capture for display radiographic images of human anatomy including both pediatric and adult patients. The device is intended for use in general projections wherever conventional screen-film systems or CR systems may be used. Excluded from the indications for use are mammography, fluoroscopy, and angiography applications".
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;"><b></b></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K203159
### 510(k) Summary
510(k) Owner: Carestream Health, Inc 150 Verona Street Rochester, New York 14608 Contact: Gina Maiolo Regulatory Affairs Manager 585.627.6543
Carestream Health, Inc. is submitting this Special 510(k) premarket notification for modifications to the DRX Plus 3543 Flat Panel Detector. Caresteam believes the modified DRX Plus 3543C Flat Panel Detector is substantially equivalent to the cleared device (K150766).
| Date Summary Prepared: August 15 2020 | | |
|---------------------------------------|---------------------------------------------------------|---------------------------------------------------------|
| | Predicate | Subject |
| Device Trade Name | DRX Plus 3543 Detector | Lux 35 Detector |
| Device Common Name | Digital Flat Panel Detector | Digital Flat Panel Detector |
| Classification Name | Solid State X-Ray Imager (Flat<br>Panel/Digital Imager) | Solid State X-Ray Imager (Flat<br>Panel/Digital Imager) |
| Device Class | Class II | Class II |
| Device Code | MQB | MQB |
| Regulation Number | 21 CFR 892.1680 | 21 CFR 892.1680 |
| Predicate Device | DRX Plus 3543GOS Flat Panel<br>Detector | DRX Plus 3543C Flat Panel<br>Detector |
#### Indications for Use
"The device is intended to capture for display radiographic images of human anatomy including both pediatric and adult patients. The device is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. Excluded from the indications for use are mammography, fluoroscopy, and angiography applications."
In addition, the indications for use of the modified device, as described in labeling does not change as a result of the device modifications.
{4}------------------------------------------------
### Device Description
The modified DRX Plus 3543C is a scintillator-photodetector device (Solid State X-ray Imager) utilizing an amorphous silicon flat panel image sensor. The modified detector is redesigned with the intent to reduce weight and increase durability, while utilizing a non-glass substrate material and cesium iodide scintillator. The modified detector, like the predicate is designed to interact with Carestream's DRX-1 System (K090318).
The modified DRX Plus 3543C Detector, like the predicate, creates a digital image from the x-rays incident on the input surface during an x-ray exposure. The flat panel imager absorbs incident x-rays and converts the energy into visible light photons. These light photons are converted into electrical charge and stored in structures called "pixels." The digital value in each pixel of the image is directly related to the intensity of the incident x-ray flux at that particular location on the surface of the detector. Image acquisition software is used to correct the digital image for defective pixels and lines on the detector, perform gain and offset correction and generate sub-sampled preview images
The modified and predicate are essentially the same as they are both digital x-ray flat panel scintillatorphotodetector devices (Solid State X-ray Imagers) utilizing an amorphous silicon flat panel image sensor. The predicate and modified are both DRX Plus 3543 flat panel detectors, the difference is that the modified device utilizes a cesium iodide (Csi) scintillator, whereas the predicate utilizes gadolinium oxysulfide (GOS). The modified device official product name is marketed as "Lux 35 Detector." Both flat panels operate primarily in a wireless state, using a battery for power and allows wireless communication for control and data transmission. This eliminates the need for cables that can hinder efficient workflow. The detectors have the ability to communicate via a wired ("tethered") connection to the console, if desired
#### Technological Characteristics
The modified detector is substantially equivalent to the predicate device currently cleared on the market (K150766) and uses the same fundamental scientific technology as the predicate device. The detectors are both digital x-ray flat panel scintillator-photodetector devices (Solid State X-ray Imager) utilizing an amorphous silicon flat panel image sensor. The predicate and modified are both DRX Plus 3543 flat panel detectors, the main difference is that the modified device utilizes a cesium iodide (Csi) scintillator and non-glass substrate whereas the predicate utilizes gadolinium oxysulfide (GOS) and a glass substrate. The modified device official product name is marketed as Lux 35 Detector.
Both flat panels operate primarily in a wireless state, using a battery for power and allows wireless communication for control and data transmission. This eliminates the need for cables that can hinder efficient workflow. The detectors have the ability to communicate via a wired ("tethered") connection to the console, if desired. The modified detector is designed to interact with Carestream's existing acquisition software that resides on the legally marketed Carestream DRX-1 System (K090318). The predicate device (DRX Plus 3543) has clearance with DRX-1 under K150766. Additionally, the DRX-1 System has clearance for the following Carestream detectors:
- K130464 (DRX 2530C)
- . K183425 (DRX Plus 2530 Detector)
- K150766 (DRX 3543 Plus Detectors) .
{5}------------------------------------------------
Carestream has obtained previous detector clearances for use with DRX-1, as listed above. This submission is for the clearance of the modified digital flat panel detector only.
After images are captured with the digital flat panel detector and can be communicated to the DRX-1 System console. The DRX-1 System contains Image Eclipse Processing Software II to allow rendering of the images. The image is processed and transformed into data for viewing and then put into DICOM format for output. Image Eclipse Processing Software II has been cleared for use (K180809) with previous Carestream DRX Plus 2530C under K183425.
The image acquisition software corrects the digital image for defective pixels and lines on the detector. perform gain and offset correction, and generate sub-sampled preview images. The image acquisition software that resides on the DRX-1 System console will be replaced with Image View Software. The ImageView Software has obtained several FDA clearances for use with Carestream Digital Radiography Systems used with the Carestream flat panels:
- 트 K163203 - DRX Evolution System with Image View Software
- . K191205 - DRX-Mobile Revolution System with Image View Software
- . K201373 - DRX-Compass
#### Summary of Guidance & Standards Compliance
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014
- . Pediatric Information for X-ray Imaging Device Premarket Notifications Guidance for Industry and Food and Drug Administration Staff Document issued on November 28, 2017
- . Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: September 1, 2016
- Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for Software contained in Medical Devices, Document issued on: May 11, 2005 Medical Devices, Document issued on: May 11, 2005
- . Guidance for Industry and FDA Radio Frequency Wireless Technology in Medical devices Guidance for Industry and Food and Drug Administration Staff Document issued on: August 14, 2013 Document issued on: August 14, 2013
{6}------------------------------------------------
Applicable Standards:
- . ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- ANSI C95 1-1999 Standard for Safety Levels with Respect to Human Exposure to Radio Frequency Electromagnetic Fields 3KHz to 300GHz
- IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment Part 1-2: General ● requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- FCC Radio Frequency Devices, Sub C (International Radiators) Sub E (Unlicensed Nationale Infrastructure Devices, 209(a) (Radiated Emissions) 30-1000MHz), 15.205 (Restricted Bands),15.203 (Antenna Requirements)
- ISO 14971:2019 Medical Devices- Application of Risk Management to medical devices
- ISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of medical devices Part 1: . Evaluation and testing within a risk management process
### Summary of Non-Clinical Testing
Clinical testing was not required to establish substantial equivalence. Bench testing was sufficient to assess the device safety and effectiveness. The performance of the modified device DRX Plus 3543C (Lux 35) detector was evaluated in non-clinical (bench) testing using a Phantom Image Study. A technical analysis (non-diagnostic) of image quality attributes such as detail, sharpness, noise and appearance of artifacts were compared for each image.
Greater than 84% of all responses were rated 0 or higher in favor of the modified DRX Plus 3543C panel. All ratings for the attributes (detail contrast, sharpness and noise) were significantly greater than 0 indicating that the modified DRX Plus 3543C images were equivalent to just noticeably better than the predicate images.
{7}------------------------------------------------
# Carestream
A comparison chart provides similarities and differences between the modified and predicate device.
| Comparison | Predicate | Modified | Risk Conclusion |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | DRX Plus 3543 Digital Flat<br>Panel Detector (K150766) | DRX Plus 3543C (Lux<br>35) Digital Flat Panel<br>Detector | |
| Indications for<br>Use | The device is intended to capture<br>for display radiographic images of<br>human anatomy including both<br>pediatric and adult patients. The<br>device is intended for use in<br>general projection radiographic<br>applications wherever conventional<br>screen-film systems or CR systems<br>may be used. Excluded from the<br>indications for use are<br>mammography, fluoroscopy, and<br>angiography applications | *Same | Device name change does not<br>impact indications for use<br>and/or safety and effectiveness |
| Image<br>Processing<br>Software | Eclipse Image Software (K060137) | Eclipse Image Software<br>II (K180809) | Predicate software used a 4-<br>frequency band decomposition<br>method, the modified device<br>uses a 4-10 frequency band<br>decomposition. Differences do<br>not impact indications for use<br>or safety and/or effectiveness |
| Substrate | Glass | Glass reinforced epoxy<br>laminate | Only the material of the<br>substrate has changed. No<br>change to device<br>functionality/operation. Does<br>not impact image quality, safety<br>and/or effectiveness |
| X-ray<br>Scintillator<br>Material | Gadolinium Oxysulfide (GOS) | Cesium Iodide (CsI) | Differences in detector material<br>does not impact safety and/or<br>effectiveness |
| DQE | DQE 26% (RQA-5, 1 cycle/mm,<br>3.1 µGy) | DQE 55% (RQA-5, 1<br>cycle/mm, 2.5 µGy) | Improved image quality |
| MTF | MTF 54% (RQA-5, 1 cycle/mm) | MTF 62% (RQA-5, 1<br>cycle/mm) | Improved image quality |
| Detector Weight | 6.75 lbs | 4.6 lbs | Difference in detector weight<br>does not impact safety and/or<br>effectiveness |
| Appearance<br>Characteristics | Squared edges | Rounded edges<br>Improved LEDs<br>New Display GUI<br>New Finger Grips | Differences in detector<br>characteristics do not impact<br>safety and/or effectiveness |
| Image Capture<br>Area (usable<br>pixel area) | 35cm x 42.1cm | *Same | No new risks |
| Pixel Pitch | 139 microns | *Same | No new risks |
| Workstation<br>Console<br>Software | DirectView Software | ImageView Software | The change in software application does not change the detector operation. It is a replacement software limited to GUI changes / Ease of Use. Changes to do not impact safety and/or effectiveness of detectors |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Carestream" in orange font. The font is sans-serif and appears to be bolded. The word is horizontally oriented and takes up most of the image space.
### Conclusion
- . The image quality of the modified device is at least as good as or better than that of the predicate device.
- Results of non-clinical testing demonstrate that the modified device is as safe and as effective as . the predicate device.
- The intended use remains unchanged. ●
- The fundamental scientific technology of the modified device is the same and is substantially . equivalent to the predicate.
- The comparison chart demonstrates that the characteristics are primarily the same. ●
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.