K203146 · Akesis, Inc. · IWB · Mar 5, 2021 · Radiology
Device Facts
Record ID
K203146
Device Name
Akesis Galaxy RTi
Applicant
Akesis, Inc.
Product Code
IWB · Radiology
Decision Date
Mar 5, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Akesis Galaxy RTi Rotating Gamma System, is a teletherapy device intended for the stereotactic irradiation of human head structures in adults
Device Story
Akesis Galaxy RTi is a teletherapy device for stereotactic irradiation of adult head structures. It utilizes 30 Cobalt-60 sources housed in a hemispheric shield with a primary collimator and an independently rotating secondary collimator hemisphere (4 sizes plus blocking). The system integrates a y-ray treatment unit, a stereotactic localization interface, and a kV x-ray positioning/verification system. Operated by clinical staff in a medical facility, the device delivers radiation therapy based on pre-planned stereotactic coordinates. The kV x-ray system provides image-based positioning and verification to ensure accurate delivery. The device benefits patients by enabling precise, targeted radiation treatment for intracranial conditions.
Clinical Evidence
No clinical data. Bench testing only. System verified for electrical safety and EMC per IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-68, 60601-2-11, 61217, 62274, and 62366. Hardware and software verification and validation conducted per 21 CFR 820, ISO 13485, and ISO 14971.
Technological Characteristics
Radionuclide radiation therapy system using 30 Cobalt-60 sources. Features hemispheric shield, primary collimator, and rotating secondary collimator. Includes kV x-ray imaging system for positioning. Complies with IEC 60601 series standards. No direct tissue-contacting components; biocompatibility testing not required.
Indications for Use
Indicated for stereotactic irradiation of human head structures in adults.
Regulatory Classification
Identification
A radionuclide radiation therapy system is a device intended to permit an operator to administer gamma radiation therapy, with the radiation source located at a distance from the patient's body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts (including beam-limiting devices), and accessories.
{0}------------------------------------------------
March 5, 2021
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
Akesis Inc % Mr. Christopher Brown Director Quality Assurance and Regulatory Affairs 5129 Commercial Circle CONCORD CA 94520
Re: K203146
Trade/Device Name: Akesis Galaxy RTi Regulation Number: 21 CFR 892.5750 Regulation Name: Radionuclide radiation therapy system Regulatory Class: Class II Product Code: IWB Dated: January 28, 2021 Received: January 29, 2021
Dear Mr. Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203146
Device Name Akesis Galaxy RTi
#### Indications for Use (Describe)
The Akesis Galaxy RTi Rotating Gamma System, is a teletherapy device intended for the stereotactic irradiation of human head structures for adults.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Premarket Notification [510(k)] Summary Akesis Galaxy Rotating Gamma System
The following information is provided following the format of 21 CFR 807.92.
# I. GENERAL INFORMATION
K203146
| Submitter's Name: | Akesis Inc.<br>5129 Commercial Circle<br>Concord, CA 94520 |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name: | Christopher Brown<br>Phone: +1 925 408 1381<br>Fax: +1 925 326 2646 |
| Date Prepared: | October 19, 2020 |
| II. DEVICE INFORMATION | |
| Proprietary Name: | Akesis Galaxy RTi |
| Classification Name: | Radionuclide radiation therapy system |
| Device Classification: | 21 CFR 892.5750 |
| Device Class: | Class II |
| Product Code: | IWB |
| Common/Usual Name: | Radionuclide radiation therapy system |
| III. PREDICATE DEVICE | Akesis Galaxy RTx Rotating Gamma System (K200050) |
| IV. DEVICE DESCRIPTION | The Akesis Galaxy RTi, is a medical teletherapy device, which contains<br>(30) Cobalt-60 sources, distributed in a single, compact source carrier or<br>source drawer, which is inside a hemispheric shield. Immediately adjacent<br>to the source unit is the "built-in" primary collimator. An independently<br>rotating secondary collimator hemisphere contains 4 sizes of collimators<br>and a blocking position. |
| | consists of a) The y-ray treatment unit, and b) interface to a stereotactic<br>localization system, and c) kV x-ray positioning and verification system |
| V. INTENDED USE | The Akesis Galaxy RTi Rotating Gamma System, is a teletherapy device<br>intended for the stereotactic irradiation of human head structures in adults |
| VI. INDICATIONS FOR USE | The Akesis Galaxy RTi Rotating Gamma System, is a teletherapy device<br>intended for the stereotactic irradiation of human head structures in adults |
{4}------------------------------------------------
# VII. COMPARISION OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Akesis Galaxy RTi has the same intended use, uses the same radioisotope, applies the same operating principle, has the same technical characteristics and meets the same performance specifications. Both the predicate and the Akesis Galaxy meet the same set of regulations and standards.
The following technological differences exists between the subject and the predicate device:
- -Addition of a kV x-ray imaging system
- Provides an interface to compatible immobilization devices -
# VIII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence:
## Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and electromagnetic compatibility (EMC) testing was conducted on this medical device. The system complies with the IEC 60601-1 standards for safety and the IEC 60601-1-2 standard for EMC. Successful testing was performed in accordance with IEC standards 60601-1-3, 60601-1-6, 60601-2-68, 60601-2-11, 60601-1-2, 61217, 62274 and 62366.
## Hardware and Software Verification and Validation Testing
Hardware and software verification and validation testing was conducted according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 Quality Management System standard, ISO 14971 Risk Management Standard and the other FDA recognized consensus standards. Test results showed conformance to applicable requirements specifications and assured hazard safeguards functioned properly. Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software in Medical Devices." The software for this device was considered as a "major" level of concern.
## Biocompatibility Testing
Biocompatibility testing was not performed since this device does not have any direct tissue contacting components / parts.
This is in accordance with the Guidance for Industry and FDA Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", September 2020.
## Animal and Clinical Studies
Animal and clinical testing was not required.
{5}------------------------------------------------
## Summarv
These tests have demonstrated that the Akesis Galaxy RTi has met its specifications, demonstrated substantially equivalent performance to the predicate device and is suitable for its intended use.
# IX. CONCLUSIONS
The Akesis Galaxy RTi has the same intended use, fundamental scientific technology and principles of operation as the predicate device.
Use of the Akesis Galaxy RTi does not raise any new or different issues of safety or effectiveness when compared with the predicate device.
Performance tests have demonstrated that the Akesis Galaxy RTi has met its specifications, demonstrated substantially equivalent performance to the predicate device and is suitable for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.