Franz Zirconia Dental Crown

K203072 · Franz Biotech, Inc. · ELL · Jan 19, 2022 · Dental

Device Facts

Record IDK203072
Device NameFranz Zirconia Dental Crown
ApplicantFranz Biotech, Inc.
Product CodeELL · Dental
Decision DateJan 19, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3920
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Franz Zirconia Dental Crown is indicated for use as main structure of an artificial dental prothesis for partially edentulous patients which require prosthetic oral reconstruction to restore chewing function. Franz Zirconia Dental Crown is indicated for use as single crown that will be cemented to an artificial or natural tooth abutment.

Device Story

Franz Zirconia Dental Crown is a Y-TZP zirconium oxide prosthetic device used to replace natural teeth. The design process involves scanning, designing, and ordering the crown via third-party CAD systems in a dental laboratory or practice. The digital design is transmitted to a centralized milling center where the crown is fabricated using additive manufacturing (3D printing). The final crown is cemented or bonded to a natural or artificial tooth abutment by a clinician. The device restores chewing function and provides a natural tooth-like appearance. It is suitable for cut-back (veneering) or stain and glaze techniques to match surrounding teeth. The device benefits patients by providing a durable, biocompatible prosthetic solution for oral reconstruction.

Clinical Evidence

No clinical data or animal studies were conducted. Substantial equivalence is supported by bench testing, including mechanical testing (biaxial flexural strength, fracture toughness, thermal expansion, radioactivity) and biocompatibility testing per ISO 10993 standards (genotoxicity, cytotoxicity, irritation, sensitization, systemic toxicity).

Technological Characteristics

Material: Y-TZP Zirconium Oxide. Manufacturing: Additive manufacturing (3D printing). Sintering temperature: 1500°C. Minimum thickness: 0.4mm (anterior), 0.7mm (pre-molar/molar). Standards: ISO 6872 (Dentistry - Ceramic materials), ISO 10993 (Biocompatibility). Connectivity: Electronic transmission of CAD designs to centralized fabrication center.

Indications for Use

Indicated for partially edentulous patients requiring prosthetic oral reconstruction to restore chewing function; intended as a single crown replacement for a natural tooth, cemented to a natural or artificial abutment.

Regulatory Classification

Identification

A porcelain tooth is a prefabricated device made of porcelain powder for clinical use (§ 872.6660) intended for use in construction of fixed or removable prostheses, such as crowns and partial dentures.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 19, 2022 Franz Biotech Inc. Anita Chen Official Applicant 5F .- 3, No. 56, Ln. 258, Ruiguang Rd., Neihu Dist. Taipei City, Taiwan 114 TAIWAN Re: K203072 Trade/Device Name: Franz Zirconia Dental Crown Regulation Number: 21 CFR 872.3920 Regulation Name: Porcelain Tooth Regulatory Class: Class II Product Code: ELL Dated: December 22, 2021 Received: December 22, 2021 Dear Anita Chen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {2}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K203072 Device Name Franz Zirconia Dental Crown Indications for Use (Describe) Franz Zirconia Dental Crown is indicated for use as main structure of an artificial dental prothesis for partially edentulous patients which require prosthetic oral reconstruction to restore chewing function. Franz Zirconia Dental Crown is indicated for use as single crown that will be cemented to an artificial or natural tooth abutment. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## K203072 # 510(k) Summary ## 510(k) Summary of Safety and Effectiveness This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92 The assigned 510(k) Number: K203072 | 1. | Submitter | | |----|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------| | | Mailing Address | Franz Biotech Inc.<br>4F, No. 167, Sec. 5, Minsheng E. Rd., Songshan<br>Dist., Taipei City 105, Taiwan (R.O.C.) | | | Factory Address | 5F.-3, No. 56, Ln. 258, Ruiguang Rd., Neihu<br>Dist., Taipei City 114,<br>Taiwan (R.O.C.)Establishment Registration No.<br>1st submission | | | Contact Person | Anita Chen | | | Phone: | +886(0) 939-855-759 | | | E-mail: | m9104303@gmail.com | | | Date Prepared | August 6, 2020 | | 2 | Device Name | | | | Common or Usual Name | Porcelain tooth | | | Regulation Description | Porcelain tooth | | | Product Code | ELL | | | Device | Franz Zirconia Dental Crown | | | CFR Classification | CFR Part 872.3920 | | | Device Class | II | | | Classification Panel | Dental | | 3 | Predicate Device Name | | | | 510(k) number: | K153534 | | | Trade or proprietary or<br>model name: | NobelProcera HT ML Full Contour Zirconia<br>Crown | | | Manufacturer: | NobelBiocare AB<br>Vastra Hamngatan 1<br>Goteborg, SE-411 17 | {5}------------------------------------------------ #### Sweden #### Device Description: 4 Franz Zirconia Dental Crown is indicated for use as main structure of an artificial dental prothesis for partially edentulous patients which require prosthetic oral reconstruction to restore chewing function. Franz Zirconia Dental Crown is indicated for use as single crown that will be cemented to an artificial or natural tooth abutment. Franz Zirconia Dental Crown is intended to be a replacement for a natural tooth. After finalizing the Franz Zirconia Dental Crown in the dental laboratory, it is cemented or bonded onto a tooth or artificial abutment, by a clinician, to provide a natural tooth like appearance and to restore chewing functionality in the patient's mouth. To achieve esthetics and required value and chroma of the surrounding natural teeth the Franz Zirconia Dental Crown is suitable for cut-back (veneering) or stain and glaze techniques. The design of the Franz Zirconia Dental Crown is determined in a dental laboratory, hospital or dental practice by scanning, designing and ordering the crown using or supported third party CAD systems. The crown, once ordered, is sent electronically to one of Franz's centralized milling centers for fabrication. Indications for Use: న. Franz Zirconia Dental Crown is indicated for use as main structure of an artificial dental prothesis {6}------------------------------------------------ for partially edentulous patients which require prosthetic oral reconstruction to restore chewing function. Franz Zirconia Dental Crown is indicated for use as single crown that will be cemented to an artificial or natural tooth abutment. Franz Dental Zirconia Crown is intended to be a replacement for a natural tooth. - 7. Technological Characteristics and Substantial Equivalence Comparison with Predicate: Intended Use A comparison of the device features, intended use, and other information demonstrates that the Franz Dental Zirconia Crown is substantially equivalent to the predicate device as summarized in Table 1. A minor technological differences between the predicate device and subject device was made by 3D print process, the predicate device was made by CNC. The materials, size, shape was totally the same with the predicate device. Both device was used the same technological for CAD. Although there is a slightly different technological for manufacturer process, as compared to the predicate, the performance data demonstrates the proposed device performs as safely and effectively as the predicate device. The differences raise no new question of safety and effectiveness. | | Candidate Device | Predicate Device | |-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | Franz biotech Inc. | Nobel Biocare AB | | Device Name | Franz Zirconia Dental<br>Crown | NobelProcera HT ML Full<br>Contour Zirconia Crown | | 510(k) Number | K203072 | K153534 | | Regulation<br>Number | 21 CFR 872.3920 | 21 CFR 872.3920 | | Regulation<br>Name | Porcelain Tooth | Porcelain Tooth | | Intended Use | Intended to be a replacement<br>for a natural tooth. | Intended for use as an aid in<br>prosthetic rehabilitation | | | Franz Zirconia Dental Crown | NobelProcera HT ML FCZ | | Indications for<br>Use | is indicated for use as main<br>structure of an artificial dental<br>prothesis for partially<br>edentulous patients which<br>require prosthetic oral<br>reconstruction to restore<br>chewing function.<br>Franz Zirconia Dental Crown<br>is indicated for use as single<br>crown that will be cemented<br>to an artificial or natural tooth<br>abutment._ | Crown is indicated for use as<br>core structure of an artificial<br>prosthesis for partially<br>edentulous<br>patients in the need of<br>prosthetic oral reconstruction<br>in order to restore chewing<br>function.<br>NobelProcera HT ML FCZ<br>Crown is indicated for use as<br>single crown that will be<br>cemented to a natural or<br>artificial tooth<br>abutment. | | Material | Y-TZP Zirconium Oxide | Y-TZP Zirconium Oxide<br>(Katana Zirconia K143439) | | Manufacturing<br>Method | Franz biotech in-house 3D-<br>printing (Additive<br>Manufacturing; AM) | Nobel Biocare in-house<br>CAM | | Minimum<br>Thickness | 0.4mm (Anterior)<br>0.7mm (Pre-molar and Molar) | 0.4mm (Anterior)<br>0.7mm (Pre-molar and<br>Molar) | | Number of units | Individual crown | Individual crown | | Single Use | Yes | Yes | | Number of units | Individual Crown | Individual crown | | Biaxial Flexural<br>Strength<br>(MPa)<br>(sintered) | Means±(SD)<br>1061(90) | Means±(SD)<br>1092(112) | | Fracture<br>Toughness<br>(MPam0.5) | 6.44 | Unknow | | Thermal<br>Expansion<br>(um/m°C) | 9.95 | 10.80 | | Chemical<br>Solubility in<br>Water<br>(ug/cm²) | <25 | <25 | | Sintering<br>Temperature | 1500°C | 1500°C | Table 1 6 {7}------------------------------------------------ {8}------------------------------------------------ ### 7. Performance Testing Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the device. - ISO 6872 Fourth edition 2015-06-01 [including AMENDMENT 1 2018-04] Dentistry - Ceramic materials ### Biocompatibility testing The biocompatibility evaluation and testing of the Product name was conducted in accordance with the following standards and guidance, as recognized by the FDA: - . ISO 10993-1:2018(en) Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process - . ISO 10993-3:2014, Biological evaluation of medical device-Part 3: Test for genotoxicity, carcinogenicity and reproductive toxicity. - . ISO 10993-5:2009, Biological evaluation of medical device-Part 5: Tests for in vitro cytotoxicity. - . ISO 10993-10:2010, Biological evaluation of medical device-Part 10: Tests for irritation and skin sensitization. - ISO 10993-11:2017, Biological evaluation of medical device-Part 11: Tests for . systemic toxicity. - ISO 10993-12:2012, Biological evaluation of medical device-Part 12: Sample . preparation and reference materials. ### Mechanical testing Franz Zirconia Dental Crown's mechanical function including Radioactivity Test, Biaxial flexural strength, Linear thermal expansion coefficient test and Fracture toughness tests which were tested and demonstrated that the design specification from design input are fulfilled. Mechanical safety tests were also conducted to demonstrate that the reliability of the device for during use no safety concern. No animal studies or clinical testing have been required for these devices. {9}------------------------------------------------ ### 8. Conclusion Based on the intended use and/or indications for use, technological characteristics, performance testing and comparison to the predicate device, the Franz Zirconia Dental Crown is substantially equivalent to the predicate device and raises no new questions of safety or effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...