CloudVue
K203058 · International Medical Solutions, Inc. · LLZ · Oct 29, 2020 · Radiology
Device Facts
| Record ID | K203058 |
| Device Name | CloudVue |
| Applicant | International Medical Solutions, Inc. |
| Product Code | LLZ · Radiology |
| Decision Date | Oct 29, 2020 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.2050 |
| Device Class | Class 2 |
| Attributes | Software as a Medical Device, 3rd-Party Reviewed |
Intended Use
CloudVue is a software application that displays medical image data to aid in diagnosis for healthcare professionals. It performs operations relating to the transfer, storage, display, and measurement of image data. CloudVue allows users to perform image manipulations, including window/level, rotation, measurement and markup. CloudVue provides 2D display, Multi-Planar Reformatting and 3D visualization of medical image data, and mobile access to images. CloudVue displays both lossy compressed images. For lossy images, the medical professional user must determine if the level of loss is acceptable for their purposes. Display monitors or mobile devices used for reading medical images for diagnostic purposes must comply with applicable regulatory approvals and with quality control requirements for their use and maintenance. For mobile diagnostic usage only when a full workstation is not available. Usage for mammography is for reference and referral only.
Device Story
CloudVue is a web-based medical image viewer; operates as a thin-client application within web browsers (Chrome, Safari, Firefox, Edge) without local installation. System comprises a web viewer and a web server; server communicates with DICOM archives (local or cloud) via DICOM/DICOMweb. Inputs: medical images (CR, CT, DX, HC, IVUS, MR, NM, OP, OPT, OT, PT, SC, US, XA). Processing: image manipulation (window/level, rotation, measurement, markup), 2D display, MPR, MIP, and 3D volume rendering. Output: rendered images displayed on user-selected monitors or mobile devices. Used by healthcare professionals in clinical settings to inform diagnostic decision-making. Benefits: provides fast, secure, portable access to full-fidelity medical images, enabling remote review and diagnostic support when full workstations are unavailable.
Clinical Evidence
No clinical testing was required. Safety and effectiveness were demonstrated through bench testing, including unit, integration, system, and benchmark testing. Portable display devices were validated for image quality.
Technological Characteristics
Software-only, zero-footprint, web-based application. Operates via web browser (thin client). Supports DICOM and DICOMweb communication protocols. Features include 2D/3D visualization (MPR, MIP, VR), image manipulation tools, and mobile access. Data security via HTTPS encryption. Access control integrated or via parent application. Risk management per ISO 14971:2007.
Indications for Use
Indicated for healthcare professionals to display, transfer, store, and measure medical image data to aid in diagnosis. Supports 2D, MPR, and 3D visualization. For use on mobile devices only when a full workstation is unavailable. Mammography use is for reference and referral only.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Related Devices
- K223048 — Medisystem Pacs · Medilab Manutencao E Sitemas Ltda · Jun 23, 2023
- K233226 — FlexView Diagnostic (v1.1.20) · Radical Imaging, LLC · Jan 17, 2024
- K182290 — IBM iConnect Access · Merge Healthcare Incorporated · Sep 25, 2018
- K212984 — MiPACS HTML5 Web Viewer · Lead Technologies, Inc. · Nov 12, 2021
- K183248 — Synapse Enterprise Viewer Version 1.0 · Fujifilm Corporation · Apr 25, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
October 29, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
International Medical Solutions, Inc. % Mr. Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL MN 55114
Re: K203058
Trade/Device Name: CloudVue Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: October 7, 2020 Received: October 8, 2020
Dear Mr. Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K203058
Device Name CloudVue
Indications for Use (Describe)
CloudVue is a software application that displays medical image data to aid in diagnosis for healthcare professionals. It performs operations relating to the transfer, storage, display, and measurement of image data.
CloudVue allows users to perform image manipulations, including window/level, rotation, measurement and markup.
CloudVue provides 2D display, Multi-Planar Reformatting and 3D visualization of medical image data, and mobile access to images.
CloudVue displays both lossy compressed images. For lossy images, the medical professional user must determine if the level of loss is acceptable for their purposes. Display monitors or mobile devices used for reading medical images for diagnostic purposes must comply with applicable regulatory approvals and with quality control requirements for their use and maintenance. For mobile diagnostic usage only when a full workstation is not available.
Usage for mammography is for reference and referral only.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for IMS. The logo consists of a circular design made of three interwoven bands on the left. To the right of the design is the text "IMS" in a bold, dark blue font. Below "IMS" is the tagline "The Power of Partnering" in a smaller, reddish-brown font.
# IMS CloudVue 510(k) Summary
K203058
#### a. Submitter
International Medical Solutions Inc. 2425 Matheson Blvd E Suite 800 Mississauga, ON, L4W 5K4 Canada
Phone: 1-949-232-9298 Email: vcolaco@imstsvc.com
Contact Person: Vernon Colaço, General Manager Date Prepared: August 27, 2020 Submission Date: September 2, 2020
#### b. Device
| Subject Device Information | |
|----------------------------|---------------------------------------------|
| Device Trade Name | CloudVue |
| Device Common Name | Medical Image Processing Software |
| Classification Name | Picture Archiving and Communications System |
| 510(k) Number | K203058 |
| Product Code | LLZ |
| Device Class | Class II |
| Regulation Number | 21 CFR 892.2050 |
| Device Panel | Radiology |
#### c. Predicate Device
| Predicate Device Information | |
|------------------------------|---------------------|
| Device Trade Name | eUnity |
| Manufacturer | Client Outlook Inc. |
| 510(k) Number | K172490 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for IMS. The logo consists of a circular design on the left and the text "IMS" on the right. Below the text "IMS" is the tagline "The Power of Partnering."
The predicate device has not been subject to a design-related recall.
#### d. Device Description
CloudVue is a software application that allows for the display of medical image data within a web browser without installing client software. Conceptually, the application comprises two components: the web viewer and the web server. The server is installed on any computer that meets the minimum system requirements and is configured to communicate with a DICOM archive. The server is capable of DICOM and DICOMweb communication, which allows the connected DICOM archive to exist on a physical server or in the cloud, giving institutions the flexibility to choose a system infrastructure that is best suited to their needs.
CloudVue provides fast and secure access to full-fidelity CR, CT, DX, HC, IVUS, MR, NM, OP, OPT, OT, PT, SC, US, and XA images using Chrome, Safari, Firefox, or Edge web browsers. With interactive features, such as: multi-study viewing, multi-monitor support, customized screen and tool layouts, annotation saving and loading, measurement drawing and calibration, reference lines, key images, multi-planar reformatting (MPR), MIP, VR rendering, and image link sharing, CloudVue is a feature-rich solution that informs diagnostic decision-making by healthcare professionals.
CloudVue also supports these same features across the following devices: Windows 10 and higher, macOS X and higher, iPad Pro 200 generation and higher, and higher. This level of portability allows healthcare professionals to access their medical image data anytime, anywhere.
The display is not part of the subject device, CloudVue, however, is an essential component of a fullyfunctional imaging system. The end user will view the medical images on the display monitor of their choice.
#### e. Indications for Use
CloudVue is a software application that displays medical image data to aid in diagnosis for healthcare professionals. It performs operations relating to the transfer, storage, display, and measurement of image data.
CloudVue allows users to perform image manipulations, including window/level, rotation, measurement and markup.
CloudVue provides 2D display, Multi-Planar Reformatting and 3D visualization of medical image data, and mobile access to images.
CloudVue displays both lossless and lossy compressed images, the medical professional user must determine if the level of loss is acceptable for their purposes. Display monitors or mobile devices used for reading medical images for diagnostic purposes must comply with applicable regulatory approvals and with quality control requirements for their use and maintenance. For mobile diagnostic usage only when a full workstation is not available.
Usage for mammography is for reference and referral only.
{5}------------------------------------------------
## f. Comparison of Technological Characteristics with the Predicate Device
| Characteristic | IMS CloudVue | Client Outlook eUnity |
|------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K203058 | K172490 |
| Regulatory Class | Class II | Class II |
| Subject to Device Recall | No | No |
| Intended Use | CloudVue is a software application that<br>displays medical image data to aid in<br>diagnosis for healthcare professionals. It<br>performs operations relating to the<br>transfer, storage, display, and<br>measurement of image data.<br>CloudVue allows users to perform image<br>manipulations, including window/level,<br>rotation, measurement and markup. | eUnity is a software application that<br>displays medical image data and<br>associated clinical reports to aid in<br>diagnosis for healthcare professionals. It<br>performs operations relating to the<br>transfer, storage, display, and<br>measurement of image data.<br>eUnity allows users to perform image<br>manipulations, including window/level,<br>rotation, measurement and markup. |
| Indications for Use | CloudVue provides 2D display, Multi-<br>Planar Reformatting and 3D visualization<br>of medical image data, and mobile<br>access to images.<br>CloudVue displays both lossless and<br>lossy compressed images. For lossy<br>images, the medical professional user<br>must determine if the level of loss is<br>acceptable for their purposes. Display<br>monitors or mobile devices used for<br>reading medical images for diagnostic<br>purposes must comply with applicable<br>regulatory approvals and with quality<br>control requirements for their use and<br>maintenance. For mobile diagnostic<br>usage only when a full workstation is not<br>available.<br>Usage for mammography is for<br>reference and referral only. | eUnity provides 2D display, Multi-Planar<br>Reformatting and 3D visualization of<br>medical image data, and mobile access<br>to images.<br>eUnity displays both lossless and lossy<br>compressed images. For lossy images,<br>the medical professional user must<br>determine if the level of loss is<br>acceptable for their purposes. Display<br>monitors or mobile devices used for<br>reading medical images for diagnostic<br>purposes must comply with applicable<br>regulatory approvals and with quality<br>control requirements for their use and<br>maintenance. For Mobile diagnostic<br>usage only when a full workstation is not<br>available.<br>Mobile usage for mammography is for<br>reference and referral only. |
| User Installation Requirements | Thin client - no install; runs within<br>browser | Thin client - no install; runs within<br>browser |
| Communications | DICOM, Non-DICOM | DICOM, Non-DICOM, IHE |
| Modalities | CR, CT, DX, HC, IVUS, MR, NM, OP, OPT,<br>OT, PT, SC, US, XA | CR, CT, DX, MR, PT, SC, US, XA, ECG, MG,<br>NM, OP, PR, RF, SR, VL |
| Tools | | |
| Window<br>Level/Rotate/Pan/Zoom/Reset,<br>Presets, Invert, Revert to Original,<br>Image Flip, Image Rotate | Yes | Yes |
| Multi-Study Viewing, Image Export,<br>Image Sharing | Yes | Yes |
| | | |
| Metadata Display/Hide | Yes | Yes |
| Orientation Labels, Keyboard<br>Shortcuts | Yes | Yes |
| Measurements, Annotations | Yes | Yes |
| Full-Screen Mode, Multi-Monitor,<br>Layouts | Yes | Yes |
| Linking Series, Image Scrolling, Linked<br>Scrolling, Reference Lines | Yes | Yes |
| Grayscale Softcopy Presentation<br>States (GSPS) | Yes | Yes |
| Key Images | Yes | Yes |
| Magnification Lens | Yes | Yes |
| MPR/MIP/3D Features | | |
| Multi-Planar Reformat (MPR) | Yes | Yes |
| Oblique and Double Oblique<br>Reformatting | Yes | Yes |
| Maximum Intensity Projection (MIP) | Yes | Yes |
| Volume Rendering (VR) | Yes | Yes |
| Opacity Presets | Yes | Yes |
| Scalpel Tool, Bone Removal | No | Yes |
| Mobile-Specific Features | | |
| Mobile Luminance Check | Yes | Yes |
| sSecurity | | |
| Data Encryption | HTTPS | HTTPS |
| Data Security | Stored on server | Stored on server |
| Access Control | Built-in access control or parent<br>application access control | Built-in access control or parent<br>application access control |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for IMS. The logo consists of a circular design on the left and the text "IMS The Power of Partnering" on the right. The circular design is made up of three curved shapes that intersect in the center. The text "IMS" is in a bold, sans-serif font, and the text "The Power of Partnering" is in a smaller, serif font.
#### g. Clinical Performance Data
No clinical testing was required to support safety and effectiveness of the subject device.
#### h. Summary of Testing
Verification and validation activities, driven by risk analysis guided by ISO 14971:2007, were conducted and documentation was provided. These activities included unit testing, integration testing, system testing, and benchmark testing. The recommended portable display devices were all tested for image quality and found to be adequate. All testing activities demonstrated that the device met all design requirements and intended use, and that it is both safe and effective.
#### i. Conclusion
In summary, the predicate device is medical image processing software that provides its users with the tools to display and manipulate medical DICOM images without altering the original image itself. The
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for IMS. The logo consists of a circular design on the left and the text "IMS" on the right. Below the text "IMS" is the tagline "The Power of Partnering".
predicate device is a zero-footprint, software-only medical device that aids in, but does not provide, a diagnosis. It also displays diagnostic quality images for both workstations and mobile platforms. These technological characteristics that comprise the predicate device are substantially similar to the subject device. Therefore, we conclude that the intended use and technological characteristics of CloudVue are substantially equivalent to those of the predicate device, and do not pose any new issues of safety or effectiveness.