InnoClip™ Disposable Clip Applier is indicated for patients undergoing laparoscopic surgical procedures involving occlusion of blood vessel, ducts and other tubular structures, and for radiographic marking. InnoClip™ Clip Applier is indicated for patients undergoing surgical procedures involving occlusion of blood vessel, ducts and other tubular structures, and for radiographic marking.
Device Story
InnoClip™ devices are surgical instruments for vessel/duct occlusion and radiographic marking during laparoscopic procedures. The system comprises a handle, trigger, 360° rotational knob, shaft, and jaws. The disposable version is a single-unit device; the reusable version features a detachable handle with a disposable clip cartridge. Surgeons operate the device through 5 mm trocar sleeves. By pulling the trigger, pre-loaded clips are advanced to the jaws to enclose and occlude target vessels. The multi-fire design allows for 10 or 16 clips per device without removal. The reusable handle is validated for 300 reprocessing cycles. The device provides surgeons with a mechanism to secure vessels, reducing surgical waste and cost.
Clinical Evidence
No human clinical data. Evidence includes in-vitro bench testing (silicone tube substrate) for pull-out force, air leakage, and clip scissoring, and in-vivo ligation testing in a swine model. Reusable handle validated for 300 reprocessing cycles.
Technological Characteristics
Mechanical clip applier; 5 mm diameter form factor; 360° rotational knob; multi-fire mechanism (10 or 16 clips). Materials evaluated per ISO 10993-1. Reusable handle validated for 300 reprocessing cycles.
Indications for Use
Indicated for patients undergoing laparoscopic surgical procedures requiring occlusion of blood vessels, ducts, or other tubular structures, and for radiographic marking.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating food and drugs in the United States.
January 4, 2022
Taiwan Surgical Corporation Ken Chen Project Director 3F., No.12, Sec.2, ShengYi Rd. Zhubei City, Hsinchu County 30261 Taiwan
Re: K202994
Trade/Device Name: InnoClip Disposable Clip Applier, InnoClip Clip Applier Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: FZP, GDO Dated: July 6, 2021 Received: July 7, 2021
Dear Ken Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Deborah Fellhauer RN, BSN Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202994
Device Name InnoClip™ Disposable Clip Applier InnoClip™ Clip Applier
### Indications for Use (Describe)
InnoClip™ Disposable Clip Applier is indicated for patients undergoing laparoscopic surgical procedures involving occlusion of blood vessel, ducts and other tubular structures, and for radiographic marking.
The InnoClip™ Clip Applier is indicated for patients undergoing laparoscopic surgical procedures involving occlusion of blood vessel, ducts and other tubular structures, and for radiographic marking.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:12pt"> "> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span style="font-size:12pt"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
The Assigned 510(k) Number: _ K202994
This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92
### 1 SUBMITTER:
| Submitter: | TAIWAN SURGICAL CORPORATION |
|------------------|------------------------------------------------------------------|
| Mailing Address: | 3F., No.12, Sec.2, ShengYi Rd., Zhubei City, Hsinchu County 302- |
| | Taiwan |
| Phone Number: | +886-3-6588129 |
| Fax Number: | +886-3-6588355 |
| Date Prepared: | December 8, 2021 |
| Contact Person: | Ken Chen |
| Email: | ra@twsc.com.tw |
| Phone Number: | +886-3-6588129 ext. 519 |
| Fax Number: | +886-3-3588355 |
### 2 DEVICE
| Trade Name: | 1. InnoClip™ Disposable Clip Applier |
|-------------------------------|--------------------------------------|
| | 2. InnoClip™ Clip Applier |
| Common Name: | Clip applier and implantable clip |
| Classification Panel: | 79 General and Plastic Surgery |
| Regulation Number: | 878.4300 |
| Classification Product Code: | FZP |
| Device Class: | II |
| Additional Regulation Number: | 878.4800 |
| Additional Product Code: | GDO |
| Additional Device Class: | I |
#### 3 PREDICATE DEVICE
ENDO CLIP™ III 5mm Clip Applier (K121194)
{5}------------------------------------------------
#### DEVICE DESCRIPTION 4
## InnoClip™ Disposable Clip Applier
InnoClip™ Disposable Clip Applier consists of a molded handle and trigger, a 360° rotational knob, a cartridge housing shaft and a pair of jaws which is easy for surgeon to hold the operation field in the natural position. This device includes models DC5T1-BX and DC5T1-B6 that are designed to be inserted through 5 mm trocar sleeves.
InnoClip™ Disposable Clip Applier can be used on vessels or tubular structures. The multi-fire design of InnoClip™ Disposable Clip Applier enables the surgeon to apply a maximum number of 10 clips for DC5T1-BX or 16 clips for DC5T1-B6 during a laparoscopic procedure without withdrawing and reinserting the device. The pre-loaded clips in the shaft are advanced to the jaw to enclose and occlude the target vessel by gently pulling the trigger.
### InnoClip™ Clip Applier
InnoClip™ Clip Applier has two parts, Reusable Handle and a Disposable Clip Cartridge. The Reusable Handle consists of a molded handle, trigger, openable cover and a 360° rotational knob. The Disposable Clip Cartridge consists of a cartridge housing shaft and a pair of jaws which provide secured of the clip to the targeted vessel. The Disposable Clip Cartridge and the Reusable Handle can be easily assembled together by inserting the Disposable Clip Cartridge through a slot located in the Reusable Handle. This device includes models RC5T1-BX and RC5T1-B6 that are designed to be inserted through 5 mm trocar sleeves.
InnoClip™ Clip Applier can be used on vessels or tubular structures. The multi-fire design of InnoClip™ Clip Applier enables the surgeon to apply a maximum number of 10 clips for RC5T1-BX or 16 clips for RC5T1-B6 during a laparoscopic procedure without withdrawing and reinserting the device. The reusable handle design also reduces the surgical cost and waste. The pre-loaded clips in the Disposable Clip Cartridge are advanced to the jaw to enclose and occlude the target vessel by gently pulling the trigger.
### INTENDED USE 5
InnoClip™ Disposable Clip Applier is indicated for patients undergoing laparoscopic surgical procedures involving occlusion of blood vessel, ducts and other tubular structures, and for radiographic marking.
InnoClip™ Clip Applier is indicated for patients undergoing surgical procedures involving occlusion of blood vessel, ducts and other tubular structures, and for radiographic marking.
{6}------------------------------------------------
#### MATERIAL 6
All patient contacting materials, including those with indirect patient contact, have been evaluated according to ISO 10993-1 and the FDA's Guidance Use of International Standard ISO 10993-1, dated June 16, 2016. All biocompatibility met the acceptance criteria.
### TECHNOLOGICAL CHARACTERISTICS 7
A comparison of the technological characteristics of the proposed InnoClip™ Disposable Clip Applier InnoClip™ Clip Applier, and the predicate has been performed. The results of this comparison demonstrate that the InnoClip™ Disposable Clip Applier and InnoClip™ Clip Applier have similar technological characteristics compared to the marketed predicate device.
### PERFORMANCE 8
In-vitro tests conducted on a silicone tube substrate including clipped pull out force test and air leakage test were performed on InnoClip™ Disposable Clip Applier and InnoClin™ Clip Applier to compared with the predicate device. The test results showed that both proposed devices have similar device performance compared to the predicate device.
In-vivo ligation tests were performed in a swine model using the InnoClip™ Disposable Clip Applier. InnoClip™ Clip Applier and the predicate device, and the results demonstrated that the proposed devices and the predicate device can ligate the target vessels and tubular tissue as its intended use.
Comparison testing was conducted on a silicone tube substrate with the InnoClip™ Disposable Clip Applier, InnoClip TM Clip Applier and predicate device to assess clip scissoring. The proposed devices met all acceptance criteria and performed as well as the predicate device.
The reusable handle of InnoClip™ Clip Applier has been validated to maintain performance functions for up to 300 times following the reprocessing methods described in the instructions for use.
### 9 CONCLUSIONS
Based on the intended use and/or indications for use, technological characteristics, performance testing and comparison to the predicate device, the InnoClip™ Disposable Clip Applier and InnoClip™ Clip Applier are substantially equivalent to the predicate device Endo Clip™ III.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.