← Product Code [MXK](/productcode/MXK) · K202989

# Myopia Master (K202989)

_Oculus Optikgeräte GmbH · MXK · Jul 14, 2021 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K202989

## Device Facts

- **Applicant:** Oculus Optikgeräte GmbH
- **Product Code:** [MXK](/productcode/MXK.md)
- **Decision Date:** Jul 14, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1850
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The Myopia Master is an interferometer indicated for measuring the axial length of the eye and is intended as an aid to eye care providers.

## Device Story

Myopia Master integrates axial length measurement, autorefractometry, and keratometry into one unit. Device uses Michelson interferometer for axial length; infrared light source and CCD camera for autorefractometry; reflected image analysis for keratometry. Used in clinical settings by eye care providers. User aligns device to patient via joystick and chin/forehead rest; software detects release position to trigger automatic examination. Output includes axial length, refractive power (sphere, cylinder, axis), corneal curvature, white-to-white distance, and pupil size. Data aids eye care providers in patient assessment. Device does not provide diagnostic readouts.

## Clinical Evidence

Bench testing only. Evaluated for electrical safety and EMC (IEC 60601-1, IEC 60601-1-2) and performance standards (ISO 15004-1, ISO 15004-2, ISO 2265, IEC 60825-1). Software verification and validation performed per FDA guidance. No clinical study data provided for diagnostic accuracy in pathological eyes.

## Technological Characteristics

AC-powered slitlamp biomicroscope. Components: Michelson interferometer (880 nm IR SLD), autorefractometer (850 nm IR LED), keratometer (940 nm IR LED), overview camera (CCD). Connectivity: Standalone unit with integrated microcomputer. Software: Moderate level of concern. Sterilization: Not applicable (non-invasive).

## Regulatory Identification

An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.

## Special Controls

*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

## Predicate Devices

- OCULUS Pentacam AXL ([K152311](/device/K152311.md))
- OCULUS PARK 1 ([K073508](/device/K073508.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

July 14, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

OCULUS Optikgeräte GmbH % Randy Prebula Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington, District of Columbia 20004

Re: K202989

Trade/Device Name: Myopia Master Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Biomicroscope Regulatory Class: Class II Product Code: MXK, HJO Dated: June 3, 2021 Received: June 3, 2021

Dear Randy Prebula:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| Indications for Use Statement |  |  |
|-------------------------------|--|--|
|                               |  |  |

510(k) Number (if known): __ K202989_

Device Name: Myopia Master

Indications for Use:

The Myopia Master is an interferometer indicated for measuring the axial length of the eye and is intended as an aid to eye care providers.

Prescription Use _ × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

### 510(k) SUMMARY

## OCULUS Myopia Master

### General Information

### Applicant:

OCULUS Optikgeräte GmbH Müncholzhäuserstr. 29 35582 Wetzlar Germany Phone: +49(0)641 2005-0 +49(0)641 2005-255 Fax:

## Contact Person:

Mr. Eckhard Loh Head of Quality and Regulatory Affairs OCULUS Optikgeräte GmbH Müncholzhäuserstr. 29 35582 Wetzlar Germany Phone: +49 (0) 641 2005-0 Fax: +49 (0) 641 2005-255

Summary Prepared: July 13, 2021

## Device Information

| Classification Name:  | Device, analysis, anterior segment<br>(AC-powered slitlamp biomicroscope,) |
|-----------------------|----------------------------------------------------------------------------|
| Trade/Propriety Name: | Myopia Master                                                              |
| Common Name:          | Refractometer                                                              |
| Regulation Number:    | 21 C.F.R. § 886.1850                                                       |
| Device class:         | II                                                                         |
| Product Code:         | MXK, HJO                                                                   |

## Predicate Devices

OCULUS Optikgeräte GmbH, Pentacam AXL (K152311) OCULUS Optikgeräte GmbH, PARK 1 (K073508)

{4}------------------------------------------------

### Intended Use / Indications for Use

The Myopia Master is an interferometer indicated for measuring the axial length of the eye and is intended as an aid to eye care providers.

## Caution:

The measurement values of the Myopia Master are not intended to be used for refractive surgery planning (e.g. IOL calculations)

Only eyes without any ocular disease were evaluated during the clinical study performed for FDA clearance of this device, so it is unknown whether accuracy and precision when used in patients with ocular pathology will yield acceptable results. Users should interpret data cautiously when assessing eyes with ocular pathology (e.g. eyes with cataracts, corneal pathology, or post surgical complications.

## Product Description/Technological Characteristics

The OCULUS Myopia Master integrates the axial length measurement function of the cleared OCULUS Pentacam AXL (K152311) into the cleared PARK 1 device (K073508), which is an ocular device that includes Scheimpflug imaging, autorefractometry and keratometry functionalities.

The Myopia Master combines the following measuring functions in one unit: Axial length, Auto-Refractometer, Keratometer.

### Auto-Refractometer

An infrared light source projects measuring light onto the retina of the eye, from which it is reflected back to the shutter location. Sensitive sensor chips, or charge-coupled device (CCD) cameras then reqister the deviation of the reflected light from the shutter location. The deviation depends on the ametropia. From that, an integrated microcomputer calculates the ametropia in diameter, based on the sphere, cylinder and cylinder axis position.

### Keratometer

To determine the curvature of the cornea, a reflected image of the cornea is captured by a camera sensor. The reflection of test marks and of a ring is used as the reflected image, which allows the central radii of the cornea to be measured.

### Axial length

The axial length of the eye is measured and displayed by a built-in Michelson interferometer.

### White-to-white

The white-to-white is measured by analyzing the overview image at the end of the alignment procedure.

The image of the eye is processed by software. Edge detection provides the transition of the white scleral tissue to darker appearing cornea/iris.

By converting the pixel positions of the detected edges into millimeters, the measured value of the "white-to-white" is achieved.

{5}------------------------------------------------

## Pupil

The pupil is measured by analyzing the overview image at the end of the alignment procedure. The image of the eye is processed by software. Edge detection provides the transition of the black pupil to brighter appearing cornea/iris.

By converting the pixel positions of the detected edges into millimeters, the measured value of the pupil is achieved.

## Performance Data

EMC and electrical safety of the subject devices were evaluated using the following consensus standards: IEC 60601-1; IEC 60601-1-2.

Testing was also performed in accordance with ISO 15004-1, ISO 15004-2, ISO 2265 and IEC 60825-1 to demonstrate compliance with these standards.

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.

## Substantial Equivalence Discussion

As explained below, the Myopia Master is substantially equivalent to other legally marketed ophthalmic devices. Specifically, the Myopia Master is substantially equivalent to the predicate devices PARK 1 (K073508) and Pentacam AXL (K152311).

The Myopia Master has the same general intended use and similar indications, technological characteristics, and principles of operation as the previously cleared predicate devices. Minor differences in the technological characteristics do not raise new questions of safety or efficacy. Bench and Clinical testing demonstrate that the Myopia Master is as safe and effective as its predicate devices. The inclusion of two predicates is appropriate, as all of the devices are generally intended for diagnosis in optometry and cleared under the same regulation 21 CFR 886.1850 addressing AC-powered slitlamp biomicroscopes. As described below, each of the measurements for the Myopia Master device is performed individually and is substantially equivalent in terms of technology, safety and performance to the two identified predicates.

A substantial equivalence chart comparing the similarities and differences between the Myopia Master and its predicate devices is provided below. Bench and clinical testing demonstrate that the Myopia Master is as safe and effective as its predicate devices.

#### 1. Intended Use/ Indications for Use

### Indications for Use Statement for Myopia Master

The Myopia Master is an interferometer indicated for measuring the axial length of the eye and is intended as an aid to eye care providers.

### Indications for Use Statement for PARK 1

The PARK 1 is designed to photograph the eye and take Scheimpflug images of the anterior segment to evaluate the thickness of the cornea. The integrated keratometer measures the

{6}------------------------------------------------

central radii of the cornea. The implanted ophthalmic refractometer measures the refractive power of the eye.

### Indications for Use Statement for Pentacam AXL

The Pentacam AXL is designed to take photos of the anterior segment of the eye which includes the cornea, pupil, anterior chamber and lens of the eye. To evaluate:

- corneal shape,
- analyze condition of the lens (opaque crystalline lens), ●
- . analyze the anterior chamber angle,
- analyze anterior chamber depth, ●
- . analyze the volume of the anterior chamber,
- analyze anterior or posterior cortical opacity,
- . analyze the location of cataracts (nuclear, sub capsular and or cortical), using cross slit imaging with densitometry,
- . corneal thickness,
- axial length (by optical biometry),
- white-to-white distance.

The Pentacam AXL also performs calculations to assist physicians in determining the power of the intraocular lens for implantation.

The Myopia Master has nearly the same intended use as its predicate devices (to photograph the eye to measure the central radii of the cornea and the refractive power of the eye (from PARK 1) and the axial length (by optical biometry) (from the Pentacam AXL). The capabilities of the device remain the same as its predicate devices, and the device does not provide any diagnostic readouts or information. Importantly, the overall intended use of the device relative to its predicates is not modified. This intended use reflects, in practicality, the manner in which the device can be used as certain measurements are taken. The Myopia Master thus meets the first requirement for establishing substantial equivalence.

#### 2. Technological Characteristics

The Myopia Master combines certain measurement capabilities of the cleared PARK 1 and the Pentacam AXL.

The combination of the predicates' two cleared technologies in one device does not impact the ability of either technology to operate per its separate intended use. The component technologies for the different measurements have already been used for the same purposes for which they are used in the subject device. In particular, the device incorporates the functionality of an autorefractometer, keratometer and Michelson interferometer for axial length measurement in a single device.

Like the PARK 1, the Myopia Master consists of a measuring head which is attached to a control unit on a cross slide to align the device to the patient via joystick by the user. The patient is positioned in front of the device in a chin- and forehead rest. Both devices also use IR LEDs for eye illumination, blue slit lamps as the light source for slit illumination, and a digital CMOS camera as the overview camera.

Minor differences, such as the observation illumination being at 810 mm rather than 840 mm, and the pulse width for axial length measurements being 520 ms rather than 400 ms, do not raise new questions of safety and effectiveness because the measurement range continues to be appropriate for the relevant patient group that is to be measured using the device.

{7}------------------------------------------------

The software for the Myopia Master is based on the software utilized with the PARK 1 predicate device. Additionally, the Myopia Master software includes the ability to measure and display axial length measurement values. The algorithms and functions for measuring, keratometry and refraction determination are unchanged from the PARK 1, while the algorithms and functions for measuring the axial length were adopted from the Pentacam AXL software.

The storage and internal communication functionalities, as well as the input and output options and functionalities, are identical to those included with the PARK 1 software, and the hardware on which the software runs is identical to that in the predicate devices. There are only minor differences in the viewing of the user interface (e.g., color changes). Certain input options have been modified to include both text and symbols.

Thus, no new or different questions of safety or effectiveness are raised by the technological characteristics of the subject device when compared to the predicates. The differences between the subject device and the cleared predicates are minor, and bench and clinical tests have confirmed the safety and effectiveness of the new device in achieving the indications for use.

#### 3. Principles of Operation

Like its predicates, the Myopia Master is intended for eye examination. The Myopia Master combines the Keratometry measurement and autorefractometry measurement functionalities of PARK 1 and the axial length measurements functionality of the Pentacam AXL.

As with the PARK 1, the alignment of the device to the patient is done using an overview camera which records the eye. The images which are recorded are then analyzed by the software. Deviations from the release position of a measurement are displayed in the scan menu of the Myopia Master software. Indicators are displayed for manual alignment of the device. If the software detects the release position, the examination starts automatically.

#### 1. PARK 1 (exempted parameters) and Myopia Master

Other Functions

Keratometer (Class I, 510(k) exempt)

Both the cleared PARK 1 and the Myopia Master determine the curvature of the cornea using a reflected cornea image which is captured by a camera sensor.

The reflections of projected spots and of a ring are in the image are analyzed by image processing. This allows the central radii of the cornea to be determined.

### Autorefractometry (Class I, 510(k) exempt)

Both the cleared PARK 1 and the Myopia Master use infrared light source projections measuring light spots onto the retina of the eye from where it is reflected back to the shutter location. A CCD camera then captures the deviation of the reflected light from the shutter location. The deviation depends on the ametropia.

From that point, an integrated microcomputer calculates the ametropia in D, based on the sphere, cylinder and cylinder axis position.

{8}------------------------------------------------

#### Pentacam AXL and Myopia Master 2.

## Axial length measurement

Both the Pentacam AXL and the Myopia Master conduct axial length measurements using a builtin Michelson interferometer (optical biometry). The Michelson interferometer is a common configuration for optical interferometry. Using a semi-transparent mirror (beam splitter), the light from an infrared super luminescence diode is split into two perpendicular beams and brought to interfere after successive reflections. Both beams are reflected back towards the beam splitter which then combines their amplitudes interferometrically.

The resulting interference pattern is directed to a photoelectric detector.

## Other Functions

White-to-white (Class I, 510(k) exempt)

The white-to-white is measured by analyzing the overview image at the end of the alignment procedure.

The image of the eye is processed by software. Edge detection provides the transition of the white scleral tissue to darker appearing cornea/iris.

By converting the pixel positions of the detected edges into millimeters, the measured value of the "white-to-white" is achieved.

Pupil (Class I, 510(k) exempt)

The pupil is measured by analyzing the overview image at the alignment procedure. The image of the eye is processed by software. Edge detection provides the transition of the black pupil to brighter appearing cornea/iris.

By converting the pixel positions of the detected edges into millimeters, the measured value of the pupil is achieved.

#### 3. Procedure for Use

The Myopia Master and both of its predicates use a primary examination device to examine and provide data regarding the eye measurements. An embedded computer is used to control the device and to display the measurement results.

#### 4. Conclusion

The Myopia Master has the same intended use as the PARK 1 and the Pentacam AXL and has similar indications, technological characteristics and principles of operation as both predicates. There are minor differences between the Myopia Master and its predicates, however, these differences do not raise any new or different questions of safety or effectiveness, because the new device does not add any technological characteristics that are not present in one or both of the predicate devices that are not interfered with as a result of the other features that are incorporated into the subject device. Furthermore, testing performed to evaluate the device demonstrated comparable safety and efficacy to the predicates. Thus, the Myopia Master is substantially equivalent to the PARK 1 and the Pentacam AXL.

{9}------------------------------------------------

## Substantial Equivalence Table (Comparing the Technological Characteristics of the Subject Device with the Relevant Predicate Device)

| Device Type                                                               | Applicant device                                                                                                                                     | Predicate device                                                                                                                                                                                                                                                                                                    |
|---------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                           |                                                                                                                                                      | (for exempted parameters)                                                                                                                                                                                                                                                                                           |
| Model                                                                     | Myopia Master                                                                                                                                        | PARK 1                                                                                                                                                                                                                                                                                                              |
| Manufacturer name                                                         | OCULUS Optikgeräte GmbH                                                                                                                              | OCULUS Optikgeräte GmbH                                                                                                                                                                                                                                                                                             |
| 510(k) Number                                                             | K202989                                                                                                                                              | K073508                                                                                                                                                                                                                                                                                                             |
| Intended use                                                              | The Myopia Master is an interferometer<br>indicated for measuring the axial length<br>of the eye and is intended as an aid to<br>eye care providers. | The PARK 1 is designed to photograph<br>the eye and take Scheimpflug images of<br>the anterior segment to evaluate the<br>thickness of the cornea. The implanted<br>keratometer measures the central radii<br>of the cornea. The implanted ophthalmic<br>refractometer measures the refractive<br>power of the eye. |
| Light source for<br>alignment system                                      | IR LED 950 nm (continuous)                                                                                                                           | IR LED 950 nm (continuous)                                                                                                                                                                                                                                                                                          |
|                                                                           | <0.1 mW/cm² (unweighted corneal and lenticular<br>infrared radiation irradiance)                                                                     | <0.1 mW/cm² (unweighted corneal and lenticular<br>infrared radiation irradiance)                                                                                                                                                                                                                                    |
|                                                                           | <0.05 W/cm² (weighted retinal visible and infrared<br>radiation thermal irradiance)                                                                  | <0.05 W/cm² (weighted retinal visible and infrared<br>radiation thermal irradiance)                                                                                                                                                                                                                                 |
| Light source for fixation                                                 | White LED (continuous)                                                                                                                               | White LED (continuous)                                                                                                                                                                                                                                                                                              |
| target illumination<br>(balloon slide background<br>light)                | <30 cd                                                                                                                                               | <30 cd                                                                                                                                                                                                                                                                                                              |
| Light source for<br>peripheral illumination /<br>Observation illumination | IR LED 810 nm (clocked)                                                                                                                              | IR LED 840 nm (clocked)                                                                                                                                                                                                                                                                                             |
|                                                                           | <50 J/cm² (weighted retinal visible and infrared<br>radiation radiant exposure for t=200s)                                                           | <50 J/cm² (weighted retinal visible and infrared<br>radiation radiant exposure t=200s)                                                                                                                                                                                                                              |
| Light source for axial<br>length measurement                              | IR SLD 880 nm (continuous)                                                                                                                           | Not included                                                                                                                                                                                                                                                                                                        |
|                                                                           | <0.1 W/cm² (unweighted anterior segment visible and<br>infrared radiation irradiance)                                                                |                                                                                                                                                                                                                                                                                                                     |
|                                                                           | <0.05 W/cm² (weighted retinal visible and infrared<br>radiation thermal irradiance)                                                                  |                                                                                                                                                                                                                                                                                                                     |
| Light source for<br>Autorefractometer                                     | IR LED 850 nm (clocked)                                                                                                                              | IR LED 880 nm (clocked)                                                                                                                                                                                                                                                                                             |
|                                                                           | <0.1 J/cm² (weighted retinal visible and infrared<br>radiation radiant exposure)                                                                     | <0.1 J/cm² (weighted retinal visible and infrared<br>radiation radiant exposure)                                                                                                                                                                                                                                    |
| Keratometer light source                                                  | IR LED 940 nm (clocked)                                                                                                                              | IR LED 940 nm (clocked)                                                                                                                                                                                                                                                                                             |
|                                                                           | <1x10-5 J/cm² (weighted retinal visible and infrared<br>radiation radiant exposure)                                                                  | <1x10-5 J/cm² (weighted retinal visible and infrared<br>radiation radiant exposure)                                                                                                                                                                                                                                 |
| Device Type                                                               | Applicant device                                                                                                                                     | Predicate device                                                                                                                                                                                                                                                                                                    |
|                                                                           | <1x10-4 J/cm² (unweighted corneal and lenticular infrared radiation radiant exposure)                                                                | <1x10-4 J/cm² (unweighted corneal and lenticular infrared radiation radiant exposure)                                                                                                                                                                                                                               |
| Camera                                                                    | CCD-Camera                                                                                                                                           | CCD-Camera                                                                                                                                                                                                                                                                                                          |
| Display                                                                   | TFT on control unit                                                                                                                                  | TFT on control unit                                                                                                                                                                                                                                                                                                 |
| Image resolution                                                          | 640 x 480 pixel                                                                                                                                      | 640 x 480 pixel                                                                                                                                                                                                                                                                                                     |
| Measuring points                                                          | 600 per image                                                                                                                                        | 600 per image                                                                                                                                                                                                                                                                                                       |
| Image size                                                                | 4.8 mm x 3.6mm                                                                                                                                       | 4.8 mm x 3.6 mm                                                                                                                                                                                                                                                                                                     |
| Power supply                                                              | External PSU                                                                                                                                         | External PSU                                                                                                                                                                                                                                                                                                        |
|                                                                           | GSM60B15-P1J Medical power adapter<br>Input: 80- 264 V AC; 47 – 63 Hz<br>Output: 15 V DC, 4 A                                                        | GSM60B15-P1J Medical power adapter<br>Input: 80- 264 V AC; 47 – 63 Hz<br>Output: 15 V DC, 4 A                                                                                                                                                                                                                       |
| Power consumption                                                         | 25 W                                                                                                                                                 | 25 W                                                                                                                                                                                                                                                                                                                |
| Protection class                                                          | 2                                                                                                                                                    | 2                                                                                                                                                                                                                                                                                                                   |
| Protection type                                                           | IP20                                                                                                                                                 | IP20                                                                                                                                                                                                                                                                                                                |
| Applied part type                                                         | B                                                                                                                                                    | B                                                                                                                                                                                                                                                                                                                   |
| Dimensions (W x D x H)                                                    | 266 x 538 x 493 – 523 mm                                                                                                                             | 266 x 538 x 493 – 523 mm                                                                                                                                                                                                                                                                                            |
| Weight                                                                    | 12,5 kg                                                                                                                                              | 12 kg                                                                                                                                                                                                                                                                                                               |
| Device Type                                                               | Applicant device                                                                                                                                     | Predicate device                                                                                                                                                                                                                                                                                                    |
|                                                                           |                                                                                                                                                      | (for exempted parameters)                                                                                                                                                                                                                                                                                           |
| Picture                                                                   | Image: Applicant device                                                                                                                              | Image: Predicate device                                                                                                                                                                                                                                                                                             |

{10}------------------------------------------------

{11}------------------------------------------------

# Substantial Equivalence Table (Comparing the Axial Length (Biometer) Technological Characteristics of the Subject Device with the Relevant Predicate Device)

| Device Type       | Applicant device                                                                                                                                     | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                   |                                                                                                                                                      | (for axial length measurement function)                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Model             | Myopia Master                                                                                                                                        | Pentacam AXL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Manufacturer name | OCULUS Optikgeraete GmbH                                                                                                                             | OCULUS Optikgeraete GmbH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| 510(k) Number     | K202989                                                                                                                                              | K152311                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                   | The Myopia Master is an interferometer<br>indicated for measuring the axial length<br>of the eye and is intended as an aid to<br>eye care providers. | The Pentacam AXL is designed to take<br>photos of the anterior segment of the<br>eye which includes the cornea, pupil,<br>anterior chamber and lens of the eye. To<br>evaluate:                                                                                                                                                                                                                                                                                                                                 |
|                   |                                                                                                                                                      | - corneal shape,<br>- analyze condition of the lens (opaque<br>crystalline lens),<br>- analyze the anterior chamber angle,<br>- analyze anterior chamber depth,<br>- analyze the volume of the anterior<br>chamber,<br>- analyze anterior or posterior cortical<br>opacity,<br>- analyze the location of cataracts<br>(nuclear, sub capsular and or<br>cortical), using cross slit imaging with<br>densitometry,<br>- corneal thickness,<br>- axial length (by optical biometry),<br>- white-to-white distance. |

{12}------------------------------------------------

| Device Type                              | Applicant device                  | Predicate device                                                                                                                                                              |
|------------------------------------------|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                          |                                   | (for axial length measurement function)<br>The Pentacam AXL also performs calculations to assist physicians in determining the power of the intraocular lens for implantation |
| Measurement range -<br>Axial length      | 14 - 40 mm                        | 14 - 40 mm                                                                                                                                                                    |
| Light source for<br>interferometer       | IR Super luminescence diode (SLD) | IR Super luminescence diode (SLD)                                                                                                                                             |
| Wavelength                               | 880 nm                            | 880 nm                                                                                                                                                                        |
| SLD-Power for<br>measurement             | 0.7 mW                            | 0.84 mW                                                                                                                                                                       |
| SLD-Power for<br>alignment               | None                              | None                                                                                                                                                                          |
| Pulse width                              | 520 ms                            | 400 ms                                                                                                                                                                        |
| IEC 60825-1<br>classification            | Class 1 laser product             | Class 1 laser product                                                                                                                                                         |
| Embedded laser class<br>(not accessible) | 3R                                | 3R                                                                                                                                                                            |

---

**Source:** [https://fda.innolitics.com/device/K202989](https://fda.innolitics.com/device/K202989)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
