ELEC Master & ELEC Master Dual are intended for use in general dental applications such as: cutting a tooth for cavity preparation, crown preparation, crown finishing, inlay, filing, polishing, prophylaxis and endodontic treatment.
Device Story
Device consists of micro motor controller and display panel; controls dental handpiece motor speed, rotation direction, and LED light. Operated by dental professionals in clinical settings. Input: user settings via display panel; Output: electrical power to motor(s) (1 motor for ELEC Master, 2 for Dual). Controller manages motor circuit; provides torque up to 3Ncm and speeds up to 40,000 RPM. Healthcare provider uses device to drive dental handpieces for various restorative and endodontic procedures. Benefits include precise control of handpiece parameters during dental treatment.
Clinical Evidence
Bench testing only. Testing included ISO 3964 (coupling dimensions), ISO 7494-1 (dental units), ISO 14457 (handpieces/motors), IEC 60601-1/1-2 (electrical safety/EMC), IEC 80601-2-60 (dental equipment safety), and IEC 62366 (usability). Sterilization and software validation were also performed. No clinical data provided.
Technological Characteristics
Electronic-micromotor system; control unit with hose and electrical motor. Max torque 3Ncm; max speed 40,000 RPM. White LED illumination. E-type handpiece coupling per ISO 3964. Sterilization: user-performed steam sterilization. Connectivity: standalone controller. Software: embedded control logic.
Indications for Use
Indicated for general dental applications including tooth cutting for cavity/crown preparation, crown finishing, inlay, filing, polishing, prophylaxis, and endodontic treatment in dental patients.
Regulatory Classification
Identification
A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 5, 2023
Mgnewton Ltd. % Priscilla Chung Official Correspondent LK Consulting Group USA, Inc. 1150 Roosevelt, Suite 200 Irvine, California 92620
Re: K202963
Trade/Device Name: ELEC Master, ELEC Master Dual Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EBW Dated: April 7, 2023 Received: April 7, 2023
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bobak Shirmohammadi -S
For Michael Adjodha, M.ChE.,COIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name ELEC Master & ELEC Master Dual
#### Indications for Use (Describe)
ELEC Master & ELEC Master Dual are intended for use in general dental applications such as: cutting a tooth for cavity preparation, crown preparation, crown finishing, inlay, filing, polishing, prophylaxis and endodontic treatment.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 10px;"> <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10px;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary K202963
This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92
Date: SEP 29, 2020
#### 1. 510K Applicant / Submitter:
MGNEWTON LTD. 472, Hanjanggun-ro, Jain-myeon, Gyeongsan-si, Gyeongsangbuk-do, Republic of Korea Tel: +82-53-214-6287 Fax: +82-70-7469-2074
### 2. Submission Contact Person
LK Consulting Group USA, Inc. 1150 ROOSEVELT, SUITE 200 Irvine , CA 92620 Priscilla Chung Phone: 714 2025789 Ext Email: Juhee.C(@)Lkconsultinggroup.com
### 3. Subject Device
- -. Trade Name : ELEC Master & ELEC Master Dual
- -. Classification Name : Controller, Foot, Handpiece and Cord
- -. Regulation Number : 21 CFR 872.4200
- -. Regulation Name : Dental handpiece and Accessories
- -. Regulatory Class : I
- -. Product Code : EBW
## 4. Predicate Device
- -. Trade Name : A-dec NLZ electric motor system
- -. 510(k) Number : K163131
- -. Regulation Number : 21 CFR 872.4200
- -. Regulation Name : Dental handpiece and Accessories
- -. Regulatory Class : I
- -. Product Code : EBW
#### 5. Description:
This product consists of a micro motor controller and display panel. Receiving rated input of AC 24V, this device controls motor speed, and rotation direction through the motor control circuit inside the controller. Motor speed, light and rotation direction functions can be set on the display panel.
| Micromotor Model Name | ELEC Master | ELEC Master Dual |
|-------------------------|-------------|------------------|
| Optic(White LED) | O | O |
| Torque | Max. 3Ncm | Max. 3Ncm |
| Number of motor(EML40W) | 1 | 2 |
| RPM(Gear 1: 1) | Max. 40,000 | Max. 40,000 |
{4}------------------------------------------------
| Direction | Clock / counter clock | Clock / counter clock |
|-----------|-----------------------|-----------------------|
|-----------|-----------------------|-----------------------|
## 6. Indications for Use
ELEC Master & ELEC Master Dual are intended for use in general dental applications such as: cutting a tooth for cavity preparation, crown preparation, crown finishing, inlay, filing, polishing, prophylaxis and endodontic treatment.
### 7. Substantial Equivalence Discussion:
The subject device is substantially equivalent to A-dec NLZ electric motor system (K163131). The subject device has the same indications for use and the technological characteristics as the predicate device. It also has the equivalent specifications as the predicate device in almost all parameters. The major difference is spray air pressure & spray water pressure. However this differences do not raise a question in substantial equivalence discussion since the user can use water and air spray hoses that are usually provided on a dental chair in the dental office. Since these are optional features, we believe these differences do not raise a new question in safety and effectiveness. Based on the performance test data, we conclude that the subject device is substantially equivalent to the predicate device.
| | Candidate Device | Predicate 1 | Comparison |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| 510(k) Number | K202963 | K163131 | - |
| Device Name | ELEC Master<br>ELEC Master Dual | A-dec NLZ electric motor<br>system | - |
| Manufacturer | MGNEWTON Ltd. | NAKANISHI INC. | - |
| Indications for Use | ELEC Master &<br>ELEC Master Dual<br>are intended for use in<br>general dental<br>applications such as:<br>cutting a tooth for<br>cavity preparation,<br>crown preparation,<br>crown finishing, inlay,<br>filing, polishing,<br>prophylaxis and<br>endodontic treatment. | A-dec NLZ electric motor<br>system is intended for use<br>in general dental<br>applications such as:<br>cutting a tooth for cavity<br>preparation, crown<br>preparation, crown<br>finishing, inlay, filing,<br>polishing, prophylaxis and<br>endodontic treatment. | Substantially<br>Equivalent |
| Drive | Electronic-micromotor | Electronic-micromotor | Substantially<br>Equivalent |
| Device components | Control unit with hose<br>and electrical motor | Control unit with hose and<br>electrical motor | Substantially<br>Equivalent |
| Light (Motor LED) | Yes | Yes | Substantially<br>Equivalent |
| Range of rotation speed | 1,000-40,000 rpm | 1,000- 40,000 rpm | Substantially<br>Equivalent |
| Rotation Direction | Forward and Reverse | Forward and Reverse | Substantially<br>Equivalent |
| Max. Torque | 3Ncm | 3Ncm | Substantially<br>Equivalent |
| Spray Air Pressure | 44 psi (0.3 MPa) | 29psi (0.2 MPa) | Different #1 |
| Spray Water Pressure | 22 psi (0.15 MPa) | 29psi (0.2 MPa) | Different #2 |
| Coolant mechanism | Coolant air | Coolant air | Substantially<br>Equivalent |
{5}------------------------------------------------
| Sterilization | Sterilized by user<br>(Steam Sterilization) | Sterilized by user<br>(Steam Sterilization) | Substantially<br>Equivalent |
|---------------------------------------------------------------------------|---------------------------------------------|---------------------------------------------|-----------------------------|
| Conformance with<br>standards for shanks<br>(Available Handpiece<br>Type) | E-type<br>Comply with<br>ISO 3964 | E-type<br>Comply with<br>ISO 3964 | Substantially<br>Equivalent |
### 8. Performance Tests (Non-clinical)
Non-clinical bench tests were performed as followings:
- ISO 3964:2016 Dental Handpieces Coupling dimensions
- ISO 7494-1:2011 Dentistry Dental units -Part 1: General requirements and test methods
- ISO 14457:2017 Dentistry Handpieces and motors
- IEC 60601-1, IEC 60601-1-2: Electrical safety and EMC
- IEC80601-2-60:2012 : Particular Requirements For The Basic Safety And Essential Performance Of Dental Equipment
- IEC 62366 : Application of usability engineering to medical devices
Along with the above tests, sterilization validation, and software validation were also conducted. None of the testing demonstrated any design characteristics that violated the requirements of the standards or resulted in any safety hazard.
#### 9. Conclusions:
Based on the information provided in this premarket notification, MGNEWTON Ltd. concludes that the ELEC Master & ELEC Master Dual are substantially equivalent to the predicate device as described herein in safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.