Uro-G Cystoscope

K202921 · Uroviu Corporation · FAJ · May 4, 2021 · Gastroenterology, Urology

Device Facts

Record IDK202921
Device NameUro-G Cystoscope
ApplicantUroviu Corporation
Product CodeFAJ · Gastroenterology, Urology
Decision DateMay 4, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

The Uro-G flexible cystoscope has been designed for endoscopic diagnosis and infusion of irrigating fluid within the bladder and urethra.

Device Story

Handheld, battery-operated portable cystoscope; consists of sterile, disposable steerable endoscopic cannula and reusable handle with integrated LCD monitor. Cannula features miniature CMOS camera, LED illumination module, and working channel for fluid infusion. Physician operates device in hospitals or clinics; inserts cannula via urethra to bladder. Handle contains video processor, rechargeable battery, and controls for brightness, image/video capture, and tip deflection. Real-time visualization on handle monitor assists physician in navigating anatomy and performing diagnostic procedures. Enables direct observation of bladder/urethral tissue; facilitates fluid infusion for irrigation. Benefits include improved ease of use via active tip deflection and integrated display, eliminating need for external monitor cables.

Clinical Evidence

No clinical data. Substantial equivalence established via bench testing, including design verification/validation, biocompatibility (ISO 10993), sterilization validation (ISO 11135), packaging/shelf-life studies, and electrical safety/EMC testing (ISO 60601-1, IEC 60601-2-18).

Technological Characteristics

Handheld system with reusable handle and disposable cannula. Cannula materials: SS 304, Polycarbonate, PTFE, Nylon, Thermoplastic polyurethane, Fluororubber TP. Sensing: CMOS camera, LED illumination. Energy: 3.7V rechargeable battery. Connectivity: Standalone (integrated LCD). Sterilization: Ethylene oxide (cannula). Standards: ISO 10993, ISO 60601-1, ISO 11135, ISO 11607.

Indications for Use

Indicated for symptomatic voiding dysfunction, hematuria, bladder tumor surveillance, recurrent lower urinary tract infection, and pelvic pain syndromes in patients requiring endoscopic diagnosis and fluid infusion in the bladder and urethra.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. May 4, 2021 UroViu Corporation Thomas Lawson, Ph.D. Vice President, Regulatory Affairs 5337 - 145th Place SE Bellevue, WA 98006 Re: K202921 Trade/Device Name: Uro-G Cystoscope Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FAJ Dated: April 1, 2021 Received: April 2, 2021 Dear Thomas Lawson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202921 Device Name Uro-G Cystoscope Indications for Use (Describe) The Uro-G flexible cystoscope has been designed for endoscopic diagnosis and infusion of irrigating fluid within the bladder and urethra. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY ## General Information | Submitter | UroViu Corporation | |-------------------------|-----------------------------------------------------------------| | Address | UroViu Corporation<br>5337 145th Place SE<br>Bellevue, WA 98006 | | FDA Registration Number | Not assigned yet | | Correspondence Person | Thomas Lawson, PhD | | Contact Information | Email:<br>thom@uroviu.com | | Date Prepared | 16 November 2020 | ### Proposed Device | Trade Name | Uro-G Cystoscope | |----------------------------------------------|--------------------------------------------| | Common Name | Uro-G | | Regulation Number and<br>Classification Name | 21 CFR§876.1500, Endoscope and Accessories | | Product Code | FAJ | | Regulatory Class | II | ## Predicate Device | Trade Name | Uro-V Cystoscope | |-----------------------------------------------------------------------------|--------------------------------------------| | Common Name | Uro-V | | Premarket Notification | K171500 | | Regulation Number<br>and Classification | 21 CFR§876.1500, Endoscope and Accessories | | Product Code | FAJ | | Regulatory Class | II | | Note: The predicate device has not been subject to a design-related recall. | | ### Reference Device | Trade Name | U-Scope 8000 System with HSC+EMB Cannula | |------------------------|------------------------------------------| | Common Name | U-Scope 8000 | | Premarket Notification | K132384 | {4}------------------------------------------------ | Regulation Number<br>and Classification | 21 CFR§884.1690, Hysteroscope and accessories | |-----------------------------------------------------------------------------|-----------------------------------------------| | Product Code | HIH | | Regulatory Class | II | | Note: The reference device has not been subject to a design-related recall. | | #### Device Description The Uro-G cystoscope is a handheld, battery-operated portable cystoscope consisting of a sterile, disposable steerable endoscopic cannula and a reusable handle with a video monitor. The disposable cannula contains a miniature CMOS camera and a light-emitting diode (LED) illumination module at its tip and one channel for infusion of irrigating fluid. The handle is lightweight and ergonomically designed. It has a connector and locking mechanism for attaching and detaching the disposable cannula. The handle contains the remaining electronics, including a power on/off button, a button to adjust the brightness of the LED, a button to allow capture of single images or to start/stop video of the procedure, a video processor, a display unit (LCD display), a rechargeable battery, management electronics, microcontrollers, and firmware. #### Indications for Use The indications for use for the Uro-G Cystoscope is: The Uro-G flexible cystoscope has been designed for endoscopic diagnosis and infusion of irrigating fluid within the bladder and urethra. #### Comparison of Technological Characteristics with the Predicate Device Uro Viu Corp. has identified the Uro-V Cystoscope (K171500, Uro Viu Corp.) as the predicate device. The Uro-V disposable cystoscope is substantially equivalent to the predicate device based upon the following similarities: - 1. The indications for use of both the predicate device and the Uro-G cystoscope are exactly the same: symptomatic voiding dysfunction, hematuria, bladder tumor surveillance, recurrent lower urinary tract infection, and pelvic pain syndromes; {5}------------------------------------------------ - 2. Both devices are introduced into the body via the urethra and then advanced to the bladder under visualization (that is, not a blind advancement); - 3. Both devices have illumination and optic components to permit visualization of the urethra and bladder; - 4. Both devices have the capability to view the urethra and bladder via video monitors; - 5. Both devices have a working channel to allow infusion of fluids; and - 6. Both devices are made from biocompatible materials. These similarities are not surprising in that the subject and predicate devices are manufactured for Uro Viu Corporation by the same contract manufacturer and both use the same handle with an integrated video screen. The U-Scope 8000 is a reference device since the same handle is used in the Uro-G cystoscope and so the electrical safety and EM compatibility testing performed in its review-K132384-is applicable for the handle of the Uro-G cystoscope. This is the only part of the submission applicable, so the U-Scope 8000 is not present in the SE comparison table. | | Subject Device<br>Uro-G Cystoscope<br>(UroViu Corp.)<br>This Submission | Predicate<br>Device<br>Uro-V Cystoscope<br>(UroViu Corp.)<br>K171500 | |----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------| | Device Class | II | II | | FDA Product Code | FAJ | SAME | | Product<br>Classification | 876.1500 | SAME | | Intended Use | The Uro-G Cystoscope has<br>been designed for endoscopic<br>diagnosis and infusion of<br>irrigating fluid within the<br>bladder and urethra. | The Uro-V cystoscope<br>has been designed for<br>endoscopic diagnosis<br>and infusion of irrigating<br>fluid within the bladder<br>and urethra. | | Indications for Use | Symptomatic voiding dysfunction<br>Hematuria<br>Bladder tumor surveillance<br>Recurrent lower urinary tract<br>infection<br>Pelvic pain syndromes | SAME | | Route of Advancement | Advanced to the bladder via the urethra. | SAME | | Site of Use | Hospitals and physician offices | SAME | | Technical Characteristics | | | | Components of the Set | Reusable handle with video screen<br><br>Attachable cannula with a working channel and an illumination source and camera at its tip | SAME | | Outer Diameter (OD) of Cannula | 5.5 mm | 4.2 mm | | Working Length of the Cannula | 380 mm | 254 mm | | Image Transmission | Image transmitted from a video camera at the tip of the cannula to a video monitor on the handle | SAME | | LCD Display Size | 3.5 inches (diagonal) on the handle | SAME | | Field of View | 140 degrees | SAME | | Focal Length | 5 to 50 mm | 3 to 50 mm | | Direction of View from Center Axis | 0 degrees (forward viewing) | SAME | | Operational Characteristics | | | | Adjust Brightness of Illumination | Adjust by depressing a button on the handle to change settings | SAME | | Images or Video | Capture still images or<br>video during a procedure<br>by depressing a camera<br>button on the handle | SAME | | Cleaning, Disinfecting, and<br>Sterilization | The handle is not<br>provided sterile. The<br>handle is cleaned and<br>disinfected following<br>instructions in the user<br>manual (IFU).<br><br>The disposable cannula is<br>provided sterile following<br>exposure to ethylene<br>oxide (EO) and is for<br>single use. It is disposed<br>after the procedure<br>following the institution's<br>procedures. | SAME | | Frequency of Use | Handle is reusable<br>Cannula is single patient<br>use | SAME | | Duration of Use | < 24 hours | SAME | | Tissue Contact Materials | Compliant with<br>ISO 10993 | SAME | #### Comparison of the Uro-G Cystoscope to the predicate device, the Uro-V Cystoscope. {6}------------------------------------------------ {7}------------------------------------------------ These devices are further noted to be substantially equivalent by their indications for use following such determinations outlined in 21 CFR 807(f): | | Uro-V Cystoscope<br>(predicate device) | Uro-G Cystoscope<br>(subject device) | |--------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | Endoscopic diagnosis and<br>infusion of irrigating fluid within<br>the bladder and urethra | EXACTLY THE SAME. | | Indications<br>for Use | Symptomatic voiding<br>dysfunction<br>Hematuria<br>Bladder tumor surveillance<br>Recurrent lower urinary tract<br>infection<br>Pelvic pain syndromes | EXACTLY THE SAME | | Method of introduction into the body | | EXACTLY THE SAME | | Characteristics of equivalence of indications for use defined in 21 CFR 807(f) | | | | Materials | Handle: not patient contact but the outer shell is plastic<br><br>Cannula:<br>Lens - Zeonex F52R Lens<br>coating - WR-110/N<br>Camera housing – SS 304<br>Camera tip - Polycarbonate<br>Fluid channel – Teflon (PTFE)<br>Fluid hub - Polycarbonate<br>Cannula shaft - Nylon<br>Adhesives - Loctite 3211 & 3218<br><br>The cannula materials passed all applicable biocompatibility tests. | Handle: Exactly the same handle as is used with the Uro-V cannula.<br><br>Cannula:<br>Lens - Zeonix F53R<br>Lens coating - WR-110/N<br>Camera housing - SS 304<br>Camera tip - Polycarbonate<br>Fluid channel – Teflon (PTFE)<br>Fluid hub - Polycarbonate<br>Cannula shaft - Thermoplastic polyurethane<br>Distal section of cannula – Fluororubber TP<br>Adhesives – Loctite 3211 & 3218<br><br>The cannula materials passed all applicable biocompatibility tests. | | Design | Handle: contains the electronics, a video processor, a rechargeable battery, and an LCD display. The handle is curve slightly so that the user can hold it easily in one hand.<br><br>Cannula:<br>1.Contains a miniature CMOS camera and LED illumination at its tip.<br>2.Contains a working channel for fluid infusion.<br>3.The tip can be deflected manually prior to use and rotated during a procedure.<br>4.Connects to the handle.<br>5.Provided sterile. | Handle: Exactly the same handle as is used with the Uro-V cannula.<br><br>Cannula:<br>1.Contains a miniature CMOS camera and LED illumination module at its tip.<br>2.Contains a working channel for fluid infusion.<br>3. The tip can be deflected and rotated during a procedure<br>4. Connects to the handle.<br>5. Provided sterile. | | Energy Used | Handle: powered by a 3.7 V rechargeable battery<br><br>Cannula: while the tip of the cannula contains both a camera | Handle: Exactly the same handle as is used with the Uro-V cannula.<br><br>Cannula: while the tip of the cannula contains both a camera | | Operational<br>Principles | Handle: Rather than run a cable<br>from the cystoscope to a separate<br>video monitor that is required of<br>all other cystoscopes, an LCD<br>display is integrated into the<br>handle of UroViu cystoscopes so<br>that the user can see directly<br>images of the tissue and<br>anatomical structures as the<br>cystoscope is being advanced to<br>the bladder. | Handle: Exactly the same handle<br>as is used with the Uro-V<br>cannula. | | | Cannula: the light at the tip<br>illuminates the field so that the<br>camera can transmit images back<br>to the LCD display on the handle.<br>It is a semi-rigid shaft with a tip<br>angle originating by the<br>company, SO the user either<br>navigates the urethra and looks<br>within the bladder by rotating the<br>cannula or can modify the angle<br>of the tip manually before<br>inserting the cannula into the<br>uretheral orifice. | Cannula: the light at the tip<br>illuminates the field so that the<br>camera can transmit images back<br>to the LCD display on the handle.<br>The angle of the tip can be<br>modified by moving a deflection<br>lever on the proximal portion of<br>the cannula, so that the user can<br>navigate the urethra and look<br>within the bladder by changing the<br>tip angle after inserting the<br>cannula into the urethral orifice. | {8}------------------------------------------------ {9}------------------------------------------------ The changes to the design of the Uro-G cystoscope centers on the cannula for this model; the handle has not changed from what it was in the Uro-V cystoscope submission (K171500). Cytoscopes are available in rigid, semi-rigid, or flexible configurations. Rigid cystoscopes typically are constructed of stainless steel and the shaft/cannula is not intended to bend or deflect. Their use requires that the physician forcethe non-flexible shaft up, down or to the side in order to navigate anatomic structures and obstructions within the urethra and bladder. The Uro-V cystoscope is considered a semi-rigid cystoscope in that its cannula comes out of the pouch in a pre-curved angle of 25° but the user has the ability to change the angle of the cannula before or during the procedure by manually bending the shaft of the cannula to a desired angle. As the Uro-V cystoscope is being advanced, the user can negotiate structures and obstructions my turning the cannula so that the tip advances away from structures/obstructions as it makes its way to the bladder. The Uro-G cystoscope builds on this by providing the user with the ability to change the angle of the cannula's tip by moving a deflection control lever on the proximal portion of the cannula. This capability to manipulate the cannula's tip during use improves ease of use for the physician but does not introduce a reduction in safety since the user can monitor the deflection with continual imaging so that structures can either be avoided or targeted directly if that was the purpose of the procedure. The specific modifications with the Uro-G cannula are: {10}------------------------------------------------ - 1. The outer diameter (OD) was increased to 5.5 mm. The OD of the Uro-G cystoscope was increased in order to permit an increase in the ID of the working channel to 2.2 mm. Since the urethral diameter of malesranges from 8 to 9 mm and the diameter in females averages 6 mm, this OD does not increase any risk to the device's use. This risk is further reduced since it is standard practice to infuse fluid as a cystoscope is advanced so that the urethra dilates ahead of the tip's position. - 2. The working length was increased to 380 mm, which is about 5 inches longer than the Uro-V model. In clinical work, the length of the Uro-V cystoscope is adequate, but in some patients does not always allow the tip to extend to the distal bladder wall in some male patients. This increase in 5 inches permits tip access to all parts of the bladder and is comparable to the lengths of other cystocopes, which range from 370 to 400 mm. - 3. The working channel has an ID of 2.2 mm. - The flexible tip can deflect up to 210 degrees up and 130 degrees down from 4. a straight, non-deflected orientation. Deflection wires embedded into the wall of the cannula that compress or extend basedupon movement of the deflection control lever accomplish this. Active tip deflection is a standard feature of some cystoscopes. - 5. The focal length of the camera is 5 to 50 mm. It is important to maintain focus of an area of interest or surrounding tissue as a cystoscope is being advanced or when targeting a specific anatomical area. This focal length is equivalent to that of the Uro-V cystoscope and also to those of cystoscopes marketed by Storz and Olympus. ## Performance Data All necessary performance testing was conducted with bench testing and included: - Design verification and validation studies; ● - . Packaging and shelf-life studies; - . Biocompatibility testing; - . Sterilization procedure validation: - Software verification and validation; and ● - Electrical safety and electromagnetic compatibility testing. ● The device was found to meet the compliance requirements for electrical safety as specified in ISO 60601-1 including provisions for EMC safety in ISO 60601-1-2 and IEC 60601-2-18 Medical electrical equipment -- Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment, including thermal safety . Mechanical characteristics were also tested, with successful results. Due to the cannula being labeled as sterile, the cannula underwent sterilization validation and shelf life testing to confirm the label shelf life complies with the following standards. {11}------------------------------------------------ K202921 - ISO 11135-1 Sterilization of health care products -- Ethylene ● oxide -Part 1 : Requirements for development, validation and routine control of a sterilization process for medical device; - ISO 11607 Packaging for Terminally Sterilized Medical Devices; - AAMI TIR12:2010 Designing, testing, and labeling . reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers; and - AAMI TIR30:2011 A compendium of processes, . materials, test methods, and acceptance criteria for cleaning reusable medical devices. ### Biocompatibility testing Testing was conducted on the cannula of the Uro-G cystoscope to verify that it is compliant with biocompatibility requirements for a short duration (<24 hours) indwelling device, as specified in ISO 10993 - Part 1, for the following tests: - о Cytotoxicity, - Irritation. o - Sensitization, and O - Systemic Toxicity (acute). O The device passed all tests. #### Animal Testing No preclinical testing of the subject device was necessary; the bench top testing was sufficient to determine the performance of the device. #### Clinical Studies No clinical testing of the subject device was necessary; the bench top testing was sufficient to determine the performance of the device. {12}------------------------------------------------ ## Conclusion The Uro-G and Uro-V cystoscopes have the same intended use, the exact same indications for use, and have equivalent technological characteristics. The minor differences between the Uro-V and the Uro-G cystoscopes do not raise any new issues of safety or effectiveness. Therefore, the Uro-G cystoscope is substantially equivalent to the identified predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order