← Product Code [KCY](/productcode/KCY) · K202919

# ATS 5000 Automatic Tourniquet Instrument (K202919)

_Mcewen and Associates Consulting , Ltd. · KCY · May 20, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K202919

## Device Facts

- **Applicant:** Mcewen and Associates Consulting , Ltd.
- **Product Code:** [KCY](/productcode/KCY.md)
- **Decision Date:** May 20, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5910
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The ATS 5000 Tourniquet System is intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities. Tourniquets have been found useful in producing a bloodless operation field in surgical procedures involving the extremities including: Reduction of certain fractures Kirschner wire removal Tumor and cyst excisions Subcutaneous fasciotomy Nerve injuries Tendon repair Bone grafts Total wrist joint replacement Replacement of joints in the fingers Knee joint replacements Amputations Replantations

## Device Story

ATS 5000 is a dual-port, microprocessor-controlled pneumatic tourniquet instrument; used by perioperative nurses, surgeons, and anesthesiologists in surgical settings. Device connects to pneumatic cuffs via tubing; operator controls via touchscreen. System monitors arterial pulsations in the limb to determine Limb Occlusion Pressure (LOP) using either a Distal Method or the 'EZ Method' (which senses pneumatic pressure pulsations directly through the cuff). Device features barcode scanning for cuff identification and automated deflation protocols (Cyclic or Step) to assist in bleeding vessel management. Output provides real-time pressure monitoring and automated LOP measurement. By standardizing LOP and automating deflation, the device aims to improve perioperative workflow efficiency and support bloodless surgical fields, potentially reducing complications associated with manual tourniquet management.

## Clinical Evidence

No clinical data was needed for this device. Safety and performance were established through non-clinical performance testing, including electrical safety (IEC 60601-1, 60601-1-2, 60601-1-8), software validation (IEC 62304), and usability testing (IEC 60601-1-6, IEC 62366-1).

## Technological Characteristics

Dual-port pneumatic tourniquet instrument; microprocessor-controlled. Features touchscreen interface, barcode reader module for cuff identification, and pneumatic pressure sensing for LOP measurement. Connectivity includes pneumatic ports for cuff hoses. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-1-8 (alarms), IEC 62304 (software lifecycle), IEC 62366-1 (usability).

## Regulatory Identification

A pneumatic tourniquet is an air-powered device consisting of a pressure-regulating unit, connecting tubing, and an inflatable cuff. The cuff is intended to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery.

## Predicate Devices

- Zimmer A.T.S. 4000 Automatic Tourniquet System ([K123553](/device/K123553.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 20, 2021

McEwen and Associates Consulting Ltd. Julie Kerr Consultant 207-1099 West 8th Avenue Vancouver, BC V6H 1C3 Canada

Re: K202919

Trade/Device Name: ATS 5000 Automatic Tourniquet Instrument Regulation Number: 21 CFR 878.5910 Regulation Name: Pneumatic Tourniquet Regulatory Class: Class I Product Code: KCY Dated: September 25, 2020 Received: September 29, 2020

Dear Julie Kerr:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

#### 510(k) Number (if known) K202919

#### Device Name

ATS 5000 Automatic Tourniquet Instrument

#### Indications for Use (Describe)

The ATS 5000 Tourniquet System is intended to be used by qualified medical professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities. Tourniquets have been found useful in producing a bloodless operation field in surgical procedures involving the extremities including:

Reduction of certain fractures Kirschner wire removal Tumor and cyst excisions Subcutaneous fasciotomy Nerve injuries Tendon repair Bone grafts Total wrist joint replacement Replacement of joints in the fingers Knee joint replacements Amputations Replantations

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## McEwen and Associates Consulting Ltd.

Suite 207, 1099 West 8th Avenue Vancouver BC Canada V6H 1C3

### K202919 - 510(k) Summary

| Sponsor:               | McEwen & Associates Consulting Ltd.<br>1099 West 8th Avenue<br>Vancouver, BC, V6H 1C3                                                                                                                                                                                                                                                                                                                                                                 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:        | Julie Kerr<br>Project Manager<br>Telephone: (604) 742-0600                                                                                                                                                                                                                                                                                                                                                                                            |
| Submission Date:       | September 25, 2020                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Trade Name:            | ATS 5000 Automatic Tourniquet System                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Common Name:           | Tourniquet device                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code / Device: | KCY - Pneumatic Tourniquet. The applicable regulation is<br>§878.5910 under the limitation defined by §878.9. We believe<br>the proposed device requires 510(k) clearance as the deflation<br>protocols, Cuff ID, and EZ LOP features falls outside the<br>exemption classification.                                                                                                                                                                  |
| Predicate Device:      | Zimmer ® A.T.S. ® 4000 Automatic Tourniquet System,<br>K123553, cleared 08/09/2013                                                                                                                                                                                                                                                                                                                                                                    |
| Device Description:    | The ATS 5000 is a non-sterile dual-port microprocessor<br>controlled pneumatic tourniquet instrument intended to be<br>used by qualified medical professionals to temporarily occlude<br>blood flow in a patient's extremities during surgical<br>procedures on those extremities.                                                                                                                                                                    |
|                        | The ATS 5000 instrument is used in conjunction with available<br>dual port tourniquet cuffs and hoses distributed/supplied by<br>Zimmer Biomet Inc. A cuff is applied to a patient prior to the<br>beginning of a procedure. Connective tubing is then attached<br>to the cuff and then plugged into the ATS 5000 connector<br>ports. The instrument is controlled via a touchscreen user<br>interface.                                               |
| Indications for Use:   | The ATS 5000 Tourniquet System is intended to be used by<br>perioperative nurses, surgeons, and anesthesiologists, to<br>temporarily occlude blood flow in a patient's extremities<br>during surgical procedures on those extremities. Tourniquets<br>have been found useful in producing a bloodless operation<br>field in surgical procedures involving the extremities including:<br>Reduction of certain fractures, Kirschner wire removal, tumor |

{4}------------------------------------------------

## McEwen and Associates Consulting Ltd.

Suite 207, 1099 West 8th Avenue Vancouver BC Canada V6H 1C3

> and cyst excisions, subcutaneous fasciotomy, nerve injuries, tendon repair, bone grafts, total wrist joint replacement, replacement of joints in the fingers, knee joint replacements, amputations, replantations.

Comparison to Predicate Device: The ATS 5000 instrument is substantially equivalent to other legally marketing tourniquet systems, specifically the Zimmer ATS 4000 Automatic Tourniquet System, in that the devices are similar in design, materials, and indications for use. A complete discussion of technological characteristics of the ATS 5000 compared to the predicate device may be found in Section 12. The below table highlights the primary technological differences that we believe

### Technological Characteristics:

| Feature                        | Discussion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Deflation<br>Protocols         | Two user-selectable deflation protocols allow the user to vary from the standard<br>rapid deflation near the end of surgery.<br>Cyclic Deflation: enables rapid detection and intervention for cauterization of<br>bleeding vessels immediately prior to wound closure.<br>Step Deflation: enables gradual cuff deflation to allow for gradual return of arterial<br>flow.<br>Both types of deflation protocols are achievable through manual changes to the<br>pressure (same between the proposed and predicate devices). Automatic protocols,<br>if selected, can simplify the perioperative workflow, and may simplify and<br>standardize the detection and management of bleeding vessels.                                                                                                                                                                                                                                                             |
| Cuff ID                        | Barcode reader module in the instrument to scan cuff packing labels. If it is a<br>barcode identified by the device the instrument will recognize the cuff as being one<br>of the cuffs listed in the Instructions for Use and will display cuff details for the user<br>to view (EZ Method enabled, dual ports, limb circumference range).<br>New feature to further personalize tourniquet usage by taking into account specific<br>details of recognized cuffs. Barcode reader module will scan any label, but will only<br>display cuff details of recognized cuffs.                                                                                                                                                                                                                                                                                                                                                                                    |
| EZ Method<br>to measure<br>LOP | The EZ Method measures LOP by employing a tourniquet cuff recognized by Cuff ID<br>to monitor arterial pulsations in an underlying limb by sensing pneumatic pressure<br>pulsations in the cuff associated with volume changes in the limb as the cuff<br>pressure is gradually increased. LOP is determined to be the cuff pressure at which<br>the monitored arterial pulsations reach a certain threshold. The EZ Method is only<br>enabled for cuffs recognized by Cuff ID.<br>The EZ Method is a supplementary method to measure LOP, in addition to the<br>Distal Method. The EZ Method enables the LOP measurement to be completed<br>without the need for a Distal Sensor, thereby increasing convenience and speed and<br>reducing disruptions of the perioperative workflow. Both methods monitor arterial<br>pulsations to determine LOP as a cuff is gradually inflated. EZ Method and Distal<br>Method provide equivalent measurements of LOP. |

{5}------------------------------------------------

## McEwen and Associates Consulting Ltd.

Suite 207, 1099 West 8th Avenue Vancouver BC Canada V6H 1C3

Performance Data:

Non-Clinical Performance:

During the development process of the ATS 5000, the following testing was completed: Electrical safety and Environmental testing in accordance with IEC 60601-1, 60601-1-2 and 60601-1-8; Software and device development was conducted in accordance with requirements of IEC 62304:2006 and AAMI/ANSI HE-75:2009/(R)2018; Device Usability testing was conducted in accordance with requirements of IEC 60601-1-6 and IEC 62366-1:2015; and Hardware and Software testing including validation. All tests passed according to predetermined acceptance criteria.

Clinical Performance – Clinical data was not needed for this device.

Conclusions:

Expanded functionality which falls outside the 510(k)exemption classification, does not require additional clinical data as the base technology safety and performance is well established. Existing literature, alongside the non-clinical performance data support the safety and performance of the proposed device.

---

**Source:** [https://fda.innolitics.com/device/K202919](https://fda.innolitics.com/device/K202919)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
