The CreoDent Solidex® Customized Abutment and screw is intended for use with an endosseous implant to support a prosthetic device in patients who are partially or completely edentulous. The device can be used for single or multiple restorations. The prosthesis can be cemented or screw retained to the abutment. An abutment screw is used to secure the abutment to the endosseous implant. The CreoDent Solidex® Customized Abutment is compatible with the following dental implants: [List of compatible implants provided in document]. All digitally designed files for CreoDent Solidex® Customized Abutment are to be sent back to a CreoDent-validated manufacturing facility for manufacture.
Device Story
Customized dental abutment; fabricated from Ti-6AL-4V ELI (ASTM F136) or Commercially Pure titanium (ASTM F67). Device supports prosthetic restorations (cement or screw-retained) on endosseous implants. Prescribing dentist specifies design requirements; CreoDent-validated facility manufactures device from digital files. Used in dental clinics; operated by dentists. Output is physical abutment; aids in restoring patient masticatory function and aesthetics. Substantial equivalence established via compatibility with previously cleared implant systems and leveraging existing mechanical/biocompatibility data.
Clinical Evidence
No clinical data included. Non-clinical performance testing included static/fatigue testing per ISO 14801:2007, reverse engineering dimensional analysis, and sterilization validation per ISO 17665-1. Biocompatibility is supported by data from predicate and reference devices.
Technological Characteristics
Materials: Ti-6AL-4V ELI (ASTM F136) or Commercially Pure titanium (ASTM F67). Energy source: N/A (mechanical). Form factor: Customized abutment with screw. Sterilization: ISO 17665-1. Connectivity: Digital file submission to manufacturing facility.
Indications for Use
Indicated for partially or completely edentulous patients requiring single or multiple dental restorations supported by endosseous implants. Compatible with specific implant systems from Altatec, Astra Tech, Biomet, Dentium, Hiossen, Lifecore, Nobel Biocare, Thommen, Straumann, and Zimmer.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
4/2/2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized depiction of an eagle, while the text reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The text is arranged in two lines, with "FDA U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.
CreoDent Prosthetics, Ltd. % Chris Brown, Manager Aclivi, LLC 6455 Farley Road Pinckney, Michigan 48169 USA
Re: K202909
Trade/Device Name: CreoDent Solidex® Customized Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: February 26, 2021 Received: March 3, 2021
Dear Chris Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-ind-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202909
#### Device Name
CreoDent Solidex® Customized Abutment
#### Indications for Use (Describe)
The CreoDent Solidex® Customized Abutment and screw is intended for use with an endosseous implant to support a prosthetic device in patients who are partially or completely edentulous. The device can be used for single or multiple restorations. The prosthesis can be cemented or screw retained to the abutment screw is used to secure the abutment to the endosseous implant. The CreoDent Solidex® Customized Abutment is compatible with the following dental implants:
| Manufacturer | Implant Line | Platform<br>Diameter (mm) | Implant Body<br>Diameter (mm) |
|------------------------------------------|-----------------------------------------------|---------------------------|-------------------------------|
| Altatec Biotechnolgies N.A. Incorporated | Camlog Screwline Implant Promote | 3.3 | 3.3 |
| Altatec Biotechnolgies N.A. Incorporated | Camlog Screwline Implant Promote | 3.8 | 3.8 |
| Altatec Biotechnolgies N.A. Incorporated | Camlog Screwline Implant Promote | 4.3 | 4.3 |
| Altatec Biotechnolgies N.A. Incorporated | Camlog Screwline Implant Promote | 5.0 | 5.0 |
| Astra Tech, Inc | Osseospeed TX Implants | 3.5 | 3.5 |
| Astra Tech, Inc | Osseospeed TX Implants | 4.5 | 4.5 |
| Biomet | 3i Osseotite Certain Dental Implants | 3.4 | 3.25 |
| Biomet | 3i Osseotite Certain Dental Implants | 4.1 | 4.0 |
| Biomet | 3i Osseotite Certain Dental Implants | 5.0 | 5.0 |
| Biomet | 3i Osseotite Certain Dental Implants | 6.0 | 6.0 |
| Dentium CO., Ltd | Dentium 3.6mm Implantium | 3.6 | 3.4 |
| Hiossen | Hiossen TS Implants | 3.5 (Mini) | 3.5 |
| Hiossen | Hiossen TS Implants | 4.0 (Regular) | 4.0 |
| Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | 3.5 | 2.4 |
| Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | 4.1 | 2.7 |
| Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | 5.0 | 3.3 |
| Nobel Biocare | NobelActive™ Internal Connection Implant | 3.5 (NP) | 3.5 |
| Nobel Biocare | NobelActive™ Internal Connection Implant | 3.9 (RP) | 4.3 |
| Nobel Biocare | NobelReplace™ TiUnite Endosseous Implant | 3.5 (NP) | 3.5 |
| Nobel Biocare | NobelReplace™ TiUnite Endosseous Implant | 4.3, 5.0 (RP) | 4.3, 5.0 |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 3.5 | 3.5 |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 4.0 | 4.0 |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 4.5 | 4.2 |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 5.0 | 5.0 |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 6.0 | 6.0 |
| Straumann | Straumann Bone Level (BL) Implants | 3.3 (NC) | 3.3 |
| Straumann | Straumann Bone Level (BL) Implants | 4.1, 4.8 (RC) | 4.1, 4.8 |
| Zimmer | Zimmer Screw-Vent, Tapered Screw-Vent Implant | 3.5 | 3.7 |
| Zimmer | Zimmer Screw-Vent, Tapered Screw-Vent Implant | 4.5 | 4.7 |
| Zimmer | Zimmer Tapered Screw-Vent Implant | 5.7 | 6.0 |
All digitally designed files for CreoDent Solidex® Customized Abutment are to be sent back to a CreoDent-validated manufacturing facility for manufacture.
{4}------------------------------------------------
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## K202909 510(k) Summary CreoDent Prosthetics, Ltd. CreoDent Solidex® Customized Abutment 4/2/2021
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | CreoDent Prosthetics, Ltd.<br>29 West 30th Street, 11th Floor<br>New York, New York 10001<br>Telephone: +1 212-302-3860 |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact<br>Email | Calvin Shim - Managing Director<br>calvin.shim@creodental.com |
| Representative/Consultant | Chris Brown, BSEE<br>Aclivi, LLC<br>3250 Brackley Drive<br>Ann Arbor, Michigan 48105<br>Telephone: +1 (810) 360-9773<br>E-mail: acliviconsulting@gmail.com |
| DEVICE NAME AND CLASSIFICATION | |
| Trade/Proprietary Name: | CreoDent Solidex® Customized Abutment |
| Common Name: | Abutment, Implant, Dental, Endosseous |
| Regulation Name: | Endosseous dental implant abutment |
| Regulation Number: | 21 CFR 872.3630 |
| Device Class: | Class II |
| Review Panel: | Dental Products Panel |
|-------------------|----------------------------------------------------------------------------------------------|
| Reviewing Branch: | Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1)<br>Dental Devices (DHT1B) |
## PREDICATE DEVICE INFORMATION
Product Code:
The Subject device in this submission is substantially equivalent in indications, intended use and design principles to the following Predicate device.
NHA
| 510(k) | Predicate Device Name | Company Name |
|---------|---------------------------------------|----------------------------|
| K162734 | CreoDent Solidex® Customized Abutment | CreoDent Prosthetics, Ltd. |
The primary intent of this submission is to expand the Predicate device clearance to also include the following legally-marketed Reference device abutments and screws within a common Indications for Use which includes the statement that "digitally designed files for CreoDent Solidex® Customized Abutment are to be sent back to a CreoDent-validated manufacturing facility for manufacture."
| 510(k) | Reference Device Name | Company Name |
|---------|---------------------------------------|----------------------------|
| K113738 | CreoDent Solidex® Customized Abutment | CreoDent Prosthetics, Ltd. |
| K150012 | CreoDent Solidex® Customized Abutment | CreoDent Prosthetics, Ltd. |
| K160436 | CreoDent Solidex® Customized Abutment | CreoDent Prosthetics, Ltd. |
{6}------------------------------------------------
## INDICATIONS FOR USE
The CreoDent Solidex® Customized Abutment and screw is intended for use with an endosseous implant to support a prosthetic device in patients who are partially or completely edentulous. The device can be used for single or multiple restorations. The prosthesis can be cemented or the abutment. An abutment. An abutment screw is used to secure the abutment to the endosseous implant. The CreoDent Solidex® Customized Abutment is compatible with the following dental implants:
| Manufacturer | Implant Line | Platform<br>Diameter<br>(mm) | Implant Body<br>Diameter (mm) |
|------------------------------------------|-----------------------------------------------|------------------------------|-------------------------------|
| Altatec Biotechnolgies N.A. Incorporated | Camlog Screwline Implant Promote | 3.3 | 3.3 |
| Altatec Biotechnolgies N.A. Incorporated | Camlog Screwline Implant Promote | 3.8 | 3.8 |
| Altatec Biotechnolgies N.A. Incorporated | Camlog Screwline Implant Promote | 4.3 | 4.3 |
| Altatec Biotechnolgies N.A. Incorporated | Camlog Screwline Implant Promote | 5.0 | 5.0 |
| Astra Tech, Inc | Osseospeed TX Implants | 3.5 | 3.5 |
| Astra Tech, Inc | Osseospeed TX Implants | 4.5 | 4.5 |
| Biomet | 3i Osseotite Certain Dental Implants | 3.4 | 3.25 |
| Biomet | 3i Osseotite Certain Dental Implants | 4.1 | 4.0 |
| Biomet | 3i Osseotite Certain Dental Implants | 5.0 | 5.0 |
| Biomet | 3i Osseotite Certain Dental Implants | 6.0 | 6.0 |
| Dentium CO., Ltd | Dentium 3.6mm Implantium | 3.6 | 3.4 |
| Hiossen | Hiossen TS Implants | 3.5 (Mini) | 3.5 |
| Hiossen | Hiossen TS Implants | 4.0 (Regular) | 4.0 |
| Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | 3.5 | 2.4 |
| Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | 4.1 | 2.7 |
| Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | 5.0 | 3.3 |
| Nobel Biocare | NobelActive™ Internal Connection Implant | 3.5 (NP) | 3.5 |
| Nobel Biocare | NobelActive™ Internal Connection Implant | 3.9 (RP) | 4.3 |
| Nobel Biocare | NobelReplace™ TiUnite Endosseous Implant | 3.5 (NP) | 3.5 |
| Nobel Biocare | NobelReplace™ TiUnite Endosseous Implant | 4.3, 5.0 (RP) | 4.3, 5.0 |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 3.5 | 3.5 |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 4.0 | 4.0 |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 4.5 | 4.2 |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 5.0 | 5.0 |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 6.0 | 6.0 |
| Straumann | Straumann Bone Level (BL) Implants | 3.3 (NC) | 3.3 |
| Straumann | Straumann Bone Level (BL) Implants | 4.1, 4.8 (RC) | 4.1, 4.8 |
| Zimmer | Zimmer Screw-Vent, Tapered Screw-Vent Implant | 3.5 | 3.7 |
| Zimmer | Zimmer Screw-Vent, Tapered Screw-Vent Implant | 4.5 | 4.7 |
| Zimmer | Zimmer Tapered Screw-Vent Implant | 5.7 | 6.0 |
All digitally designed files for CreoDent Solidex® Customized Abutment are to be sent back to a CreoDent-validated manufacturing facility for manufacture.
## DEVICE DESCRIPTION
The CreoDent Solidex® Customized Abutment is fabricated of Ti-6AL-4V ELI titanium alloy meeting the requirements of ASTM F136 or Commercially Pure (CP Ti) titanium meeting the requirements of ASTM F67 (Grade 04-06). The CreoDent Solidex® Customized Abutment retention screws are fabricated of Ti-6AL-4V ELI titanium alloy meeting the requirements of ASTM F136 or Commercially Pure (CP Ti) titanium meeting the requirements of ASTM F67 (Grade 04-06). The CreoDent Solidex® Customized Abutment is screw-retained for use with endosseous dental implants to provide support for a prosthetic restoration. These abutments are indicated for cement or screw-retained restorations. CreoDent Solidex® Customized Abutments are compatible with the following implant systems and the listed implant and restorative platform diameters:
{7}------------------------------------------------
| Manufacturer | Implant Line | Platform<br>Diameter<br>(mm) | Implant Body<br>Diameter (mm) | Table 5A: Indications for Use Comparison with Predicate K162734 | | | | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|------------------------------|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|------------------------|----------------------------|----------------------|
| Altatec Biotechnolgies N.A. Incorporated | Camlog Screwline Implant Promote | 3.3 | 3.3 | Subject Device<br>CreoDent Solidex® Customized Abutment | | Predicate Device<br>CreoDent Solidex® Customized Abutment<br>K162734 | | | | |
| Altatec Biotechnolgies N.A. Incorporated | Camlog Screwline Implant Promote | 3.8 | 3.8 | The CreoDent Solidex® Customized Abutment and screw is intended for use with an endosseous implant to support a prosthetic device in patients who are partially or completely edentulous. The device can be used for single or multiple restorations. The prosthesis can be cemented or screw retained to the abutment. An abutment screw is used to secure the abutment to the endosseous implant. The CreoDent Solidex® Customized Abutment is compatible with the following dental implants: | | The CreoDent Solidex® Customized Abutment and screw is intended for use with an endosseous implant to support a prosthetic device in patients who are partially or completely edentulous. The device can be used for single or multiple restorations. The prosthesis can be cemented or screw retained to the abutment. An abutment screw is used to secure the abutment to the endosseous implant. The Solidex® Customized Abutment is compatible with the following dental implants: | | | | |
| Altatec Biotechnolgies N.A. Incorporated | Camlog Screwline Implant Promote | 4.3 | 4.3 | Manufacturer | Implant Line | Manufacturer | Implant Line | Platform Diameter (mm) | Implant Body Diameter (mm) | Implant Lengths (mm) |
| Altatec Biotechnolgies N.A. Incorporated | Camlog Screwline Implant Promote | 5.0 | 5.0 | Altatec Biotechnolgies N.A.<br>Incorporated | Camlog Screwline Implant Promote | Altatec Biotechnolgies N.A.<br>Incorporated | Camlog Screwline Implant Promote | 3.3 | 3.3 | 11,13,16 |
| Astra Tech, Inc | Osseospeed TX Implants | 3.5 | 3.5 | Altatec Biotechnolgies N.A.<br>Incorporated | Camlog Screwline Implant Promote | Altatec Biotechnolgies N.A.<br>Incorporated | Camlog Screwline Implant Promote | 3.8 | 3.8 | 9,11,13,16 |
| Astra Tech, Inc | Osseospeed TX Implants | 4.5 | 4.5 | Altatec Biotechnolgies N.A.<br>Incorporated | Camlog Screwline Implant Promote | Altatec Biotechnolgies N.A.<br>Incorporated | Camlog Screwline Implant Promote | 4.3 | 4.3 | 9,11,13,16 |
| Biomet | 3i Osseotite Certain Dental Implants | 3.4 | 3.25 | Altatec Biotechnolgies N.A.<br>Incorporated | Camlog Screwline Implant Promote | Altatec Biotechnolgies N.A.<br>Incorporated | Camlog Screwline Implant Promote | 5.0 | 5.0 | 9,11,13,16 |
| Biomet | 3 i Osseotite Certain Dental Implants | 4.1 | 4.0 | Astra Tech, Inc | Osseospeed TX Implants | Dentium CO., Ltd | Dentium 3.6mm Implantium | 3.6 | 3.4 | 8,10,12,14 |
| Biomet | 3 i Osseotite Certain Dental Implants | 5.0 | 5.0 | Astra Tech, Inc | Osseospeed TX Implants | Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 3.5 | 2.8 | 8,9.5,11,12.5,14,17 |
| Biomet | 3i Osseotite Certain Dental Implants | 6.0 | 6.0 | Biomet | 3i Osseotite Certain Dental Implants | Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 4.0 | 3.5 | 8,9.5,11,12.5,14,17 |
| Dentium CO., Ltd | Dentium 3.6mm Implantium | 3.6 | 3.4 | Biomet | 3i Osseotite Certain Dental Implants | Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 4.2 | 3.5 | 8,9.5,11,12.5,14,17 |
| Hiossen | Hiossen TS Implants | 3.5 (Mini) | 3.5 | Biomet | 3i Osseotite Certain Dental Implants | Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 5.0 | 4.3 | 8,9.5,11,12.5,14,17 |
| Hiossen | Hiossen TS Implants | 4.0 (Regular) | 4.0 | Biomet | 3i Osseotite Certain Dental Implants | Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 6.0 | 5.3 | 8,9.5,11,12.5,14,17 |
| Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | 3.5 | 2.4 | Dentium CO., Ltd | Dentium 3.6mm Implantium | Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | 3.5 | 2.4 | 10,11.5,13,15 |
| Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | 4.1 | 2.7 | Hiossen | Hiossen TS Implants | Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | 4.1 | 2.7 | 10,11.5,13,15 |
| Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | 5.0 | 3.3 | Hiossen | Hiossen TS Implants | Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | 5.0 | 3.3 | 8,10,11.5,13 |
| Nobel Biocare | NobelActive™ Internal Connection Implant | 3.5 (NP) | 3.5 | Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | | | | | |
| Nobel Biocare | NobelActive™ Internal Connection Implant | 3.9 (RP) | 4.3 | Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | | | | | |
| Nobel Biocare | NobelReplace™ TiUnite Endosseous Implant | 3.5 (NP) | 3.5 | Lifecore Biomedical, Inc | PrimaConnex Implants Internal (KeyStone) | | | | | |
| Nobel Biocare | NobelReplace™ TiUnite Endosseous Implant | 4.3, 5.0 (RP) | 4.3, 5.0 | Nobel Biocare | NobelActive™ Internal Connection Implant (NP) | | | | | |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 3.5 | 3.5 | Nobel Biocare | NobelActive™ Internal Connection Implant (RP) | | | | | |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 4.0 | 4.0 | Nobel Biocare | NobelReplace™ TiUnite Endosseous Implant | | | | | |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 4.2 | 4.5 | Nobel Biocare | NobelReplace™ TiUnite Endosseous Implant | | | | | |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 5.0 | 5.0 | Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | | | | | |
| Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | 6.0 | 6.0 | Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | | | | | |
| Straumann | Straumann Bone Level (BL) Implants | 3.3 (NC) | 3.3 | Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | | | | | |
| Straumann | Straumann Bone Level (BL) Implants | 4.1, 4.8 (RC) | 4.1, 4.8 | Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | | | | | |
| Zimmer | Zimmer Screw-Vent, Tapered Screw-Vent Implant | 3.5 | 3.7 | Thommen Medical AG | SPI Element Dental Implants RC INCELL PF | | | | | |
| Zimmer | Zimmer Screw-Vent, Tapered Screw-Vent Implant | 4.5 | 4.7 | Straumann | Straumann Bone Level (BL) Implants | | | | | |
| Zimmer | Zimmer Tapered Screw-Vent Implant | 5.7 | 6.0 | Straumann | Straumann Bone Level (BL) Implants | | | | | |
| Zimmer | Zimmer Screw-Vent, Tapered Screw-Vent Implant | | | | | | | | | |
| Zimmer | Zimmer Screw-Vent, Tapered Screw-Vent Implant | | | | | | | | | |
| Zimmer | Zimmer Tapered Screw-Vent Implant | | | | | | | | | |
| All digitally designed files for Creodent Solidex® Customized Abutment are to be sent back to a CreoDent-validated manufacturing facility for manufacture. | | | | | | | | | | |
The design of subject device is customized to the requirements of each patient as may be specified by the prescribing dentist. Customization is limited by the minimum dimensions for wall thickness, diameter, post height, collar height and angulation. All digitally designed files for CreoDent Solidex® Customized Abutment are to be sent back to a CreoDent-validated manufacturing facility for manufacture.
This submission contains no new compatibilities.
{8}------------------------------------------------
## EQUIVALENCE TO MARKETED DEVICE
The Subject device is substantially equivate device with respect to Indications for Use and technological principles. The comparison table below compare the Indications for Use and Technological Characteristics of the Subject and Predicate devices.
# ludications for Use Camparia
{9}------------------------------------------------
The Subject and Predicate devices are Substant, only differing in the list of compatible implant systems supporting a finding of Substantial Equivalence. The implant length column the compatibility table, as the implant and restorative platform diameters are the primary determining factors in compatibility. Differences in the list of compatible implant systems for each clearanded use of the device.
| Subject Device | Reference Device | Reference Device | Reference Device | Technical Characteristics Substantial Equivalence Comparison Table | | | | | | | | | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.