← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K202889

# Able Medical Devices Valkyrie Thoracic Fixation System (K202889)

_Jm Longyear Manufacturing, LLC D.B.A. Able Medical Devices · HRS · Mar 2, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K202889

## Device Facts

- **Applicant:** Jm Longyear Manufacturing, LLC D.B.A. Able Medical Devices
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Mar 2, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Valkyrie Thoracic Fixation System is intended for use in the stabilization and fixation of fractures in the chest wall including sternal reconstructive surgical procedures, trauma, or planned osteotomies. The system is intended for use in patients with normal and/or poor bone quality.

## Device Story

Valkyrie Thoracic Fixation System is a bone fixation system consisting of plates and screws used for chest wall stabilization. The system is used by surgeons in clinical settings to stabilize fractures, reconstruct the sternum, or secure bone following planned osteotomies. The device provides mechanical support to the chest wall, aiding in bone healing and structural integrity. It is designed to accommodate varying patient anatomy through a range of screw diameters (3.0mm and 3.5mm) and lengths (7-20mm). The system is suitable for patients with both normal and poor bone quality.

## Clinical Evidence

Bench testing only. Performance data includes static and dynamic construct tension testing. Biocompatibility testing was performed per ISO 10993 standards (cytotoxicity, systemic toxicity, genotoxicity, intracutaneous irritation, pyrogenicity, sensitization, and chemical characterization). Sterilization, packaging shelf-life, and packaging integrity were validated per ISO 11137 and ASTM standards.

## Technological Characteristics

Plates are manufactured from PEEK-Optima (ASTM F2026); screws are manufactured from Ti-6Al-4V (ASTM F136). The system utilizes a mechanical bone fixation design. Sterilization is validated per ISO 11137-1:2006. Packaging validated per ASTM 1980, F88, F1886, F1929, and F2096.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Biomet Microfixation Sternalock 360 Sternal Closure System ([K163007](/device/K163007.md), [K151019](/device/K151019.md), [K121302](/device/K121302.md), [K110574](/device/K110574.md))
- Tritium Sternal Cable Plate System ([K193468](/device/K193468.md), [K133785](/device/K133785.md))
- KLS Martin LSS Plating System ([K151983](/device/K151983.md))
- Stainless Steel Suture ([K170767](/device/K170767.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 2, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

JM Longyear Manufacturing, LLC D.B.A. Able Medical Devices % Nathan Wright Engineer & Regulatory Specialist Empirical Testing Corp 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K202889

Trade/Device Name: Valkyrie Thoracic Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: January 25, 2021 Received: January 26, 2021

Dear Nathan Wright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For; Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement on last page. |
|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <b>Indications for Use</b>                                              |                                                                                                    |
| 510(k) Number ( <i>if known</i> )<br>K202889                            |                                                                                                    |
| Device Name                                                             |                                                                                                    |
| Valkyrie Thoracic Fixation System                                       |                                                                                                    |
| Indications for Use ( <i>Describe</i> )                                 |                                                                                                    |

The Valkyrie Thoracic Fixation System is intended for use in the stabilization and fixation of fractures in the chest wall including sternal reconstructive surgical procedures, trauma, or planned osteotomies. The system is intended for use in patients with normal and/or poor bone quality.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span>                      Prescription Use (Part 21 CFR 801 Subpart D)                 </span> <span>                     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

{3}------------------------------------------------

# 510(K) SUMMARY - K202889

| Submitter's Name:          | Able Medical Devices                                                                                            |
|----------------------------|-----------------------------------------------------------------------------------------------------------------|
| Submitter's Address:       | 512 4th Street<br>Gwinn, MI 49841                                                                               |
| Submitter's Telephone:     | 906-372-3213                                                                                                    |
| Contact Person:            | Nathan Wright MS<br>Empirical Testing Corp.<br>719-351-0248<br>nwright@empiricaltech.com                        |
| Date Summary was Prepared: | September 25, 2020                                                                                              |
| Trade or Proprietary Name: | Valkyrie Thoracic Fixation System                                                                               |
| Common or Usual Name:      | Bone Plate                                                                                                      |
| Classification:            | Class II per 21 CFR §888.3030 (primary) and 21 CFR §888.3040                                                    |
| Regulation Name:           | Single/multiple component metallic bone fixation appliances and<br>accessories (primary); Screw, Fixation, Bone |
| Product Code:              | HRS (primary), HWC                                                                                              |
| Classification Panel:      | Orthopedics                                                                                                     |

## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The Valkyrie Thoracic Fixation System consists of a variety of screws and plates intended for use in the stabilization and fixation of fractures in the chest wall including sternal reconstructive surgical procedures, trauma, or planned osteotomies. The system is intended for use in patients with normal and/or poor bone quality.

To accommodate varying patient anatomy and surgeon preference, the Valkyrie Thoracic Fixation System includes screws in 3.0mm and 3.5mm diameters and lengths from 7-20mm. The system also includes various styles of plates. The Valkyrie Thoracic Fixation System plates are made from PEEK-Optima™ per ASTM F2026, and the screws are made from Ti-6Al-4V per ASTM F136.

## INDICATIONS FOR USE

This Valkyrie Thoracic Fixation System is intended for use in the stabilization and fixation of fractures in the chest wall including sternal reconstructive surgical procedures, trauma, or planned osteotomies. The system is intended for use in patients with normal and/or poor bone quality.

## TECHNOLOGICAL CHARACTERISTICS

The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates:

{4}------------------------------------------------

- Indications for Use ●
- Materials of manufacture ●
- Structural support mechanism ●

| 510k Number                           | Trade or Proprietary or Model Name                               | Manufacturer                         | Predicate Type |
|---------------------------------------|------------------------------------------------------------------|--------------------------------------|----------------|
| K163007, K151019,<br>K121302, K110574 | Biomet Microfixation<br>Sternalock 360 Sternal<br>Closure System | Biomet Microfixation                 | Primary        |
| K193468, K133785                      | Tritium Sternal Cable Plate<br>System                            | Pioneer Surgical<br>Technology, Inc. | Additional     |
| K151983                               | KLS Martin LSS Plating<br>System                                 | KLS Martin L.P.                      | Additional     |
| K170767                               | Stainless Steel Suture                                           | Ethicon Incorporated                 | Additional     |

### Table 5-1 Predicate Devices

## PERFORMANCE DATA

The Valkyrie Thoracic Fixation System has been tested in the following test modes:

- Static construct tension ●
- Dynamic construct tension
- Biocompatibility ●
	- Cytotoxicity (ISO 10993-5:2009) o
	- Systemic toxicity (ISO 10993-11:2017) o
	- Genotoxicity (ISO 10993-3:2014) O
	- Intracutaneous irritation (ISO 10993-10:2010) O
	- Pyrogenicity (ISO 10993-11:2017) o
	- Sensitization (ISO 10993-10:2010) o
	- o Chemical characterization (ISO 10993-18:2020)
- Sterilization validation (ISO 11137-1:2006) .
- Packaging Shelf-life (ASTM 1980) .
- Packing Peel Strength (ASTM F88)
- Packaging integrity validation (ASTM F1886, ASTM F1929, ASTM F2096) .

The results of this non-clinical testing show that the Valkyrie Thoracic Fixation System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

## CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the Valkyrie Thoracic Fixation System is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K202889](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K202889)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
